The Discovery 1 - Non-invasive Hemodynamic Blood Pressure Monitor device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory. The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.
Device Story
Portable, handheld, non-invasive hemodynamic blood pressure monitor; powered by internal rechargeable lithium polymer battery. Device uses pulse wave probe with two LEDs to collect PPG signals from capillary fingertip tissue. Principle of operation involves calculating blood pressure via pulse wave conduction time and pulse wave characteristics; systolic/diastolic pressures derived through fitting operation; pulse rate calculated from pulse wave signal. Intended for home use by adults 18+; results stored in memory. Provides users with blood pressure trends and spot-check pulse data to assist in monitoring hemodynamic status.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and ISO 81060-2:2013 (clinical validation of automated measurement) demonstrated.
Technological Characteristics
Handheld device; internal 3.7V 2200mAh Li-Pol battery; DC 5V adapter. Sensing via two-LED pulse wave probe. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 81060-2, and IEC 62133. IP22 ingress protection.
Indications for Use
Indicated for adults aged 18+ for non-invasive measurement and trending of systolic/diastolic blood pressure, peripheral pulse rate, and peripheral pulse wave using capillary fingertip tissue.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 7, 2019
Vita-Course Technologies Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. No. A415, Block A. NanShan Medical Devices Industrial Park, Nanshan District, Shenzhen, 518067 Cn
Re: K190893
Trade/Device Name: Discovery 1 - Non-invasive Hemodynamic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 9, 2019 Received: May 9, 2019
Dear Kevin Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190893
### Device Name
Discovery 1 - Non-invasive Hemodynamic Blood Pressure Monitor
### Indications for Use (Describe)
The Discovery 1 - Non-invasive Hemodynamic Blood Pressure Monitor device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory. The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
## Prepared in accordance with the requirements of 21 CFR Part 807.92
## Prepared Date: 2019/04/01
#### 1. Submission sponsor
Name: Vita-Course Technologies Co., Ltd.
Address: 3F, Block B, Dajiahao Plaza, No.362, Yu'an 2nd Road, Bao'an District, Shenzhen, Guangdong, China
Contact person: Kezheng Ma
Title: Management Representative
E-mail: makezheng@vita-course.com
Tel: +86-755-23188226
#### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District, Shenzhen,
Guangdong, P.R. China 518067
Contact person: Kevin Wang
E-mail: kevin(@chonconn.com
Tel: +86-755-33941160
| Trade/Device Name | Non-invasive Hemodynamic Blood Pressure Monitor |
|-------------------|---------------------------------------------------------------------------|
| Model | Discovery 1 |
| Common Name | Non-invasive Blood Pressure Monitor |
| Regulatory Class | Class II |
| Classification | 21CFR §870.1130 / System, measurement, blood-pressure, non-invasive / DXN |
| Submission type | Traditional 510(K) |
#### Subject Device Information 3.
#### 4. Predicate Device
By submission of the Traditional 510(k), Vita-Course Technologies Co., Ltd. is requesting clearance for Discovery 1 Non-invasive Hemodynamic Blood Pressure Monitor. It is comparable to the following legally marketed system:
- Cnoga Medical ACAL, TensorTip ""- Non-invasive Hemodynamic Blood Pressure Monitor 1. under K093981.
#### ട്. Intended use & Indication for use
The Discovery 1 Non-invasive Hemodynamic Blood Pressure Monitor device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary
{4}------------------------------------------------
finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory. The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.
#### 6. Device Description
Discovery 1 is a portable non-invasive Hemodynamic Blood Pressure Monitor powered by an internal rechargeable lithium polymer battery. It contains a pulse wave probe incorporating two LEDs, which can collect the pulse wave to the device.
#### 7. Principle of Operation
The product is designed to calculate the blood pressure in terms of conduction time of pulse and wave parameters of pulse. This method is proposed based on interdependent characteristics between conduction time of the pulse wave along the arterial as well as characteristic parameters of the pulse wave and arterial pressure. Through the determination of the PPG signal, conduction time of the pulse wave along thee arterial and other correlation variables would be obtained and accordingly the systolic and diastolic pressures of the body are gotten by fitting operation. The pulse rate can be obtained by calculating the pulse wave signal.
