The StealthStation® System, with Synergy® Cranial software is designed as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The StealthStation® System, with StealthStation® Cranial software is designed as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.
Device Story
StealthStation System is an image-guided surgical navigation platform. It integrates diagnostic images (CT, MR, Nuclear Medicine) to track surgical instruments relative to patient anatomy. The system uses optical (infrared) or electromagnetic localization to provide real-time guidance. Surgeons use the software for preoperative planning, target selection, and intraoperative visualization. The system displays 2D/3D anatomical views, trajectory guidance, and instrument position. It is used in neurosurgical settings (OR) by surgeons to improve precision in procedures like biopsies, resections, and lead placements. The output aids clinical decision-making by providing visual confirmation of instrument location relative to planned targets, potentially reducing surgical invasiveness and improving accuracy.
Clinical Evidence
Bench testing only. Performance validated using anatomically representative phantoms under worst-case configurations. Primary endpoints were 3D positional accuracy and trajectory angle accuracy. Results demonstrated mean positional error ≤ 2.0 mm and mean trajectory angle error ≤ 2.0 degrees. Software verification and validation performed in laboratory and simulated use settings.
Technological Characteristics
Software-based surgical navigation system. Localization via optical (infrared) or electromagnetic tracking. Interfaces with imaging modalities (CT, MR, Nuclear Medicine) via DICOM. Supports various registration methods (PointMerge, Tracer, Touch-N-Go, StealthAiR, O-arm, Stereotactic Localizer, StarFix). Programming language: C++. Connectivity: Network, CD, DVD, USB.
Indications for Use
Indicated for patients undergoing open or percutaneous neurosurgical procedures where reference to rigid anatomical structures can be identified relative to images. Includes cranial biopsies, tumor resections, craniotomies/craniectomies, skull base procedures, transsphenoidal procedures, thalamotomies/pallidotomies, pituitary tumor removal, CSF leak repair, and ventricular catheter/lead/electrode placement.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
StealthStation System with Synergy Cranial v2.2.7 software (K150216)
StealthStation System with StealthStation Cranial v3.0 Software (K153660)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 31, 2019
Medtronic Navigation Inc. Elizabeth Waite Principal Regulatory Affairs Specialist 826 Coal Creek Circle Louisville. Colorado 80027
Re: K190672
Trade/Device Name: StealthStation Synergy Cranial S7 Software v.2.2.8, StealthStation Cranial Software v3.1.1 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: March 14, 2019 Received: March 15, 2019
Dear Elizabeth Waite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190672
Device Name Synergy® Cranial Software
### Indications for Use (Describe)
The Synergy® Cranial software is surgical navigation software that, when used with the StealthStation® System as a planning and intraoperative guidance system, is intended to aid in precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The system is indicated for any medical condition in which reference to a rigid anatomical structure can be identified relative to images of the anatomy.
This can include, but is not limited to, the following cranial procedures:
- Cranial Biopsies
- Tumor Resections
- Craniotomies/Craniectomies
- Skull Base Procedures
- Transsphenoidal Procedures
- Thalamotomies/Pallidotomies
- Pituitary Tumor Removal
- CSF Leak Repair
- Pediatric Catheter Shunt Placement
- General Catheter Shunt Placement
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190672
Device Name StealthStation Cranial Software
### Indications for Use (Describe)
The StealthStation® System, with StealthStation® Cranial software, is intended to aid in precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The system is indicated for any medical condition in which reference to a rigid anatomical structure can be identified relative to images of the anatomy.
This can include, but is not limited to, the following cranial procedures (including stereotactic frame-based and stereotactic frame alternatives-based procedures):
- · Cranial biopsies (including stereotactic)
- · Deep brain stimulation (DBS) lead placement
- · Depth electrode placement
- Tumor resections
- · Craniotomies/Craniectomies
- Skull Base Procedures
- · Transsphenoidal Procedures
- Thalamotomies/Pallidotomies
- · Pituitary Tumor Removal
- · CSF leak repair
- · Pediatric Ventricular Catheter Placement
- · General Ventricular Catheter Placement
Type of Use (Select one or both, as applicable)
| <div style="display:inline-block;"><input checked="true" type="checkbox"/></div> | |
|----------------------------------------------------------------------------------|--|
| <div style="display:inline-block;"><input type="checkbox"/></div> | |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# April 18, 2019
- I. Company: Medtronic Navigation, Inc. 826 Coal Creek Circle Louisville, Colorado 80027 USA Telephone Number: 720-890-3200 Fax Number: 720-890-3500
- Contact: K. Elizabeth Waite (Primary) Principal Regulatory Affairs Specialist Telephone Number: 720-890-2182 Fax Number: 720-890-3500
Rishi Sinha (Alternate) Regulatory Affairs Manager Telephone Number: 720-890-2485 Fax Number: 720-890-3500
- II. Proprietary Trade Name: StealthStation® Cranial Software
- III. Classification Name: Stereotaxic Instrument (21 CFR 882.4560)
- IV. Classification: Class II, Stereotaxic Instrument
- V. Product Codes: HAW
- VI. Predicates: Medtronic Navigation, Inc. manufactured software;
- K150216, StealthStation System with Synergy Cranial v2.2.7 software ●
- K153660, StealthStation System with StealthStation Cranial v3.0 Software .
