This Mei Ceramic Bracket is intended for the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. The devices are intended to be single use only.
Device Story
Mei Ceramic Bracket consists of translucent polycrystalline aluminum oxide (Al2O3) brackets bonded to teeth; used by orthodontists to facilitate tooth movement. Brackets are linked by arch wire, which transmits steady, gentle pressure to teeth. Device is used in clinical settings; applied by dental professionals. Output is mechanical force applied to teeth to achieve desired alignment. Benefits include correction of malocclusion and aesthetic improvement during treatment.
Clinical Evidence
No clinical data. Bench testing only, including dimension verification (ISO 27020), bracket removal, shear bonding, torque, and wire drag tests. Biocompatibility assessed per ISO 10993-1.
Technological Characteristics
Material: Translucent polycrystalline aluminum oxide (Al2O3, 99.99%). Dimensions/Design: MBT and ROTH designs; slot sizes .018" and .022". Mechanical principle: Passive force transmission via arch wire. Non-sterile packaging. Compliance: ISO 27020 (dimensions), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for orthodontic tooth movement in patients requiring corrective dental treatment. Single use only.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
Predicate Devices
Sapphire Ceramic Bracket / Ortho Technology Inc. (K073045)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 4, 2019
Orthosun Co., Ltd. % Peter Chung Representative Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K190608
Trade/Device Name: Mei Ceramic Bracket Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NJM Dated: August 5, 2019 Received: August 13, 2019
Dear Peter Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K190608
Device Name Orthodontic Ceramic Bracket
Indications for Use (Describe)
This Mei Ceramic Bracket is intended for the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. The devices are intended to be single use only.
| Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size: 16px;"> ☒ </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="font-size: 16px;"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains a logo for Orthosun.Co., Ltd. The logo consists of a blue abstract shape on the left, followed by the company name "Orthosun.Co., Ltd." in black text. Below the company name is the website address "www.orthosun.co.kr", also in black text. The logo appears to be for a company in the medical or healthcare industry.
# 510(k) Summary [as required by 807.92(c)]
#### l. SUBMITTER
- a) Company: Orthosun Co., Ltd.
- b) Address: 2F, 112, Pyeongni-gil, Gyeyang-gu, Incheon, 21008, Republic of Korea
- Tel. : 82-32-545-2825 C)
- d) Fax : 82-32-554-2879
- e) President of Company: Mr. Woo-sik Kim
- f) Contact Person : Mr. Peter Chung, US designated agent
- Contact Person Telephone: 412-687-3976 g)
- Contact Person Address: 300, Atwood Street, Pittsburgh, PA, 15213, USA h)
- i) Submission Date: February 18, 2019
## II. DEVICE
- a) Trade Name : Mei Ceramic Bracket b) Common Name : Orthodontic Ceramic Bracket c) Classification Name : bracket, ceramic, orthodontic d) Product Code : NJM Regulation Number : 872.5470 e) 4) Class of device : Class II Panel : Dental g)
### III. PREDICATE DEVICE
- a) Primary Predicate Device: K073045, Sapphire Ceramic Bracket / Ortho Technology Inc.
### IV. DEVICE DESCRIPTION
The proposed device, Mei Ceramic Bracket, consists of ceramic orthodontic brackets which are bonded to teeth to apply pressure to the tooth, transmitted through a flexible orthodontics wire, to alter the tooth position. The ceramic bracket is produced using Al203, translucent polycrystalline aluminum oxide (99.99%). The brackets are bonded to the teeth with commercially available materials and linked together by "arch wire" that applies steady, gentle pressure to produce desired tooth movement.
### V. INDICATIONS FOR USE
This Mei Ceramic Bracket is intended for the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. The devices are intended to be single use only.
### VI. SUBSTANIAL EQUIVALNECE COMPARISON
The Mei Ceramic Bracket is similar designs and dimensions, and has the same material, intended use, and technological characteristics as the identified primary predicate device (K073045). When compared with predicate device, no new questions of substantial equivalence have been
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for Orthosun.Co., Ltd. The logo consists of a blue abstract shape on the left, followed by the company name in black text. Below the company name is the website address, www.orthosun.co.kr, also in black text.
raised for the Mei Ceramic Bracket.
| | SUBJECT Device | Primary PREDICATE Device<br>K073045 |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Orthosun Co., Ltd. | Ortho Technology Inc. |
| Common Name | Orthodontic Ceramic Bracket | Orthodontic Ceramic Bracket |
| Trade Name | Mei Ceramic Bracket | Sapphire Ceramic Bracket |
| Indications for Use/ Intended<br>Use | Mei Ceramic Bracket is<br>indicated for the orthodontic<br>movement of teeth. It is used<br>temporarily and is removed<br>after orthodontic treatment has<br>been completed. The devices<br>are intended to be single use<br>only. | This device is indicated for the<br>orthodontic movement of<br>teeth. It is used temporarily and<br>is removed after orthodontic<br>treatment has been completed.<br>The devices are intended to be<br>single use only. |
| Clarity | Translucent | Translucent |
| Bracket Design | MBT, ROTH designs without<br>hook | MBT, ROTH designs with and<br>without hook |
| Torque | -21° to 17° | -22° to +17° |
| Angulation | Up to +9° | Up to +11° |
| Available Slot Size | .018″, .022″ | .018″, .022″ |
| Material | Alumina (Al2O3) | Alumina (Al2O3) |
| Colour | White, same as tooth color | White, same as tooth color |
| Indication System | Colored-dot | Colored-dot |
| Biocompatibility | Biocompatible according to ISO<br>10993-1 | Biocompatible according to ISO<br>10993-1 |
| Single Use | yes | yes |
| Non-Sterile Packaging | yes | yes |
### VII. NON-CLINICAL TEST DATA
Non-clinical performance data included testing results for dimension, Bracket removal test, Shear bonding test, Torque test, and Wire drag test. The testing analysis shows that Mei Ceramic Bracket perform comparably to the predicate devices.
Biocompatibility assessment and testing was performed using standard risk assessment techniques and in consideration of FDA and internationally recognized guidance's.
Mei Ceramic Bracket was inspected and met the dimensions itemized in ISO 27020, Dentistry -Brackets and tubes for use in orthodontics.
#### VIII. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification concludes that the Mei Ceramic Bracket is safe and effective and substantially equivalent to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.