K190682 · Permobil AB · IPL · Feb 27, 2020 · Physical Medicine
Device Facts
Record ID
K190682
Device Name
Explorer Mini
Applicant
Permobil AB
Product Code
IPL · Physical Medicine
Decision Date
Feb 27, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3900
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Explorer Mini is a pediatric powered wheelchair with the intention to provide mobility to pediatric users weighing up to 35 pounds and maximum length of up to 39 inches tall, between 12-36 months of age, who position themselves in a sitting position in the wheelchair and have the capacity to operate a joy stick hand control.
Device Story
Pediatric powered wheelchair for children 12-36 months with mobility impairments; provides mobility via joystick-controlled gear-motor wheel assemblies. Input: user joystick activation; Output: motorized propulsion and deceleration. Used in home, clinical, or preschool environments under adult supervision. Device features adjustable vertical column, 360-degree torso support, and saddle seat for sitting or straddle standing. Benefits include early mobility and postural support. Healthcare providers/caregivers use device to facilitate child movement; output directly enables independent navigation for the child.
Clinical Evidence
Usability study conducted on 33 children with mobility impairments to evaluate compensatory motor/sensory skills and postural control. Label comprehension study performed on 15 therapists and 15 parents. Human factors validation confirmed device safety and effectiveness for intended users.
Technological Characteristics
Steel frame with molded plastic overlays; aluminum tower/seat components. Powered by 2x 12V AGM lead-acid batteries. 24V PG nVR2 controller. Central joystick. ISO 7176 series testing for stability, braking, and durability. Manual sit-to-stand adjustment.
Indications for Use
Indicated for pediatric users aged 12-36 months, weighing up to 35 lbs, max height 39 inches, with mobility impairments requiring assistance; must be able to operate a joystick hand control and sit in the wheelchair.
Regulatory Classification
Identification
A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
February 27, 2020
Permobil AB Ivan Fernandez Director of Regulatory Compliance Box 120 S-861 23 Timra, Sweden
Re: K190682
Trade/Device Name: Explorer Mini Regulation Number: 21 CFR 890.3900 Regulation Name: Standup Wheelchair Regulatory Class: Class II Product Code: IPL Dated: January 28, 2020 Received: January 28, 2020
## Dear Ivan Fernandez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, Ph.D. Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190682
Device Name
Explorer Mini
Indications for Use (Describe)
The Explorer Mini is a pediatric powered wheelchair with the intention to provide mobility to pediatric users weighing up to 35 pounds and maximum length of up to 39 inches tall, between 12-36 months of age, who position themselves in a sitting position in the wheelchair and have the capacity to operate a joy stick hand control.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary - K190682
| Submitter | Permobil AB<br>Box 120<br>S-861 23 Timrå<br>Sweden |
|---------------------------------|--------------------------------------------------------------------------------------------------|
| Phone:<br>Facsimile: | +46 60 595900<br>+46 60 575250 |
| Contact Person: | Ivan Fernandez, Director of Regulatory Compliance |
| Email Address: | ivan.fernandez@permobil.com |
| Telephone number: | 800-736-0925 |
| Date Prepared: | February, 2020 |
| Trade Name: | Explorer Mini |
| Common/<br>Classification Name: | Wheelchair, Standup,<br>Standup Wheelchair |
| Product Code: | IPL<br>21CFR 380.3900 |
| Predicate Device: | WHEELCHAIR MODEL STS<br>Innovative Products Inc<br>071390<br>Product code: IPL<br>21CFR 380.3900 |
## Description of the Explorer Mini
This submission covers the Explorer Mini device.
### Indication for use:
The Explorer Mini is a pediatric powered wheelchair with the intention to provide mobility to pediatric users weighing up to 35 pounds and maximum length of up to 39 inches tall, between 12-36 months of age, who position themselves in a sitting position in the wheelchair and have the capacity to operate a joy stick hand control.
{4}------------------------------------------------
### General Description:
The Explorer Mini is designed for children with mobility impairments that will never walk, inefficient mobility or lose the ability to walk efficiently, need mobility assistance in early childhood.
