K190802 · Civco Medical Solutions · ITX · May 24, 2019 · Radiology
Device Facts
Record ID
K190802
Device Name
Envision ultrasound pad and cover
Applicant
Civco Medical Solutions
Product Code
ITX · Radiology
Decision Date
May 24, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The Envision cover and Envision pad are intended as a microbial barrier between the patient and medical imaging electronics. The Envision cover and Envision pad are used, without the need for ultrasound coupling gel, in adult and pediatric patients, for diagnostic ultrasound imaging, in sterile and nonsterile fields, that currently use an ultrasound coupling gel or fluid alone or in combination with a protective transducer cover, including ultrasound guided venous access, ultrasound imaging over surgical wounds, during transcutaneous biopsy; for intraoperative, endocavity, or transcutaneous imaging procedures; or to enhance acoustic coupling to difficult geometries.
Device Story
Envision Ultrasound cover and pad function as microbial barriers for diagnostic ultrasound procedures. Device eliminates need for traditional ultrasound coupling gel by utilizing a proprietary coating layer activated by hydrating media (e.g., saline). Pad consists of four layers: release liner, adhesive, film carrier, and acoustic coupling coating. Cover is a standard ultrasound transducer sheath with the pad integrated at a cutout. Used in healthcare facilities by clinicians for various imaging procedures (venous access, biopsy, intraoperative). Clinician applies hydrated pad to transducer or patient to facilitate sound wave transmission. Benefits include reduced mess compared to traditional gels and maintenance of sterile field. Device is single-use and sterilized via Ethylene Oxide.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including biocompatibility (ISO 10993), viral penetration (ASTM F1671-13), water leak testing, acoustic performance (impedance, velocity), physical/chemical characterization (density, pH), simulated use, packaging/shelf-life validation (ISO 11607-2, ASTM F1980-16), and sterilization validation (ISO 11135).
Technological Characteristics
Pad materials: polycarbonate, silicone elastomer, polyether polyurethane, and proprietary coating. Cover material: polyurethane. Sensing principle: acoustic coupling via hydrated coating. Connectivity: standalone. Sterilization: Ethylene Oxide (ISO 11135). Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic ultrasound imaging in sterile or non-sterile fields, including ultrasound-guided venous access, imaging over surgical wounds, transcutaneous biopsy, intraoperative, endocavity, or transcutaneous procedures, where a microbial barrier and acoustic coupling are required.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
EcoVue® Sterile and Non-Sterile Ultrasound Gels (K181363)
Submission Summary (Full Text)
{0}------------------------------------------------
May 24, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
CIVCO Medical Solutions % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114
Re: K190802
Trade/Device Name: Envision™ Ultrasound Cover and Pad Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: Class II Product Code: ITX Dated: May 21, 2019 Received: May 22, 2019
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number K190802
#### Device Name
Envision™ Ultrasound cover and pad
#### Indications for Use (Describe)
The Envision cover and Envision pad are intended as a microbial barrier between the patient and medical imaging electronics.
The Envision cover and Envision pad are used, without the need for ultrasound coupling gel, in adult and pediatric patients, for diagnostic ultrasound imaging, in sterile fields, that currently use an ultrasound coupling gel or fluid alone or in combination with a protective transducer cover, including ultrasound guided venous access, ultrasound imaging over surgical wounds, during transcutaneous biopsy; for intraoperative, endocavity, or transutaneous imaging procedures; or to enhance acoustic coupling to difficult geometries.
| | <span> <span style="font-size: 10pt;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> | | <span> <span style="font-size: 10pt;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
|--|--------------------------------------------------------------------------------------------------------------|--|--------------------------------------------------------------------------------------------------|
|--|--------------------------------------------------------------------------------------------------------------|--|--------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
{3}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k) is being prepared in accordance with the requirements of 21 CFR 807.92.
