K190667 · Paragon Medical · KCT · Jun 13, 2019 · General Hospital
Device Facts
Record ID
K190667
Device Name
Sonopet iQ Sterilization Tray
Applicant
Paragon Medical
Product Code
KCT · General Hospital
Decision Date
Jun 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Sonopet iQ Sterilization Tray is intended to enclose and organize Stryker Sonopet iQ reusable medical instruments during processing at a healthcare facility. The trays are not intended to maintain sterility; they are intended to be used in conjunction with a legally marketed, FDA-cleared sterilization wrap or rigid container in order to maintain sterility of the enclosed devices. The Sonopet iQ Sterilization Tray is intended for use in any of the following standard sterilization cycles: [Table of cycles omitted for brevity, see source].
Device Story
Reusable sterilization tray; stainless steel base and lid with perforated pattern for steam penetration; includes stainless steel and silicone brackets for device fixation; features two retractable handles and two stainless steel latches. Used in healthcare facilities to organize and protect Stryker Sonopet iQ instruments (handpiece, cleaning wire, torque wrench) during pre-vacuum steam sterilization. Tray is non-sterile upon receipt; requires sterilization before use. Does not maintain sterility independently; requires external sterilization wrap or rigid container. Benefits include organized, secure processing of instruments, ensuring compatibility with standard autoclave cycles and maintaining structural integrity over repeated use cycles.
Clinical Evidence
No clinical data. Bench testing only. Validations included sterilization efficacy (Geobacillus stearothermophilus BI), dry time, manual/automated cleaning (AAMI TIR 30:2011), biocompatibility (ISO 10993-5 MEM Elution), and design verification (handle strength, latch force, stacking, weight) per ANSI/AAMI ST77 and DIN EN 868-8. Life cycle testing confirmed integrity after repeated processing cycles.
Technological Characteristics
Materials: Stainless steel, silicone, thermoplastic polymers. Design: Perforated base/lid, retractable handles, latching mechanism. Sterilization: Pre-vacuum steam (132°C/4 min or 134°C/3 min). Standards: ANSI/AAMI ST77, DIN EN 868-8, AAMI TIR 30:2011, ISO 10993-5, ANSI/AAMI/ISO 17665-1. Connectivity: None.
Indications for Use
Indicated for enclosing and organizing Stryker Sonopet iQ reusable medical instruments during processing at a healthcare facility. Not intended to maintain sterility; must be used with FDA-cleared sterilization wrap or rigid container. Intended for use in pre-vacuum steam sterilization cycles.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Advanced Imaging Modality (AIM) Sterilization Tray, Endoscope and Camera Sterilization Tray and Camera Sterilization Tray (K152951)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 13, 2019
Paragon Medical C/o Rebecca Walker Regulatory Affairs Manager 8 Matchett Dr Pierceton, Indiana 46562
Re: K190667
Trade/Device Name: Sonopet iQ Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: May 7, 2019 Received: May 8, 2019
Dear Rebecca Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth F. Claverie-Williams, MS Assistant Director. THT4B2: Disinfection Reprocessing and Personal Protection Acting Assistant Director, THT4B1: Sterility Devices DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K190667
Device Name
### Sonopet iQ Sterilization Tray
Indications for Use (Describe)
The Sonopet iQ Sterilization Tray is intended to enclose and organize Stryker Sonopet iQ reusable medical instruments during processing at a healthcare facility. The trays are not intended to maintain sterility; they are intended to be used in conjunction with a legally marketed, FDA-cleared sterilization wrap or rigid container in order to maintain sterility of the enclosed devices. The Sonopet iQ Sterilization Tray is intended for use in any of the following standard sterilization cycles:
| Method | Parameter | Cycle | |
|---------------------|--------------------|-----------------------------------|-----------------------------------|
| | | Cycle 1 | Cycle 2 |
| Pre-vacuum<br>Steam | Enclosure | Double Wrap or<br>Rigid Container | Double Wrap or<br>Rigid Container |
| | Temperature | 132°C<br>(270°F) | 134°C<br>(273°F) |
| | Sterilization Time | 4 minutes | 3 minutes |
| | Dry Time | 30 minutes | 30 minutes |
The maximum product load for the tray is as follows:
| Tray Description | Catalog Number | Maximum Tray<br>Product Load | Maximum Tray<br>Weight |
|-------------------------------|----------------|---------------------------------------------------------------------|------------------------|
| Sonopet iQ Sterilization Tray | 5500-800-278 | Handpiece (Qty 1)<br>Cleaning Wire (Qty 1)<br>Torque Wrench (Qty 1) | 6.85 lb |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
| X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Paragon Medical. The logo consists of a red pentagon with a white star in the center, and the words "PARAGON MEDICAL" in black text to the right of the pentagon. There are two black lines above and below the text.
