K190984 · Xiamen Ants Bro Technology Co., Ltd. · FLL · Apr 22, 2020 · General Hospital
Device Facts
Record ID
K190984
Device Name
Infrared Forehead Thermometer
Applicant
Xiamen Ants Bro Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Apr 22, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Infrared Forehead Thermometer (Models: HA-610, HA-620, HA-630, HA-640, HA-650) is non-sterile, reusable, clinical thermometer intended for the intermination of human body temperature in a no touch mode using the center of the forehead as the measurement site on people of all ages. The devices are reusable for home use only.
Device Story
Hand-held, battery-powered infrared thermometer; measures infrared energy emitted from forehead; converts input to oral equivalent temperature. Operates in no-touch mode at 1-5 cm distance. Components include thermopile sensor, integrated thermistor for ambient readings, buzzer, LCD, and ABS plastic enclosure. Used by patients at home for self-monitoring or by caregivers. Output displayed on LCD; provides temperature reading and high-temperature alarm. Assists users in monitoring body temperature; diagnosis remains responsibility of professional physician.
Clinical Evidence
Clinical testing performed on 240 subjects across three age groups (A: 0-1 year; B: 1-5 years; C: >5 years). Protocol and analysis conducted per ISO 80601-2-56. Results confirmed compliance with clinical performance requirements for body temperature measurement.
Technological Characteristics
ABS plastic enclosure; infrared thermopile sensor with integrated thermistor; battery-powered (1.5V AAA x 2); LCD display. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98 (2016). Biocompatibility per ISO 10993.
Indications for Use
Indicated for intermittent determination of human body temperature in people of all ages using the center of the forehead as the measurement site. Intended for home use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Braun BNT400 No Touch + Forehead Thermometer (K181242)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 22, 2020
Xiamen Ants Bro Technology Co., Ltd. % Cassie Lee Official Correspondent Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road Huangpu District Guangzhou, Guangdong 510000 China
Re: K190984
Trade/Device Name: Infrared Forehead Thermometer (Models: HA-610, HA-620, HA-640, HA-650) Regulation Number: 21 CFR 880.2910
Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: March 17, 2020 Received: March 24, 2020
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tina Kiang, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices and Human Factors OHT3: Office of Gastrorenal, OB/GYN, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190984
Device Name
Infrared Forehead Thermometer (Models: HA-610, HA-620, HA-630, HA-640, HA-650)
Indications for Use (Describe)
Infrared Forehead Thermometer (Models: HA-610, HA-620, HA-630, HA-640, HA-650)
is a non-sterile, reusable, clinical thermometer intent determination of human body temperature in a touch and no touch mode using the center of the forehead as the measurement site on people of all ages. The devices are reusable for home use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K190984
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
## 1. Submitter's Information
510(k) Owner's Name: Xiamen Ants Bro Technology Co., Ltd. Establishment Registration Number: Applying Address: 4th and 5th floor, No. 5 Building, Tech.&Innovation center, No. 289 Wengjiao Road, Haicang District, Xiamen China Tel: +86-134-5902-0349 Fax: +86-0592-6537633 Contact Person: Jane Xu Email: Sales1@asxd.com.cn
Date: April 22, 2020
## Application Correspondent:
Contact Person: Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
## 2. Subject Device Information
Trade Name: Infrared Forehead Thermometer (Models: HA-610, HA-620, HA-630, HA-640, Common Name: Clinical electronic thermometeHA-650) Classification name: Thermometer, electronic, clinical Review Panel: General Hospital Product Code: FLL Regulation Class: II Regulation Number: 21 CFR 880.2910
### 3. Predicate Device Information
Sponsor: Kaz USA, Inc., a Helen of Troy Company Trade Name: Braun BNT400 No Touch + Forehead Thermometer Common Name: Clinical electronic thermometer Classification Name: Thermometer, Electronic, Clinical 510(K) Number: K181242 Review Panel: General Hospital Product Code: FLL Regulation Number: 880.2910 Regulation Class: II
{4}------------------------------------------------
## 4. Device Description
The Infrared Forehead Thermometer (Models: HA-610, HA-630, HA-640, HA-640, HA-650) is a hand-held, battery powered, infrared Thermometer that coverts a user's forehead temperature, using the infrared energy emitted from the area around the user's forehead, to an oral equivalent temperature when placed within 1-5 cm to the subject's forehead with no contact. It uses a thermopile sensor with integrated thermistor for ambient temperature readings.
It composed by a measuring sensor, set button, a start button, battery compartment, Buzzer, a LCD and a ABS plastic enclosure, and measuring without probe cover.
## 5. Intended Use / Indications for Use
Infrared Forehead Thermometer (Models: HA-610, HA-620, HA-630, HA-640, HA-650) is non-sterile, reusable, clinical thermometer intended for the intermination of human body temperature in a no touch mode using the center of the forehead as the measurement site on people of all ages. The devices are reusable for home use only.
