SurgicalAR

K190764 · Medivis, Inc. · LLZ · May 13, 2019 · Radiology

Device Facts

Record IDK190764
Device NameSurgicalAR
ApplicantMedivis, Inc.
Product CodeLLZ · Radiology
Decision DateMay 13, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

Indications for Use

SurgicalAR is a software device for display of medical images and other healthcare data. It includes functions for image review image manipulation, basic measurements and 3D visualization (MPR reconstructions and 3D volume rendering). Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA cleared display that meets technical specifications reviewed and accepted by FDA or displays accepted by the appropriate regulatory agency for the country in which it is used. Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance. SurgicalAR software is indicated for use by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians and technologists. When accessing SurgicalAR software from a wireless stereoscopic head-mounted display (HMD) or mobile device, images viewed are for informational purposes only and not intended for diagnostic use.

Device Story

SurgicalAR is a software platform for 3D visualization of pre-acquired 2D medical images (CT/MR DICOM data) to assist surgical planning. Clinicians use the software on desktop PCs (mouse/keyboard), touchscreen monitors, or wireless stereoscopic head-mounted displays (HMDs) (hand gestures). The device reconstructs 3D models from DICOM inputs, enabling MPR and 3D volume rendering. Used in operating rooms and hospital offices; operated by surgeons, radiologists, and technologists. Output is viewed on monitors or HMDs to support clinical decision-making and surgical planning. HMD/mobile viewing is restricted to informational purposes. Benefits include enhanced visualization of patient anatomy for surgical preparation.

Clinical Evidence

Bench testing only. Design verification confirmed design outputs met input requirements. Design validation, including human factors and usability engineering, was conducted via simulated use to identify potential use errors and assess risk management effectiveness. All testing met predetermined acceptance criteria with no identified risks to safety or efficacy.

Technological Characteristics

Software-only platform for DICOM image processing. Features MPR and 3D volume rendering. Supports traditional monitors, touchscreens, and wireless stereoscopic HMDs. Connectivity via standard network protocols. Data security includes HTTPS and SSL encryption. No hardware components; no essential performance characteristics identified.

