Neuro-IOM system with Neuro-IOM.NET software is a medical device intended for intraoperative neurophysiologic monitoring: the device provides information to assess a patient's neurophysiological status. The system allows to monitor the functional integrity and/or mapping of central and peripheral nervous system including motor and sensory pathways. It is provided in III different configurations: I. 32/B II. 32/S III. 16/S The system ensures the following IOM modalities: free-run EMG (electromyography), direct nerve stimulation including pedicle screw test, SSEP (somatosensory evoked potentials), MEP (motor evoked potentials), EEG (electroencephalography), AEP (auditory evoked potentials), VEP (visual evoked potentials), direct cortical stimulation. Also, the train-of-four (TOF) stimulation is performed. The system is not intended to measure the vital signs. It records the data to be interpreted by the neuromonitoring specialist.
Device Story
Neuro-IOM system is an intraoperative neurophysiologic monitoring (IOM) device; records electrical signals from central/peripheral nervous systems; provides electrical stimulation (SSEP, MEP, TOF, direct nerve/cortical stimulation) to elicit responses. Inputs: physiological signals via electrodes; outputs: waveforms, spectral arrays (CSA/DSA), and trend plots displayed on a PC. Used in operating rooms/ICUs by trained neuromonitoring specialists. Device transforms raw signals into visual data for real-time assessment of neural integrity during surgery. Clinicians use output to monitor for acute dysfunction (e.g., mechanical trauma, vascular insufficiency) and adjust surgical approach. Benefits include real-time feedback to prevent permanent neurological damage.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993), electrical safety (AAMI/ANSI ES 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), and performance testing per IEC 60601-2-40 and IEC 80601-2-26. Software verification and validation conducted per FDA guidance for moderate level of concern.
Technological Characteristics
PC-based IOM system; 16/32 channel configurations; USB connectivity. Stimulators: electrical (monophasic/biphasic, 0-200mA, 1-1000V), auditory, visual. Filters: 0.2-2000Hz (low freq), 10Hz-4kHz (high freq). Input impedance >1000 MOhm. CMRR >90 dB. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-40, ISO 10993, ISO 14971.
Indications for Use
Indicated for intraoperative neurophysiologic monitoring to assess neurophysiological status and monitor functional integrity/mapping of central and peripheral nervous systems (motor and sensory pathways) in patients undergoing surgery. Modalities include free-run EMG, direct nerve stimulation, SSEP, MEP, EEG, AEP, VEP, direct cortical stimulation, and TOF. Not for vital signs measurement. For use by trained personnel.
Regulatory Classification
Identification
An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
May 22, 2021
Neurosoft Ltd % Barry Ashar Official Correspondent Makromed, Inc. 88 Stiles Road Salem, NH 03079 USA
Re: K190703
Trade/Device Name: Neuro-IOM system with Neuro-IOM.NET software, models - 32/B - 32/S - 16/S Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked Response Electrical Stimulator Regulatory Class: Class II Product Code: GWF, GWE, GWJ, OLT, PDQ Dated: April 9, 2021 Received: April 20, 2021
Dear Barry Ashar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190703
Device Name
Neuro-IOM system with Neuro-IOM.NET software, models - 32/B - 32/S - 16/S
Indications for Use (Describe)
Neuro-IOM system with Neuro-IOM.NET software is a medical device intended for intraoperative neurophysiologic monitoring: the device provides information to assess a patient's neurophysiological status.
The system allows to monitor the functional integrity and/or mapping of central and peripheral nervous system including motor and sensory pathways.
It is provided in III different configurations:
I. 32/B
II. 32/S
III. 16/S
The system ensures the following IOM modalities: free-run EMG (electromyography), direct nerve stimulation including pedicle screw test, SSEP (somatosensory evoked potentials), MEP (motor evoked potentials), EEG (electroencephalography), AEP (auditory evoked potentials), VEP (visual evoked potentials), direct cortical stimulation. Also the train-of-four (TOF) stimulation is performed.
The system is not intended to measure the vital signs. It records the data to be interpreted by the neuromonitoring specialist.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
Company Name: Neurosoft Ltd Company Address: 5, Voronin str. 153032 Ivanovo Company Phone: Company Fax: Company e-mail: Official Contact for Correspondence: Phone: E-mail: secondary contact: Phone: Email: Date Summary Prepared:
Russian Federation +7 4932 59-33-44 +7 4932 24-04-80 info@neurosoft.com Mr. Barry Ashar (Consultant) 603.674.9074 bashar@makromed.com Mr. Eugene Polezhaev +7 4932 58-45-84 polezhaev@neurosoft.ru
April 9, 2021
DEVICE IDENTIFICATION Trade name:
Generic/ Common Name: Regulation number:
Regulation name: Product Code: Subsequent Product Codes: Panel:
Neuro-IOM system with Neuro-IOM.NET software, models - 32/B - 32/S - 16/S
intraoperative neurophysiologic system 21 CFR § 882.1870 Class II Evoked response electrical stimulator GWF GWE, GWJ, OLT Neurology
#### PREDICATE DEVICES:
Neurosoft identified the following legally marketed devices as substantially equivalent: - Xltek Protektor 32, NATUS MEDICAL, INC., K093304
#### DEVICE DESCRIPTION:
Neuro-IOM system with Neuro-IOM.NET software is a medical device intended for intraoperative neurophysiologic monitoring: the device provides information to assess a patient's neurophysiological status.
