K190782 · Ldi Corporation · KGO · Jun 27, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190782
Device Name
FlexGard Cut-Resistant Glove Liners
Applicant
Ldi Corporation
Product Code
KGO · General, Plastic Surgery
Decision Date
Jun 27, 2019
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The FlexGard® Cut-Resistant Glove Liners are single use glove liners intended to help with protection against cuts when used with an inner and outer surgical glove. The Glove Liner must be worn between two surgical gloves.
Device Story
FlexGard Cut-Resistant Glove Liners are single-use, ambidextrous, knitted high-density polyethylene liners. Worn between two surgical gloves, they provide supplemental cut resistance during surgical procedures. The device is terminally sterilized via ethylene oxide and packaged in a Tyvek pouch. It is intended for use by healthcare professionals in clinical settings to reduce the risk of cuts or slashes to the hands during surgery. The device does not provide puncture resistance or full cut-proof protection.
Clinical Evidence
No clinical data. Bench testing only. Performance demonstrated via ANSI/ISEA 105-2011 Level 2 cut resistance testing (ASTM F1790-05), sterility validation (ANSI/AAMI/ISO 11135-1:2007), sterile barrier integrity (ASTM F88/F88M-15, ASTM F2096-11), and biocompatibility (ISO 10993-5, ISO 10993-10).
Indicated for use as a single-use glove liner to provide cut protection when worn between two surgical gloves.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 27, 2019
LDI Corporation % Gary Syring Principal Consultant Quality & Regulatory Associates, LLC 800 Levanger Lane Stoughton, Wisconsin 53589
Re: K190782
Trade/Device Name: FlexGard Cut-Resistant Glove Liners Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I Product Code: KGO Dated: March 22, 2019 Received: March 27, 2019
Dear Gary Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190782
Device Name FlexGard® Cut-Resistant Glove Liners
#### Indications for Use (Describe)
The FlexGard® Cut-Resistant Glove Liners are single use glove liners intended to help with protection against cuts when used with an inner and outer surgical glove. The glove liner must be worn between two surgical gloves.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for LDI Environmental and Safety Solutions. The letters "LDI" are in a large, bold, blue font. Below the letters is the text "ENVIRONMENTAL AND SAFETY SOLUTIONS" in a smaller font. To the right of the letters is a green recycling symbol with a red cross in the center.
#### 510(k) Summary - K190782
This summary is provided to support the 510(k) pre-market notification for the FlexGard® Cut-Resistant FlexGard Liners, as accessories to surgical gloves.
| Company Name: | LDI Corporation<br>3560 Lafayette Road<br>Building 2, Suite C<br>Portsmouth, NH 03801 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Lou LaMarca, CEO<br>LDI Corporation<br>Phone: 603-436-0077 |
| Date Summary Prepared: | June 20, 2019 |
| Trade Name: | FlexGard® Cut-Resistant Glove Liners |
| Common Name: | Surgical Glove Accessory, Glove Liners |
| Classification Name: | Surgeon's Glove (Accessory: Cut Resistant Glove Liners)<br>21 CFR 878.4460 Surgical Glove (Accessory), Class I<br>Product Code: KGO<br>Class I |
| Predicate Devices: | K160974<br>Gammex® Cut Resistant Glove Liner<br>Ansell Healthcare Products LLC |
#### Product Description
The FlexGard Cut Resistant Glove Liners provide an optional method of providing cut resistance to surgical gloves. The FlexGard Cut Resistant Glove Liners are applied between layers of surgical gloves. The FlexGard Cut Resistant Glove Liners are placed over a surgical glove and subsequently covered by a surgical glove.
The FlexGard Cut Resistant Glove Liners are ambidextrous, there is no left or right handed Glove Liner variations.
The FlexGard Cut Resistant Glove Liners are provided to the user terminally sterilized by a validated ethylene oxide process. The FlexGard Cut Resistant Glove Liners are labeled for single use.
{4}------------------------------------------------
The FlexGard Cut Resistant Glove Liners provide protection against cuts and slashes. The FlexGard Cut Resistant Glove Liners are not puncture resistant or cut proof.
# Intended Use of the Device
The FlexGard Cut-Resistant Glove Liners are single use glove liners intended to help with protection against cuts when used with an inner and outer surgical glove. The Glove Liner must be worn between two surgical gloves.
