RIGIDLOOP Titanium Button , RIGIDLOOP Cortical Fixation System XL Implant
K190774 · Medos International SARL · MBI · Jul 25, 2019 · Orthopedic
Device Facts
Record ID
K190774
Device Name
RIGIDLOOP Titanium Button , RIGIDLOOP Cortical Fixation System XL Implant
Applicant
Medos International SARL
Product Code
MBI · Orthopedic
Decision Date
Jul 25, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RIGIDLOOP Titanium Button is intended for fixation of soft-tissue to bone for the following indications when used with PERMATAPE™ 2.5mm Suture, #2 ORTHOCORD® Suture, #2 DYNACORD™ Suture, or equivalent nonabsorbable or partially-absorbable high strength operative suture: Shoulder: Acromioclavicular repair Pectoralis major / minor repair(1) Biceps tenodesis Elbow: Ulnar or radial collateral ligament reconstruction Biceps tendon reattachment Knee: Cruciate ligament reconstruction (1) Extracapsular repair: MCL, LCL, and Posterior oblique ligament Patellar tendon repair Meniscal root repair(2) (1) PERMATAPE 2.5mm suture must be used for pectoralis major/minor repair and cruciate ligament reconstruction (2) Size 0 or size 2/0 ORTHOCORD Suture may be used for meniscal root repair. RIGIDLOOPTM Cortical Fixation System XL Implant When used in conjunction with the RIGIDLOOP Cortical Fixation System, the RIGIDLOOP Cortical Fixation System XL Implant is intended for fixation of soft tissue to bone in orthopaedic procedures such as ACL repairs. When used in conjunction with the RIGIDLOOP Titanium Button, the RIGIDLOOP Cortical Fixation System XL Implant is intended for fixation of soft-tissue to bone for the following indications: Shoulder: Pectoralis major / minor repair Biceps tenodesis Elbow: Ulnar or radial collateral ligament reconstruction Biceps tendon reattachment Knee: Cruciate ligament reconstruction Extracapsular repair: MCL, LCL, and Posterior oblique ligament Patellar tendon repair Meniscal root repair
Device Story
Cortical fixation system for soft-tissue to bone attachment in orthopedic reconstructive surgery; includes titanium button implant and pre-attached non-implantable utility sutures (UHMWPE and Ethibond). XL implant version used for cortical breaching cases. Used in clinical settings by surgeons; device provides mechanical fixation of soft tissue to bone. Benefits include secure fixation during healing. Sterile, single-use.
Clinical Evidence
Bench testing only. Evaluation of fixation strength performed. Biological, sterility, packaging, and shelf-life testing conducted. Bacterial Endotoxin Testing met limits. No clinical or animal studies required.
Technological Characteristics
Titanium button implant; UHMWPE and Ethibond non-implantable utility sutures. Sterile, single-use. Mechanical fixation principle. No software or energy source.
Indications for Use
Indicated for soft-tissue to bone fixation in orthopedic reconstructive surgeries including shoulder (acromioclavicular, pectoralis major/minor, biceps tenodesis), elbow (ulnar/radial collateral ligament reconstruction, biceps tendon reattachment), and knee (cruciate ligament reconstruction, MCL/LCL/posterior oblique ligament repair, patellar tendon repair, meniscal root repair).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medos International SARL % LeeAnn Walosin Regulatory Affairs Specialist DePuy Mitek, Inc. 325 Paramount Drive Raynham, Massachusetts 02767
Re: K190774
Trade/Device Name: RIGIDLOOP™ Titanium Button, RIGIDLOOP™ Cortical Fixation System XL Implant Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: June 27, 2019 Received: June 28, 2019
Dear LeeAnn Walosin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190774
Device Name RIGIDLOOP™ Titanium Button RIGIDLOOP™ Cortical Fixation System XL Implant
#### Indications for Use (Describe)
The RIGIDLOOP Titanium Button is intended for fixation of soft-tissue to bone for the following indications when used with PERMATAPE™ 2.5mm Suture, #2 ORTHOCORD® Suture, #2 DYNACORD™ Suture, or equivalent nonabsorbable or partially-absorbable high strength operative suture:
Shoulder: Acromioclavicular repair Pectoralis major / minor repair(1) Biceps tenodesis
Elbow: Ulnar or radial collateral ligament reconstruction Biceps tendon reattachment
Knee: Cruciate ligament reconstruction (1) Extracapsular repair: MCL, LCL, and Posterior oblique ligament Patellar tendon repair Meniscal root repair(2)
(1) PERMATAPE 2.5mm suture must be used for pectoralis major/minor repair and cruciate ligament reconstruction (2) Size 0 or size 2/0 ORTHOCORD Suture may be used for meniscal root repair.
