The Bio-TackS Orthodontic Mini Implant is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.
Device Story
Bio-TackS Orthodontic Mini Implant is a screw-shaped device used as a temporary fixed anchorage point in the jawbone to facilitate orthodontic tooth movement. The device is inserted by a professional orthodontist into the patient's jawbone; it provides a stable mechanical source for attaching orthodontic appliances. After treatment completion, the device is removed. The device is manufactured from Ti-6Al-4V ELI titanium alloy. It is provided sterile for single use. The device functions by providing a stationary point against which orthodontic forces can be applied to move specific teeth while preventing unwanted movement of others. Clinical benefit is derived from improved control over tooth positioning during complex orthodontic treatments.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including insertion/removal torque, rotational fracture torque, and axial pull-out strength tests conducted per ASTM F543-07. Biocompatibility was verified per ISO 10993-1, -5, -10, and -11.
Technological Characteristics
Material: Ti-6Al-4V ELI (ASTM F136). Sensing/Actuation: Mechanical screw fixation. Dimensions: Diameters 1.4mm, 1.6mm, 1.8mm; Lengths 6mm, 8mm, 10mm. Sterilization: Radiation (ISO 11137). Design: Tapered body with various head configurations (Simple, Through Hole, Small, Bracket, Hex, Hex Cross, Hex Open Cross, Stopper types).
Indications for Use
Indicated for patients in need of dental orthodontics or maxillofacial surgery requiring a fixed anchorage point for orthodontic appliances. For professional use only by qualified, trained orthodontists.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 17, 2020
BIO CETEC CO., LTD. % Dave Kim Medical Device Regulatory Affairs MTech Group Inc. 8310 Buffalo Speedway Houston, Texas 77025
Re: K190871
Trade/Device Name: Bio-TackS Orthodontic Mini Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: OAT Dated: December 17, 2019 Received: December 17, 2019
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190871
Device Name Bio-TackS Orthodontic Mini Implant
Indications for Use (Describe)
The Bio-TackS Orthodontic Mini Implant is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-1.05z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K190871 510(k) Summary
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92
Date: January 17, 2020
# 1. 510K Applicant / Submitter:
BIO CETEC CO., LTD. RM 1001-1002 New T Castle, 108, Gasan digital 2-ro, Geumcheon-gu, Seoul, REPUBLIC OF KOREA Tel: +82-2-704-2876 Fax: +82-2-704-2877
# 2. Submission Contact Person
MTech Group Inc. 7707 Fannin St. Ste 200, V111, Houston, TX 77054 Mr. Dave Kim Phone: 713-467-2607 Email: davekim@mtech-inc.net
### 3. Device
- Trade / Device Name : Bio-TackS Orthodontic Mini Implant
- Classification Name : Implant, Endosseous, Orthodontic
- Regulation Number : 21 CFR 872.3640
- Regulation Name : Endosseous dental implant
- Regulatory Class : II
- Product Code : OAT
# 4. Predicate Device
- Trade / Device Name : Orthodontic Screw
- 510(k) Number : K161197
- Regulation Number : 21 CFR 872.3640
- Regulation Name : Endosseous dental implant
- Regulatory Class : II
- Product Code : OAT
# 5. Description:
Fixed on jawbone the Bio-TackS Orthodontic Mini Implant is a screw-shaped product used as a fixed point for orthodontic treatment and it is applied in order to secure sufficient physical and mechanical fixing source for treatment.
#### 6. Indications for Use
The Bio-TackS Orthodontic Mini Implant is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.
{4}------------------------------------------------
# 7. Substantial Equivalence Discussion:
The Bio-TackS Orthodontic mini implant is substantially equivalent to Osstem Orthodontic Screw (K161197). The following comparison table is presented to demonstrate substantial equivalence.
