K190772 · Olympus Medical Systems Corp. · IZI · Oct 10, 2019 · Radiology
Device Facts
Record ID
K190772
Device Name
SURGICAL MICROSCOPE SYSTEM ORBEYE with IR
Applicant
Olympus Medical Systems Corp.
Product Code
IZI · Radiology
Decision Date
Oct 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1600
Device Class
Class 2
Indications for Use
This instrument has been designed to be used for viewing intra-operative blood flow in the cerebral vascular area and is intended for use in adults only.
Device Story
The ORBEYE with IR is a surgical microscope system used for intra-operative visualization of cerebral blood flow. It utilizes an LED light source, infrared imaging unit, and infrared filter unit to excite and resolve fluorescence from intravenously administered Indocyanine Green (ICG). The system captures images via two UHD resolution sensors, which are displayed on a 4K monitor for stereoscopic observation by surgeons. Operated in an OR setting, the system assists clinicians in assessing vascular patency and flow during cerebral procedures. The device includes an auto-focus unit and foot switch for hands-free control. By providing real-time, high-definition fluorescence imaging, the system aids in intra-operative decision-making, potentially improving surgical outcomes by allowing surgeons to verify blood flow in the cerebral vascular area.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing (IR illumination wavelength, optical properties, patient/user safety, wireless interference) and an animal study (5 beagle dogs) comparing the subject device to a predicate system for cerebral surface observation using ICG. No gross pathological abnormalities were observed.
Technological Characteristics
System includes OME-V200 microscope, OME-L200 LED light source, MAJ-2304 IR imaging unit, and MAJ-2307 filter unit. Features UHD-3D imaging, IR fluorescence capability, and auto-focus. Complies with ANSI/AAMI ES 60601-1:2005/(R)2012 and IEC 60601-1-2:2014. Software-controlled; wireless connectivity for foot switch.
Indications for Use
Indicated for viewing intra-operative blood flow in the cerebral vascular area in adult patients.
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
SPY SCOPE INTRA-OPERATIVE OPERATING SYSTEM (K091515)
SURGICAL MICROSCOPE SYSTEM ORBEYE (510(k) exempt)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 10, 2019
Olympus Medical Systems Corp. % Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, Pennsylvania 18034-0610
Re: K190772
Trade/Device Name: SURGICAL MICROSCOPE SYSTEM ORBEYE with IR Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic X-Ray System Regulatory Class: Class II Product Code: IZI, EPT Dated: September 10, 2019 Received: September 11, 2019
Dear Sheri Musgnung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K190772
Device Name
SURGICAL MICROSCOPE SYSTEM ORBEYE with IR
Indications for Use (Describe)
This instrument has been designed to be used for viewing intra-operative blood flow in the cerebral vascular area and is intended for use in adults only.
Type of Use (Select one or both, as applicable):
| ☐ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------|-----------------------------------------------|
|------------------------------------------------|-----------------------------------------------|
__ Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K190772
Device Name INFRARED FILTER UNIT MAJ-2307
Indications for Use (Describe)
This instrument has been designed to be used for viewing intra-operative blood flow in the cerebral vascular area and is intended for use in adults only.
Type of Use (Select one or both, as applicable)
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### б K190772 Page 1 of
Image /page/4/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are a sans-serif font and are closely spaced together. There is a thin, gold line underneath the word, which appears to be a design element.
Traditional 510(k) Notification SURGICAL MICROSCOPE SYSTEM ORBEYE with IR
Date Prepared: September 10, 2019
# 510(k) Summary
### 5.1 GENERAL INFORMATION
■ 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
- Contact Person: Sheri L. Musgnung Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung(@olympus.com
- Manufacturing Site Olympus Medical Systems Corp. Hinode Plant 34-3 Hirai, Hinode-machi, Nishitama-gun, Tokyo 190-0182, Japan
## 5.2 DEVICE IDENTIFICATION
| ■ Device Name<br>■ Model Name | SURGICAL MICROSCOPE SYSTEM ORBEYE with IR |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| [System components]<br>ORBEYE SURGICAL MICROSCOPE OME-V200<br>LED LIGHT SOURCE OME-L200<br>INFRARED IMAGING UNIT MAJ-2304<br>INFRARED FILTER UNIT MAJ-2307<br>AUTO FOCUS UNIT MAJ-2303<br>FOOT SWITCH MAJ-2301<br>4K COMBINED CABLE 10M MAJ-2309<br>RECORDER REMOTE CABLE 3P MAJ-2310<br>POWER CORD US MAJ-2312<br>Indocyanine Green for Injection, USP | |
| ■ Common Name | Surgical microscope system |
| ■ Regulation Number | 892.1600 |
{5}------------------------------------------------
#### б K190772 Page 2 of
Image /page/5/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are a sans-serif font and are closely spaced together. Below the word "OLYMPUS" is a horizontal yellow bar. The logo is clean and modern.