## 510(k) Summary for Sonopet iQ Sterilization Tray, K190667
In accordance with 21 CFR 807.92, the following 510(k) summary is provided:
K152951
| 1. Applicant Information: | Paragon Medical<br>8 Matchett Drive<br>Pierceton, IN 46562 USA<br>Phone: 574-594-2140<br>Fax: 574-594-2154 |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| 2. Correspondent Contact Information: | Jill Bouillon<br>Quality Manager<br>Jill.bouillon@nninc.com<br>Phone: 574-594-2140 ext. 10403 |
| 3. Date Prepared: | June 12, 2019 |
| 4. Trade Name: | Sonopet iQ Sterilization Tray |
| 5. Common Name: | Sterilization Tray |
| 6. Classification Name: | Sterilization wrap containers, trays, cassettes and other accessories<br>(21 CFR 880.6850, Product Code KCT) |
| 7. Predicate Device: | Advanced Imaging Modality (AIM) Sterilization Tray, Endoscope<br>and Camera Sterilization Tray and Camera Sterilization Tray. |
- 8. Device Description: The Sonopet iQ Sterilization Tray is a reusable device manufactured of perforated stainless steel for the base and lid to allow for sterilization of the enclosed devices. The tray is used to enclose and organize Stryker Sonopet iQ reusable medical instruments during processing at a healthcare facility. The tray is compatible with pre-vacuum steam sterilization. The tray is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, FDA-cleared sterilization wrap or rigid container in order to maintain sterility of the enclosed devices. The tray is provided in a non-sterile condition, and must be sterilized prior to use.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Paragon Medical. The logo consists of a red pentagon with a white star in the center. To the right of the pentagon is the text "PARAGON" in black, with the text "MEDICAL" below it, also in black. There are two black lines above and below the text.
- 9. Indications for Use: The Sonopet iQ Sterilization Tray is intended to enclose and organize Stryker Sonopet iQ reusable medical instruments during processing at a healthcare facility. The trays are not intended to maintain sterility; they are intended to be used in conjunction with a legally marketed, FDA-cleared sterilization wrap or rigid container in order to maintain sterility of the enclosed devices. The Sonopet iQ Sterilization Tray is intended for use in any of the following standard sterilization cycles:
| Method | Parameter | Cycle | |
|---------------------|--------------------|-----------------------------------|-----------------------------------|
| | | Cycle 1 | Cycle 2 |
| Pre-vacuum<br>Steam | Enclosure | Double Wrap or<br>Rigid Container | Double Wrap or<br>Rigid Container |
| | | | |
| | Temperature | 132°C<br>(270°F) | 134°C<br>(273°F) |
| | | | |
| | Sterilization Time | 4 minutes | 3 minutes |
| | Dry Time | 30 minutes | 30 minutes |
The maximum product load for the tray is as follows:
| Tray Description | Catalog Number | Maximum Tray Product Load | Maximum Tray Weight |
|-------------------------------|----------------|---------------------------------------------------------------------|---------------------|
| Sonopet iQ Sterilization Tray | 5500-800-278 | Handpiece (Qty 1)<br>Cleaning Wire (Qty 1)<br>Torque Wrench (Qty 1) | 6.85 lb |
- 10. Technological Characteristics: The following pages contain a detailed comparison of the subject device to the predicate comparison of the technological (Table 1) and performance (Table 2) characteristics of the subject and predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Paragon Medical. The logo consists of a red pentagon with a white star in the center on the left. To the right of the pentagon are the words "PARAGON" and "MEDICAL" stacked on top of each other, with a black line above and below the words.
