The Eye Prime is intended for ophthalmic imaging when the following conditions are present or suspected: •Cataracts •Retinal Detachment (separation of the retina from the middle coat of the eyeball) •Orbital Lesions •Tumors •Foreign Bodies •Inflammation • Vascular Irregularities
Device Story
Eye Prime is a diagnostic ophthalmic ultrasound system used in clinics and hospitals by ophthalmologists, sonographers, or trained echographers. It utilizes a console with interchangeable probes (12/18 MHz posterior B-scan; 35/50 MHz anterior UBM) and a laptop for image processing, user interface, and data management. The device emits acoustic pulses via a transducer; echoes from eye structures are converted into voltages, amplified, and processed to generate 2D cross-sectional B-scan images. The system supports DICOM connectivity for EMR integration. By visualizing internal ocular structures, the device assists clinicians in diagnosing conditions like retinal detachment, tumors, and foreign bodies, facilitating clinical decision-making and patient management.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including acoustic output measurement, biocompatibility (ISO 10993), and electrical/mechanical safety (IEC 60601 series). Measurement accuracy was verified as equivalent to the predicate device.
Technological Characteristics
B-mode ultrasound system; mechanical sector (1 element) and annular array (6 elements) transducers; 12-50 MHz frequency range. Laptop-based processing; Windows 10 OS; 512GB storage. Connectivity: DICOM. Power: 100-240VAC. Non-sterile. Compliance: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, IEC 62366, ISO 10993-1, NEMA UD2-2.
Indications for Use
Indicated for adult male and female patients requiring ophthalmic imaging for cataracts, retinal detachment, orbital lesions, tumors, foreign bodies, inflammation, or vascular irregularities.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Eye Cubed Version 4 Diagnostic Ultrasound (K103608)
Submission Summary (Full Text)
{0}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K190786
Device Name Eye Prime
Indications for Use (Describe)
The Eye Prime is intended for ophthalmic imaging when the following conditions are present or suspected:
·Cataracts
- ·Retinal Detachment (separation of the retina from the middle coat of the eyeball)
·Orbital Lesions
·Tumors
·Foreign Bodies
•Inflammation
· Vascular Irregularities
Type of Use (Select one or both, as applicable)| X | Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{1}------------------------------------------------
| DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM | |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Eye Prime System<br>Probes : Posterior B-Scan Probe (12 MHz), Posterior B-Scan Probe (18 MHz), Anterior<br>B-Scan (UBM) Probe ( 35 or 50 MHz) |
| Intended Use | The Eye Prime is intended for ophthalmic imaging when the following conditions are<br>present or suspected:<br>•Cataracts<br>•Retinal Detachment (separation of the retina from the middle coat of the eyeball)<br>•Orbital Lesions<br>•Tumors<br>•Foreign Bodies<br>•Inflammation<br>•Vascular Irregularities |
# Intended Use: Diagnostic ultrasound imaging of the human body as follows:
| Intended Use: Diagnostic ultrasound imaging of the human body as follows: | | | | | | | | |
|---------------------------------------------------------------------------|---|---|---|------------|------------|---------------|---------------|----------------|
| Mode of Operation | | | | | | | | |
| Clinical Application | A | B | M | PW Doppler | CW Doppler | Color Doppler | Power Doppler | Combined Modes |
| Ophthalmic | | N | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | | | | | | | | |
| Intraoperative | | | | | | | | |
| Intraoperative Neurological | | | | | | | | |
| Pediatric | | | | | | | | |
| Small Organ | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Cardiac | | | | | | | | |
| Transesophageal | | | | | | | | |
| Transrectal | | | | | | | | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Transcranial | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Laparoscopic | | | | | | | | |
| MSK Conventional | | | | | | | | |
| MSK Superficial | | | | | | | | |
| Vascular Access | | | | | | | | |
| Nerve Block | | | | | | | | |
| Other | | | | | | | | |
{2}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
| Probe | Ophthalmic Anterior Segment B-scan (UBM) probe |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Imaging anterior segment of the eye. The probe may operate at a nominal frequency of 50 MHz or 35 MHz, depending on the transducer selected. |
| | Mode of Operation | | | | | | | |
|-----------------------------|-------------------|---|---|---------------|---------------|------------------|------------------|-------------------|
| Clinical Application | A | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes |
| Ophthalmic | | N | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | | | | | | | | |
| Intraoperative | | | | | | | | |
| Intraoperative Neurological | | | | | | | | |