Traditional 510(k) Notification SURGICAL MICROSCOPE SYSTEM ORBEYE with IR
#### 878.4700
- Regulation Name Angiographic x-ray system Surgical microscope and accessories
- Regulatory Class Class II
- Product Code IZI, EPT
- Classification Panel Radiology General & Plastic Surgery
## 5.3 PREDICATE DEVICE
#### Predicate device
| Device name | 510(k) Submitter | 510(k) No. |
|-------------|--------------------------------|------------|
| Leica FL800 | LEICA BIOSYSTEMS RICHMOND, Inc | K141136 |
#### ■ Reference device
| No. | Device name | 510(k) Submitter | 510(k) No. |
|-----|-----------------------------------------------|----------------------------------|---------------------------------------------|
| 1 | SPY SCOPE INTRA-OPERATIVE<br>OPERATING SYSTEM | NOVADAQ<br>TECHNOLOGIES, INC. | K091515 |
| 2 | SURGICAL MICROSCOPE<br>SYSTEM ORBEYE | OLYMPUS MEDICAL<br>SYSTEMS CORP. | 510(k) exempt<br>(under<br>878.4700<br>EPT) |
### 5.4 DEVICE DESCRIPTION
The SURGICAL MICROSCOPE SYSTEM ORBEYE with IR is used for viewing intra-operative blood flow in the cerebral vascular area.
When the INFRARED IMAGING UNIT MAJ-2304 and the INFRARED FILTER UNIT MAJ-2307 are mounted on the LED LIGHT SOURCE OME-L200 and the ORBEYE SURGICAL MICROSCOPE OME-V200 respectively, it allows the surgical microscope to produce excitation light and resolve florescence light from the fluorescent ICG. The subject device of this 510(k) submission, ORBEYE with IR system, enables the microscopic observation with IR (Infrared) Imaging.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, creating a strong visual impact. A thin, curved, yellow line is located beneath the word.
Traditional 510(k) Notification SURGICAL MICROSCOPE SYSTEM ORBEYE with IR
The ORBEYE with IR system will be marketed as a System and will also be sold as individual components.
The subject system consists of the following primary components:
- ORBEYE SURGICAL MICROSCOPE OME-V200 -
- LED LIGHT SOURCE OME-L200 -
- INFRARED IMAGING UNIT MAJ-2304 -
- -INFRARED FILTER UNIT MAJ-2307
- AUTO FOCUS UNIT MAJ-2303 -
- FOOT SWITCH MAJ-2301 -
- -4K COMBINED CABLE 10M MAJ-2309
- RECORDER REMOTE CABLE 3P MAJ-2310 -
- -POWER CORD US MAJ-2312
- Indocyanine Green for Injection, USP -
#### ORBEYE SURGICAL MICROSCOPE OME-V200
OME-V200 is a surgical microscope to enlarge the images of the surgical site or enable stereoscopic observation. In the microscope unit, two image sensors in UHD (ultra high definition) resolution are incorporated. Connecting the microscope with an UHD-3D monitor enables displaying enlarged images or viewing stereoscopic images in UHD resolution.
#### LED LIGHT SOURCE OME-L200
OME-L200 is an LED light source to supply illumination light to the surgical site through the OME-V200.
#### INFRARED IMAGING UNIT MAJ-2304
MAJ-2304 is a unit that enables IR Imaging.
#### INFRARED FILTER UNIT MAJ-2307
MAJ-2307 is a unit that switches filters to enable IR Imaging.