| | Subject Device | Predicate 1 | SE<br>Analysis |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| 510(k) Number | K190871 | K161197 | - |
| Device Name | Bio-TackS Orthodontic Mini Implant | Orthodontic Screw | - |
| Common Name | Orthodontic mini implant | Orthodontic mini implant | - |
| Manufacturer | BIOCETEC CO., LTD. | OSSTEM Implant Co., Ltd. | - |
| Design | Details refer to below Table(*) | - | - |
| Indication for Use | The Bio-TackS Orthodontic mini implant is indicated for<br>use as a fixed anchorage point for attachment of<br>orthodontic appliances to facilitate the orthodontic<br>movement of teeth. It is used temporarily and is removed<br>after orthodontic treatment has been completed. Screws are<br>intended for single use only. | The Orthodontic Screw is indicated for use<br>as a fixed anchorage point for attachment of<br>orthodontic appliances to facilitate the<br>orthodontic movement of teeth. It is used<br>temporarily and is removed after<br>orthodontic treatment has been completed.<br>Screws are intended for single use only. | SE |
| Head Structure | Simple Head, Through Hole, Small Head, Bracket Head,<br>Hex Head, Hex Cross, Hex Open Cross, Simple Head Hex<br>Stopper, Small Head Hex Stopper, Through Hole Stopper | Simple Head, Through Hole, Small Head,<br>Bracket Head | Different |
| Body Diameter | Ø 1.4mm,<br>Ø 1.6mm, Ø 1.8mm | Ø1.2mm, Ø 1.4mm,<br>Ø 1.6mm, Ø 1.8mm | SE |
| Length | 6mm, 8mm, 10mm | 6mm, 8mm, 10mm | SE |
| Material of Fixture | Ti-6Al-4V ELI<br>[ASTM F136] | Ti-6Al-4V ELI<br>[ASTM F136] | SE |
| Surface Treatment | Machined | Machined | SE |
| Sterilization | Radiation | Radiation | SE |
| Shelf-life | 5 years | 8 years | - |
| Target Population | Professional use only - qualified dentists. Strictly reserved<br>to specialized and trained users. | Professional use only - qualified dentists.<br>Strictly reserved to specialized and trained<br>users. | SE |
| Principles of<br>operation | Orthodontic screw is inserted into either jaw to help the<br>orthodontist move the correct teeth and stop the wrong<br>teeth from moving in the wrong direction. | Orthodontic screw is inserted into either jaw<br>to help the orthodontist move the correct<br>teeth and stop the wrong teeth from moving<br>in the wrong direction. | SE |
| Single use | Yes | Yes | SE |
| Sterile Packaging | Yes | Yes | SE |
| Target Population | Patients in need of Dental orthodontics, maxillofacial<br>surgery | Patients in need of Dental orthodontics,<br>maxillofacial surgery | SE |
| Anatomical Site | Teeth | Teeth | SE |
| Location of Use | Use only by professional orthodontists | Use only by professional orthodontists | SE |
| Bio-compatibility | All user directly contacting materials are compliance with<br>ISO10993 requirements. | All user directly contacting materials are<br>compliance with ISO10993 requirements. | SE |
(※) Images(Photos) of each type
{5}------------------------------------------------
| No | Type | External Appearance |
|----|----------------------------------------|---------------------------------------|
| 1 | Orthodontic Screw Simple Head<br>type | Image: Orthodontic Screw Simple Head |
| 2 | Orthodontic Screw Through Hole<br>type | Image: Orthodontic Screw Through Hole |
| 3 | Orthodontic Screw Small head<br>type | Image: Orthodontic Screw Small head |
| 4 | Orthodontic Screw Bracket Head<br>type | Image: Orthodontic Screw Bracket Head |
| 5 | Orthodontic Screw Hex Head<br>type | Image: Orthodontic Screw Hex Head |
| 6 | Orthodontic Screw Hex Cross<br>type | Image: Orthodontic Screw Hex Cross |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a close-up of an orthodontic screw. The screw has a hex open cross type head. The number 7 is also present in the image, likely indicating a product or item number.
Diameter-Length Combinations (in mm) for the Proposed Device:
Ø 1.6mm, Ø 1.8mm
- 1. 1.4x6. 1.4x8
- 2. 1.6x6, 1.6x8, 1.6x10
- 3. 1.8x6. 1.8x8. 1.8x10
Bio-TackS Orthodontic Mini Implant and the predicate device (Osstem Orthodontic Screw (K161197)) have identical indication for use statements and the same intended use.
It shows equivalent specifications with the predicate devices in most of parameters.
| | However, there are differences in the device design characteristics (Head structure). | |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Bio-TackS | Orthodontic Screw (K161197) |
| Body Diameter | Ø 1.4mm.<br>011 1 1 1 0 1 0 1 0 1 1 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 | Ø1.2mm. Ø 1.4mm.<br>011 1 1 1 0 1 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 |
The different head structure for the subject device include Hex Head, Hex Cross, Hex Open Cross, Simple Head Hex Stopper, Small Head Hex Stopper, Through Hole Stopper.
Ø 1.6mm, Ø 1.8mm
# 8. Performance Tests (Non-clinical)
Non-clinical performance tests were performed according to the test standard, ASTM F543 Standard Specification and Test Methods for Metallic Medical Bone Screws.
The following tests for performance of the subject device have been conducted;
# Sterilization and Shelf Life:
Sterilization validation was performed according to ISO 11137-1. ISO 11137-2, and ISO 11137-3 using the VDmax25 method. Shelf life of the packaging was demonstrated by accelerated aging of the product with evaluations for seal peeling (ASTM F88), dye penetration (ASTM F1929), and sterility testing.
# Insertion-Removal torque test
According to the test standard (ASTM F543-07 A1), Insertion torque / removal torque testing is performed to evaluate the mechanical stability of orthodontic mini implants. The largest diameter (1.8mm) and longest (10mm) screws are tested as the worst case; BSH1810S (BioTackS), OSSH1810 (Osstem Mini Implant). With the same diameter and length, both samples demonstrated almost equivalent torque values when inserted and removed at 1440 degrees at 1 rpm. A slight difference in the results is due to the different in drill-bit design and thread design.