The system allows to monitor the functional integrity and/or mapping of central and peripheral nervous system including motor and sensory pathways. It is provided in III different configurations:
I.32/B 32/8 II. III. ાર્ભદ્ર
{4}------------------------------------------------
The system ensures the following IOM modalities: free-run EMG (electromyography), direct nerve stimulation including pedicle screw test, SSEP (somatosensory evoked potentials), MEP (motor evoked potentials), EEG (electroencephalography), AEP (auditory evoked potentials), VEP (visual evoked potentials, direct cortical stimulation. Also, the train-of-four (TOF) stimulation is performed.
The system is not intended to measure the vital signs. It records the data to be interpreted by the neuromonitoring specialist.
The systems can be used in operating rooms, intensive care units of different health care facilities (including clinics, hospitals, health centers, ambulance centers, etc.), specialized medical facilities (including prevention centers, medicine centers for emergency, military and medical expertise centers), research and educational medical and biological facilities where the neuromonitoring is required, only by qualified operators who have received training on these devices.
Description of stimuli
| Stimulator | Picture | Modalities |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Electrical Stimulator Built-in<br>Electronic Unit | Single pulse monophasic normal<br>Image: [Single pulse monophasic normal waveform with labels A and D]<br>A – amplitude (0 – 200 mA, 0 – 400 V)<br>D – duration (50 – 5000 μs) | SSEP, TOF |
| | Single pulse monophasic inverse<br>Image: [Single pulse monophasic inverse waveform with labels A and D]<br>A – amplitude (0 – 200 mA, 0 – 400 V)<br>D – duration (50 – 5000 μs) | SSEP, TOF |
| | Single pulse biphasic<br>Image: [Single pulse biphasic waveform with labels A and D] | SSEP, TOF |
| | A - amplitude (0 - 200 mA, 0 – 400 V)<br>D - duration (50 – 5000 μs) | |
| | Train of pulses<br>Image: Train of pulses | MEP |
| | A - amplitude (0 – 200 mA, 0 – 400 V)<br>D - duration (50 – 5000 μs)<br>ISI - interstimulus interval (3 – 10 ms)<br>N - number of pulses (2 - 200) | |
| Low Current Stimulator Built-in<br>Electronic Unit | Single pulse monophasic normal<br>Image: Single pulse monophasic normal<br>A - amplitude (0-20 mA, 0 – 30 V)<br>D - duration (50 – 500 μs) | Direct nerve<br>stimulation |
| | Single pulse monophasic inverse<br>Image: Single pulse monophasic inverse<br>A - amplitude (0-20 mA, 0-30 V)<br>D-duration (50 – 500 μs) | Direct nerve<br>stimulation |
| | Single pulse biphasic | Direct nerve<br>stimulation |
| Image: Pulse diagram with amplitude A and duration D<br><br>A – amplitude (0 – 20 mA, 0 – 30 V)<br><br>D – duration (50 – 500 μs) | | |
| Image: Train of pulses diagram with amplitude A, duration D, and interstimulus interval ISI<br><br>Train of pulses<br><br>A – amplitude (0 – 20 mA, 0 – 30 V)<br><br>D – duration (50 – 500 μs)<br><br>ISI – interstimulus interval (2 - 10 ms)<br><br>N – number of pulses (1 - 200) | Direct nerve<br>stimulation | |
| Electrical Stimulator Built-in<br>Amplifier Unit | Image: Single pulse normal diagram with amplitude A and duration D<br><br>Single pulse normal<br><br>A – amplitude (0 – 200 mA)<br><br>D – duration (50 – 5000 μs) | SSEP, TOF |
| | Image: Single pulse inverse diagram with amplitude A and duration D<br><br>Single pulse inverse<br><br>A – amplitude (0 – 200 mA) | SSEP, TOF |
| | D - duration (50 – 5000 μs) | |
| | Train of pulses<br>Image: Train of pulses diagram with A, ISI, and D labels<br>ISI<br>D | MEP |
| | A – amplitude (0 - 200 mA) | |
| | D - duration (50 – 5000 μs) | |
