K190937 · Ethicon Endo-Surgery, LLC · OXC · Oct 14, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190937
Device Name
Echelon Endopath Staple Line Reinforcement
Applicant
Ethicon Endo-Surgery, LLC
Product Code
OXC · General, Plastic Surgery
Decision Date
Oct 14, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ECHELON ENDOPATH Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. ECHELON ENDOPATH Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel and colorectal procedures.
Device Story
Echelon Endopath Staple Line Reinforcement is an absorbable buttress used in surgical environments to reinforce staple lines during tissue transection. The device consists of a Vicryl (polyglactin 910) multifilament fiber layer laminated with Polydioxanone (PDO) film and a synthetic attachment material. It is applied to the anvil and reload of Echelon 60 mm endoscopic linear cutters. During surgery, the stapler places titanium staples through the tissue and the reinforcement material, simultaneously dividing both. The buttress provides mechanical support to soft tissue during the healing process and is absorbed by the body via hydrolysis within approximately 120 days. The device is intended for use by surgeons in clinical settings to improve staple line integrity and reduce bleeding at the transection site.
Clinical Evidence
No clinical data. Substantial equivalence supported by ex-vivo design verification (thickness, porosity, tensile strength, staple line integrity) and in-vivo animal studies evaluating hemostasis, tissue healing, and device usability in gastric, thoracic, small bowel, and colorectal models.
Technological Characteristics
Absorbable implant consisting of knitted Vicryl (polyglactin 910) fibers laminated with PDO film and a synthetic polymer/poloxamer adhesive. Dimensions designed for Echelon 60 mm endoscopic linear cutters. Biocompatibility per ISO 10993-1. Sterile, single-use. Absorption via hydrolysis (complete at ~120 days).
Indications for Use
Indicated for patients undergoing surgical procedures requiring soft tissue transection or resection where staple line reinforcement is needed, including lung resection, bariatric, gastric, small bowel, and colorectal procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Gore Seamguard Bioabsorbable Staple Line Reinforcement (K032865)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
October 14, 2019
Ethicon Endo-Surgery LLC Emily Nesbitt Associate Director, Regulatory Affairs 4545 Creek Rd Cincinnati, Ohio 45242
Re: K190937
Trade/Device Name: Echelon Endopath Staple Line Reinforcement Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OXC Dated: September 10, 2019 Received: September 13, 2019
Dear Emily Nesbitt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190937
Device Name ECHELON ENDOPATH Staple Line Reinforcement
#### Indications for Use (Describe)
ECHELON ENDOPATH Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. ECHELON ENDOPATH Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel and colorectal procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:12px;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span style="font-size:12px;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
#### Company
Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969
#### Contact
Emily Nesbitt (née Kruetzkamp), Associate Director, Regulatory Affairs Ethicon Endo-Surgery, Inc. Telephone: 513.337.1546 Email: enesbitt@its.jnj.com
| Date Prepared: | April 9, 2019 |
|----------------------|------------------------------------------------------------------------------|
| Trade Name: | Echelon Endopath Staple Line Reinforcement |
| Common Name: | Staple Line Reinforcement Material |
| Classification Name: | Surgical mesh |
| Classification: | 21CFR 878.3300 |
| Product Code: | OXC |
| Device Class: | Class II |
| Panel: | General and Plastic Surgery |
| Predicate Device | Gore Seamguard Bioabsorbable Staple Line Reinforcement cleared under K032865 |
| Reference Device | Vicryl (Polyglactin 910) Mesh, K810428 |
### Device Description
Echelon Endopath Staple Line Reinforcement is a staple line reinforcement, also known as a buttress, for use in the surgical environment for the purpose of reinforcing a staple line.
The Subject Device is to be used Echelon Endopath GST 60 mm reloads in conjunction with the Echelon family of 60 mm endoscopic linear cutters. The Echelon 60 mm Stapler in conjunction with the Echelon Endopath GST60 mm reload places staggered rows of titanium staples with a reinforcement material, and simultaneously divide the tissue and the reinforcement material between the stapled rows. The Subject Device is an absorbable staple line reinforcement material which is secured to the stapler anvil and reload with a synthetic attachment material. The Subject Device contains an applicator and the implantable device which consists of 3 materials: the Vicryl material, the Polydioxanone film and the attachment material. Echelon Endopath Staple Line Reinforcement is an implanted material which works as an adjunct to surgical staples after transection, to provide support to soft tissue during the healing process.
The Echelon Endopath Staple Line Reinforcement Subject Device and Predicate Device have identical intended use: to support staple line reinforcent in soft tissue. The Subject Device has a
{4}------------------------------------------------
smaller list of anatomical sites than the Predicate Device, and the Subject Device is a subset of the Predicate Indications.