| Table 1: Comparison of Technological Characteristics | | | | | |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|-------------------------------------------|---------------------------|
| | Predicate Device (K152951) | Subject Device (K190667) | | | |
| Device Name | Advanced Imaging Modality (AIM) Sterilization Tray, Endoscope and<br>Camera Sterilization Tray and Camera Sterilization Tray | Sonopet iQ Sterilization Tray | | | |
| Manufacturer | Paragon Medical<br>8 Matchett Dr<br>Pierceton, IN 46562 | Paragon Medical<br>8 Matchett Dr<br>Pierceton, IN 46562 | | | |
| 510(k)<br>Number | K152951 | K190667 | | | |
| Product Code | KCT | KCT | | | |
| Design | The trays (stainless steel base with a removable stainless steel lid)<br>have an evenly distributed perforated steam hole pattern to facilitate<br>sterilant/steam penetration.<br>The base includes silicone brackets for Stryker endoscopy device<br>fixation within the tray and two (2) latches to affix the lid to the base.<br>Handles on the sides of the base facilitate carrying of the tray. The<br>trays are reusable and provided in a non-sterile condition. | The trays (stainless steel base with a removable stainless steel lid)<br>has an evenly distributed perforated steam hole pattern to facilitate<br>sterilant/steam penetration.<br>The base includes stainless steel and silicone brackets for device<br>fixation within the tray and two (2) stainless steel latches to affix<br>the lid to the base. Two retractable handles on the sides of the base<br>facilitate carrying of the tray. The trays are reusable and provided in<br>a non-sterile condition. | | | |
| Materials | Stainless Steel, Silicone | Stainless Steel, Silicone, Thermoplastic Polymers | | | |
| Indications for<br>Use | The Advanced Imaging Modality (AIM) Sterilization Tray, Endoscope<br>and Camera Sterilization Tray, and Camera Sterilization Tray are<br>intended to enclose and organize Stryker Endoscopy reusable medical<br>instruments during sterilization at a healthcare facility. The trays are<br>optional accessories to the Stryker Endoscopy instruments for which<br>they are designed. The trays are not intended to maintain sterility;<br>they are intended to be used in conjunction with legally marketed,<br>FDA-cleared sterilization wrap in order to maintain sterility of the<br>enclosed devices. The Advanced Imaging Modality (AIM) Sterilization<br>Tray, Endoscope and Camera Sterilization Tray, and Camera<br>Sterilization Tray are intended for use in any of the following standard<br>sterilization machines / cycles: | The Sonopet iQ Sterilization Tray is intended to enclose and<br>organize Stryker Sonopet iQ reusable medical instruments during<br>processing at a healthcare facility. The tray is not intended to<br>maintain sterility; it is intended to be used in conjunction with a<br>legally marketed, FDA-cleared sterilization wrap or rigid container in<br>order to maintain sterility of the enclosed devices. The Sonopet iQ<br>Sterilization Tray is intended for use in any of the following standard<br>sterilization cycles: | | | |
| | Method | Cycle | | | |
| | | | | | |
| | Ethylene<br>Oxide (EtO) | Preconditioning Parameters | | | |
| | | Temperature (°C) 55°C (131°F) | | | |
| | | Chamber Humidity 70% RH | | | |
| | | Vacuum Set Points 1.3 psia | | | |
| | | Time 30 minutes | | | |
| | | Exposure | | | |
| | | Concentration 725 mg/L, 100% EO | | | |
| | | Temperature (°C) 55°C +2°C (131°F +5°F) | | | |
| | | Time 1 hour | | | |
| | | Chamber Humidity (%) 70% RH (50-80%) +5% | | | |
| | | Aeration | | | |
| | | Aeration Time 12 hours | | | |
| | | Temperature 35°C - 54°C (95°F - 129°F) | | | |
| | Pre-vacuum<br>Steam | Cycle 1 | Cycle 2 | | |
| | | Wrapping Double | Double | | |
| | | Temperature 132°C<br>(270°F) | 134°C<br>(273°F) | | |
| | | Sterilization Time 4 minutes | 3 minutes | | |
| | Pre-vacuum<br>Steam | Parameter | Cycle | | |
| | | Enclosure | Cycle 1 Double Wrap<br>or Rigid Container | Cycle 2 Double Wrap<br>or Rigid Container | |
| | | Temperature | 132°C<br>(270°F) | 134°C<br>(273°F) | |
| | | Sterilization Time | 4 minutes | 3 minutes | |
| | | Dry Time | 30 minutes | 30 minutes | |
| The maximum product load for the tray is as follows: | | | | | |
| Catalog<br>Number | Maximum Tray Product<br>Load | Maximum Tray<br>Weight | | | |
| 5500-800-278 | Handpiece (Qty 1)<br>Torque Wrench (Qty 1)<br>Cleaning Wire (Qty 1) | 6.85 lb. | | | |
| | | Dry Time | 30 minutes | 30 minutes | |
| | Gravity | | Cycle 1 | Cycle 2 | |
| | Steam | Wrapping | Double | Double | |
| | | | 132°C | 134°C | |
| | | Temperature | (270°F) | (273°F) | |
| | | Sterilization Time | 15 minutes | 10 minutes | |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.