The Explorer Mini is intended for children weighing up to 35 pounds versus the Wheelchair Model STS (K071390) which is intended for users weighing up to 150 pounds. The Explorer Mini is designed specifically for the pediatric users. This is a limit in indications for use to the larger population of the predicate device that includes children and young adults. Clinical and non-clinical performance testing has demonstrated this difference does not affectiveness of the device when used as labeled.
### Use environment:
Explorer Mini is intended to be used in everyday life, in safe environments where the child is. Typical environments are at the clinical training site, at home, at preschool, at friends and extended family. The design of the device is small and compact, making it possible for the child to drive the device in these environments.
Like any other infant or toddler at play, Explorer Mini is intended to be used under supervision by an adult, and it is the care-givers responsibility to ensure that the surrounding environment is safe and appropriate for the child to drive.
### Principle of operation:
Explorer Mini includes a base to which the wheels and casters are attached. An adjustable height, vertical column is attached to the base. The driver control (joystick) is integrated at the top of the column and the seating system is attached to the column.
The Explorer Mini speed and direction are controlled via a control system comprising a power module and joystick.
The Explorer Mini has the appearance of a "ride on toy" and weighs between 60% and 90% less than typical power chairs available to this population today. Both appearance and weight are factors critical to acceptance by the intended user population.
Explorer Mini seat is configured with a permanent 360degree support assembly positioned around the upper torso for added user safety and stability. This assembly supports both sitting and standing. An adjustable-position saddle shaped seat is used for sitting. The saddle shape allows for 'straddle' standing or alternately the seat can be removed to open the area for full active standing. The 360degree support assembly adjusts to accommodate and support standing while driving.
The Explorer Mini is powered by two 5 Amp. batteries providing an approximate driving range up to 4.6 miles.
The base provides the propulsion which is derived from two (left and right) front mounted gear-motor wheel assemblies supported by two rear mounted 360degree swiveling casters. Also incorporated in the base are the batteries and control module.
When the user activates the joystick, the controller receives a signal to move the device in the direction the joystick is pointed. Simultaneously, the controller directs the gear-motors to respond appropriately. When the user releases the joystick, the chair decelerates to a stop. The inherent gear ratio holds the device in place like a park brake.
{5}------------------------------------------------
#### Predicate comparison of major technical features:
The Explorer Mini is substantially equivalent to WHEELCHAIR MODEL STS (Innovative Products Inc., K071390, product code: IPL). The Explorer Mini has similar intended use and indications, technological characteristics, and principles of operation. The differences between the Explorer Mini and the predicate device are:
- The total weight of the of the device is considerably less at 40 pounds. This make the Explorer Mini easier to handle for secondary users.
- The speed of the Explorer Mini is 1.5mph while the Wheelchair Model STS (K071390) has a speed of up to 5mph. The speed is set to correlate to the speed a child within the intended age has when or running slowly. The low speed also ensures that the caregiver can keep up with child. This ensures that Explorer Mini has a safer speed for the pediatric end user as well as care giver.
The minor technological differences between Explorer Mini and its Predicate device Wheelchair Model STS raise no new issues of safety or effectiveness. Performance data demonstrates that the Explorer Mini is as safe and effective as the Wheelchair Model STS. Thus, the Explorer Mini is substantially equivalent.
Table 1 provides a comparison of the predicate device (Wheelchair Model STS) and the Explorer Mini in order to demonstrate substantial equivalence.
