The MIS Bunion Plating System is intended for fixation of osteotomies and corrective procedures of the hallux and associated disorders such as hallux valgus.
Device Story
MIS Bunion Plating System; metallic bone fixation appliance for hallux valgus correction. Implant inserted via minimally invasive surgical technique into metatarsal intramedullary canal; acts as buttress to hold metatarsal head. Features tapered proximal end with interrupted surfaces for cancellous bone traction; distal head seats against osteotomy surface. System includes various plate sizes and configurations; accepts Ø3.0mm/Ø3.5mm MotoBAND CP screws, Ø3.0mm polyaxial locking screws, and optional Ø2.7mm non-locking screws for fixation to metatarsal shaft. Used by orthopedic surgeons in clinical/OR settings. Provides mechanical stability for bone healing; benefits patient through minimally invasive approach for bunion correction.
Clinical Evidence
Bench testing only. Mechanical analysis performed on worst-case cross-sections and screw pull-through construct testing validated that the modified system performs equivalently to the predicate device.
Technological Characteristics
Metallic bone fixation plate; material: implant grade titanium (Ti-6Al-4V ELI) per ASTM F136. Features tapered proximal end for intramedullary insertion; distal head with screw holes for locking/non-locking screws. Mechanical fixation principle. No software or energy source.
Indications for Use
Indicated for patients requiring fixation of osteotomies and corrective procedures of the hallux and associated disorders, including hallux valgus.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
April 12, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
CrossRoads Extremity Systems, LLC % Theresa Leister Senior Consultant Mrc-X, LLC 6075 Poplar Ave Memphis, Tennessee 38119
Re: K190658
Trade/Device Name: MIS Bunion Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: March 6, 2019 Received: March 14, 2019
Dear Theresa Leister:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190658
Device Name MIS Bunion Plating System
Indications for Use (Describe)
The MIS Bunion Plating System is intended for fixation of osteotomies and corrective procedures of the hallux and associated disorders such as hallux valgus.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
MIS Bunion Plating System March 6, 2019
| Company: | CrossRoads Extremity Systems, LLC<br>6055 Primacy Parkway, Suite 140<br>Memphis, TN 38119 |
|--------------------------------|-------------------------------------------------------------------------------------------|
| Establishment<br>Registration: | 3011421599 |
| Primary Contact: | Theresa Leister<br>Phone: 901-489-1715 |
| Company Contact: | Chad Hollis<br>Phone: 901-221-8406 |
| Trade Name: | MIS Bunion Plating System |
| Common Name: | Plate, Fixation, Bone |
| Classification: | Class II |
Regulation Number: 21 CFR 888.3030 (Single/multiple component metallic bone fixation appliances and accessories)
Panel: 87- Orthopedic
Product Code: HRS
Predicate Devices: CrossRoads Extremity Systems, LLC, MIS Bunion System (K181872)
#### Device Description:
The CrossRoads MIS Bunion System contains a buttressing implant which can be surgically implanted using a minimally invasive surgical technique to correct hallux valgus. The MIS Bunion implant is to be made of implant grade titanium (Ti-6Al-4V ELI) per ASTM F136 with features allowing it to accept a CrossRoads MotoBAND CP screw (previously cleared in K152306), the subject Ø3.0mm MIS Bunion locking screw, or the optional subject Ø2.7mm non-locking screw in the implant. Instrumentation required for performing this surgery include a k-wire, for use with an osteotomy guide, osteotome guide and osteotome, a suture passing loop, drill bit and handle. Additional general surgical instrumentation utilized include reamers, drill guide/depth gauge. The MIS Bunion implant is tapered on the proximal end so that it can be easily inserted into the intramedullary canal of the metatarsal and the distal head of the implant will seat directly against the osteotomy surface. The MIS bunion implant wedges firmly within the intramedullary canal and acts as a buttress to hold the metatarsal head in place. To accommodate a range of anatomical sizes and severities of corrections, the
{4}------------------------------------------------
implant will be offered in an array of configurations; wherein each incremental size will increase in thickness on the distal end, resulting in different buttressing offsets. The tapered proximal end of the implant will also have interrupted surfaces to gain traction within the cancellous bone, thus resisting pullout. Finally, the MIS bunion implant will accept Ø3.0mm or Ø3.5mm CrossRoads MotoBAND CP screws (K152306) or the subject MIS Bunion Ø 3.0mm polyaxial locking screws in the most distal hole, in addition to the optional subject Ø2.7mm non-locking screw accepted only in the angular cross hole for additional fixation of the implant to the metatarsal shaft.
This Special 510(k) submission seeks to modify the design of the previously cleared plate to add an additional angular distal screw hole to accept optional placement of the subject Ø2.7mm non-locking screw for additional plate stability. Additionally, this submission seeks to expand the size range of the plates to include a shorter length plate to better accommodate patient anatomy.
## Indications for Use:
The MIS Bunion Plating System is intended for fixation of osteotomies and corrective procedures of the hallux and associated disorders such as hallux valgus.
## Substantial Equivalence:
The subject MIS Bunion System components are substantially equivalent to the following predicate devices to the previously cleared MIS Bunion System (K181872).
The subject components are identical to the predicate devices in terms of indications for use, materials, biocompatibility, and sterility. The only changes being made are the following design changes:
- . Design changes to the MIS Bunion System buttressing plate to provide greater fixation and implant stability
- . A shorter plate has been added for all plate configurations, to better accommodate patient anatomy.
- . Ø2.7mm non-locking screws as well as Ø3.0mm polyaxial locking screws have been added to the system.
Mechanical analysis was performed on the subject design changes discussed above. This analysis verified that the design changes included in the subject devices do not present a new worst-case scenario. Thus, it can be concluded that the subject devices raise no new questions of safety and effectiveness and are substantially equivalent to the predicate devices.
# Performance Testing:
Mechanical analysis was performed utilizing the worst-case cross sections the subject MIS Bunion System. This analysis showed the subject device to be substantially equivalent in terms of performance to the predicate MIS Bunion System. Thus, it was determined that no additional mechanical testing is required.
{5}------------------------------------------------
Additional mechanical analysis was performed on the subject MIS Bunion System screws. Screw pull through construct testing was also performed on the subject Ø3.0mm Locking Screw validating that the subject does not present a new worst case. All mechanical analysis performed showed the subject screws to be substantially equivalent in terms of performance to previously cleared Crossroads screws.
#### Conclusion:
There are no substantial differences between the subject and predicate MIS Bunion Plating System with respect to intended use and technological characteristics, including basic design, materials of manufacture, mechanical properties, and intended effect.
Therefore, the MIS Bunion System can be found substantially equivalent to the cited predicate, as it does not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.