#### AUTO FOCUS UNIT MAJ-2303
MAJ-2303 is a unit that enables Auto Focus.
#### FOOT SWITCH MAJ-2301
MAJ-2301 is a foot switch and receiver that operates a surgical microscope.
#### 4K COMBINED CABLE 10M MAJ-2309
MAJ-2309 is a video cable for displaying 4K video of a surgical microscope on a monitor.
#### RECORDER REMOTE CABLE 3P MAJ-2310
MAJ-2310 is a cable for operating the recorder from a surgical microscope.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly shadowed, giving them a three-dimensional appearance. A thin, horizontal, gradient line in shades of yellow and orange is located beneath the word.
Traditional 510(k) Notification SURGICAL MICROSCOPE SYSTEM ORBEYE with IR
### POWER CORD US MAJ-2312
MAJ-2312 is the power cable for supplying power to the surgical microscope.
#### Indocvanine Green for Iniection. USP
Fluorescent agent for IR imaging, supplied from Diagnostic Green LLC.
### 5.5 INDICATIONS FOR USE
### SURGICAL MICROSCOPE SYSTEM ORBEYE with IR
This instrument has been designed to be used for viewing intra-operative blood flow in the cerebral vascular area and is intended for use in adults only.
### INFRARED FILTER UNIT MAJ-2307
This instrument has been designed to be used for viewing intra-operative blood flow in the cerebral vascular area and is intended for use in adults only.
# 5.6 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE
The SURGICAL MICROSCOPE SYSTEM ORBEYE with IR has the same technological characteristics and design as the predicate device except for the following new features:
- 1. Image sensor
- 2. Software
- 3. Excitation light
- 4. UHD (Ultra High Definition) technology
All other technological characteristics of both the subject and predicate devices are identical.
Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.
### 5.7 PERFORMANCE DATA
510(k) Summary Page 4 of 6
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The letters are closely spaced and fill most of the frame. A thin, horizontal yellow bar is visible beneath the word, adding a subtle color contrast to the composition. The registered trademark symbol is visible to the right of the word.
The following performance data were provided in support of the substantial equivalence determination.
### 1) Software verification and validation testing
Software verification and validation testing for the SURGICAL MICROSCOPE SYSTEM ORBEYE with IR were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".
### 2) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the SURGICAL MICROSCOPE SYSTEM ORBEYE with IR. The system complies with the ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012 standards for safety and the IEC 60601-1-2:2014 standards for EMC.
### 3) Performance testing - Bench
Bench testing for the SURGICAL MICROSCOPE SYSTEM ORBEYE with IR as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.
- । Verification for the IR illumination light wavelength
- Verification for the optical property of the objective IR optical systems ।
- Verification for patient safety of the IR illumination light (focusing on patient's । thermal injury)
- Verification for patient safety of the IR illumination light (focusing on users' eye । injury)
- Verification for the wireless connection mechanism not to be interfered by the । radio waves from other equipment
## 4) Performance testing - Animal
In the animal study, indocyanine green (ICG) solution was intravenously administered to 5 male beagle dogs of 7-8 months old. The cerebral tissue, the cerebral surface arteries and veins were observed under near-infrared light (IR) using the subject device system and Leica M525 F50 in combination with the predicate device. The animal testing is performed in accordance to 21 CFR Part 58 - GOOD LABORATORY PRACTICE FOR NONCLINICAL LABORATORY STUDIES.
Neither the subject system nor predicate system induced any gross pathological abnormalities in the observation.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly shadowed, giving them a three-dimensional appearance. A thin, horizontal gold line is underneath the word. A registered trademark symbol is located to the right of the word.
#### Traditional 510(k) Notification SURGICAL MICROSCOPE SYSTEM ORBEYE with IR
#### 5) Performance testing - Clinical
No clinical study was performed to demonstrate substantial equivalence.
#### 6) Risk analysis
Risk analysis for the SURGICAL MICROSCOPE SYSTEM ORBEYE with IR was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
### 5.8 CONCLUSIONS
Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate devices, the SURGICAL MICROSCOPE SYSTEM ORBEYE with IR raises no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, efficacy and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.