### The assigned 510(k) numberis: K190802.
- 1. Submitter's Identifications:
| Establishment: | CIVCO Medical Instruments Co., Inc. |
|--------------------------------------------|--------------------------------------------|
| Address: | 102 First Street South<br>Kalona, IA 52247 |
| Registration Number: | 1937223 |
| Operations Manufacturer<br>Owner/Operator: | CIVCO Medical Instruments Co., Inc. |
| Owner/Operator Number: | 1937223 |
| Contact Person: | Jim Leong |
| Phone: | 319-656-4447 |
| e-mail: | James.Leong@civco.com |
### 2. Date 510(k) Summary Prepared: May 15, 2019
# 3. Name of the Subject Device and Classification Information:
| Trade/Device Name | Envision™ Ultrasound cover and pack |
|---------------------|-------------------------------------|
| Regulation Number | 21 CFR 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | Class II |
| Product Code | ITX |
# 4. Information for the Predicate Device:
| Tradename/Device Name | General Purpose Transducer Cover |
|-----------------------|-------------------------------------|
| Manufacturer | CIVCO Medical Instruments Co., Inc. |
| 510(k) Number | K970513 |
| Regulation Number | 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | II |
| Product Code | ITX |
{4}------------------------------------------------
#### 5. Information for Reference Device:
| Tradename/Device Name | EcoVue® Sterile and Non-Sterile Ultrasound Gels |
|-----------------------|-------------------------------------------------|
| Manufacturer | HR Pharmaceuticals, Inc. |
| 510(k) Number | K181363 |
| Regulation Number | 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | II |
| Product Code | MUI |
### 6. Device Description:
The Envision Ultrasound pad and Envision Ultrasound cover (Envision) are microbial barriers for use during ultrasound procedures. The Envision device contains a coating that facilitates the acoustic coupling between the patient and ultrasound equipment without the need for traditionally ultrasound gel. This enables the clean transmission of sound waves between the transducer and patient without the mess of gels.
The Envision Ultrasound pad is comprised of four layers. The first layer is a crack and peel release liner that is removed by the customer to expose the adhesive layer (second layer) that sticks to the patient or the transducer within the cover. The third layer is a film carrier layer to hold the adhesive and coating layers together. The final layer is the coating layer, which is the acoustic couplant layer and is activated by using hydrating media, such as saline.
The Envision pad is sold as both a standalone pad and also sold as an integrated ultrasound cover (cover). The cover is a traditional CIVCO ultrasound cover, where one end of the cover is cut out, and the Envision pad, cut down to size, is sealed to the cover at the location of the cut out.
- 7. Intended Use / Indications for Use:
The Envision cover and Envision pad are intended as a microbial barrier between the patient and medical imaging electronics.
The Envision cover and Envision pad are used, without the need for ultrasound coupling gel, in adult and pediatric patients, for diagnostic ultrasound imaging, in sterile and nonsterile fields, that currently use an ultrasound coupling gel or fluid alone or in combination with a protective transducer cover, including ultrasound guided venous access, ultrasound imaging over surgical wounds, during transcutaneous biopsy; for intraoperative, endocavity, or transcutaneous imaging procedures; or to enhance acoustic coupling to difficult geometries.