#### VII. Product Description
The StealthStation® System, with StealthStation Cranial software helps quide surgeons during cranial surgical procedures such as biopsies, tumor resections, and shunt and lead placements. The StealthStation® Cranial software works in conjunction with an Image Guided System (IGS) which consists of clinical software, surgical instruments, a referencing system and platform/computer hardware. Image quidance, also called navigation, tracks the position of instruments in relation to the surqical anatomy and identifies this position on diagnostic or intraoperative images of the patient. StealthStation® Cranial
{5}------------------------------------------------
software functionality is described in terms of its feature sets which are categorized as imaging modalities, registration, planning, interfaces with medical devices, and views. Feature sets include functionality that contributes to clinical decision making and are necessary to achieve system performance.
#### VIII. Indications for Use
## Cranial Software v2.2.8
The StealthStation System, with Synerqy Cranial software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The system is indicated for any medical condition in which reference to a rigid anatomical structure can be identified relative to images of the anatomy.
This can include, but is not limited to, the following cranial procedures:
- Cranial Biopsies
- Tumor Resections
- Craniotomies/Craniectomies
- Skull Base Procedures
- Transsphenoidal Procedures
- Thalamotomies/Pallidotomies
- Pituitary Tumor Removal
- CSF Leak Repair
- Pediatric Catheter Shunt Placement
- General Catheter Shunt Placement
## Cranial Software v3.1.1
The StealthStation® System, with StealthStation® Cranial software, is intended to aid in precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The system is indicated for any medical condition in which reference to a rigid anatomical structure can be identified relative to images of the anatomy.
This can include, but is not limited to, the following cranial procedures (including stereotactic frame-based and stereotactic frame alternatives-based procedures):
- · Cranial biopsies (including stereotactic)
- · Deep brain stimulation (DBS) lead placement
- Depth electrode placement
- · Tumor resections
- · Craniotomies/Craniectomies
- · Skull Base Procedures
{6}------------------------------------------------
- Transsphenoidal Procedures
- Thalamotomies/Pallidotomies
- Pituitary Tumor Removal
- CSF leak repair
- Pediatric Ventricular Catheter Placement
- General Ventricular Catheter Placement
#### IX. Summary of the Technological Characteristics
## StealthStation Cranial Software v2.2.8 as compared to Predicate Device
| | Subject Device | Predicate Device |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | StealthStation System with Synergy Cranial | K150216 StealthStation System with |
| Item | v2.2.8 Software | Synergy Cranial v2.2.7 Software |
| Intended Use | The StealthStation® System, with<br>Synergy® Cranial software is designed as<br>an aid for precisely locating anatomical<br>structures in either open or percutaneous<br>neurosurgical procedures. | The StealthStation® System, with<br>Synergy® Cranial software is designed as<br>an aid for precisely locating anatomical<br>structures in either open or percutaneous<br>neurosurgical procedures. |
| Indications for Use | The StealthStation System, with Synergy<br>Cranial software, is intended as an aid for<br>precisely locating anatomical structures in<br>either open or percutaneous neurosurgical<br>procedures. The system is indicated for<br>any medical condition in which reference to<br>a rigid anatomical structure can be<br>identified relative to images of the<br>anatomy.<br>This can include, but is not limited to, the<br>following cranial procedures:<br>- Cranial Biopsies<br>- Tumor Resections<br>- Craniotomies/Craniectomies<br>- Skull Base Procedures<br>- Transsphenoidal Procedures<br>- Thalamotomies/Pallidotomies<br>- Pituitary Tumor Removal<br>- CSF Leak Repair<br>- Pediatric Catheter Shunt Placement<br>- General Catheter Shunt Placement | The StealthStation System, with Synergy<br>Cranial software, is intended as an aid for<br>precisely locating anatomical structures in<br>either open or percutaneous neurosurgical<br>procedures. The system is indicated for<br>any medical condition in which reference to<br>a rigid anatomical structure can be<br>identified relative to images of the<br>anatomy.<br>This can include, but is not limited to, the<br>following cranial procedures:<br>- Cranial Biopsies<br>- Tumor Resections<br>- Craniotomies/Craniectomies<br>- Skull Base Procedures<br>- Transsphenoidal Procedures<br>- Thalamotomies/Pallidotomies<br>- Pituitary Tumor Removal<br>- CSF Leak Repair<br>- Pediatric Catheter Shunt Placement<br>- General Catheter Shunt Placement |