#### Comparison to the Predicate Device 8.
The Discovery 1 non-invasive Hemodynamic Blood Pressure Monitor and TensorTip™ - Non-invasive Hemodynamic Blood Pressure Monitor (K093981), are identical in purpose, function, core technology and method of operation. Only minor differences exist between the Discovery 1 and predicate, which do not affect the safety or effectiveness of the subject device. Table 1 provides a comparison of the subject and predicate devices.
| Features | Subject Device | Predicate Device K093981 | Remark |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| | Discovery 1 non-invasive<br>Hemodynamic Blood Pressure<br>Monitor | TensorTipTM - Non-invasive<br>Hemodynamic Blood Pressure Monitor | |
| Indication<br>for use | The Discovery 1 - Non-invasive<br>Hemodynamic Blood Pressure Monitor<br>device is a small, lightweight,<br>handheld, device intended for<br>measuring and display of Blood<br>Pressure trending (systolic and<br>diastolic) and spot-check of Peripheral<br>pulse rate (PPR) and Peripheral pulse<br>wave (PPW). Measurement is<br>performing on capillary finger tip tissue<br>(other than the thumb). The ring finger<br>is the recommended site. The results of | The TensorTip - Non-invasive blood<br>pressure trending device is a small,<br>lightweight, handheld, device intended for<br>measuring and display of Blood Pressure<br>trending (systolic and diastolic) and spot-<br>check of Peripheral pulse rate (PPR) and<br>Peripheral pulse wave (PPW).<br>Measurement is performing on capillary<br>fingertip tissue (other than the thumb). The<br>ring finger is the recommended site. The<br>results of each measurement are stored in<br>the system memory. The device is intended | Same |
| Features | Subject Device<br>Discovery 1 non-invasive<br>Hemodynamic Blood Pressure<br>Monitor<br>system memory. The device is intended<br>for use in the home environment. It is<br>intended to be used by any person aged<br>above 18 years old. | Predicate Device K093981<br>TensorTipTM- Non-invasive<br>Hemodynamic Blood Pressure Monitor<br>for use in the home environment. It is<br>intended to be used by any person aged<br>above 18 years old. | Remark |
| Patient<br>Population | >18 years old | >18 years old | Same |
| Principle of<br>Operation | Hemodynamics | Hemodynamics | Same |
| Power<br>source | DC adaptor: 5Vd.c.,<br>Internal lithium polymer (Li-Pol)<br>battery: 3.7Vd.c., 2200 mAh | DC adaptor: 5Vd.c.<br>Internal Li-Polymer battery: 3.7Vd.c.,<br>740mAh | Different 1) |
| Blood<br>pressure | Systolic pressure range: 50 mmHg ~200 mmHg<br>Diastolic pressure range: 40 mmHg ~160 mmHg<br>Accuracy: ±5 mmHg | Systolic pressure range: 90 mmHg ~ 240 mmHg<br>Diastolic pressure range: 30 mmHg ~ 130 mmHg<br>Accuracy: <10% | Different 2) |
| Pulse rate | Measurement range: 40 bpm ~ 240 bpm<br>Accuracy: ±2% or ±2 bpm, whichever<br>is greater | Measurement range: 30 bpm ~200 bpm<br>Accuracy: Mean Square Error 0.5digits | Different 3) |
| Operating<br>environment | Temperature: 5 °C ~ 45 °C<br>Humidity: 10% ~ 95% | Temperature: 6 °C ~ 45 °C<br>Humidity: 10 ~ 80% | Different 4) |
| IP degree | IP22 | IP00 | Different 5) |
| Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 81060-2 | IEC 60601-1<br>IEC 60601-1-2<br>ISO 81060-2 | Different 6) |
Table 1: Comparison to Predicate Device
each measurement are stored in the
{5}------------------------------------------------
Justifications for differences between Discovery 1 and the predicate device are shown as below:
Different (1): The capacity of battery is different. However, the battery of subject device is in compliance with IEC 62133: 2012- Secondary cells and batteries containing alkaline or other non-acid electrolytes -Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications. Thus, this difference will not raise any safety or effectiveness issue.
Different (2): The measurement range and the accuracy of blood pressure is different. The subject device is compliant with FDA recognized consensus standard ISO 81060-2: 2013. This difference will not raise any safety or effectiveness issue.
Different (3): The measurement range and the accuracy of pulse rate is different. The subject device has been verified all the full claimed range. This difference will not raise any safety or effectiveness issue.
{6}------------------------------------------------
Different (4): The operating environment is minor different. However, the difference will not raise any safety or effectiveness issue.
Different (5): The degree of protection is different and the subject device is better than predicate device. The subject device is compliant with FDA recognized consensus standard IEC 60601-1-11: 2015. This difference will not raise any safety or effectiveness issue.
Different (6): The compliance standards are different. The subject device is compliant with FDA recognized consensus standard IEC 60601-1-11: 2015. This difference will not raise any safety or effectiveness issue.
#### 9. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
| NO. | Item | Description |
|-----|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Safety | Ensures the device meet the requirements of IEC 60601-1: 2012 and<br>IEC 60601-1-11: 2015 |
| 2 | EMC | Ensures the temperature probes meet the requirements of IEC<br>60601-1-2: 2014, IEC 60601-1-11: 2015 clause 12 and ISO 80601-<br>2-61: 2017 clause 201.17 and clause 202 |
| 3 | Clinical accuracy | ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2:<br>Clinical validation of automated measurement type |
## 10. Conclusion
It has been shown in this 510(k) submission that the difference between the proposed devices and the predicate devices do not raise any questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards demonstrate that the proposed are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, performance characteristics, and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.