| | ISI – interstimulus interval (2 - 10 ms) | |
| | N - number of pulses (1 - 255) | |
| | | |
| Low Current Stimulator Built-in<br>Amplifier Unit | Single pulse normal<br>Image: Single pulse normal diagram with A and D labels<br>A – amplitude (0 - 20 mA)<br>D - duration (50 – 500 μs) | Direct nerve<br>stimulation |
| | Single pulse inverse<br>Image: Single pulse inverse diagram with A and D labels<br>A – amplitude (0 - 20 mA)<br>D - duration (50 – 500 μs) | Direct nerve<br>stimulation |
| | Train of pulses | Direct nerve<br>stimulation |
| | | |
| | Image: waveform with ISI and D labels<br>A - amplitude (0 - 20 mA)<br>D - duration (50 – 500 μs)<br>ISI - interstimulus interval (2 - 10 ms)<br>N - number of pulses (1 - 255) | |
| Transcranial Electrical<br>Stimulator | Single pulse monophasic normal<br>Image: waveform with A and D labels<br>A - amplitude (1 – 1000 V)<br>D - duration (40 – 200 μs) | MEP |
| | Single pulse monophasic inverse<br>Image: waveform with A and D labels<br>A - amplitude (1 – 1000 V)<br>D - duration (40 – 200 μs) | MEP |
| | Single pulse biphasic<br>Image: waveform with A and D labels | MEP |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| A – amplitude (1 – 1000 V) | | MEP |
|------------------------------------------------------------------------------------------------------------------------------------|--|-----|
| D – duration (40 – 200 μs) | | |
| Train of pulses | | |
| <img alt="Train of pulses diagram" src="pulse_train_1.png"/> | | |
| A - amplitude (1 - 1000 V)<br>D - duration (40 – 200 μs)<br>ISI - interstimulus interval (1 - 10 ms)<br>N - number of pulses (1-9) | | |
| Double train | | MEP |
| <img alt="Double train diagram" src="pulse_train_2.png"/> | | |
| A - amplitude (1 – 1000 V) | | |
| D - duration (40 – 200 μs) | | |
| ISI - interstimulus interval (1 - 10 ms)<br>ITI - intertrain interval (10 – 100 ms) | | |
| Train + Pulse | | MEP |
| <img alt="Train + Pulse diagram" src="pulse_train_3.png"/> | | |
| A - amplitude (1 - 1000 V) | | |
| D - duration (40 – 200 μs) | | |
| ISI - interstimulus interval (1 – 10 ms) | | |
{10}------------------------------------------------
| ITI – interval between train and pulse (10 – 100 ms) | | |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-----|
| Auditory Stimulator | Click<br>Image: Click waveform<br>A<br>D<br>A – amplitude (0 – 120 dB)<br>D – duration (100 – 5000 μs) | AEP |
| | Tone<br>Image: Tone waveform<br>A<br>D<br>A – amplitude (0 – 120 dB)<br>D – duration (0.1 – 90 ms)<br>Tone frequency – 100 – 5000 Hz | AEP |
| Visual Stimulator | Image: Visual Stimulator waveform<br>A<br>D<br>D – duration (2 - 1500 ms)<br>Max. luminance - 1500 cd/m² | VEP |
## INDICATIONS FOR USE:
Neuro-IOM system with Neuro-IOM.NET software is a medical device intended for intraoperative neurophysiologic monitoring: the device provides information to assess a patient's neurophysiological status.
The system allows to monitor the functional integrity and/or mapping of central and peripheral nervous system including motor and sensory pathways.
{11}------------------------------------------------
It is provided in III different configurations: I. 32/B II. 32/S III. 16/S
The system ensures the following IOM modalities: free-run EMG (electromyography), direct nerve stimulation including pedicle screw test, SSEP (somatosensory evoked potentials), MEP (motor evoked potentials), EEG (electroencephalography), AEP (auditory evoked potentials), VEP (visual evoked potentials), direct cortical stimulation. Also, the train-of-four (TOF) stimulation is performed.
The system is not intended to measure the vital signs. It records the data to be interpreted by the neuromonitoring specialist.
