VitalPatch 7D Biosensor - single, VitalPatch 7D Biosensor - bag of 20
K190916 · Vitalconnect, Inc. · DRG · Jun 7, 2019 · Cardiovascular
Device Facts
Record ID
K190916
Device Name
VitalPatch 7D Biosensor - single, VitalPatch 7D Biosensor - bag of 20
Applicant
Vitalconnect, Inc.
Product Code
DRG · Cardiovascular
Decision Date
Jun 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Indications for Use
The VitalConnect Platform is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This can include heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, body temperature, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). Data are transmitted wirelessly from the VitalConnect Sensor for storage and analysis. The VitalConnect Platform can include the ability to notify healthcare professionals when physiological data fall outside selected parameters. The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information. The data from the VitalConnect Platform are intended for use by healthcare professionals as an aid to diagnosis and treatment. The device is not intended for use on critical care patients.
Device Story
Wearable biosensor (VitalPatch) collects physiological signals (ECG, skin/ambient temperature, tri-axial accelerometer data); transmits data wirelessly via Bluetooth Low Energy to central server for storage/analysis; used by healthcare professionals in home/clinical settings to monitor general care patients. Device processes inputs to calculate heart rate, respiratory rate, HRV, R-R interval, step count, and posture; detects falls; notifies clinicians of out-of-parameter data. Output aids diagnosis and treatment decisions. Modification extends wear duration from 5 to 7 days (168 hours).
Clinical Evidence
No clinical data. Substantial equivalence established via benchtop performance testing, including wear duration, battery life, and EOL detection testing for 168-hour operation. Biocompatibility, electrical safety (IEC 60601-1 series), and EMC (IEC 60601-1-2) testing previously conducted.
Technological Characteristics
Wearable patch with ECG electrodes, skin/ambient thermistors, and tri-axial accelerometer. Wireless connectivity via Bluetooth Low Energy (2.4GHz ISM band). Data encrypted via AES-CCM. Dimensions: 120 x 40.6 x 9.4 mm. Firmware v3.2.1.0. Complies with IEC 60601-1, 60601-1-11, 60601-2-25, 60601-2-47, and FCC Part 15.
Indications for Use
Indicated for general care patients 18 years or older for continuous collection of physiological data (ECG, heart rate, HRV, R-R interval, respiratory rate, body/skin temperature, activity, posture) in home and healthcare settings. Not for critical care patients.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
K190196 % Cynthia Merrell Vice-President QA &RA VitalConnect Inc. Contact Address
Re: K190196
Trade/Device Name: VitalPatch® Biosensor Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter And Receiver Regulatory Class: Class II Product Code: DRG Dated: June 7, 2019 Received: April 9, 2019
Dear Cynthia Merrell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Matthew Hillebrenner Deputy Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190196
Device Name VitalPatch Biosensor
#### Indications for Use (Describe)
The VitalConnect Platform is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This can include heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, body temperature, activity (including step count), and posture (body position relative to gravity including fall). Data are transmitted wirelessly from the VitalConnect Sensor for storage and analysis. The VitalConnect Platform can include the ability to notify healthcare professionals when physiological data fall outside selected parameters.
The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information. The data from the VitalConnect Platform are intended for use by healthcare professionals as an aid to diagnosis and treatment. The device is not intended for use on critical care patients.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
#### 1. SUBMITTER VitalConnect, Inc. Address 224 Airport Parkway, Suite 300 City, State, Zip USA San Jose, CA, 95110, USA Tel: +1.408.963.4600 Fax: +1.408.963.2828
Contact Person: Cynthia Merrell, VP QA&RA Date Prepared: April 4, 2019
| II. DEVICE | |
|------------------------------|-------------------------------------|
| Name of Device: | VitalPatch Biosensor |
| Classification Name: | Cardiovascular Monitoring Devices 2 |
| Regulation: | CFR §870.2910, 21 CFR 870.1025 |
| Regulatory Class: | Class II |
| Product Classification Code: | DRG, DSI, MHX |
PREDICATE DEVICE |||.
| Predicate Manufacturer: | Vital Connect, Inc. |
|-------------------------|-----------------------------------|
| Predicate Trade Name: | VitalPatch® VitalConnect Platform |
| Predicate 510(k): | K183078 |
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
VitalPatch is a wearable biosensor designed to measure a patient's vital signs, including heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). VitalPatch was most recently cleared in K183078.
The following modifications are proposed for the VitalPatch wearable biosensors:
- . Increasing the duration for use from 5 days (120 hours) to 7 days (168 hours). This includes a minor software change and minor changes to the IFU.
All other vital signs, including heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, body temperature, skin temperature, activity (including step count), and posture (body position relative to gravity including fall) cleared in K183078 remain the same.
{4}------------------------------------------------
#### INDICATIONS FOR USE V.
VitalConnect, Inc.