RIGIDLOOPTM Cortical Fixation System XL Implant
When used in conjunction with the RIGIDLOOP Cortical Fixation System, the RIGIDLOOP Cortical Fixation System XL Implant is intended for fixation of soft tissue to bone in orthopaedic procedures such as ACL repairs.
When used in conjunction with the RIGIDLOOP Titanium Button, the RIGIDLOOP Cortical Fixation System XL Implant is intended for fixation of soft-tissue to bone for the following indications:
Shoulder: Pectoralis major / minor repair Biceps tenodesis
Elbow: Ulnar or radial collateral ligament reconstruction Biceps tendon reattachment
Knee: Cruciate ligament reconstruction Extracapsular repair: MCL, LCL, and Posterior oblique ligament Patellar tendon repair Meniscal root repair
{3}------------------------------------------------
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DePuy Synthes Mitek Sports Medicine. The logo consists of a circular design with interconnected shapes in a gold color, followed by the text "DePuy Synthes" in a dark blue color. Below that, the text "MITEK SPORTS MEDICINE" is written in a smaller font size, also in dark blue. Finally, the text "COMPANIES OF Johnson & Johnson" is written in a smaller font size.
## 510(k) SUMMARY
# RIGIDLOOP Titanium Button & RIGIDLOOP XL Implant
| Date Summary<br>Prepared | March 25, 2019 | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Submitter's<br>Name and<br>Address | DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 | |
| Legal<br>Manufacturer | Medos International SARL<br>Chemin-Blanc 38, Case Postale<br>CH 2400<br>Le Locle, Switzerland | |
| Contact<br>Person | LeeAnn Walosin<br>Regulatory Affairs Specialist<br>DePuy Mitek<br>a Johnson & Johnson<br>company<br>325 Paramount Drive<br>Raynham, MA 02767, USA | Telephone: 612-230-9909<br>Facsimile: 508-977-6409<br>e-mail: lswiftwa@its.jnj.com) |
| Name of<br>Medical Device | Proprietary Name:<br>Common Name: | RIGIDLOOP™ Titanium Button;<br>RIGIDLOOP™ Cortical Fixation<br>System XL Implant<br>Fixation Device |
| Device<br>Classification | Smooth or threaded metallic bone fixation fastener, classified as Class<br>II, product code MBI regulated per 21 CFR 888.3040. (Orthopedic<br>panel) | |
| Substantial<br>Equivalence | The proposed RIGIDLOOP Titanium Button and the RIGIDLOOP<br>Cortical Fixation System XL Implant (when used with the proposed<br>RIGIDLOOP Titanium Button) are substantially equivalent to the<br>following predicate devices:<br><br>PREDICATE DEVICES:<br>ARTHREX PEC REPAIR BUTTON, LARGE PEC<br>BUTTON, BICEP BUTTON, PROXIMAL BICEPS<br>BUTTON, (Arthrex: K123341) SWIVELOCK (Arthrex: K173845) REFERENCE DEVICE: RIGIDLOOP CORTICAL FIXATION SYSTEM, (Depuy<br>Mitek a Johnson & Johnson Company: K130814) | |
| RIGIDLOOP Titanium Button | | |
| Indications for<br>Use | The RIGIDLOOP Titanium Button is intended for fixation of soft-<br>tissue to bone for the following indications when used with<br>PERMATAPE™ 2.5mm Suture, #2 ORTHOCORD® Suture, #2<br>DYNACORD™ Suture, or equivalent nonabsorbable or partially-<br>absorbable high strength operative suture:<br><br>Shoulder:<br>Acromioclavicular repair<br>Pectoralis major / minor repair1<br>Biceps tenodesis<br><br>Elbow:<br>Ulnar or radial collateral ligament reconstruction<br>Biceps tendon reattachment<br><br>Knee:<br>Cruciate ligament reconstruction1<br>Extracapsular repair: MCL, LCL, and Posterior oblique ligament<br>Patellar tendon repair<br>Meniscal root repair2<br><br>1 PERMATAPE 2.5mm suture must be used for pectoralis major/minor<br>repair and cruciate ligament reconstruction<br>2 Size 0 or size 2/0 ORTHOCORD Suture may be used for<br>meniscal root repair. | |
| | RIGIDLOOP Cortical Fixation System XL Implant<br>When used in conjunction with the RIGIDLOOP Cortical Fixation<br>System, the RIGIDLOOP Cortical Fixation System XL Implant is<br>intended for fixation of soft tissue to bone in orthopedic procedures such<br>as ACL repairs. | |
| | When used in conjunction with the RIGIDLOOP Titanium Button, the<br>RIGIDLOOP Cortical Fixation System XL Implant is intended for<br>fixation of soft-tissue to bone for the following indications: | |
| | Shoulder:<br>Pectoralis major / minor repair<br>Biceps tenodesis | |
| | Elbow:<br>Ulnar or radial collateral ligament reconstruction<br>Biceps tendon reattachment | |
| | Knee:<br>Cruciate ligament reconstruction<br>Extracapsular repair: MCL, LCL, and Posterior oblique ligament<br>Patellar tendon repair<br>Meniscal root repair | |