{12}------------------------------------------------
Page 10
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL E Q U I V A L E N C E:
In support of a substantial equivalence determination, hereunder are comparison charts with the submitted devices.
| ATTRIBUTE /<br>CHARACTERISTICS | Neuro-IOM 16S, 32S, 32B<br>Neurosoft Ltd<br>(Submitted Product) | Xltek Protektor 32<br>NATUS MEDICAL, INC | Why the differences do not adversely<br>affect the safety and effectiveness |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| "K" numbers | NA | K093304 | |
| Proprietary / Trade<br>Name | Neurosoft Ltd/Neuro-IOM 16S, 32S, 32B<br>with Neuro-IOM.NET software | Xltek Protector 32 | |
| CFR Section | 882.1870 | 882.1870 | Same |
| Pro-code | GWF, GWE, GWJ, OLT | GWF, GWE, GWJ, OLT | Same |
| Classification name | Stimulator, Electrical, Evoked Response | Stimulator, Electrical, Evoked Response | Same |
| Intended Use | Neuro-IOM system with Neuro-IOM.NET<br>software is a medical device intended for<br>intraoperative neurophysiologic monitoring: the<br>device provides information to assess a patient's<br>neurophysiological status.<br>The system allows to monitor the functional<br>integrity and/or mapping of central and<br>peripheral nervous system including motor and<br>sensory.<br>It is provided in III different configurations:<br>I. 32/B<br>II. 32/S<br>III. 16/S<br>The system ensures the following IOM<br>modalities: free-run EMG (electromyography),<br>direct nerve stimulation including pedicle screw<br>test, SSEP (somatosensory evoked potentials),<br>MEP (motor evoked potentials), EEG<br>(electroencephalography), AEP (auditory evoked<br>potentials), VEP (visual evoked potentials),<br>direct cortical stimulation. Also, the train-of-four<br>(TOF) stimulation is performed.<br>The system is not intended to measure the vital<br>signs. It records the data to be interpreted by the<br>neuromonitoring specialist. | The Protektor32 channel system, composed<br>of both hardware and software, is intended to<br>be used for intraoperative neurological<br>monitoring. The instrument uses<br>Electroencephalography (EEG), Evoked<br>Potentials (EP), Electromyography (EMG)<br>and Transcranial Motor Evoked Potential<br>(TcMEP) stimulation techniques to provide<br>the healthcare professionals with information<br>to help assess a patient's neurological status<br>during surgery.<br>The TcMEP mode is intended for<br>intraoperative diagnosis of acute dysfunction<br>in corticospinal axonal conduction brought<br>about by mechanical trauma (traction,<br>shearing, laceration, or compression) or<br>vascular insufficiency.<br>The EPWorks software, an integral part of<br>the system, is intended to allow a medical<br>professional to manually configure<br>stimulation and acquisition parameters and<br>to manually create EEG, EP, EMG and<br>TcMEP protocols according to their own<br>requirements. The intended use, for each of<br>the software's output, is as follows:<br>● The EEG, EP, and EMG waveforms are | |
| | | | |
| | | intended to help the user assess a<br>patient's neurological status during<br>surgery.<br>- Simple waveform parameters (e.g.,<br>amplitude. latency), and user-defined<br>Fast Fourier transform (FFT) displays<br>(compressed spectral array or CSA,<br>density spectral array or OSA) are<br>intended to help the user analyse the<br>EEG and EP waveforms.<br>This device is intended to be used by<br>qualified medical practitioners, trained in<br>EEG, EP and EMG who will exercise<br>professional judgment when using the<br>information. | |
| Intended User | The device is to be used by trained<br>personnel only | The device is to be used by trained<br>personnel only | Same |
| Device Hardware<br>Setup | Connected to PC, not standalone | Connected to PC, not standalone | Same |
| Standards | AAMI/ANSI ES 60601-1:2005/(R)2012<br>IEC 60601-1-2:2014<br>IEC 60601-1-6:2013<br>IEC 60601-2-40:2016<br>IEC 62366-1:2015<br>AAMI/ANSI 62304:2006<br>ISO series 10993<br>IEC 60068-2-31:2008<br>IEC 60068-2-80:2005<br>ISO 14971:2007<br>ISO 13485:2012 | IEC 60601-1:2000<br>IEC 60601-1-1:2000<br>IEC 60601-1-2:2005<br>IEC 60601-2-40:1998<br>IEC 61000-4-2:2001<br>IEC 61000-4-3:2002<br>IEC 61000-4-4:2004<br>IEC 61000-4-5:2005<br>IEC 61000-4-6:2006<br>IEC 61000-4-8:2001<br>IEC 61000-4-11:2004<br>61000-3-3:2002<br>ISO 15223:2000 | |
| Workflow, Menu | PC-controlled | PC-controlled | Same |
| Interface to<br>Computer | USB | USB | Same |
| Amplifiers | | | |
| Channels | 16/32 | 16/32 | Same |
| 1.5 mm touch-proof<br>input jacks on pods | Same | Same | Same |
| Cable Length | 5 m | 10 ft., 20 ft. | Same |
| Input impedance | >1000 MOhm | >100 MOhm | Since the higher input impedance in our<br>device improves the quality of signal |
| Common Mode<br>Rejection (CMRR) | >90 dB…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.