Indications for Use

Indicated for qualified healthcare professionals (radiologists, non-radiology specialists, physicians, technologists) for display, review, manipulation, measurement, and 3D visualization of medical images. Not for primary interpretation of lossy compressed mammography or digitized film screen images. HMD/mobile access is for informational purposes only, not diagnostic use.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ May 13, 2019 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. MEDIVIS, Inc. % Ms. Amy Lynn Regulatory Affairs Consultant 20 Jay Street, Suite 312 BROOKLYN NY 11201 Re: K190764 Trade/Device Name: SurgicalAR Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: March 15, 2019 Received: March 25, 2019 Dear Ms. Lynn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190764 Device Name SurgicalAR #### Indications for Use (Describe) SurgicalAR is a software device for display of medical images and other healthcare data. It includes functions for image review image manipulation, basic measurements and 3D visualizations and 3D volume rendering). Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA cleared display that meets technical specifications reviewed and accepted by FDA or displays accepted by the appropriate regulatory agency for the country in which it is used. Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance. SurgicalAR software is indicated for use by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians and technologists. When accessing SurgicalAR software from a wireless stereoscopic head-mounted display (HMD) or mobile device, images viewed are for informational purposes only and not intended for diagnostic use. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------| | <div> <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." ============================================================================================================================================================================== {3}------------------------------------------------ # Section 5 - 510(k) Summary | Company: | MEDIVIS, Inc. | |----------------------------|-------------------------------------------------------| | Address: | 20 Jay Street, Suite 312<br>Brooklyn, New York, 11201 | | Contact: | Amy Lynn | | | Regulatory Affairs Consultant | | Telephone Number: | 954.600.8299 | | Email: | amylynncicatello@gmail.com | | Preparation Date: | March 15, 2019 | | Trade Name: | SurgicalAR | | Device Common Name: | System, Image Processing, Radiological | | Classification Regulation: | 21 CFR 892.2050 | | Classification Name: | Picture archiving and communications system | | Class: | II | | Panel: | Radiology | | Product Code: | LLZ | | Predicate Device: | VitreaView (K163232) | ## 5.1 Device Description SurgicalAR is a software platform to be used by clinicians for the visualization of medical images in 3D to allow for surgical planning activities. The device takes pre-acquired 2D medical images and reconstructs 3D models that a clinician can then view on a stereoscopic, holographic display. The software application is used to: • Load patient CT/MR DICOM data • View DICOM data using a traditional computer monitor or in Augmented Reality (AR) using a head-mounted display, HMD. ## 5.2 Intended Use SurgicalAR is a software device for display of medical images and other healthcare data. It includes functions for image review image manipulation, basic measurements and 3D visualization (MPR reconstructions and 3D volume rendering). Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA cleared display that meets technical specifications reviewed and accepted by FDA or displays accepted by the appropriate regulatory agency for the country in which it is used. #### K190764 {4}------------------------------------------------ Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance. SurgicalAR software is indicated for use by qualified healthcare professionals including, but not restricted to radiologists, non-radiology specialists, physicians and technologists. When accessing SurgicalAR software from a wireless stereoscopic head-mounted display (HMD) or mobile device, images viewed are for informational purposes only and not intended for diagnostic use. ## 5.3 Operator profile Qualified healthcare professionals, including but not restricted to surgeons, radiologists, nonradiology specialists, physicians, and technologists. # 5.4 Patient population The device is software which allows for viewing of DICOM data. Therefore, there is no specific patient population. 5.5 Intended use environment The software is intended to be used: - In operating rooms - In office environments within hospitals or at any other location with a computer - . For informational only purposes at any location using the head-mounted display (HMD) ## 5.6 Operating principle There are different operating principles such as viewing: - On desktop PCs with a traditional monitor the interaction with the software is performed with a mouse and/or keyboard; - On PCs with a touchscreen monitor the interaction with the software is performed with a touchscreen interface; or - On a head-mounted display the interaction with the software is performed using hand ● gestures. ## 5.7 Part of the body or type of tissue applied or interacted with The device is software only which allows for the viewing of DICOM data. Therefore, it does not interact with any body and/or tissue part and the viewing is also not limited to any body and/or tissue part. ## 5.8 Comparison to predicate device The SurgicalAR proposed device is substantially equivalent to the predicate device. K163232. as both devices have the same intended use and technological characteristics. While both devices do not have the exact same software features, the difference does not pose any safety and/or effectiveness questions. A full substantial equivalent discussion is provided in Section 12. The exact feature differences and omissions are detailed in the tables contained within Section 12. {5}------------------------------------------------ The proposed device slightly differs from the predicate device in two ways as listed below: First, the proposed device is not available for use through internet browsers like the predicate device. The predicate device allows access via an internet browser in addition to traditional installation on a PC's hard drive. This design difference does not create new or additional safety risks regarding the use of the device but rather limits the scope of accessibility of SurgicalAR. Secondly, the proposed device does not contain all of the same features as the predicate device. The lack of features are not key features nor do they impact the function or intended use of the device. The table below demonstrates the similarities in technology and features between SurgicalAR and the predicate device. | Criteria | Description | Subject Device | Predicate Device | Comparison | |-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------|-------------------------------------------------------------------------------------------------| | Annotation and<br>Measurement Tools | • Line<br>• Angle<br>• Ruler<br>• Arrow | Yes | Yes | Same | | User Installation<br>Requirements | • Runs within browser<br>using HTML and<br>JavaScript only<br>• No installation is<br>required on user's machine | No | Yes | Different.