← Product Code [LZA](/productcode/LZA) · K190876

# mCare®Powder-free Nitrile Blue Examination Gloves with Chemotherapy Drugs Labeling Claimed (K190876)

_Mercator Medical (Thailand), Ltd. · LZA · Dec 2, 2019 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190876

## Device Facts

- **Applicant:** Mercator Medical (Thailand), Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Dec 2, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

mCare® Powder Free Nitrile Blue Examination Gloves with Chemotherapy Drugs Labeling Claimed are disposable devices intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05, The following drugs had no breakthrough detected up to 240 minutes 1. Cisplatin, 1.0 mg/ml 2. Cyclophosphamide (Cytoxan), 20 mg/ml 3. Dacarbazine (DTIC), 10.0 mg/ml 4. Doxorubicin Hydrochloride, 2.0 mg/ml 5. Etoposide (Toposar), 20.0 mg/ml 6. Fluorouracil, 50.0 mg/ml 7. Methotrexate, 25 mg/ml 8. Mitomycin C,0.5 mg/ml 9. Paclitaxel (Taxol), 6.0 mg/ml 10. Vincristine 1.0 mg/ml The following drugs showed breakthrough detected in less than 240 minutes: 1. Carmustine (BCNU), 3.3 mg/ml: 24.6 minutes 2. Thiotepa (THT) 10.0 mg/ml: 68.9 minutes Warning-Not for Use with Carmustine and Thiotepa

## Device Story

mCare® Powder-Free Nitrile Blue Examination Gloves are disposable, non-sterile nitrile gloves. Designed for use by healthcare professionals or in clinical settings to provide a barrier against contamination between patient and examiner. The gloves are tested for permeation resistance against specific chemotherapy agents per ASTM D6978-05. The device provides a protective barrier; however, it is not intended for use with Carmustine or Thiotepa due to breakthrough times below 240 minutes. The gloves are intended for over-the-counter use.

## Clinical Evidence

Bench testing only. Permeation testing performed per ASTM D6978-05 to determine breakthrough times for ten chemotherapy drugs. Results indicate no breakthrough for eight drugs up to 240 minutes; breakthrough detected for Carmustine and Thiotepa in less than 240 minutes.

## Technological Characteristics

Material: Nitrile. Form factor: Disposable examination glove. Testing standard: ASTM D6978-05 for chemotherapy drug permeation. Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are displayed side-by-side.

Mercator Medical (Thailand) LTD % Kevin Walls Principal Consultant Regulatory Insight, Inc. Contact Address

Re: K190876

Trade/Device Name: mCare® Powder-Free Nitrile Blue Examination Gloves with Chemotherapy Drugs Labeling Claimed Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: October 26, 2019 Received: October 28, 2019

Dear Kevin Walls:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

## K190876

Device Name

mCare® Powder-Free Nitrile Blue Examination Gloves with Chemotherapy Drugs Labeling Claimed

### Indications for Use (Describe)

mCare® Powder Free Nitrile Blue Examination Gloves with Chemotherapy Drugs Labeling Claimed are disposable devices intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner.

In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05, The following drugs had no breakthrough detected up to 240 minutes

- 1. Cisplatin, 1.0 mg/ml
- 2. Cyclophosphamide (Cytoxan), 20 mg/ml
- 3. Dacarbazine (DTIC), 10.0 mg/ml
- 4. Doxorubicin Hydrochloride, 2.0 mg/ml
- 5. Etoposide (Toposar), 20.0 mg/ml
- 6. Fluorouracil, 50.0 mg/ml
- 7. Methotrexate, 25 mg/ml
- 8. Mitomycin C,0.5 mg/ml
- 9. Paclitaxel (Taxol), 6.0 mg/ml
- 10. Vincristine 1.0 mg/ml

The following drugs showed breakthrough detected in less than 240 minutes:

- 1. Carmustine (BCNU), 3.3 mg/ml: 24.6 minutes
2. Thiotepa (THT) 10.0 mg/ml: 68.9 minutes

Warning-Not for Use with Carmustine and Thiotepa

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K190876](https://fda-staging.innolitics.com/device/K190876)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).