| Feature | Predicate Device: | Subject Device: | Comparison |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Wheelchair Model STS | Explorer Mini | |
| 510(k) number | K071390 | K190682 | N/A |
| Intended Use | The intended use of the model<br>STS pediatric powered<br>wheelchair is to provide<br>mobility to children and<br>young adults, weighting up to<br>150 pounds, with the ability to<br>place themselves in a sitting<br>position in the wheelchair and<br>have the capacity to operate a<br>standard joy stick hand<br>control. | The Explorer Mini is<br>a pediatric powered<br>wheelchair with the<br>intention to provide<br>mobility to pediatric<br>users weighing up to<br>35 pounds and<br>maximum length of<br>up to 39 inches tall,<br>between 12-36<br>months of age, who<br>position themselves<br>in a sitting position in<br>the wheelchair and<br>have the capacity to<br>operate a joy stick | Permobil has designed<br>the Explorer Mini<br>specifically for the<br>pediatric users<br>weighing up to 35<br>pounds and<br>maximum length of<br>up to 39 inches tall,<br>between 12-36<br>months of age, which<br>is a narrower scope of<br>the predicate device.<br>The predicate device<br>has a broader target<br>group "…provide<br>mobility to children |
| | | | weighting up to 150<br>pounds...”. Several of<br>the functions in the<br>Explorer Mini are<br>designed to be<br>appropriate for the<br>young target group,<br>such as the lower<br>speed, seating area,<br>table area, etc. |
| Product code | IPL 890.3900 | IPL 890.3900 | Identical |
| User age ranges<br>and maximum<br>weight<br>capacities | Children or young adults<br>weight up to 150 lbs. | 12-36 months -35lbs<br>(16kg) or 39”<br>(100cm) | Differences is that<br>STS does specify<br>upper weight limit but<br>does not specify<br>intended user age<br>population. The<br>intended user of<br>Explorer Mini is<br>described by a lower<br>age limit and a<br>maximum weight and<br>physical size limit.<br>Regardless of user<br>weight or age, no new<br>or increased risks are<br>identified. The<br>Explorer Mini has<br>been subjected to all<br>applicable ISO 7176<br>series tests |
| Device<br>description | STS includes a base to which<br>the wheels and casters are<br>attached with the option of<br>attaching a seat or standing<br>frame. | Explorer Mini<br>powered wheelchair<br>includes a base to<br>which wheels and<br>casters are attached<br>with the option of<br>utilizing the seat or<br>configuring the<br>system for standing | Functionally<br>Equivalent |
| | When the joystick is activated<br>a signal is directed through the<br>power module to the gear<br>motors which propel the chair | When the joystick is<br>activated a signal is<br>directed through the<br>power module to the<br>gear motors which<br>propel the chair | Identical |
| | & control both speed and direction | control both speed and direction | |
| | STS uses a motor lock system<br>for park brakes. A park brake<br>for safety is attached to each<br>gear-motor which is activated<br>when the chair comes to a stop | A park brake is<br>provided by the gear<br>ratio which is<br>designed to prevent<br>movement when no<br>signal is detected. | Functionally<br>Equivalent |
| Total width | 24" | 19" | Functionally<br>Equivalent |
| Total length | 32" | 25" | Functionally<br>Equivalent |
| Total height | 35" (to backrest) | Minimum: 28.7"<br>Maximum 36.6" | Functionally<br>Equivalent |
| Turning<br>diameter | 50" | 42.5" | Functionally<br>Equivalent |
| Speed Range | 0-5 mph | 0-1.5 mph | Appropriate for the<br>intended user group |
| Driving range | 25 miles | 3.4 miles | Appropriate for the<br>intended user group |
| Brake distance-<br>Normal<br>operation<br>(Horizontal-<br>Forward-Max<br>speed) | Information regarding the<br>braking distance for the STS<br>is not available in the labeling | Minimum braking<br>distance from<br>maximum speed<br>11.8" | The lower speed of the<br>Explorer Mini and the<br>gear ratio which is<br>designed to prevent<br>movement when no<br>signal is detected, and<br>thereby ensuring that |