| Device<br>Description | RIGIDLOOP Titanium Button<br>The proposed RIGIDLOOP Titanium Button is a cortical fixation device<br>that provides a means of fixation of soft tissue to bone in orthopedic<br>reconstructive surgery. The device consists of a titanium button implant<br>that is provided with pre-attached leading and trailing sutures that are<br>non-implantable. The white leading utility suture is composed of<br>UHMWPE (device high molecular weight polyethylene) and the green<br>trailing suture is composed of #5 Ethibond®. This device is provided<br>sterile and is for single use only. | |
| | RIGIDLOOP Cortical Fixation System XL Implant<br>The RIGIDLOOP XL Implant is an extra-large titanium button implant<br>that is currently intended for use with the already cleared RIGIDLOOP<br>Cortical Fixation System devices (K130814). The RIGIDLOOP XL<br>Implant can now also be used with the proposed RIGIDLOOP Titanium<br>Button for the proposed indications. The RIGIDLOOP XL Implant is<br>used in cases of cortical breaching. This device is provided sterile and is<br>for single use only. | |
| Technological<br>Characteristics | The proposed device, the RIGIDLOOP Titanium button (when used as a<br>standalone system and when used in conjunction with the currently | |
| Although there are differences between the proposed device and the<br>predicate in the implant design, implant size, implant material, operative<br>sutures, and principle in operation, performance testing has<br>demonstrated that the proposed device is as safe and effective as the<br>predicates and that the differences do not raise any different questions<br>concerning safety and efficacy. Based on this comparison with the<br>predicate devices, the proposed device is considered substantially<br>equivalent to the predicate devices: ARTHREX PEC REPAIR<br>BUTTON, LARGE PEC BUTTON, BICEP BUTTON, PROXIMAL<br>BICEPS BUTTON, (Arthrex: K123341) and SWIVELOCK (Arthrex:<br>K173845).<br>The implant design, implant materials, and implant materials are<br>identical to the reference device; the principle of operation is similar<br>when compared to the reference device: RIGIDLOOP CORTICAL<br>FIXATION SYSTEM, (Depuy Mitek a Johnson & Johnson Company:<br>K130814). | | |
| Non-clinical<br>Testing | Verification activities have been performed on the proposed<br>RIGIDLOOP Titanium button, when used as a standalone device, when<br>used in conjunction with the currently marketed RIGIDLOOP Cortical<br>Fixation System XL Implant, and also on the chosen predicates.<br>Performance testing included an evaluation of fixation strength. The<br>testing demonstrated substantial equivalence of device performance.<br>Safety evaluations were conducted to address biological, sterility,<br>packaging and shelf-life testing. Bacterial Endotoxin Testing has been<br>completed and results have demonstrated that the proposed devices meet<br>the endotoxin limits. No animal or clinical studies were required. The<br>proposed device has raised no new issues of safety and efficacy. | |
| Safety and<br>Performance | Results of the performance testing have demonstrated that the proposed<br>devices are suitable for their intended use.<br>Based on similarities in the indications for use, technological<br>characteristics, and performance in comparison to the predicate devices,<br>the proposed device is considered substantially equivalent to the<br>predicate devices. | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image displays the logo for DePuy Synthes Mitek Sports Medicine. The text "DePuy Synthes" is in a bold, sans-serif font, with "MITEK SPORTS MEDICINE" appearing in a smaller font size below it. Underneath this, the phrase "COMPANIES OF Johnson-Johnson" is present, with the Johnson & Johnson logo in a stylized script.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for DePuy Synthes Mitek Sports Medicine. The logo consists of a circular graphic to the left of the text "DePuy Synthes" in a dark blue font. Below that, the text "MITEK SPORTS MEDICINE" is present. Underneath a thin line, the text "COMPANIES OF Johnson-Johnson" is written in a smaller font size.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for DePuy Synthes Mitek Sports Medicine. The logo consists of a circular graphic to the left of the text "DePuy Synthes" in a dark blue font. Below that is the text "MITEK SPORTS MEDICINE" in a smaller, lighter blue font. Underneath that is the text "COMPANIES OF Johnson & Johnson" in a small font.
has the same intended use and indications as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.