K190757 · Siemens Medical Solution USA, Inc. · LNH · May 31, 2019 · Radiology
Device Facts
Record ID
K190757
Device Name
MAGNETOM Avanto Fit
Applicant
Siemens Medical Solution USA, Inc.
Product Code
LNH · Radiology
Decision Date
May 31, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Your MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. Your MAGNETOM MR system may also be used for imaging during interventional procedures with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Device Story
MAGNETOM Avanto is an MR diagnostic system; inputs include radiofrequency signals and magnetic field gradients; system transforms these into cross-sectional images, spectroscopic images, or spectra. Operated by trained physicians/technicians in clinical settings. New software version (syngo MR E11E) adds 'Compressed Sensing (CS) Cardiac Cine' feature migrated from previously cleared systems. Output is displayed for physician interpretation to assist in diagnosis. Benefits include diagnostic visualization of internal anatomy and function; supports interventional procedures via MR-safe device compatibility.
Clinical Evidence
No clinical tests were conducted. Substantial equivalence is supported by non-clinical software verification and validation testing performed in accordance with FDA guidance documents and recognized international standards (IEC 62304, IEC 60601 series).
Technological Characteristics
Magnetic resonance diagnostic device; software-controlled imaging system. Conforms to IEC 60601-1 (safety/EMC), IEC 60601-2-33 (MR safety), ISO 14971 (risk management), and IEC 62304 (software lifecycle). Supports DICOM connectivity. Features Compressed Sensing (CS) Cardiac Cine application.
Indications for Use
Indicated for patients requiring magnetic resonance diagnostic imaging of the head, body, or extremities to visualize internal structure and function. Used for diagnostic imaging and during interventional procedures with MR-compatible devices.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
MAGNETOM Aera with Software syngo MR E11C-AP02 (K163312)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
Siemens Medical Solution USA, Inc. % Cordell L Fields, Esq. Regulatory Affairs Specialist 40 Liberty Boulevard, Mail Code 65-1A MALVERN PA 19355
Re: K190757
Trade/Device Name: MAGNETOM Avantofft Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI, MOS Dated: March 22, 2019 Received: March 25, 2019
Dear Mr. Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good
May 31, 2019
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and the word is horizontally oriented.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|-----------------------------------------|----------------------------------|
| Food and Drug Administration | Expiration Date: 06/30/2020 |
| Indications for Use | See PRA Statement below. |
510(k) Number (if known)
K190757 Device Name MAGNETOM Avantofit
#### Indications for Use (Describe)
Your MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.
Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Your MAGNETOM MR system may also be used for imaging during interventional procedures with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
# 510(k) Summary MAGNETOM Avantofit with software syngo MR E11E
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355, USA Establishment Registration Number: 2240869
March 22, 2019 Date Prepared
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR ട്ട 807.92.
## 1. General Information
#### Importer/Distributor:
Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355, USA Establishment Registration Number: 2240869
Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen Germany Establishment Registration Number: 3002808157
#### 2. Contact Information
Cordell L. Fields, Esq. Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355, USA Phone: (610) 448-6545 Fax: (610) 448-6469 E-mail: cordell.fields@siemens-healthineers.com
{4}------------------------------------------------
## 3. Device Name and Classification
| Trade Name | MAGNETOM Avantofit |
|-----------------------|---------------------------------------------|
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH<br>Secondary: LNI, MOS |
#### 4. Legally Marketed Predicate Device
| Trade Name | MAGNETOM Avantofit |
|------------|--------------------|
|------------|--------------------|
| 510(k) Number | K173592, Cleared February 13, 2018 |
|-----------------------|---------------------------------------------|
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH |
| | Secondary: LNI, MOS |
### 5. Device Description
The subject device, MAGNETOM Avanto® with software syngo MR E11E. is a modification of the previously cleared predicate device, MAGNETOM Avanto" with software syngo MR E11C-AP04 (K173592). The software version syngo E11E for MAGNETOM Avanto™ has been modified to include the software application "Compressed Sensing (CS) Cardiac Cine." This software application was migrated unchanged from the previously cleared MAGNETOM Skyra and Aera systems with syngo MR E11C-AP02 (K163312).
#### 6. Indication for Use
The indications for use for the subject device is the same as the predicate device and is as follows:
Your MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used.