| System Accuracy<br>Requirement | Under representative worst-case<br>configuration, the StealthStation® System<br>with Synergy Cranial Software, has<br>demonstrated performance in 3D positional<br>accuracy with a mean error ≤ 2.0 mm and<br>in trajectory angle accuracy with a mean<br>error ≤ 2.0 degrees.<br>Specific Mean Accuracy Values | Under representative worst-case<br>configuration, the StealthStation® System<br>with Synergy® Cranial Software, has<br>demonstrated performance in 3D positional<br>accuracy with a mean error ≤ 2.0 mm and<br>in trajectory angle accuracy with a mean<br>error ≤ 2.0 degrees.<br>Specific Mean Accuracy Values |
| | Positional Error - 0.70 mm | Position Error - ≤ 2.00 mm |
| Item | Subject Device<br>StealthStation System with Synergy Cranial<br>v2.2.8 Software | Predicate Device<br>K150216 StealthStation System with<br>Synergy Cranial v2.2.7 Software |
| | Trajectory Error - 0.46 degrees | Trajectory Error - ≤ 2.0° |
| Imaging Modalities | X-Ray based,<br>MR based<br>Nuclear Medicine based | X-Ray based,<br>MR based<br>Nuclear Medicine based |
| Registration<br>Features | Exam-to-Exam Registration: Identity Merge<br>Registration, Manual Merge Registration<br>and Automatic Merge Registration.<br>Patient Registration: PointMerge<br>registration, Tracer registration, Touch-N-<br>Go registration, StealthAiR registration and<br>O-arm registration | Exam-to-Exam Registration: Identity Merge<br>Registration, Manual Merge Registration<br>and Automatic Merge Registration.<br>Patient Registration: PointMerge<br>registration, Tracer registration, Touch-N-<br>Go registration, StealthAiR registration and<br>O-arm registration |
| Planning Features | Plan Entry and Target Selection<br>3D Model Building<br>Advanced Visualization | Plan Entry and Target Selection<br>3D Model Building<br>Advanced Visualization |
| Medical Device<br>Interfaces | Microscope Navigation: Zeiss, Leica<br>Ultrasound Navigation: Aloka and Sonosite<br>Medtronic O-arm | Microscope Navigation: Zeiss, Leica<br>Ultrasound Navigation: Aloka and Sonosite<br>Medtronic O-arm |
| View (Display)<br>Features | Ultrasound Video In, Ultrasound Overlay,<br>3D, 2D Anatomic Orthogonal, Trajectory 1<br>and 2, Target Guidance, Trajectory<br>Guidance, Probes Eye, Look Ahead,<br>Microscope Injection, Video Input | Ultrasound Video In, Ultrasound Overlay,<br>3D, 2D Anatomic Orthogonal, Trajectory 1<br>and 2, Target Guidance, Trajectory<br>Guidance, Probes Eye, Look Ahead,<br>Microscope Injection, Video Input |
| Software Interface<br>(GUI) | Blue style with chronological next/back<br>task flow at the top of the screen. Image<br>controls on the left. Planning information<br>on the right. | Blue style with chronological next/back<br>task flow at the top of the screen. Image<br>controls on the left. Planning information<br>on the right. |
| Programming<br>Language | C++ | C++ |
| Scanner Interface<br>Technology (to<br>imaging devices) | Network Connectivity<br>CD, DVD, USB<br>DICOM or Stealth format Import<br>Export in Stealth format | Network Connectivity<br>CD, DVD, USB<br>DICOM or Stealth format Import<br>Export in Stealth format |
| Localization<br>Technology | Optical (infra-red)<br>Electromagnetic | Optical (infra-red)<br>Electromagnetic |
{7}------------------------------------------------
# StealthStation Cranial Software v3.1.1 as compared to Predicate Device
| Item | Subject Device | Predicate Device |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | StealthStation System with Cranial v3.1.1<br>Software<br>The StealthStation® System, with<br>StealthStation® Cranial software is<br>designed as an aid for locating anatomical<br>structures in either open or percutaneous<br>neurosurgical procedures. | K153660 StealthStation System with<br>StealthStation Cranial v3.0 Software<br>The StealthStation® System, with<br>StealthStation® Cranial software is<br>designed as an aid for locating anatomical<br>structures in either open or percutaneous<br>neurosurgical procedures. |
