K190685 · Xuzhou Kernel Medical Equipment Co., Ltd. · OAP · Jul 26, 2019 · Physical Medicine
Device Facts
Record ID
K190685
Device Name
Hair Growth System
Applicant
Xuzhou Kernel Medical Equipment Co., Ltd.
Product Code
OAP · Physical Medicine
Decision Date
Jul 26, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hair Growth System (Model: KN-8000A) is intended for the promotion of hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenetic alopecia who have Norwood Hamilton Classifications IIa-V; and both genders having Fitzpatrick Classification of Skin Phototypes I-IV.
Device Story
Hair Growth System (KN-8000A) is a transportable, non-invasive, low-level laser therapy (LLLT) device for androgenetic alopecia. Device features a host machine, lifting bracket, and hood array containing 650nm diode lasers (max 5mW output). Operated by clinicians in a clinic setting; 8-inch rotatable touch screen interface allows selection of continuous or pulse irradiation modes. Integrated access sensors automatically power off lasers if the patient's scalp deviates from the work area; working distance positioning equipment ensures 4cm ± 1cm distance. Device uses military-class quick connectors for light source replacement. Double switch protection (key switch and power-on password) prevents unauthorized use. Output is red laser light; intended to stimulate hair growth. Benefits include non-invasive treatment for hair loss.
Clinical Evidence
Bench testing only. Evaluated via electrical safety (IEC 60601-1, IEC 60825-1), electromagnetic compatibility (IEC 60601-1-2), and software verification/validation per FDA guidance. Biocompatibility testing performed on silicon rubber loop bar sucker per ISO 10993-5 and ISO 10993-10.
Indicated for females with androgenic alopecia (Ludwig-Savin I-II) and males with androgenetic alopecia (Norwood-Hamilton IIa-V), Fitzpatrick skin types I-IV, to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Xuzhou Kernel Medical Equipment Co., Ltd. % Jet Li Regulation Manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, CN
### Re: K190685
Trade/Device Name: Hair Growth System Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: June 18, 2019 Received: June 28, 2019
July 26, 2019
## Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil R. P. Ogden, M.S. Acting Team Assistant Director Light Based Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190685
Device Name
Hair Growth System Device (Model: KN-8000A)
Indications for Use (Describe)
The Hair Growth System (Model: KN-8000A) is a prescription use device intended for the prowth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenic alopecia who have Norwood Hamilton Classifications IIa-V; and both genders having Fitzpatrick Classification of Skin Phototypes I to IV.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 100%;"> <b> <span style="text-decoration: overline;">X</span> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 100%;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Date of the summary prepared: July 26, 2019
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information
### Sponsor
- Company Name: Xuzhou Kernel Medical Equipment Co., Ltd. �
- Address: Kernel Mansion, Economic Development District, Xuzhou City, Jiangsu Province, � China
- Phone: +86 1831 685 8036 �
- Fax: +86 0516-87732208 13776587162 �
- Contact Person (including title): Wang Jing (Management Representative) �
- E-mail: wjkernel@126.com �
### Application Correspondent:
Guangzhou KEDA Biological Tech Co., Ltd. Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou City, China Contact Person: Jet Li Tel: +86-18588874857 Email: med-il@foxmail.com
#### Subject Device Information 2.
- � Trade Name: Hair Growth System (Model: KN-8000A)
- Lamp, non-heating, for promotion of hair growth � Common Name:
- Infrared lamp per 21 CFR 890.5500surgery and in Classification name: � dermatology
- General & Plastic Surgery � Review Panel:
- OAP (Laser, comb, hair) � Product Code:
2
- Requlation Class: �
- 21 CFR 890.5500 � Regulation Number:
### 3. Predicate Device Information
| Sponsor | Capillus, LLC. | Sunetics International<br>Marketing Group LLC |
|-------------|------------------|-----------------------------------------------|
| Device Name | Capillus devices | Sunetics Clinical Laser<br>"G" |
{4}------------------------------------------------
| | Capillus352;<br>Capillus302 | |
|----------------------|-----------------------------|-----------------|
| 510(k) Number | K162994 | K132646 |
| Product Code | OAP | OAP |
| Regulation<br>Number | 21 CFR 890.5500 | 21 CFR 890.5500 |
| Regulation Class | 2 | 2 |
#### 2. Device Description
The Hair Growth System Model: KN-8000A is a transportable, non-invasive, low-level laser device intended to treat Androgenetic Alopecia (Hair Loss) and to promote hair growth in both Males and Females. It produces a red light, diode lasers operating at a 650-nanometer wavelength (maximum output power of each is 5 mW). The device mainly includes a host machine, light source and lifting bracket. The light source applicator was equipped with the access sensor device, it can power off the light source when the patient's scalp deviates from the work area.