The Echelon Endopath Staple Line Reinforcement Subject Device and Predicate Device are both synthetic, absorbable staple line reinforcement materials to be applied to staplers to provide support to the staple line during the tissue healing process. The Subject Device is to be applied to the Ethicon Echelon Flex Powered Plus Stapler. Both materials degrade via hydrolysis and have been evaluated for biocompatibility in alignment with ISO 10993-1 and are deemed biocompatible.
Echelon Endopath Staple Line Reinforcement is composed of undyed multifilament fibers identical to those used in Vicryl surgical sutures. The polyglactin 910 implant is knitted by a process which interlocks each fiber juncture to prevent unraveling. The implanted material consists of a Vicryl layer laminated with a PDO film on each side, and a Buttress Attachment Material which is a synthetic material applied to the top outer surface of each implant. The staple line reinforcement is used as a buttress to provide support during the healing process. Both Subject and Predicate materials are hydrolyzed by body fluids and absorbed after serving the intended function. The subject Device absorption is essentially complete at 120 days.
The primary difference between the Subject and Predicate Devices is the feature for loading and delivery of the device to the implantation site. A thin layer of bioabsorbable adhesive is coated onto one surface of the implantable device and is composed of a synthetic polymer/poloxamer blend to attach the Subject Device to the stapler. Whereas the Predicate Device utilized a sleeve construction and suture pullcord to secure the reinforcement material on the stapler.
### Indications for Use
ECHELON ENDOPATH Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection with staple line reinforcement is needed. ECHELON ENDOPATH Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel and colorectal procedures.
### Summary of Similarities and Differences in Technological Characteristics
Staple line reinforcement is the technological principle for both the Subject Endopath Echelon Staple Line Reinforcement and the Predicate Gore® Seamguard® Reinforcement device. It is based on the use of applying an absorbable material, also known as a buttress, as a reinforcement to the transected surgical staple line in soft tissue. At a high level, the Subject and Predicate Devices are based on the following same technological elements:
- Staple line reinforcement device is applied to a surgical stapler
- Staple line reinforcement and stapler are positioned on the desired anatomical area to be transected
- Stapler is fired, knife proceeds to transect tissue, and staples with buttress are applied to ● the tissue
- Absorbable implant reinforces tissue and helps to secure the staple line ●
{5}------------------------------------------------
The Endopath Echelon Staple Line Reinforcement differs from Predicate Gore® Seamguard® Reinforcement device in the Steps for Use to load the buttress device onto the surgical stapler. The Subject Device uses an attachment material to permit temporary attachment to the surgical stapler.
Both devices are sterile, single use, absorbable implants that absorb over time. The Subject Device absorbs over approximately 120 days, the Predicate Device absorbs over approximately 180 days.
# Performance Data
Performance tests were performed ex-vivo and in-vivo, as appropriate for the required endpoint. to demonstrate the ECHELON ENDOPATH Staple Line Reinforcement is substantially equivalent to the Predicate Device.
Ex-vivo evaluations include design verification testing of the ECHELON ENDOPATH Staple Line Reinforcement, including:
- Device Implant Characterization ●
- Thickness O
- O Porosity
- Density O
- Tensile O
- Stiffness O
- Tear Resistance O
- Burst strength o
- Staple pull out strength O
- Device Compatibly with Staplers and Reloads ●
- Force to Close o
- Staple Height O
- Staple Form Quality O
- Staple Line Integrity O
- Release of stapled reinforcement material from Endocutter O
- Manipulation of device on tissue, buttress security on surgical stapler .
Preclinical in-vivo evaluations include studies with animal models which support the intended use of the ECHELON ENDOPATH Staple Line Reinforcement, demonstrating performance in the indicated gastric, thoracic, small bowel, and colon/colorectal procedures. The surgical functionality of the Subject Device, such as device attachment to the surgical stapler and usability, was demonstrated in animal studies. In-vivo testing evaluations include:
- Hemostasis, bleeding at the staple line
- . Tissue healing response, survival studies: Gastric, thoracic tissue absorption, Perpendicular firings, L-shaped firings, Sequential Firings
{6}------------------------------------------------
- Usability of the Subject Device
- Abrasion evaluation on tissue, survival study ●
Animal tests were performed to verify that the Subject Device meets the definition of substantial equivalence to the Predicate Device.
Stability studies were conducted, and the shelf life of the device was demonstrated. Biocompatibility evaluation was performed and confirmed that the proposed device is biocompatible for the intended patient contact profile. Subject Device is a medical device implant for use on tissue with long-term (>30 days) patient exposure. The patient-contacting materials used in the Subject Device are categorized as Long-term patient contacting materials.
## Consensus Standards
All components of the Echelon Endopath Staple Line Reinforcement are comprised of materials which are in accordance with ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.
### Clinical
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
### Conclusion
The conclusions of the testing criteria demonstrate that ECHELON ENDOPATH Staple Line Reinforcement is as safe and effective and performs as well as the legally marketed Predicate Device, Gore Seamguard K032865 in terms of intended use, design, materials, biocompatibility, sterility, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.