| Item | Subject Device | Predicate Device |
| | StealthStation System with Cranial v3.1.1 | K153660 StealthStation System with |
| | Software | StealthStation Cranial v3.0 Software |
| Indications for Use | The StealthStation® System, with<br>StealthStation® Cranial software, is<br>intended to aid in locating anatomical<br>structures in either open or percutaneous<br>neurosurgical procedures. The system is<br>indicated for any medical condition in<br>which reference to a rigid anatomical<br>structure can be identified relative to<br>images of the anatomy. | The StealthStation® System, with<br>StealthStation® Cranial software, is<br>intended to aid in locating anatomical<br>structures in either open or percutaneous<br>neurosurgical procedures. The system is<br>indicated for any medical condition in<br>which reference to a rigid anatomical<br>structure can be identified relative to<br>images of the anatomy. |
| | This can include, but is not limited to, the<br>following cranial procedures (including<br>stereotactic frame-based and stereotactic<br>frame alternatives-based procedures):<br>- Cranial Biopsies<br>- Deep brain stimulation (DBS)<br>lead placement<br>- Depth electrode placement<br>- Tumor Resections<br>- Craniotomies/Craniectomies<br>- Skull Base Procedures<br>- Transsphenoidal Procedures<br>- Thalamotomies/Pallidotomies<br>- Pituitary Tumor Removal<br>- CSF Leak Repair<br>- Pediatric Ventricular Catheter Placement<br>- General Ventricular Catheter Placement<br>The user should consult the "Navigational<br>Accuracy" section of the User Manual to | This can include, but is not limited to, the<br>following cranial procedures (including<br>stereotactic frame-based and stereotactic<br>frame alternatives-based procedures):<br>- Cranial Biopsies<br>- Deep brain stimulation (DBS)<br>lead placement<br>- Depth electrode placement<br>- Tumor Resections<br>- Craniotomies/Craniectomies<br>- Skull Base Procedures<br>- Transsphenoidal Procedures<br>- Thalamotomies/Pallidotomies<br>- Pituitary Tumor Removal<br>- CSF Leak Repair<br>- Pediatric Ventricular Catheter Placement<br>- General Ventricular Catheter Placement<br>The user should consult the "Navigational<br>Accuracy" section of the User Manual to |
| | assess if the accuracy of the system is<br>suitable to their needs. | assess if the accuracy of the system is<br>suitable to their needs. |
| System Accuracy<br>Requirement | Under representative worst-case<br>configuration, the StealthStation® System<br>with StealthStation Cranial v3.0 Software,<br>has demonstrated performance in 3D<br>positional accuracy with a mean error ≤<br>2.0 mm and in trajectory angle accuracy<br>with a mean error ≤ 2.0 degrees.<br>Specific Mean Accuracy Values | Under representative worst-case<br>configuration, the StealthStation® System<br>with StealthStation® Cranial v3.0 Software,<br>has demonstrated performance in 3D<br>positional accuracy with a mean error ≤<br>2.0 mm and in trajectory angle accuracy<br>with a mean error ≤ 2.0 degrees.<br>Specific Mean Accuracy Values |
| | Positional Error – 1.16 mm<br>Trajectory Error – 0.41 degrees | Position Error - ≤ 2.00 mm<br>Trajectory Error - ≤ 2.0° |
| Imaging Modalities | X-Ray based,<br>MR based<br>Nuclear Medicine based | X-Ray based,<br>MR based<br>Nuclear Medicine based |
| | Subject Device | Predicate Device |
| | StealthStation System with Cranial v3.1.1<br>Software | K153660 StealthStation System with<br>StealthStation Cranial v3.0 Software |
| Item | | |
| Registration<br>Features | Exam-to-Exam Registration: Identity Merge<br>Registration, Manual Merge Registration<br>and Automatic Merge Registration.<br><br>Patient Registration: PointMerge<br>registration, Tracer registration, Touch-N-<br>Go registration, StealthAiR registration, O-<br>arm registration, Stereotactic Localizer<br>Registration and StarFix Bone Anchor<br>Registration | Exam-to-Exam Registration: Identity Merge<br>Registration, Manual Merge Registration<br>and Automatic Merge Registration.<br><br>Patient Registration: PointMerge<br>registration, Tracer registration, Touch-N-<br>Go registration, StealthAiR registration, O-<br>arm registration, Stereotactic Localizer<br>Registration and StarFix Bone Anchor<br>Registration |