### 5. Intended Use / Indications for Use
The Hair Growth System (Model: KN-8000A) is intended for the promotion of hair growth in females with androgenic alopecia who have Ludwig-Savin Classifications I-II, and in males with androgenetic alopecia who have Norwood Hamilton Classifications Ila-V; and both genders having Fitzpatrick Classification of Skin Phototypes I-IV.
### 6. Design
The Hair Growth System (Model: KN-8000A) is designed to be free to stretch and be able to take care of every part of the scalp in consideration of the characteristics of the human head. The device laser light source is arranged in matrix structure with dot scope to balance the light energy. Light source shall achieve the fixation of light source and connections with electrical through the military-class quick connector, which makes it more convenient and faster to replace the light source. Free lift shall be designed of cantilever, so that the light source can stay at any angle. 8 "rotatable touch screen design, easy to operate, which does not need the professional training. According to the difference of the treatment parts, each treatment hood array can be controlled individually . According to treatment needs, the continuous irradiation or pulse irradiation mode can be selected. There is the function of auxiliay display in the light source, to facilitate doctors to timely grasp the treatment time information. There is the access sensor in the treatment applicator of the hood array. When the patient's scalp deviate from the work area, the light source can be stopped in time to ensure patient safety. Working distance positioning equipment, to facilitate users to quickly adjust to the appropriate working distance. Double switch protections of key switch and power-on password, to prevent the use of unauthorized personnel.
### 7. Materials
There is one kind of patient- directly contacting component in the subject device as the following list.
| Component of<br>Device Requiring<br>Biocompatibility | Material of<br>Component | Body Contact<br>Category<br>(ISO 10993-1) | Contact Duration<br>(ISO 10993-1) |
|------------------------------------------------------|--------------------------|-------------------------------------------------------|-----------------------------------|
| loop bar sucker | silicon rubber | Surface-contacting<br>device: skin of patient<br>head | Maximum 30<br>minutes(<24hours) |
{5}------------------------------------------------
|--|--|--|--|
We conduct biocompatibility test on the material of loop bar sucker. So the parts' biocompatibility is safe.
### 8. Physical characteristics
| Basic Unit Characteristics | |
|---------------------------------------------------------|-------------------------------------------------------------|
| Compliance* with 21 CFR 898 | No |
| Main Unit Weight | 3.0-4.0 kg |
| Housing Materials of main unit | ABS |
| Environment for operation | Temperature: 5 ~ 40 °C |
| | Relative humidity: ≤85% |
| | Atmospheric pressure: 700hPa ~ 1060hPa |
| Storage and Transport Conditions | Temperature: -5 ~ 40 °C |
| | Relative humidity: 10-95% |
| | Atmospheric pressure: 700hPa ~ 1060hPa |
| Compliance with Voluntary<br>Standards | Yes, Comply with IEC 60601-1, IEC 60601-1-2,<br>IEC 60825-1 |
| Patient leakage current | Comply with IEC 60601-1 |
| Power Source | AC 100-240V, 50 / 60Hz ± 2% |
| Software/Firmware/Microprocessor<br>Control? | Yes |
| Specification | |
| Input power | 120VA |
| The specification and rating value of<br>fuse protector | T1.5AL / 250V Φ5 * 20 |
| Structure style of stand | Cart type |
| Radiation area | 900 cm²±10% |
| Irradiation distance | 4cm±1cm |