# Rotational fracture torque strength test
According to the test standard (ASTM F543-07 A1), Rotational fracture torque testing is performed to determine the material yield strength. The smallest diameter (1.4mm) with a 10mm length; BSH1408S (BioTacks), OSSH1408 (Osstem mini implant) were tested. The tapered implant fracture occurred in the same tread section. Both implants are made from the same raw material (titanium alloy-ELI grade 23), so the material yield strengths of both products are the same.
{7}------------------------------------------------
In addition, all screw-head types, with the same diameter and length, are tested.
# Axial pull-out strength testing
According to the test standard (ASTM F543-07 A1), Pull-out strength testing is performed for the effectiveness of the mechanical design of the orthodontic mini implant and its elimination. The screw samples with the shortest (6mm) and smallest diameter (1.4mm) were tested: BSH1406S (BioTackS). OSSH1406 (Osstem mini implant). The test data demonstrated similar results due to the similar pitch tread thickness of the samples.
| Test | Type | Diameter | Length | Shape |
|------------------------------------|------------------------|----------|--------|-------|
| Insertion-Removal torque testing | BSH1810S<br>(BioTackS) | 1.8 mm | 10 mm | Taper |
| | OSSH1810<br>(Osstem) | 1.8 mm | 10 mm | Taper |
| Rotational fracture torque testing | BSH1408S<br>(BioTackS) | 1.4 mm | 8 mm | Taper |
| | OSSH1408<br>(Osstem) | 1.4 mm | 8 mm | Taper |
| Axial Pull-out strength testing | BSH1406S<br>(BioTackS) | 1.4 mm | 6 mm | Taper |
| | OSSH1406<br>(Osstem) | 1.4 mm | 6 mm | Taper |
Table 1. Description of orthodontic mini implant Used in This Study
Bench Test included rotational fracture torque strength testing for all screw-head types, with the same diameter and length. The test results were similar. Differences of Head structure do not affect effectiveness or safety of the Bio-TackS Orthodontic Mini Implant and Predicate Device.
The results of the bench testing demonstrate similar performance for Bio-TackS Orthodontic Mini Implant and the predicate device.
Biocompatibility testing accordance with ISO 10993-1 has been conducted for Bio-TackS Orthodontic Mini Implant.
- Category : Implant Device for Tissue / Bone
- Contact duration: C-permanent (>30d)
| Human<br>Contact Part | Test Item | Test Report Number | Test Standard | Test Result |
|-----------------------|--------------------------------------------|-----------------------------|---------------|----------------------------------|
| Tissue / Bone | Cytotoxicity | 180600062 PPT-18-0028-01(E) | ISO 10993-5 | Non-cytotoxic |
| | Acute Systemic Toxicity | 180600062 PPT-18-0028-02(E) | ISO 10993-11 | Non-acute systemic toxicity |
| | Intracutaneous<br>(Intradermal) Reactivity | 180600062 PPT-18-0028-03(E) | ISO 10993-10 | None Irritation |
| | Local Lymph Node<br>Assay (LLNA) | 180600062 PPT-18-0028-04(E) | ISO 10993-10 | Do not show any hypersensitivity |
Biocompatibility testing including Cytotoxicity, Sensitization or Intracutaneous Reactivity, Systemic Toxicity was completed according to the following standards:
ISO 10993-1:2009 Biological Evaluation of Medical Devices -Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009 Biological Evaluation of Medical Devices – Part 5 Cytotoxicity
ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin sensitization
ISO 10993-11:2017 Biological Evaluation of Medical Devices Part 11: Tests for systemic toxicity test
{8}------------------------------------------------
The biocompatibility test results demonstrated that there is no new concern in the cytotoxicity, acute systemic toxicity, intracutaneous reactivity, local lymph node assay (LLNA).
Subchronic Toxicity Test, Genotoxicity Test, Implantation were not conducted due to the reasons as follows;
Products with titanium materials tested according to the contents of ASTM F 136, X2. BIOCOMPATIBILITY in Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications of ASTM F136 do not have side effects on safety after a procedure.
X2. BIOCOMPATIBILITY
- X2.1
The alloy composition covered by this specification has been employed successfully in human implant annlications in contact with soft tissue and bone for over a decade. Due to the well-characterized level of biological response exhibited by this alloy, it has been used as a control material in Practice F 981.
#### SUMMARY OF CLINICAL TESTS 9.
Clinical testing was not required to demonstrate the substantial equivalence of the Bio-TackS Orthodontic Mini Implant to its predicate device.
# 10. Conclusions:
Based on the information above, Bio-TackS Orthodontic Mini Implant has same indications for use and intended use. Both the subject device and the predicate / reference devices have shown similar performance results in these bench tests
In conclusion, Bio-TackS Orthodontic Mini Implant is substantially equivalent to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.