The Guide Sheath is intended to introduce therapeutic devices into the vasculature, excluding the coronary and neuro vasculature.
Device Story
The Surmodics Guide Sheath is a braid-reinforced catheter introducer designed to facilitate the delivery of therapeutic devices into the peripheral vasculature. It features a hydrophilic coating on the working length to reduce friction during insertion and an embedded radiopaque marker band near the distal tip for visualization. The device includes a dilator with an atraumatic tip, a cross-cut valve integrated into the hub, and a 3-way stopcock with sidearm extension tubing. It is used by clinicians in a clinical setting to provide a pathway for therapeutic devices. The sheath and dilator assembly allows for simultaneous movement. The device is supplied sterile and is compatible with .018" or .035" guidewires.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including visual inspection, dimensional evaluation, radiopacity (F640), tensile strength (ISO 11070), liquid leak (ISO 11070), connector compatibility (ISO 594), simulated use, kink radius (BS EN 13868), torque strength, 3-point bend (ASTM F2606), coating integrity/coverage, pinch test, removal force, lamination, and particulate testing (USP 788). Biocompatibility testing was performed per ISO 10993-1, and packaging testing per ISO 11607-1.
Technological Characteristics
Braid-reinforced catheter introducer with hydrophilic coating. Features include radiopaque marker band, cross-cut valve, and 3-way stopcock. Compatible with .018" or .035" guidewires. Sterilized via Ethylene Oxide (EtO) per ISO 11135:2014. Mechanical testing standards include ISO 11070:2014 (tensile/leak), ISO 594-1/2 (connectors), BS EN 13868:2002 (kink radius), and ASTM F2606-8 (3-point bend).
Indications for Use
Indicated for the introduction of therapeutic devices into the peripheral vasculature.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 18, 2019
Surmodics, Inc. Peggy Ju Regulatory Affairs Specialist 9924 West 74th Street Eden Prairie, MN 55344
Re: K190644
Trade/Device Name: Guide Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: March 12, 2019 Received: March 13, 2019
Dear Ms. Peggy Ju:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{3}------------------------------------------------
# 510(K) Summary - K190644
Image /page/3/Picture/3 description: The image shows the word "SURMODICS" in a sans-serif font. To the left of the word is a blue and green geometric shape. The blue part of the shape is on top, and the green part is on the bottom.
# Date Prepared: 3/12/2019
#### Submitters Name / Contact Person
#### 510k Submitter Address
Surmodics, Inc. 9924 W 74th St Eden Prairie MN 55344 Phone - (952) 500 -7000
#### Contact for Official/Routine Correspondence
Peggy Ju Regulatory Affair Associate 9924 W 74th St, Eden Prairie, MN 55344 Phone - (651) 319-3602 Fax - (763) 557-6674
| General Information | |
|----------------------|-------------------------------|
| Trade Name: | Guide Sheath |
| Common / Usual Name: | Catheter Introducer |
| Classification: | Class 2 per 21 CFR 870.1340 |
| Product Code: | DYB - Catheter Introducer |
| Predicate Device: | Cook Flexor Shuttle (K142819) |
## Device Description
The Surmodics Guide Sheath is designed to perform as a guide sheath and/or introducer sheath. The sheath is braid reinforced with an embedded radiopague marker band to identify the distal tip. It also has a hydrophilic coating over the working length to provide a lubricious surface to ease insertion. The device comes packaged with a dilator, which is either .018" or .035" guidewire compatible.
This highly flexible, braid reinforced sheath is designed to resist kinking and maximize strength while keeping a minimized profile. Incorporated in the sheath is a radiopaque marker band located approximately 5mm from the distal tip. The Guide Sheath has an atraumatic tip and has a hydrophilic coating over the working length.
The dilator hub is designed to snap into the Guide Sheath cross cut valve cap. Additionally, the dilator shaft design allows for simultaneous movement of the sheath and dilator assembly. The dilator features an atraumatic tip, which is designed to extend beyond the sheath tip by approximately 5 cm. The dilator shaft is radiopaque.
The Guide Sheath features a custom designed cross cut valve that is integrated into the sheath hub and includes a 3-way stopcock with sidearm extension tubing. No assembly is required.
{4}------------------------------------------------
The sheath is packaged in a dual sterile barrier configuration. The sheath and dilator will be placed individually in high density polyethylene (HDPE) coiled hoops, which are then placed in a poly-tyvek pouch within a shelf box.
## Indication for Use
The Guide Sheath is intended to introduce therapeutic devices into the vasculature, excluding the coronary and neuro vasculature.
## Comparison of Technological Characteristics
The Surmodics Guide Sheath device is substantially similar to the legally marketed predicate device in design, intended use, principles of use, materials, sizes and sterility. The Surmodics Guide Sheath and the predicate device are intended for assisting with introduction of therapeutic devices into the peripheral anatomy. Both devices have similar dimensions and similar accessory compatibility. The devices are made from similar materials and both have a lubricious coating.
# Substantial Equivalence and Summary of Studies
Results of design verification testing demonstrate that the technological differences identified do not raise new questions of safety or effectiveness compared to the predicate device. The Guide Sheath is substantially equivalent to the predicate device based on intended use/indications for use and technological characteristics. The subject device has been evaluated through the following bench tests:
## Sheath & Dilator
- . Visual inspections
- Dimensional evaluations ●
- Radiopacity, F640 ●
- Tensile strength, ISO 11070:2014 .
- Tip tensile strength ●
- Liquid Leak Test, ISO 11070:2014 ●
- Hub/Luer Connector Compatibility, . ISO 594-1:1986 and ISO 594-2:1998
- Simulated use ●
- Kink Radius, BS EN 13868:2002
- . Tracking Forces
- Torque Strength ●
- Coating Length .
- . 3 Point Bend, ASTM F2606-8
- Coating Integrity ●
- Coating Coverage .
- . Vertical Pinch Test
- Removal Force ●
- Lamination ●
- Particulate, USP 788 ●
## Packaging Testing
The following packaging tests were performed in accordance with ISO 11607-1:2016/(R)2010:
- Visual Inspection per ASTM F1886-09
- Seal Strength per ASTM F88M-15 ●
- Bubble Leak per ASTM F2096-11
- Environmental Monitoring per ISTA P2A (2011)
- Distribution Simulation per ASTM D4169-16 ●
- Shelf Life ●
{5}------------------------------------------------
## Biocompatibility
The following biocompatibility tests were performed in accordance with ISO 10993-1:2009:
- Cytotoxicity
- Sensitization ●
- Irritation/Intracutaneous reactivity ●
- Acute Systemic toxicity ●
- Hemolysis
- Pyrogencity ●
- Thrombogenicity ●
- Unactivated Partial Thromboplastin Time (UPTT) ●
- Complement Activation ●
All test results met documented acceptance criteria and did not raise new questions of safety or effectiveness. The Surmodics Guide Sheath is substantially equivalent to the predicate device.
#### Sterilization
The results of the sterilization product testing demonstrate that the Ethylene Oxide (EtO) sterilization method for the Guide Sheath meet the requirements of ISO 11135:2014, and that the sterility of the device will be maintained over the entirety of the shelf life.
## Clinical Data
No clinical data is being submitted for the Surmodics Guide Sheath.
## Conclusions
Based upon the device description, indications for use, technological characteristics & performance data it can be concluded that the Guide Sheath is substantially equivalent to the predicate device and is appropriate for the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.