| Laser Category | 3R laser products |
| Laser output wavelength | 650nm±10nm |
| Maximum output laser radiation | <5mW |
| Technology | Laser |
### 10. Test Summary
The Hair Growth System has been evaluated for its safety and performance by lab bench testing as following:
- Electrical safety test according to IEC 60601-1 and IEC 60825-1 standards �
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for PreMarket Submissions and for Software Contained in Medical Devices"
{6}------------------------------------------------
## 11. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, and intended use of Hair Growth System, model: KN-8000A is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Candidate | K162994 | K132646 |
| Device Name | Hair Growth System | Capillus Devices<br>Capillus352;<br>Capillus 302 | Sunetics Clinical Laser<br>"G" |
| LLLT Device Type | LLLT | LLLT | LLLT |
| Prescription | Prescription Use | Prescription Use | Prescription Use |
| Intended Use | intended for the<br>promotion of hair<br>growth in females with<br>androgenic alopecia<br>who have Ludwig-<br>Savin Classifications I-<br>II, and in males with<br>androgenetic alopecia<br>who have Norwood<br>Hamilton<br>Classifications IIa-V;<br>and both genders<br>having Fitzpatrick<br>Classification of Skin<br>Phototypes I-IV. | intended to treat<br>Androgenetic<br>Alopecia and<br>promote hair growth<br>in males who have<br>Norwood-Hamilton<br>Classifications of IIa<br>to V patterns of<br>hair loss and to treat<br>Androgenetic<br>Alopecia and promote<br>hair growth in females<br>who have Ludwig<br>(Savin) Scale I-1 to I-<br>4, II-1, II-2, or frontal;<br>both with Fitzpatrick<br>Skin Types I to IV. | The Sunetics Clinical<br>Laser (model 'G" &<br>model "W2326") is<br>indicated to<br>treat Androgenetic<br>Alopecia (Hair Loss)<br>and to promote hair<br>growth in Males<br>who have Norwood<br>Hamilton Classifications<br>of IIa to V and<br>Fitzpatrick Skin<br>Types I to IV and also<br>in Females who have<br>Ludwig (Savin) 1-4.11I-<br>1, 11-2, or<br>frontal patterns of hair<br>loss & Fitzpatrick Skin<br>Types I to IV. |
| Laser Diode Class | 3R | 3R | 3R |
| Design | Hood array | Helmet/Cap | Hood array |
| Laser Diode Power | $4.5mW \pm 10%$ | 5mW max | 5mW max |
| Wavelength | 650nm | 650nm | 650nm |
| Irradiance (power<br>per area) | 2.332mW/cm2 | Capillus352: 2.745<br>mW/cm2<br>Capillus302: 2.256<br>mW/cm2 | |
| Classification | OAP | OAP | OAP |
| Classification Name | Infrared Lamp | Infrared Lamp | Infrared Lamp |
| Common Usage Name | Lamp, Non-Heating | Lamp, Non-Heating | Lamp, Non-Heating |
| Review Panel | General & Plastic<br>Surgery Committee | General & Plastic<br>Surgery Committee | General & Plastic<br>Surgery Committee |
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II |
| Fitzpatrick Skin<br>Phototypes | I- IV | I- IV | I- IV |
| Ludwig-Savin Scale<br>(women) | Ludwig-Savin I-II<br>(females) | Ludwig-Savin I-II<br>(females) | Ludwig-Savin I-II<br>(females) |
| Norwood-Hamilton<br>(men) | Norwood Hamilton IIA-V (males) | Norwood Hamilton IIA-V (males) | Norwood Hamilton IIA-V (males) |
| Treatment Frequency | Treatment- 17weeks, every other day (indefinite) | Treatment- 17weeks, every other day (indefinite) | every other day (indefinite) |
| Device Class | II | II | II |
| Biocompatibility | ISO 10993-5;<br>ISO10993-10 | ISO 10993-5;<br>ISO10993-10 | ISO 10993-5;<br>ISO10993-10 |
| Safety and EMC<br>standard | IEC 60601-1;<br>IEC60601-1-2 | IEC 60601-1;<br>IEC60601-1-2 | IEC 60601-1;<br>IEC60601-1-2 |
| Laser performance<br>standard | IEC60825-1 | IEC60825-1 | IEC60825-1 |
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.