| Time to brake | Information regarding the<br>braking distance for the STS<br>has not been found. The STS<br>User Manual states:<br>The STS is equipped with two<br>powerful brake systems.<br>• Regenerative – uses<br>electricity to rapidly<br>slow the vehicle when<br>the joystick returns to<br>the center/stop<br>position.<br>• Disc park brake –<br>activates mechanically<br>after regenerative<br>braking slows the | A park brake is<br>provided by the gear<br>ratio which is<br>designed to prevent<br>movement when no<br>signal is detected.<br>When the user<br>releases the joystick,<br>the chair immediately<br>decelerates to a stop. | the Explorer Mini<br>stops immediately<br>make the Explorer<br>Mini safer than the<br>STS in regard to<br>braking distance |
| | or when power is<br>removed from the<br>system for any reason<br>The STS uses a motor lock<br>system for park brakes. A park<br>brake for safety is attached to<br>each gear-motor which is<br>activated when the chair<br>comes to a stop. | | |
| Operating<br>surface &<br>environment | The STS is designed to<br>operate both indoors and<br>outdoors. It maneuvers on all<br>terrains except soft sand. | Explorer Mini is<br>designed for indoor<br>use and limited<br>outdoor use on firm,<br>flat and dry surfaces | Explorer Mini is<br>intended to be used on<br>safer<br>surfaces/surroundings.<br>Even though the<br>Explorer Mini is<br>designed for limited<br>outdoor use it has<br>been subjected to the<br>same stability test<br>requirements as an<br>outdoor chair. |
| Manageable<br>gradient | 5° | 6° | Functionally<br>Equivalent |
| Max. curb height<br>Obstacle<br>Climbing Ability | 2.5" | 1.0" | Appropriate for the<br>intended use<br>environment |
| Max. chair<br>weight | 160 lbs. | 52 lbs. | Appropriate for the<br>intended user group |
| Frame material,<br>frame tube size | The STS is constructed<br>primarily of steel tubing. The<br>base has a steel frame with a<br>molded Plastic Overlay. The<br>seat and standing frame are<br>constructed of steel tubing. A<br>version of the standing system<br>uses a wooden backboard. | The Explorer Mini<br>base is constructed<br>with a steel frame<br>with molded Plastic<br>Overlays. The tower<br>is constructed with<br>aluminum extrusions<br>while the seat and<br>table are a<br>combination of<br>aluminum and steel<br>components with<br>molded plastic<br>overlays. The back<br>and lateral arms are<br>molded plastic | There are differences<br>in the materials used<br>but both products are<br>appropriately<br>constructed for the<br>intended user and the<br>intended use<br>environment and has<br>been tested according<br>to the requirements<br>and to ensure safety of<br>the device. The<br>Explorer Mini has<br>been tested per ISO<br>7176 test<br>requirements. No new<br>risk identified. Both |
| | | | Mini are appropriately<br>constructed for the<br>intended use |
| Controller | Controller 24V PG | Controller 24V PG nVR2 | Functionally Equivalent |
| Joystick mounting | Left or right mounted | Central mounted | Functionally Equivalent |
| Batteries | 2 x 28 AH Gel cell batteries | 2 x 12 V AGM lead-acid batteries in series | Functionally Equivalent |
| Stand-up feature | Power to go from sit to stand | Manual adjustment to go from sit to stand | Functionally Equivalent |
#### Table 1 Comparison to predicate device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "permobil" in blue font. The font is sans-serif and the letters are all lowercase. The word is slightly italicized, and the "o" is smaller than the other letters. The background is white.
## Non-Clinical Testing:
The Explorer Mini has been tested according to ISO 7176 evaluates the strength, stability (static and dynamic), and durability of the powered wheelchair. Tests are performed on the whole wheelchair to determine the performance, safety, and effectiveness of the device. All performed testing revealed that the Explorer Mini successfully passed all requirements and was found to be substantially equivalent to the predicate device mentioned in this submission.
The predicate device, the STS device, was tested according to FDA Recognized standards. The Explorer Mini was tested to the current recognized standards; ISO 7176 Series.