| Pediatric | | | | | | | | |
| Small Organ | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Cardiac | | | | | | | | |
| Transesophageal | | | | | | | | |
| Transrectal | | | | | | | | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Transcranial | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Laparoscopic | | | | | | | | |
| MSK Conventional | | | | | | | | |
| MSK Superficial | | | | | | | | |
| Vascular Access | | | | | | | | |
| Nerve Block | | | | | | | | |
| Other | | | | | | | | |
### Intended Use: Diagnostic ultrasound imaging of the human body as follows:
{3}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
| Probe | Ophthalmic 12 MHz Posterior Segment B-scan<br>transducer |
|--------------|----------------------------------------------------------|
| Intended Use | Imaging posterior segment of the eye |
| | Mode of Operation | | | | | | | |
|-----------------------------|-------------------|---|---|---------------|---------------|------------------|------------------|-------------------|
| Clinical Application | A | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes |
| Ophthalmic | | N | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | | | | | | | | |
| Intraoperative | | | | | | | | |
| Intraoperative Neurological | | | | | | | | |
| Pediatric | | | | | | | | |
| Small Organ | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Cardiac | | | | | | | | |
| Transesophageal | | | | | | | | |
| Transrectal | | | | | | | | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Transcranial | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Laparoscopic | | | | | | | | |
| MSK Conventional | | | | | | | | |
| MSK Superficial | | | | | | | | |
| Vascular Access | | | | | | | | |
| Nerve Block | | | | | | | | |
| Other | | | | | | | | |
### Intended Use: Diagnostic ultrasound imaging of the human body as follows:
{4}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
| Probe | Ophthalmic 18 MHz Posterior Segment B-scan |
|--------------|--------------------------------------------|
| | transducer |
| Intended Use | Imaging posterior segment of the eye |
| Mode of Operation | | | | | | | | |
|-----------------------------|---|---|---|---------------|---------------|------------------|------------------|-------------------|
| Clinical Application | A | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Power<br>Doppler | Combined<br>Modes |
| Ophthalmic | | N | | | | | | |
| Fetal | | | | | | | | |
| Abdominal | | | | | | | | |
| Intraoperative | | | | | | | | |
| Intraoperative Neurological | | | | | | | | |
| Pediatric | | | | | | | | |
| Small Organ | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Cardiac | | | | | | | | |
| Transesophageal | | | | | | | | |
| Transrectal | | | | | | | | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Transcranial | | | | | | | | |
| Peripheral Vascular | | | | | | | | |
| Laparoscopic | | | | | | | | |
| MSK Conventional | | | | | | | | |
| MSK Superficial | | | | | | | | |
| Vascular Access | | | | | | | | |
| Nerve Block | | | | | | | | |
| Other | | | | | | | | |
## Intended Use: Diagnostic ultrasound imaging of the human body as follows:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Ellex Medical Pty Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW BUFFALO MN 55313
April 11, 2019
Re: K190786
Trade/Device Name: Eye Prime Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX Dated: March 25, 2019 Received: March 27, 2019
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{6}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Summary 21 CFR 807.92
K190786
#### I. Submitter
Ellex Medical Pty Ltd.
3-4 Second Avenue, Mawson Lakes,
South Australia 5095, AUSTRALIA
Contact Person: Andrew Pfeiffer
Telephone: 61 8 70748104
Email: apfeiffer@ellex.com
Date of preparation: 20/02/2019
# II. Proposed Device
Trade Name: Eye Prime
Manufacturer: Ellex Medical Pty Ltd
Common or Usual Name: Diagnostic ophthalmic ultrasound system
Classification:
| Name | 21 CFR Number | Product Code |
|------------------------------------------|---------------|--------------|
| Ultrasonic Pulsed Echo Imaging<br>System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
## III. Legally marketed Device /Predicate Device
Device name: Eye Cubed Version 4 Diagnostic Ultrasound
Premarket Notification: K103608
Manufacturer: Ellex Innovative Imaging
{8}------------------------------------------------
# IV. Device Technological characteristics
The Eye Prime is a new, diagnostic ophthalmic ultrasound Track 1 device. Its function is to perform B-mode ultrasound scans. The system consists of 1) A console which supports 12 and 18MHz Posterior B scan probes and Anterior B- Scan. Ultrasound Bio-microscopy (UBM) probes with either 35 or 50MHz transducers. 2) A laptop which performs the image processing requirements of the system; provides the user interface for operating the device and viewing the resultant images. The laptop also provides data storage, export, archiving and networking for the device and includes the connection for footswitch.
