The Aspiron(TM) Aspiration Catheter is used for removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature system.
Device Story
Aspiron Aspiration Catheter is a dual-lumen, rapid-exchange catheter designed for mechanical thrombectomy. It features a guide wire lumen (compatible with 0.014" wires) and a larger extraction lumen for thrombus removal. The device is delivered via 6F guide catheters; a pre-loaded stylet prevents kinking during navigation. A hydrophilic coating on the distal shaft facilitates tracking. Used in clinical settings by physicians, the device is operated with accessories including 30ml suction syringes, a one-way stopcock, extension tubing, and 70μm filter caps. By applying manual suction, the clinician aspirates emboli/thrombi through the extraction lumen. This process restores blood flow in occluded vessels, providing a therapeutic benefit for patients suffering from acute vascular blockages.
Clinical Evidence
Bench testing only. Performance testing included dimensional verification, tensile strength, flexibility/kink resistance, torque strength, radiopacity, extraction rate, leak testing, and coating integrity. Biocompatibility testing (ISO 10993-1) included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogenicity, complement activation, and thromboresistance for both the catheter and accessories. All predetermined acceptance criteria were met.
Technological Characteristics
Dual-lumen, rapid-exchange catheter; 140cm working length; 6F guide catheter compatible; 0.014" guide wire compatibility. Features hydrophilic coating on distal shaft. Includes stylet for kink resistance. Accessories: 30ml suction syringes, stopcock, extension tube, flushing needle, 70μm filter caps. Sterilization method not specified.
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from coronary and peripheral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 29, 2019
Meril Life Sciences Private Limited % H. Semih Oktay President CardioMed Device Consultants, LLC 1783 Forest Drive, #254 Annapolis, Maryland 21401
Re: K190619
Trade/Device Name: Aspiron(TM) Aspiration Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEZ Dated: March 8, 2019 Received: March 11, 2019
Dear Mr. Oktay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190619
Device Name Aspiron(TM) Aspiration Catheter
Indications for Use (Describe)
The Aspiron(TM) Aspiration Catheter is used for removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature system.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Meril" in a sans-serif font against a light blue background. The letters are white, except for a small orange dot above the "i". The logo is simple and clean, with a focus on the brand name.
## 510(k) SUMMARY
# 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92.
## 5.1 Applicant:
Meril Life Sciences Private Limited Bilakhia house, Survey No. 135/139 Muktanand Marg, Chala, Vapi Gujarat 396191 INDIA Phone: +91 - 2603052100 Mobile: +91 – 9601260306
### 5.2 Contact Person:
Utpal Thakor Director Meril Life Sciences Private Limited Bilakhia house, Survey No. 135/139 Muktanand Marg, Chala, Vapi Gujarat 396 191 INDIA Phone: +91 - 2603052100 Mobile: +91 – 9601260306 E mail: utpal.thakor@merillife.com
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Meril" in a sans-serif font against a light blue background. The letters are white, except for a small yellow circle above the "i". The word is centered in the image and appears to be a logo or brand name.
# 5.3 Date prepared: Jan 11, 2019
## 5.4 Device information:
Proprietary Name: Aspiron™ Aspiration Catheter Common / Usual Name: Aspiration Thrombectomy Catheter Classification Name: Embolectomy Catheter (21 CFR 870.5150, Product Code QEZ) Device Class: Class II
## 5.5 Predicate device:
- . Export Advance™ Aspiration Catheter (K152335)
## 5.6 Device description:
The Aspiron™ Aspiration Catheter is a dual lumen, rapid exchange catheter consisting of a distal tip, distal shaft, and proximal shaft with luer hub at the proximal end. The smaller of the two lumens is the Guide wire lumen (distal tip) and is compatible with 0.014" (0.36mm) guide wires. The larger diameter lumen is the extraction lumen and when used with the Aspiron™ accessories, allows for removal of thrombus. The extraction lumen is provided pre-loaded with a stylet that resists kinking during delivery. A hydrophilic coating is applied to the distal shaft. The Aspiron™ is compatible with 6F guide catheters and has a working length of 140cm. The Aspiron™ accessories include 30ml suction syringes, a one way stopcock, an extension tube, a flushing needle and 70μm filter caps.
## 5.7 Indication for use:
The Aspiron™ Aspiration Catheter is used for removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature system.
{5}------------------------------------------------
# 5.8 Comparison of Technological characteristics:
The Aspiron™ Aspiration Catheter is similar to the predicate devices with respect to device design, materials and method of sterilization.
Comparative testing between the Aspiron™ Aspiration Catheter and Export Advance™ Aspiration Catheter found similar performance characteristics with respect to Catheter Extraction Rate, Tensile strength and Radiopacity.
The Aspiron™ Aspiration Catheter indication differs from the predicate Export Advance™ Aspiration Catheter in that the Aspiron™ Aspiration Catheter is limited to removal of fresh, soft emboli and thrombi from vessels. This difference does not raise new concerns because the more limited use of the Aspiron™ Aspiration Catheter is a more conservative indication.
## 5.9 Non clinical Performance data:
To ensure that the device design and construction are suitable for the intended use, the Aspiron™ Aspiration Catheter was subjected to the following performance testing:
- . Dimensional verification
- Tensile Strength ●
- Flexibility and kink ●
- Torque strength
- Radiopacity ●
- . Catheter Preparation, Extraction and Removal
- . Catheter Extraction Rate (Aspiration Rate )
- . Aspiration Catheter Leak
- Coating Integrity & Acute Particulate Evaluation ●
- Biocompatibility testing of Aspiron™ Aspiration Catheter and its accessories ● are in compliance with the ISO 10993-1. Aspiron "" was subjected to following tests.
- In Vitro Cytotoxicity Study o
- Skin Sensitization Study o
- Intracutaneous reactivity test o
- Acute Systemic Toxicity Study O
- In vitro Hemolysis test Study O
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Meril" in a sans-serif font against a light blue background. The letters are white, except for a small yellow dot above the "i". The logo is simple and clean, with a focus on the company name.
K190619 Page 4
- Material Mediated Pyrogen Test o
- In vitro C3a and SC5b-9 complement activation Test O
- In Vivo Thromboresistance O
The Accessories of Aspiron™ Aspiration Catheter were subjected to the following tests.
- O In Vitro Cytotoxicity Study
- Skin Sensitization Study O
- Intracutaneous reactivity test O
- Acute Systemic Toxicity Study O
- In vitro Hemolysis test Study O
- Material Mediated Pyrogen Test O
## 5.10 Conclusion
The Aspiron™ Aspiration catheter met all predetermined acceptance criteria as specified by applicable standards, FDA guidance documents and test protocols. No safety and efficacy issues were raised during the testing program.
The Aspiron™ Aspiration catheter is substantially equivalent to the predicate Export Advance™ Aspiration Catheter
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.