| Technological Comparison Table | | | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Feature | FlexGard® Cut-Resistant Glove<br>Liners<br>K190782 | Gammex® Cut Resistant Glove Liner<br>(Primary Predicate K160974) | Discussion of<br>Feature<br>Comparison |
| Product Code, | KGO | KGO | Identical |
| Classification | 21 CFR 878.4460 Surgical Glove<br>(Accessory), Class I | 21 CFR 878.4460 Surgical Glove<br>(Accessory), Class I | |
| Indications for<br>Use | The FlexGard® Cut-Resistant Glove<br>Liners are single use glove liners<br>intended to help with protection against<br>cuts when used with an inner and outer<br>surgical glove. The glove liner must be<br>worn between two surgical gloves. | The Gammex(R) Cut Resistant Glove<br>Liner is a single use glove liner intended<br>to provide ANSI/ISEA 105 Cut Level<br>Protection 2 against cuts when used<br>with an inner and outer surgical glove.<br>The glove liner should be worn between<br>two surgical gloves. | Similar |
| Design | Knitted glove, ambidextrous | Knitted glove, ambidextrous | Similar |
| Cuff size feature | Color coded cuff band to reflect size | Color coded cuff band to reflect size | Similar |
| Material | High density polyethylene | High strength polyethylene | Similar |
| Cut Protection | ANSI / ISEA 105, Level 2 cut resistance<br>tested to<br>ASTM F1790-05. | ANSI Cut Level 2 tested to<br>ASTM F1790-97 | Same |
| Single use,<br>disposable | Yes | Yes | Same |
| Provided sterile | Yes | Yes | Same |
| Sterilization<br>method | Ethylene Oxide | Gamma Irradiation | Different |
| Sterility<br>Assurance Level<br>(SAL) | 10-6 | 10-6 | Same |
| Sterile barrier<br>package | Tyvek Pouch | Double Tyvek Pouch | Different |
| Shelf life | 2 years | Unknown | Different |
| Biocompatibility<br>evaluation | Cytotoxicity, Irritation, Sensitization | Cytotoxicity, Irritation, Sensitization | Same |
| Sizes | Six (6):<br>Small (S) | Five (5):<br>Extra Small (XS) | Similar |
| Technological Comparison Table | | | |
| Feature | FlexGard® Cut-Resistant Glove<br>Liners<br>K190782 | Gammex® Cut Resistant Glove Liner<br>(Primary Predicate K160974) | Discussion of<br>Feature<br>Comparison |
| | Medium (M) | Small (S) | |
| | Large (L) | Medium (M) | |
| | Extra-Large (XL) | Large (L) | |
| | Two-Extra-Large (2XL) | Extra Large (XL) | |
| | Three-Extra-Large (3XL) | | |
## Summary of Technological Characteristics
{5}------------------------------------------------
## Summary of Nonclinical Performance Test
Non-clinical testing was performed to demonstrate the subject device met the acceptance criteria of the standards used to evaluate the functionality of the device.
| Test | Test Method Summary | Acceptance<br>Criteria | Result |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------|
| Cut Protection to<br>ANSI/ISEA<br>Level 2 cut<br>resistance. | Cut protection of the FlexGard Cut Resistant Glove<br>Liners has been demonstrated to ANSI/ISEA 105-<br>2011, Level 2 by application of the ASTM F1790-05<br>test method after ethylene oxide sterilization,<br>simulated distribution and two-year real time aging. | Meet<br>requirements for<br>ANSI/ISEA 105-<br>2011, Level 2 cut<br>resistance. | Pass |
| Sterility | FlexGard Cut Resistant Glove Liners are terminally<br>ethylene oxide sterilized devices, by a validated EtO<br>sterilization cycle, per ANSI/AAMI/ISO 11135-1:<br>2007. | Sterility<br>assurance level<br>(SAL) of 10-6. | Pass |
| | The FlexGard Cut-Resistant Glove Liner sterile<br>barrier package is a Tyvek pouch. The methods<br>applied to evaluate the Tyvek pouch integrity post<br>terminal sterilization by simulated distribution in<br>compliance with ASTM 4169:2016, Tyvek pouch<br>seal strength evaluated per ASTM F88/F88M-15,<br>and Tyvek pouch integrity evaluated by ASTM<br>F2096-11, Standard Test Method for Detecting Gross<br>Leaks in Porous Medical Packaging by Internal<br>Pressurization (Bubble Leak). | The sterile<br>barrier Tyvek<br>pouch must have<br>a seal strength<br>$\geq$ 1 lbf and not<br>leak. | Pass |
| Biocompatibility | The following test methods were applied to final<br>finished form FlexGard Cut Resistant Glove Liner,<br>with passing results:<br>• Cytotoxicity by ANSI/AAMI/ISO 10993-5:2009<br>(R)2014 | Cytotoxicity<br>passes for a<br>Grade of $\leq$ 2. | Pass<br>Grade = 1 |
| | • Sensitization by ANSI/AAMI/ISO 10993-10:2010<br>(R)2014 | Sensitization<br>passes for a<br>Grade = 0. | Pass<br>Grade = 0 |
| | • Irritation by ANSI/AAMI/ISO 10993-10:2010<br>(R)2014 | Irritation passes<br>for a Primary<br>Irritation Index<br>$\leq$ 0.4 | Pass<br>Primary Irritation<br>Index = 0 |
{6}------------------------------------------------
| Test | Test Method Summary | Acceptance<br>Criteria | Result |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Shelf life two (2)<br>years (real time<br>aging) | Shelf life testing of FlexGard Cut Resistant Glove<br>Liners cut resistance to ANSI/ISEA 105-2011, Level<br>2 by application of the ASTM F1790-05 test method<br>after two years of real time aging.<br>Tyvek sterile barrier pouch integrity after two years<br>of real time aging evaluated for seal strength per<br>ASTM F88/F88M-15, and Tyvek pouch integrity<br>evaluated by ASTM F2096-11, Standard Test<br>Method for Detecting Gross Leaks in Porous<br>Medical Packaging by Internal Pressurization<br>(Bubble Leak). | Meet<br>requirements for<br>ANSI/ISEA 105-<br>2011, Level 2 cut<br>resistance.<br>The sterile<br>barrier Tyvek<br>pouch must have<br>a seal strength<br>$\u2265$ 1 lbf and not<br>leak. | Pass |
Clinical data is not needed for this submission.
## Conclusion
The conclusions drawn from nonclinical tests demonstrate that the FlexGard Cut Resistant Glove Liners are as safe, as effective and performs well as or better than the legally marketed device Gammex® Cut Resistant Glove Liner, cleared to market by K160974.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.