The VitalConnect Platform is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This can include heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, body temperature, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). Data are transmitted wirelessly from the VitalConnect Sensor for storage and analysis. The VitalConnect Platform can include the ability to notify healthcare professionals when physiological data fall outside selected parameters.
The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information. The data from the VitalConnect Platform are intended for use by healthcare professionals as an aid to diagnosis and treatment. The device is not intended for use on critical care patients.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVCE
The intended use and technological features of the proposed Vital Connect Platform do not substantially differ from the legally marketed predicate device. The Vital Connect Platform and the predicate device have substantially equivalent intended uses and methods of operation.
| Predicate (K183078) | Subject Device |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The VitalConnect Platform is a wireless remote monitoring<br>system intended for use by healthcare professionals for<br>continuous collection of physiological data in home and<br>healthcare settings. This can include heart rate,<br>electrocardiography (ECG), heart rate variability, R-R<br>interval, respiratory rate, body temperature, skin<br>temperature, activity (including step count), and posture<br>(body position relative to gravity including fall). Data are<br>transmitted wirelessly from the VitalConnect Sensor for<br>storage and analysis. The VitalConnect Platform can<br>include the ability to notify healthcare professionals when<br>physiological data fall outside selected parameters.<br><br>The device is intended for use on general care patients who<br>are 18 years of age or older as a general patient monitor, to<br>provide physiological information. The data from the<br>VitalConnect Platform are intended for use by healthcare<br>professionals as an aid to diagnosis and treatment. The<br>device is not intended for use on critical care patients. | The VitalConnect Platform is a wireless remote monitoring<br>system intended for use by healthcare professionals for<br>continuous collection of physiological data in home and<br>healthcare settings. This can include heart rate,<br>electrocardiography (ECG), heart rate variability, R-R<br>interval, respiratory rate, body temperature, skin<br>temperature, activity (including step count), and posture<br>(body position relative to gravity including fall). Data are<br>transmitted wirelessly from the VitalConnect Sensor for<br>storage and analysis. The VitalConnect Platform can<br>include the ability to notify healthcare professionals when<br>physiological data fall outside selected parameters.<br><br>The device is intended for use on general care patients who<br>are 18 years of age or older as a general patient monitor, to<br>provide physiological information. The data from the<br>VitalConnect Platform are intended for use by healthcare<br>professionals as an aid to diagnosis and treatment. The<br>device is not intended for use on critical care patients. |
#### Table 1 Comparison of Indications for Use (the indications for use are identical)
{5}------------------------------------------------
#### Table 2 Size Comparison of the adhesive patch
| Dimension | Predicate (K183078) | Subject Device |
|-----------|---------------------|----------------|
| Length | 115 mm | 120 mm |
| Width | 35.6 mm | 40.6 mm |
| Height | 7.5 mm | 9.4 mm |
#### Table 3 Component Comparison of the Predicate and Subject Devices
| Component | Predicate (K183078) | Proposed Device |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| ECG electrodes | Allow the recording of a single-lead<br>bipolar ECG at a sampling rate of 125<br>Hz | Identical |
| Skin Thermistor | Mounted on the patch. Designed to<br>monitor skin temperatures when the<br>patch is attached to the skin. | Identical |
| Ambient Thermistor | Designed to monitor ambient<br>temperature to more accurately calculate<br>body temperature using the skin<br>temperature input. | Identical |
| Flexible Assembly | Provides a connection between the ECG<br>electrodes, thermistors and patch<br>electronics. The flexible circuit is sealed<br>within the patch. | Identical |
| Firmware | Version 3.1.0.1 to support functions<br>listed in Table 4 for Predicate<br>(K183078) | Version 3.2.1.0 of firmware fixes a bug<br>and allows configuration for 120-hour<br>patch life timer or 168-hour patch-life<br>timer. |
#### Table 4 – Device Performance Characteristics Comparison
| Performance<br>Characteristic | Predicate (K183078) | Proposed Device |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Wireless Transmission | Bluetooth Low Energy (BT4.1) technology | Identical |
| Data encryption | Advanced Encryption Standard-CCM mode | Identical |
| Radio Frequency | 2.4GHz ISM band<br>FCC Part 15 Complied | Identical |
| ECG Dynamic Range | The Sensor electrodes make contact with<br>the skin to measure the differential voltage<br>generated from the heart. The resulting<br>analog ECG waveform is digitized so that<br>the data can be streamed to a display or<br>recording device.<br>Range: -10mV to +10mV | Identical |
| Heart Rate (stationary<br>and ambulatory) | Heart rate is measured and calculated in<br>Beats Per Minute (BPM) both for stationary<br>and ambulatory use.<br>Range: 30 - 200 BPM | Identical |