| Planning Features | Plan Entry and Target Selection<br>3D Model Building<br>Advanced Visualization<br>Create Patient Based Anatomical<br>Coordinate Space<br>Stereotactic Frame Settings<br>Brain Atlas: Schaltenbrand-Wahren Atlas<br>with Talairach Grid<br>STarFix Designer<br>Annotations | Plan Entry and Target Selection<br>3D Model Building<br>Advanced Visualization<br>Create Patient Based Anatomical<br>Coordinate Space<br>Stereotactic Frame Settings<br>Brain Atlas: Schaltenbrand-Wahren Atlas<br>with Talairach Grid<br>STarFix Designer<br>Annotations |
| Medical Device<br>Interfaces | Microscope Navigation: Zeiss, Leica<br>Ultrasound Navigation: Aloka and Sonosite<br>Medtronic O-arm<br>Stereotactic Frame Systems: Fischer ZD,<br>Fischer RM, Integra CRW and Leksell<br>Nexframe® Stereotactic System<br>STarFix™Platform | Microscope Navigation: Zeiss, Leica<br>Ultrasound Navigation: Aloka and Sonosite<br>Medtronic O-arm<br>Stereotactic Frame Systems: Fischer ZD,<br>Fischer RM, Integra CRW and Leksell<br>Nexframe® Stereotactic System<br>STarFix™Platform |
| View (Display)<br>Features | Ultrasound Video In, Ultrasound Overlay,<br>3D, 2D Anatomic Orthogonal, Trajectory 1<br>and 2, Target Guidance, Trajectory<br>Guidance, Probes Eye, Look Ahead,<br>Microscope Injection, Video Input | Ultrasound Video In, Ultrasound Overlay,<br>3D, 2D Anatomic Orthogonal, Trajectory 1<br>and 2, Target Guidance, Trajectory<br>Guidance, Probes Eye, Look Ahead,<br>Microscope Injection, Video Input |
| Software Interface<br>(GUI) | Blue style with chronological next/back<br>task flow at the top of the screen. Image<br>controls on the left. Planning information<br>on the right. | Blue style with chronological next/back<br>task flow at the top of the screen. Image<br>controls on the left. Planning information<br>on the right. |
| Programming<br>Language | C++ | C++ |
| Scanner Interface<br>Technology (to<br>imaging devices) | Network Connectivity<br>CD, DVD, USB<br>DICOM Import<br>DICOM Export | Network Connectivity<br>CD, DVD, USB<br>DICOM Import<br>DICOM Export |
| Localization<br>Technology | Optical (infra-red)<br>Electromagnetic | Optical (infra-red)<br>Electromagnetic |
{8}------------------------------------------------
{9}------------------------------------------------
### X. Identification of Legally Marketing Devices
StealthStation® System with Synergy Cranial Software (K150216) StealthStation® System with Cranial Software (K150663)
{10}------------------------------------------------
#### XI. Discussion of the Performance Testing
The following table summarizes the testing conducted on the StealthStation® System with StealthStation® Cranial Software:
## Description
Under representative worst-case configuration, the StealthStation® System with StealthStation® Cranial Software, has demonstrated performance in 3D positional accuracy with a mean error ≤ 2.0 mm and in traiectory angle accuracy with a mean error ≤ 2.0 degrees. This performance was determined using anatomically representative phantoms and utilizing a subset of system components and features that represent the worst-case combinations of all potential system components.
The test configurations included CT images with slice spacing and thickness ranging between 0.6 mm to 1.25 mm and T1-weighted MR images with slice spacing and thickness ranging between 1.0 mm to 3.0 mm. In the imaging protocol, we recommend slice spacing and thickness for CT and MR imaging to be 1.0 mm or less.
Software verification and validation testing for each requirement specification. Design verification and validation was performed using the StealthStation Cranial software in laboratory and simulated use settings. The results support the safety of the device and demonstrate that the software should perform as intended in the specified use conditions.
System integration performance testing for cranial surgical procedures using anatomical phantoms.
The following table summarizes the quality assurance measures that were applied during development of the software component of the system:
## Description
Software Development Life Cycle
Software Risk Assessment
Software Configuration Management and Version Control
#### XII. Conclusions
The StealthStation Cranial software has been shown through testing and comparison to be substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.