{5}------------------------------------------------
# 8. Comparison to Legally Marketed Device
| Item | | Subject Device<br>CIVCO Envision Cover and Pad<br>K190802 | Predicate Device<br>CIVCO General Purpose<br>Transducer Cover<br>K970513 |
|-------------------------|--------------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Material | | Pad: Polycarbonate, silicone elastomer,<br>polyether polyurethane, coating<br>formulation | Pad: N/A |
| | | Cover: Polyurethane | Cover: Same |
| Material Integrity | | All acceptance criteria met related to<br>the strength and elasticity of the cover<br>sheath component. | Equivalent testing and results |
| Microbial Barrier | | Meets requirements of ASTM F1671-<br>13 for prevention of blood-borne<br>pathogens | Equivalent testing |
| Acoustic<br>Performance | Acoustic<br>Impedance | Primary coating: $1.69 ± 0.18 x 10^5$<br>g/(cm² sec)<br>Secondary coating: $1.68 ± 0.29 x 10^5$<br>g/(cm² sec) | $1.68 ± 0.24 x 10^5$ g/(cm² sec) |
| | Acoustic/<br>Sound<br>Velocity | Primary Coating: 1594 ± 170 m/s<br>Secondary Coating: 1631 ± 282 m/s | Reference Device Data:<br>1398-1750 m/s |
| | Density of<br>Coating | Primary: 1.06 g/mL<br>Secondary: 1.03 g/mL | Reference Device Data:<br>0.85-1.15g/mL |
| | pH of<br>coating | 7.11 (both primary and secondary<br>coatings) | Reference Device Data:<br>5.5-7.8 |
| Acoustic Coupling | | Hydrated coating used to facilitate<br>coupling | Uses gel to facilitate coupling |
| Sterilization | | Ethylene Oxide | Ethylene Oxide |
| Shelf-life | | 1 year | 3 years |
| Accessory | | Saline wipe to facilitate coupling | Ultrasound gel packet |
| Environment of Use | | Healthcare facility/hospital | Healthcare facility/hospital |
{6}------------------------------------------------
# 9. Comparison of Indications to the Legally Marketed Device:
The proposed Envision devices have the same intended use, of being a microbial barrier, as the legally marketed device. Any differences in the indications for use statement would not affect the safety or effectiveness of the device because the safety and efficacy related to the indicated procedures is dictated by the use of the underlying ultrasound equipment, and the ultrasound pad and cover is secondary to the ultrasound use. Any questions related to safety and effectiveness of the Envision pad and cover have been addressed using the same testing performed by the legally marketed device, including biocompatibility, viral penetration, and material strength and elasticity studies.
# 10. Summary of Non-Clinical Tests Performed:
- Biocompatibility:
The Envision devices met ISO 10993-1 biocompatibility requirements for limited contact duration for surface devices of breached or compromised surface and external communicating indirect blood path/tissue contact:
- -Cytotoxicity - ISO 10993-5
- -Sensitization - ISO 10993-10
- Irritation ISO 10993-10 -
- Acute Systemic Toxicity ISO 10993-11 -
- -Hemolysis - ASTM F 756
- -Subcutaneous Implantation - ISO 10993-6
- -Material Mediated Pyrogenicity - ISO 10993-11
- Viral Penetration:
The Envision pad's ability to withstand viral penetration was performed to evaluate the barrier performance of the Envision pad material in accordance with ASTM F1671-13.
- Water leak testing: ●
Water leak testing was performed to demonstrate material strength and elasticity of the Envision device.
- Acoustic Impedance: ●
The measurement of acoustic impedance was performed to compare the acoustic characteristics of the Envision pad to the predicate cover with ultrasound coupling gel.
- Acoustic/Sound Velocity: ●
The measurement of the acoustic velocity was calculated from the acoustic impedance and density to compare the acoustic velocity of the Envision pad to the predicate cover with ultrasound coupling gel.
{7}------------------------------------------------
- Physical and Chemical Characteristics:
The density and pH of the Envision pad coating was measured for comparison to the ultrasound coupling gel, which would be used with the predicate device.
- Simulated Use Testing
Simulated use testing was performed to ensure the design of the Envision conforms to the user needs and intended use.
- Packaging and Shelf-life:
The Envision packaging was validated per ISO 11607-2 and ISTA 2A to ensure the packaging maintains its integrity over the course of the device's shelf-life. Accelerated aging in support of a one-year shelf life was completed in accordance with ASTM F1980-16.
- Rehydration Testing
The Envision devices were subjected to multiple rounds of rehydration and tested to ensure the devices would withstand multiple rounds of rehydration and perform as intended.
- Ethylene Oxide Sterilization Validation and Residuals Testing
The Envision devices are validated per ISO 11135 to ensure that the devices meet the required sterility assurance level, so the devices remain adequately free from viable microorganisms following sterilization.
The Envision devices are 2x Ethylene Oxide sterilized and tested to ensure that residuals from the Ethylene Oxide cycle do not exceed the limits per ISO 10993-7.
#### 11. Clinical Test Performed:
Clinical tests were not required to demonstrate substantial equivalence.
# 12. Conclusions
The Envision device has the same intended use and its technological characteristics do not raise any different questions of safety or effectiveness, as compared to the legally marketed device. Therefore, the Envision pad and cover are substantially equivalent to the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.