| Recognized standard | Standard name | Outcome |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| ISO 7176-1:2014<br>[ANSI/RESNA WC-1/1] | Wheelchairs - Part 1: Determination of static stability | Pass |
| ISO 7176-2:2001<br>[ANSI/RESNA WC-1/1] | Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs | Pass |
| ISO 7176-3:2012<br>[ANSI/RESNA WC-2/3] | Wheelchairs - Part 3: Determination of effectiveness of brakes | Pass |
| ISO 7176-4:2008<br>[ANSI/RESNA WC-2/4] | Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range | Pass |
| ISO 7176-5:2008<br>[ANSI/RESNA WC-1/5] | Wheelchairs - Part 5: Determination of overall dimensions, mass and maneuvering space | Pass |
| ISO 7176-6:2001<br>[ANSI/RESNA WC-2/6] | Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs | Pass |
| ISO 7176-7:1998<br>[ANSI/RESNA WC-<br>1/7] | Wheelchairs - Part 7: Measurement of seating and<br>wheel dimensions | Pass |
| ISO 7176-8:2014<br>[ANSI/RESNA WC-<br>1/8] | Wheelchairs - Part 8: Requirements and test<br>methods for static, impact and fatigue strengths | Partial<br>conformance<br>provided<br>alternative<br>evidence to fulfill<br>requirement |
| ISO 7176-9:2009<br>[ANSI/RESNA WC-<br>2/9] | Wheelchairs - Part 9: Climatic tests for electric<br>wheelchairs | Pass |
| ISO 7176-11:2012 | Wheelchairs - Part 11: Test dummies | N/A - This<br>standard sets<br>requirements for<br>testing |
| ISO 7176-10:2008<br>[ANSI/RESNA WC-<br>2/10] | Wheelchairs - Part 10: Determination of obstacle-<br>climbing ability of electrically powered<br>wheelchairs | Pass |
| ISO 7176-13:1989 | Wheelchairs - Part 13: Determination of<br>coefficient of friction of test surfaces | N/A - This<br>standard sets<br>requirements for<br>testing |
| ISO 7176-14:2008<br>[ANSI/RESNA WC-<br>2/14] | Wheelchairs - Part 14: Power and control systems<br>for electrically powered wheelchairs and scooters -<br>Requirements and test methods | Pass |
| ISO 7176-15:1996<br>[ANSI/RESNA WC-<br>1/15] | Wheelchairs - Part 15: Requirements for<br>information disclosure, documentation and<br>labeling | Pass |
| ISO 7176-16:2012<br>[ANSI/RESNA WC-<br>1/16] | Wheelchairs - Part 16: Resistance to ignition of<br>postural support devices | Pass |
| ISO 7176-21:2009<br>[ANSI/RESNA WC-<br>2/21] | Wheelchairs - Part 21: Requirements and test<br>methods for electromagnetic compatibility of<br>electrically powered wheelchairs and scooters, and<br>battery chargers | Pass |
| ISO 7176-25:2013 | Wheelchairs - Part 25: Batteries and chargers for<br>powered wheelchairs | Partial<br>conformance<br>provided<br>alternative<br>evidence to fulfill<br>requirement |
| RESNA WC-1:2009<br>Section 20 | Determination of the Performance of Stand-up<br>Type Wheelchairs | Pass |
| ISO 10993-1 | Biological evaluation of medical devices | Pass |
Table 2 List of recognized standards to which the Explorer Mini complies
{10}------------------------------------------------
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "permobil" in a sans-serif font. The word is written in all lowercase letters and is a dark blue color. The letters are closely spaced together, giving the word a compact appearance.
## Clinical Testing:
The device is designed to provide the compensatory motor skills that work in concert with the sensory skills to bring about postural control, upper extremity stability in this young population where several developmental milestones remain to be met. A usability study was performed on 33 children with mobility impairments justifying a need for a device as the Explorer Mini. A label comprehension study of the User's manual was performed on 15 physiotherapists/occupational therapists and 15 parents. The human factor validation testing revealed that the Explorer Mini is a safe and effective medical device for the intended user, and all studied use-related risks were deemed acceptable.
### Conclusions:
The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as the legally marketed device. The Explorer Mini and Predicate device, Wheelchair Model STS are substantially equivalent. Explorer Mini has the same general intended use and similar indications, principles of operation, and similar technological characteristics as the previously cleared WHEELCHAIR MODEL STS; Innovative Products Inc
- K071390 ●
- Product code: IPL
- 21CFR 380.3900
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.