## 6. Test Summary
6.1 Infrared Forehead Thermometer has been evaluated the safety and performance by lab bench testing as following:
- Electrical safety test according to IEC 60601-1, IEC 60601-1-11 and ISO 80601-2-56 standards �
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- � ISO 80601-2-56 Medical Electrical Equipment - Part 2-56: Particular Requirements For Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement
- ◆ Particular Requirements for Basic Safety and Essential Performance of Clinical Thermometers for Body Temperature Measurement according to ASTM E1965-98 (2016): Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
## 6.2 Discussion of Clinical Tests Performed
Clinical tests were conducted on the Model HA-610. The clinical tests evaluated 240 of subjects. and the thermometer was evaluated in four groups A1 - 0 up to three months to one year; B1 older than one years and younger than five years; and C - older than five years old. The clinical performance test protocol and data analysis were conducted in accordance with the requirement of ISO80601-2-56. The test report showed the clinical performance of the subject devices complied with the requirement of ISO 80601-2-56.
### 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Infrared Forehead Thermometer is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Remark |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Company | Xiamen Ants Bro | Kaz USA, Inc., a Helen of | -- |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Remark |
| | Technology Co., Ltd. | Troy Company | |
| Device Name and<br>Model | Infrared Forehead<br>Thermometer (Models: HA-<br>610, HA-620, HA-630, HA-<br>640, HA-650) | Braun BNT400 No Touch +<br>Forehead Thermometer | -- |
| 510(k) Number | Applying | K181242 | -- |
| Product Code | FLL | FLL | SE |
| Thermometer Type | Infrared Forehead<br>Thermometer | Infrared Forehead<br>Thermometer | SE |
| Intended Use &<br>Indications for Use | Infrared Forehead<br>Thermometer (Models: HA-<br>610, HA-620, HA-630, HA-<br>640, HA-650)<br>is non-sterile, reusable,<br>clinical thermometer<br>intended for the intermittent<br>determination of human<br>body temperature in a no<br>touch mode using the<br>center of the forehead as<br>the measurement site on<br>people of all ages. The<br>devices are reusable for<br>home use only. | The Braun BNT400 No<br>Touch + Forehead<br>Thermometer is non-sterile,<br>reusable, clinical<br>thermometer intended for<br>the intermittent<br>determination of human<br>body temperature in a touch<br>and no touch mode using<br>the center of the forehead<br>as the measurement site on<br>people of all ages. | SE<br>Difference<br>in name<br>and<br>included<br>home<br>use. |
| Measurement<br>method | Infrared radiation detection | Infrared radiation detection | SE |
| Measurement<br>mode | Forehead measure mode | Forehead measure mode | SE |
| Features | High Temperature Alarm<br>feature and memory<br>feature | Fever Insight temperature<br>guidance feature and<br>memory feature | SE<br>Note 2 |
| Measuring<br>range | 34.0 - 43.0°C (93.2 –<br>109.4°F) | 34.0 - 43.0°C (93.2 –<br>109.4°F) | SE |
| Measurement<br>Accuracy (Body) | $\pm$ 0.2°C (35.0~42.0°C) / $\pm$<br>0.4°F (95.0~107.6°F)<br>$\pm$ 0.3°C (34.0~34.9°C)<br>(42.1~43.0°C) | $\pm$ 0.2°C (35.0~42.0°C) / $\pm$<br>0.4°F (95.0~107.6°F)<br>$\pm$ 0.3°C (34.0~34.9°C)<br>(42.1~43.0°C) | SE |
| Display Resolution | 0.1 °C/°F | 0.1 °C/°F | SE |
| C/F switchable | Yes | Yes | SE |
| Display | LCD display | LCD display | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Remark |
| Battery type | 1.5V AAA x 2 | 1.5V AAA x 2 | SE |
| Low battery<br>indication | Yes | Yes | SE |
| Materials | ABS | User contacting materials<br>include ABS (device<br>housing / handle, power<br>button), TPR (temperature<br>button & forehead touch<br>bumper), & PMMA (LCD<br>lens & protective scanner<br>cap). | SE<br>Note 3 |
| Measurement<br>distance | 1-5 cm | 0-2.5cm | Note 4 |
| Operating condition | Temperature: 10 - 40°C<br>(50.0 - 104.0°F)<br>Humidity: ≤ 85%RH | Temperature: 15 - 40°C (59<br>- 104.0°F)<br>Humidity: 15 ~ 95%RH | SE<br>Note 1 |
| Biocompatibility | ISO 10993 | ISO 10993 | SE |
| Electric Safety and<br>EMC | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-1-11 | IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-1-11 | SE |
| Performance | ISO 80601-2-56 | ISO 80601-2-56 | SE |
{5}------------------------------------------------
{6}------------------------------------------------
## Comparison in Detail(s):
## Note 1:
The "Operating condition" of subject device is a little different from the predicate devices, but all of them are meet the safety standards IEC 60601-1 and IEC 60601-1-11, the differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
## Note 2:
Although the "High Temperature Alarm feature" of subject device is a little different from the predicate devices, but both measuring temperature of subject device and predicate device are for indicate only diagnosis should be made by a professional physician, so this difference will not cause safety and effectiveness issues.
## Note 3:
The subject medical device in its final finished form is identical to Digital Thermometer (K190990) in formulation, processing, sterilization, and geometry and no other chemicals have been added (e.g. plasticizers, fillers, additives, cleaning agents, mold release agents).
## Note 4:
Although the "Measurement distance" of subject device is a little different from the predicate devices, but both of subject device and predicate device are meet the clinical accuracy requirements of the standards ISO 80601-2-56, the differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
{7}------------------------------------------------
# Finial Conclusion:
The subject device Infrared Forehead Thermometer (Models: HA-610, HA-630, HA-640, HA-650) have all features of the predicate devices K181242. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.