These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Your MAGNETOM MR system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
{5}------------------------------------------------
# 7. Substantial Equivalence
The MAGNETOM Avanto® with software syngo MR E11E is substantially equivalent to the following device:
| Predicate Device | FDA<br>Clearance<br>Number | FDA<br>Clearance<br>Date | Product<br>Code |
|-----------------------------------------------------------|----------------------------|--------------------------|------------------|
| MAGNETOM Avantofit<br>with Software syngo MR<br>E11C-AP04 | K173592 | February 13,<br>2018 | LNH, LNI,<br>MOS |
As described above, the MAGNETOM Avanto" with software syngo MR E11E includes one new feature that is already cleared with the following reference device:
| Reference Device | FDA<br>Clearance<br>Number | FDA<br>Clearance<br>Date | Product<br>Code |
|------------------------------------------------------|----------------------------|--------------------------|------------------|
| MAGNETOM Aera with<br>Software syngo MR<br>E11C-AP02 | K163312 | January 27, 2017 | LNH, LNI,<br>MOS |
## 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device
The subject device MAGNETOM Avanto®" with software syngo MR E11E is substantially equivalent to the predicate device, MAGNETOM Avanto™ with software syngo MR E11C-AP04, with regard to the operational environment, programming language, operating system, and performance.
The subject device, MAGNETOM Avanto" with software syngo MR E11E, conforms to the IEC 62304, Edition 1.1, 2015-06, standard for software medical devices and other relevant IEC and NEMA standards.
While there is a difference in technological characteristics between the subject device and predicate device (the difference being the software application migrated from the previously cleared MAGNETOM Aera system with syngo MR E11C-AP02 (K163312)), this difference has been tested and the conclusions from the nonclinical data suqgests that the feature bear an equivalent safety and performance profile as that of the predicate device.
## 9. Nonclinical Performance Testing
The following performance testing was conducted on the subject device:
- Software verification and validation testing was completed in accordance with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", dated May 11, 2005.
{6}------------------------------------------------
# SIEMENS
- Performance testing was completed in accordance with the FDA guidance । document, "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices", dated November 18, 2016
The results from each set of tests demonstrate that the device performs as intended and is therefore substantially equivalent to the predicate device to which it has been compared.
MAGNETOM Avanto™ with software syngo MR E11E conforms to the following FDA recognized and international IEC and ISO standards:
| Recogniton<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>date | Standards<br>Development<br>Organization |
|----------------------|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------|
| 19-4 | General II<br>(ES/EMC) | C1:2009/(R)2012 and A2:2010/(R)2012<br>(Consolidated Text) Medical electrical<br>equipment - Part 1: General<br>requirements for basic safety and<br>essential performance (IEC 60601-<br>1:2005, MOD) | ES60601-<br>1:2005/(R)2012<br>and A1:2012, | AAMI ANSI |
| 19-8 | General | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety<br>and essential performance - Collateral<br>Standard: Electromagnetic<br>disturbances - Requirements and tests | 60601-1-2 Edition<br>4.0 2014-02 | IEC |
| 12-295 | Radiology | Medical electrical equipment - Part 2-<br>33: Particular requirements for the<br>basic safety and essential performance<br>of magnetic resonance equipment for<br>medical diagnostic | 60601-2-33 Ed.<br>3.2 B:2015 | IEC |
| 5-40 | General I<br>(QS/RM) | Medical devices - Application of risk<br>management to medical devices | 14971 Second<br>Edition 2007-03-<br>01 | ISO |
| 5-114 | General | Medical devices - Application of<br>usability engineering to medical devices | 62366-1:2015 | AAMI ANSI<br>IEC |
| 13-32 | Software/Info<br>rmatics | Medical device software - Software life<br>cycle processes | IEC 62304 Edition<br>1.1 2015-06 | AAMI ANSI<br>IEC |
| 12-300 | Radiology | Digital Imaging and Communications in<br>Medicine (DICOM) Set 03/16/2012<br>Radiology | PS 3.1 - 3.20<br>(2016) | NEMA |
No clinical tests were conducted to support the claim of substantial equivalence between the subject and predicate device.
{7}------------------------------------------------
# 10. General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk management is ensured via compliance with ISO 14971:2007 to identify and provide mitigation of potential hazards in a risk analysis early in the design phase and continuously throughout the development of the product. These risks are controlled via measures realized in hardware and software development, testing, and product labeling. To minimize risks. Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risk. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and-post processing of magnetic resonance images.
## 11.Conclusion as to Substantial Equivalence
There are no changes to the indications for use for the subject device as compared to that of the legally marketed predicate device, MAGNETOM Avanto™ with software syngo MR E11C-AP04 (K173592).
While the new software feature provides an additional capability compared to the predicate device, the additional capability is currently cleared features of the reference device MAGNETOM Aera with software syngo MR E11C-AP02 (K163312) and do not raise new questions of safety and effectiveness. All features have been verified and validated to support the claim of substantial equivalence to the predicate device.
For the aforementioned mentions Siemens believes that the subject device, MAGNETOM Avanto™ with software syngo MR E11E, is substantially equivalent to the predicate device, MAGNETOM Avanto10 with software syngo MR E11C-AP04.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.