| Respiratory Rate | From a combination of ECG and tri-axial<br>accelerometer sensor signals, the Vital<br>Connect Sensor can accurately measure the<br>respiratory rate of the person, irrespective of<br>whether the person is stationary or<br>ambulatory.<br>Range: 10-30 breaths per minute | Identical |
| Skin Temperature | Using a thermistor sensor on the Patch,<br>which is in close proximity to the skin, the<br>Vital Connect Sensor accurately measures<br>the temperature of the skin, reporting the<br>temperature in degrees Centigrade. A<br>known quantity of current generated by the<br>sensor is converted to a voltage that is<br>accurately matched to skin temperature.<br>Range: 15°C - 50 °C (61°F - 113°F) | Identical |
| Ambient Temperature | Using an additional thermistor sensor on the<br>Patch, which is away from the skin, the<br>VitalConnect Biosensor accurately<br>measures patch ambient temperature,<br>reporting the temperature in degrees<br>Centigrade. A known quantity of current<br>generated by the sensor is converted to a<br>voltage that is accurately matched to the<br>ambient temperature.<br>Range: 15°C - 50 °C (61°F - 113°F) | Identical |
| Body Temperature | A calculated value as a function of the skin<br>and ambient temperatures.<br>Range: 32°C - 42°C | Identical |
| Fall Detection | Using a tri-axial accelerometer and digital<br>signal processing techniques, the Vital<br>Connect Sensor detects falls while<br>minimizing false notifications.<br>Range: Fall or No Fall | Identical |
| Step Count | The Vital Connect Sensor also uses the tri-<br>axial accelerometer to compute step count.<br>As the person walks, the Vital Connect<br>Sensor is capable of distinguishing steps<br>from other movements. The detected steps<br>are accumulated to provide an accurate step<br>count.<br>Range: 0 - 65535 steps | Identical |
| Posture Detection | Using the built-in tri-axial accelerometer,<br>the Vital Connect Sensor can determine the<br>posture of the person and provide wireless<br>real-time updates to a central server.<br>Postures detected include lying down,<br>upright, walking, running and leaning. | Identical |
| R-R Interval | The R-R interval is the measurement of the<br>interval from the R wave peak of one QRS<br>complex to the next R-wave peak on the<br>electrocardiogram. The device has enhanced<br>QRS performance during motion. | Identical |
| Heart Rate Variability | HRV quantifies the variation in the beat-to-<br>beat interval time series obtained from the<br>ECG waveform. Using advanced signal<br>processing algorithms, the Sensor detects<br>each QRS peak with high temporal<br>accuracy. Measurement of the R wave to R<br>wave (RR) interval allows short and long-<br>term variability analysis to determine<br>analytics such as mean, median, standard<br>deviation, frequency content, etc. | Identical |
| Low Power Mode | When patch is off the body, to conserve<br>battery power, the processor goes into a low<br>power mode, resulting in reduced current<br>draw | Identical |
{6}------------------------------------------------
{7}------------------------------------------------
### VII. PERFORMANCE DATA
Verification and validation activities established the safety and performance characteristics of the proposed device with respect to the predicate. The following performance data have been provided in support of the substantial equivalence determination:
# Sterilization & Shelf-life Testing
The Vital Connect Platform is provided non-sterile, and therefore sterilization data is not provided. Accelerated aging data was provided to support a shelf-life of 9 months as previously cleared.
# Biocompatibility Testing
Biocompatibility testing, previously conducted, included in-vitro cytotoxicity, irritation and sensitization, according to the recommendations of ISO 10993-1:2009, Biological evaluation of medical devices – Part 1: Evaluation and testing.
{8}------------------------------------------------
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the VitalPatch. The device complies with the IEC 60601-1, IEC 60601-1-11. IEC 60601-2-25, and the IEC 60601-2-47 standards for safety. The device also complies with the IEC 60601-1-2 standard for EMC, IEC/TS 62657-2 and FCC CRF47 Part 15 Subpart C standards for wireless communication.
## Software Verification and Validation Testing
Software verification and validation testing was conducted, and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The software for this device is determined as a "moderate" level of concern because a failure or latent flaw could lead to a minor injury to the patient through incorrect information or through the action of the care provider.
# Benchtop & Simulated Use Testing
The following testing was provided to demonstrate the safe and effective performance of the subject device for 168 hours:
- Wear Duration Testing .
- . Battery operating life testing
- . EOL Detection after 168 hours
# Animal Study
Animal performance testing was not required to demonstrate safety and effectiveness of the device.
# Clinical Studies
Clinical testing was not required to demonstrate the safety and effectiveness of the VitalPatch. Instead, substantial equivalence is based upon benchtop performance testing.
### VIII. CONCLUSIONS
The subject device is substantially equivalent in design and intended use to the predicate device. Any differences between the subject device and the predicate device have no significant influence on safety or effectiveness as established through performance testing. The subject device raises no new issues of safety or effectiveness from the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.