## V. Principle of Operation
In ophthalmic ultrasound imaging, a transducer emits a short acoustic pulse in response to a voltage transient generated by a pulser. Echoes are generated as the pulse propagates through eye tissue and encounters ophthalmic structures differing in acoustic impedance from their surroundings. When these echoes are received by the transducer they are converted back into voltages and amplified. The range of each echo is proportional to the time delay between pulse emission and echo return. Each pulse/echo provides information along a line of sight. By mechanically sweeping the transducer across an arc, information from a series of lines-of-sight is obtained. Converting the echo amplitude into pixel intensity allows a 2D cross-sectional B-Scan image to be displayed. This is the basis for B-Scan imaging.
## VI. Indications for Use
Eye Prime is used for ophthalmic imaging when the following conditions are present or suspected:
- . cataracts
- retinal detachments (a separation of the retina from the middle coat of the eyeball) .
- . orbital lesions
- . tumours
- foreign bodies
- . inflammation
- vascular irregularities. .
{9}------------------------------------------------
| Model | Eye Prime<br>(proposed device) | Eye Cubed Version 4<br>(predicate device) |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) number | N/A | K(103608) |
| Clinical application | Diagnostic ophthalmic ultrasound<br>for use in Clinics/Hospitals by<br>ophthalmologists, sonographers or<br>trained echographers. | Diagnostic ophthalmic ultrasound<br>for use in Clinics/Hospitals by<br>ophthalmologists, sonographers or<br>trained echographers. |
| Sterility | Non Sterile | Non Sterile |
| Patient population | Adult male and female | Adult male and female |
| Intended Use | Ophthalmic imaging when the<br>following conditions are present or<br>suspected:<br>•cataracts<br>•retinal detachments (a separation<br>of the retina from the middle coat<br>of the eyeball)<br>•orbital lesions<br>•tumours<br>•foreign bodies<br>•inflammation<br>•vascular irregularities. | Ophthalmic imaging when the<br>following conditions are present or<br>suspected:<br>•cataracts<br>•retinal detachments (a separation<br>of the retina from the middle coat<br>of the eyeball)<br>•orbital lesions<br>•tumours<br>•foreign bodies<br>•inflammation<br>•vascular irregularities. |
| Modes of Operation | B-Mode (Posterior and Anterior) | A-Mode (Biometry and Diagnostic)<br>B-Mode (Posterior and Anterior) |
| Probe /Transducer type | Mechanical Sector (1 element)<br>Annular Array (6 elements) | Mechanical Sector (1 element) |
| | N/A<br>N/A | N/A<br>Static Biometry A-Scan<br>Static Diagnostic A-Scan |
| Transducer Frequency Range<br>(MHz) | 12-50 | 8-40 |
| Image Acquisition Rate | 10 FPS | 12.5 - 50 FPS |
| Operating System | Windows 10 | Windows XP |
| Internal Storage | 512GB | 160 GB |
| Image processing unit and<br>User Interface | Laptop | A desktop computer with an<br>external monitor is used to<br>process image data and as the user<br>interface |
| Electronic Medical Record<br>(EMR) connectivity | DICOM(Digital Imaging and<br>Communications in Medicine) | DICOM(Digital Imaging and<br>Communications in Medicine) |
| Model | Eye Prime (proposed device) | Eye Cubed V4 |
| Dimensions (inches) | 3.4 x 6.9x8.9 | 6.5x15.3x16.9 |
| Power requirements | 100-240VAC, 50/60HZ, 50VA<br>(console only) | 100-240VAC, 50/60HZ,220VA |
| Acoustic output power<br>MI < 0.23<br>Ispta.3 < 17 mW/cm | Track 1 | Track 1 |
| Compliance standards | * US Federal Performance *<br>Standards 21 CFR 892.1560 and<br>21CFR 892.1570 for Class II<br>Ultrasonic products<br>-AAMI / ANSI ES60601-<br>1:2005/(R)2012 and A1:2012,<br>-IEC 60601-1-2:2014<br>-IEC 60601 1-6:2010 + AMD1: 2013<br>-IEC 60601-2-37:2007 +<br>AMD1:2015<br>-IEC 62366:2007+AMD1:2014<br>-ISO 10993-1<br>-NEMA UD2-2:2004 | * US Federal Performance *<br>Standards 21 CFR 892.1560 and<br>21CFR 892.1570 for Class II<br>Ultrasonic products<br>-IEC 60601-1<br>-IEC 60601-1-2<br>-IEC 60601-2-37<br>-ISO 10993-1<br>-NEMA UD2-2-2004 |