<br>Predicate allows<br>for access via Internet. | | Data Type<br>Supported | • DICOM<br>• Non-DICOM | Yes | Yes | Same | | Image<br>View/Manipulation | • Image Zoom<br>• Pan<br>• Window Level<br>• AutoWindow<br>• Level<br>• Reset<br>• Scout Lines<br>• Image Rotate<br>• Image Flip<br>• Magnify<br>• Image Invert<br>• Image Cine | No | Yes | Different.<br>Predicate has<br>"image invert"<br>and "image cine" features. | | Data Encryption | • HTTPS<br>• SSL | Yes | Yes | Same | | | Description | Subject Device | Predicate<br>Device | Comparison | | Criteria | SurgicalAR | Vitrea View<br>(K163232) | | | | Patient<br>Demographic<br>Display | Capable of displaying<br>patient demographic<br>information | Yes | Yes | Same | | Linking | Co-planar linking:<br>• Autolink<br>• Manual | Yes | Yes | Same | | User and Password<br>Control | Users can be managed<br>via an internal<br>database, active<br>directory, or parent<br>application | Yes | Yes | Same | | Data Security | Stored on server | Yes | Yes | Same | | Audit Trail | Audit trail logged | Yes | Yes | Same | | User Management | Database structure<br>allows mapping users<br>to groups internally or<br>mapping external<br>groups (AD, parent<br>application) to internal<br>groups and role | | | | | Transmission<br>Modes | Via the web with<br>Internet browsers | No | Yes | Different.<br>Predicate allows<br>for access via<br>Internet. | | File Type Used | • JPEG for<br>Lossy<br>data<br>• PNG for<br>Lossless data | Yes | Yes | Same | | MPR Viewing | This viewing feature<br>enables the display of<br>reformatted CT and<br>MR images into axial,<br>coronal and sagittal<br>orientations | Yes | Yes | Same | | 3D Volume<br>Rendered Viewing | This viewing feature<br>enables the display of<br>3D perspective views<br>of CT and MR image<br>sets that have been<br>transformed into | Yes | Yes | Same | | Criteria | Description | Subject Device | Predicate Device | Comparison | | | volumes. It also<br>provides presets to<br>enable users to alter<br>the visualization<br>parameters<br>of the 3D views to<br>highlight features. | SurgicalAR | Vitrea View<br>(K163232) | | | Active Target Tool | This viewing feature<br>provides a facility to<br>view a single target<br>location within<br>multiple images. | Yes | Yes | Same | | Crosshair<br>Navigation and<br>Synchronization | This viewing feature<br>provides a facility to<br>synchronize and scroll<br>through multiple<br>views at the same<br>time. | Yes | Yes | Same | | Ability to clone<br>images side by side | Ability to clone<br>images side by side. | No | Yes | Different.<br>Predicate<br>contains a<br>"clone"<br>copy/paste<br>feature. | | Ability to close an<br>image by clicking<br>an "X" in the<br>upper-left portion<br>of the view port | Ability to close an<br>image by clicking an<br>"X" in the upper-left<br>portion of the<br>viewport. | Yes | Yes | Same | | Ability to select<br>locale and language<br>settings on the<br>login screen | Ability to select locale<br>and language settings<br>on the login screen. | No | Yes | Different.<br>Predicate has<br>support for<br>multiple<br>languages. | | Ability to<br>customize the<br>columns in the<br>study directory by<br>selecting the<br>dropdown arrow on | Ability to customize<br>the columns in the<br>study directory by<br>selecting the<br>dropdown arrow on<br>the right side of each<br>column. | No | Yes | Different.<br>Predicate has<br>feature to<br>customize the<br>view of the<br>directory. | | Criteria | Description | Subject Device | Predicate<br>Device | Comparison | | the right side of<br>each column. | | | | | | Help Tips | Proactive help tips<br>appear for 10-15<br>seconds to educate<br>users on certain<br>functionality that may<br>not be obvious to a<br>new user. | No | Yes | Different.<br>Predicate has<br>feature to view<br>"help tips". | | Support for TIF<br>Files | Vitrea View can<br>display TIF files. | Yes | Yes | Same | | Tablet<br>support for<br>information<br>purpose<br>only<br>(Not for diagnostic<br>use) | This viewing feature<br>provides access of<br>Vitrea View software<br>on various iOS and<br>Android tablet<br>devices through the<br>default internet<br>browser. Key<br>features are:<br>• Two-finger<br>pinch to<br>zoom and<br>pan<br>• Touch and<br>drag to<br>scroll<br>• Double-tap to<br>access<br>Gesture<br>menu<br>• Tap<br>Carousel<br>thumbnail,<br>then tap<br>Image Pane<br>to swap<br>images<br>• Ambient<br>Lighting<br>Check | No | Yes | Different.<br>Predicate has<br>support for iOS<br>and Android for<br>non-diagnostic<br>viewing. | | HMD support for<br>information | This viewing feature<br>provides access of<br>SurgicalAR software | No | Yes | Different.<br>Proposed device<br>has support for | | Criteria | Description | Subject Device | Predicate Device | Comparison | | purpose only (not<br>for diagnostic use) | on consumer, off-the-<br>shelf-wireless, Wi-Fi<br>enabled, stereoscopic<br>head-mounted display<br>with minimum of 2GB<br>RAM | SurgicalAR | Vitrea View<br>(K163232) | HMDs for non-<br>diagnostic<br>viewing. | | Diagnostic<br>quality medical<br>image review | Ability to provide<br>diagnostic quality<br>medical image review<br>for multi-dimensional<br>digital images<br>acquired from a<br>variety of imaging<br>devices | Yes | Yes | Same | {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ # 5.9 Tests conducted Both design verification and design validation were successfully conducted as part of the testing for the SurgicalAR device. The purpose of the design verification activity was to ensure that the design output specifications met the design input requirements. Deviations associated with the written protocol were captured within the design verification report. These deviations did not negatively impact the device, it's intended use, nor did they introduce additional risks. Therefore, it is concluded at the design verification was successful in that the design output specifications satisfactorily met the design input requirements. Design validation, including human factors and usability engineering, was conducted. Human factors and usability engineering testing was performed to assess user interactions with the device's user interface. The purpose is to identify use errors that would or could result in serious harm to the patient or user. Human factors validation testing is also used to assess the effectiveness of risk management measures. The testing performed was simulated use. The design validation was successfully completed, and testing met all predetermined acceptance criteria. Furthermore, no additional risks to the safety or efficacy of the device were identified. #### 5.10 Essential performance characteristics SurgicalAR is software for medical image visualization. A hardware shutdown, power failure or other hardware issue that makes the software inoperable does not cause harm. Therefore, no unacceptable risk arises if the hardware loses performance. The hardware does not have any essential performance characteristics. The software risks were analyzed to find risks in the intolerable range. No risks have been identified in the intolerable range. As such, no measures are considered essential performance characteristics. In summary, the SurgicalAR software does not have any essential performance characteristics.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...