← Product Code [MAX](/productcode/MAX) · K190708

# AccelFix Lumbar Interbody Fusion Cage System (K190708)

_L&K BIOMED Co., Ltd. · MAX · Sep 16, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190708

## Device Facts

- **Applicant:** L&K BIOMED Co., Ltd.
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Sep 16, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

AccelFix Lumbar Expandable Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft and/or allogenous bone graft composed of cancellous and/or corticocancellous bone. AccelFix Lumbar Expandable Cage System is to be used with supplemental fixation systems. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

## Device Story

Hollow, rectangular intervertebral body fusion cage; manufactured from titanium alloy; implanted via transforaminal, posterior, anterior-to-psoas, or lateral approaches. Device inserted in unexpanded state; expanded in situ using system instrumentation to restore disc height. Features serrated superior/inferior surfaces for fixation. Used in conjunction with autogenous or allogenous bone graft and supplemental spinal fixation. Operated by surgeons in clinical/OR settings. Provides structural support to vertebral bodies to facilitate fusion; benefits patients by stabilizing spinal segments and alleviating discogenic pain.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated per ASTM F2077 (static/dynamic axial compression, compression-shear, torsion) and ASTM F2267 (subsidence). Results demonstrate mechanical properties similar to or higher than predicate devices.

## Technological Characteristics

Material: Ti-6Al-4V ELI titanium alloy (ASTM F136). Form factor: Hollow, rectangular, expandable cage with serrated surfaces. Approaches: Transforaminal, posterior, anterior-to-psoas, lateral. Mechanical testing standards: ASTM F2077, ASTM F2267.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- ProLift® Expandable System ([K173182](/device/K173182.md))
- Concorde Lift Expandable Cage ([K171425](/device/K171425.md))
- LnK Lumbar Interbody Fusion Cage System ([K110783](/device/K110783.md), [K120063](/device/K120063.md), [K121096](/device/K121096.md), [K151140](/device/K151140.md), [K181380](/device/K181380.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 16, 2019

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

L&K Biomed Co., Ltd. Kihyang Kim RA Vice President #201, 202 16-25, Dongbaekjungang-ro 16 beon-gil, Giheung-gu Yongin-si, Republic of Korea

## Re: K190708

Trade/Device Name: AccelFix Lumbar Interbody Fusion Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: August 13, 2019 Received: August 15, 2019

## Dear Kihyang Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Melissa Hall Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190708

### Device Name

AccelFix Lumbar Expandable Cage System

### Indications for Use (Describe)

AccelFix Lumbar Expandable Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft and/or allogenous bone graft composed of cancellous and/or corticocancellous bone. AccelFix Lumbar Expandable Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

| Type of Use (Select one or both, as applicable)                                       |                                                                                      |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92:

#### 1 Manufacturer

| Submitter:      | Gook Jin Kang                                    |
|-----------------|--------------------------------------------------|
|                 | L&K BIOMED Co., Ltd.                             |
|                 | #201, 202 16-25, Dongbaekjungang-ro 16 beon-gil  |
|                 | Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea |
|                 | Phone. 82-2-6717-1983/ FAX .82-2-6717-1949       |
| Contact Person: | Kihyang Kim                                      |
|                 | khkim@lnkbiomed.com/ khkim3747@gmail.com         |

#### 2. Device Identification

| Trade Name:          | AccelFix Lumbar Expandable Cage System |
|----------------------|----------------------------------------|
| Common Name:         | Intervertebral Body Fusion Cage        |
| Product Code:        | MAX                                    |
| Classification:      | Class II                               |
| Classification Name: | Intervertebral body fusion device      |
| Regulation No.       | 21 CFR 888.3080                        |
| Classification Pane  | Orthopedic                             |

#### 3. Predicate or legally marketed devices which are substantially equivalent.

The AccelFix Lumbar Expandable Cage System has undergone a comprehensive battery of non-clinical testing, including chemical and physical testing. Testing has shows the device is as safe and effective for its intended use as a predicate device and supports a determination of substantial equivalence.

The AccelFix Lumbar Expandable Cage System met all pre-defined acceptance criteria and, in tests where it was compared to either AccelFix Lumbar Expandable Cage System the predicate devices, was found to not represent a new worst case.

Overall, the results of the performance bench tests support the substantial equivalence of the Subject device.

The design feature and indications for use for the subject device 'AccelFix Lumbar Expandable Cage System' is substantially equivalent to the following predicate(s);

- Primary Predicate Device: ProLift® Expandable System (K173182) 1)
- 2) Additional Predicate Devices: Concorde Lift Expandable Cage(171425) LnK Lumbar Interbody Fusion Cage System (K110783, K120063, K121096, K151140, K181380)

#### 4 Materials

AccelFix Lumbar Expandable Cage System is manufactured from Ti-6Al-4V ELI titanium alloy (ASTM F136). This this is the same material used in the predicate devices.

{4}------------------------------------------------

#### 5. Description of the Device

The AccelFix Lumbar Expandable Cage System's implants are interbody fusion devices intended for use as an aid in spinal fixation. They are made of Titanium 6AL-4V Alloy (ASTM F136). These hollow, rectangular implants are offered in a variety of widths, lengths, heights and lordotic angles designed to adapt to a variety of patient anatomies. The implants can be expanded in height after insertion in the unexpanded state using the system instrumentation. The implants have serrations on the superior and inferior surfaces designed for fixation.

- AccelFix-XT Expandable Cage is to be implanted via transforaminal and posterior approach.
- AccelFix-XTP Expandable Cage is to be implanted via Anterior to Psoas approach. ।
- AccelFix-XL Expandable Cage is to be implanted via lateral approach. ।

#### 6. Intended use

AccelFix Lumbar Expandable Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft and/or allogenous bone graft composed of cancellous and/or corticocancellous bone. AccelFix Lumbar Expandable Cage System is to be used with supplemental fixation systems. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

#### 7. Performance -Bench testing

The AccelFix Lumbar Expandable Cage System was tested according to the ASTM F 2077, and ASTM F 2267 as noted below;

Static compression, dynamic compression, static and dynamic shear testing according to ASTM F2077, was presented to demonstrate the substantial equivalency of the AccelFix Lumbar Expandable Cage System to the predicate devices.

- Static Axial Compression Test ASTM F 2077 -17 .
- . Static Compression-Shear Test - ASTM F 2077 -17
- " Static Torsion Test - ASTM F 2077 -17
- . Static Expulsion Test
- . Static Subsidence Test - ASTM F 2267 - 04 (Reapproved 2018)/F 2077-17
- . Dynamic Axial Compression Test- ASTM F 2077 -17
- יי Dynamic Compression-Shear Test - ASTM F 2077 -17

Bench testing to evaluate the mechanical properties of the AccelFix Lumbar Expandable Cage System showed a higher or similar mechanical value than predicate marketed devices.

{5}------------------------------------------------

#### 8. Summary of Technology Characteristics

AccelFix Lumbar Expandable Cage System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing.

#### 9. Substantial Equivalence:

AccelFix Lumbar Expandable Cage System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance.

#### 10. Conclusion

The information presented demonstrates the substantial equivalency of the AccelFix Lumbar Expandable Cage System to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K190708](https://fda-staging.innolitics.com/device/K190708)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).