# VII.Comparison of Intended use and Technological Characteristics with Predicate Device
{10}------------------------------------------------
### Summary of Non-Clinical Tests VIII.
The device has been evaluated for acoustic output, biocompatibility, as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform to the applicable medical device safety standards as below:
### Safety and Product Specific testing: ●
- AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, Medical Electrical Equipment Part . 1: General Requirements for Basic Safety and Essential Performance
- IEC 60601-1-2:2014, Medical Electrical Equipment Part 1-2: General Requirements for . Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
{11}------------------------------------------------
- IEC 60601 1-6:2010 + AMD1: 2013, Medical Electrical Equipment Part 1-6: General 트 Requirements for Safety - Collateral Standard: Usability
- . IEC 60601-2-37:2007 + AMD1:2015, Medical Electrical Equipment – Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
- IEC 62366:2007+AMD1:2014, Medical devices Application of usability engineering to I medical devices
- ISO 10993-5:2009, Biological Evaluation of Medical Devices Part 5: Tests For In Vitro . Cytotoxicity
- ISO 10993-10:2010, Biological evaluation of Medical Devices- -Part 10: Tests for irritation . and skin sensitization
- NEMA UD2-2:2004, Acoustic Output Measurement Standard for Diagnostic Ultrasound . Equipment
### Biocompatibility Testing .
The Eye Prime has two patient-contacting elements: the Posterior B-Scan Probes, the single use ClearScan transducer cover / Scleral Shells for the Anterior UBM probe transducers. In all cases, the patient-contacting materials were found to be biocompatible.
#### Software Verification and Validation Testing .
Verification and Validation activities were designed and performed to demonstrate that the Eye Prime met pre-determined performance specifications.
#### . Measurement accuracy
Eye Prime System has been verified and has been found to be equivalent in measurement accuracy to Eyecubed Version 4.
{12}------------------------------------------------
#### Risk Analysis Method Used .
Ellex Medical Pty Ltd applied ISO-14971 to the design and development of the Eye Prime system. The conclusion from the risk analysis was the device was safe for its intended use and does not pose any unacceptable risks.
# IX. Summary of Clinical Tests
Since the Eye Prime device uses the same technology and principles as the predicate device, clinical data is not required.
# X. Substantial Equivalence Discussion
The Indications for Use for Eye Prime are exactly the same as predicate Eye Cubed Version 4 Diagnostic Ultrasound.
Technological characteristics of the Eye Prime are substantially equivalent with respect to the basic design and function of the predicate device. Both the subject and predicate device are based on the same ultrasound imaging and measurement principles. The major technological aspects they have in common are the following:
- · Ophthalmic B-mode scanning for imaging and measurement.
- · Ultrasound frequencies ranging from 8MHz to 50MHz.
- Windows PC or laptop for operator controls and display of measurements and images. Database for storage of measurements and images.
- The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The systems have acoustic power levels which are below the applicable FDA limits.
- · The systems have been designed in compliance with approved electrical and physical safety standards.
{13}------------------------------------------------
There are differences in the probes in the proposed device and there is neither the Biometry A-Scan nor the Diagnostic A-scan available for Eye Prime. However the modifications do not raise concerns for safety and effectiveness of the device based on the device safety and performance testing.
# XI. Conclusion
Based on the indications for use, technological characteristics, and safety and performance testing, the subject Eye Prime met the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, principles of operation and indications for use to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.