The PIVO™ device attaches to a peripheral IV catheter system for use as a direct blood draw device into a vacuum tube or a syringe.
Device Story
PIVO is a sterile, single-use, needle-free blood collection device. It attaches to a peripheral IV (PIV) catheter system via a clip-to-connect mechanism. The device consists of an inner tube with a pusher/slider, a proximal flexible tube with a female luer, and an outer housing. To operate, the clinician attaches the device to the PIV, advances the inner tube through the PIV into the vein using the slider, and collects blood into a vacuum tube or syringe attached to the female luer. After collection, the inner tube is retracted and the device is removed. It is used in clinical settings to facilitate blood draws, potentially reducing the need for additional venipunctures and improving patient comfort. Available in 20, 22, and 24 gauge sizes.
Clinical Evidence
Bench testing only. Performance testing included leak testing, dimensional testing, joint strength, PIV compatibility, flow rate, hemolysis, shelf life, and biocompatibility per ISO 10993-1. No clinical data was required or provided.
Technological Characteristics
Sterile, single-use, needle-free blood collection device. Materials include Pebax/Pebax Vestimide. Features a clip-to-connect attachment for PIV systems, a slider-actuated inner tube, and a female luer for vacuum tube/syringe connection. Available in 20G, 22G, and 24G sizes with color-coded features (pink, blue, yellow). Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for use as a direct blood draw device into a vacuum tube or syringe by attaching to a peripheral IV catheter system.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 13, 2019
Velano Vascular Tiffini Wittwer Consulting Director of Regulatory Affairs 221 Pine St #200 San Francisco, California 94104
Re: K190604
Trade/Device Name: PIVO™M Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: April 12, 2019 Received: April 15, 2019
Dear Tiffini Wittwer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tina Kiang, Ph.D Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190604
Device Name PIVOTM
Indications for Use (Describe)
The PIVO™ device attaches to a peripheral IV catheter system for use as a direct blood draw device into a vacuum tube or a syringe.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitter: | Velano Vascular<br>221 Pine St #200<br>San Francisco CA 94104 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | April 3, 2019 |
| Contact Person: | Tiffini Wittwer, MPH<br>Consulting Director of Regulatory Affairs<br>Phone: 707.799.6732<br>E-mail: tdiage@raechelon.com |
| Trade Name: | PIVOTM |
| Common Name: | Blood specimen collection device |
| Classification: | Class II |
| Product Code: | JKA per 21CFR 862.1675 |
| Predicate Device(s): | The subject device is equivalent to the following devices:<br>• K163508 – PIVOTM |
| Device Description: | The PIVOTM device is a sterile, single use device. It is a<br>needle- free blood collection device that attaches to a<br>peripheral IV (PIV) catheter system. The device is comprised<br>of an inner tube with a pusher/slider, proximal flexible tube<br>with female luer, outer housing and clip-to-connect distal<br>end. The clip-to-connect attaches to the PIV system. The<br>female luer attaches to a blood transfer device or syringe.<br>The inner tube is then advanced to collect a blood sample.<br>Once complete, the inner tube is retracted, and the device is<br>removed from the PIV.<br>The device is available in three sizes, 20, 22, and 24 gauge. |
| Indication for Use: | The PIVOTM device attaches to a peripheral IV catheter system<br>for use as a direct blood draw device into a vacuum tube or a<br>syringe. |
| Reason For Submission: | Line extension to include a 24G compatible size device.<br>Additional changes since last 510(k) clearance include<br>dimensional change to 20G device and material change to non-<br>patient contacting component in 20G and 22G device. |
| Technological<br>Characteristics: | The PIVOTM device attaches to a PIV system via a clip-to-<br>connect attachment. Once attached, a slider is moved forward<br>and advances the inner tube into the PIV system. The<br>differences between the predicate and the subject device, are the<br>different colored features to visually separate the different sizes |
## 510(k) Summary (K190604):
{4}------------------------------------------------
and the diameter of the inner tube for compatibility to various PIV gauges, dimensional change to the 20G inner tube to reduce stiffness and the change in material in the proximal tube. The 20G has pink features and the largest inner tube, the 22G has blue features, and the 24G has yellow features and the smallest inner tube.
| | PIVOTM 24G<br>(Subject Device) | PIVOTM 20G and22G<br>(Predicate Device) | Analysis of<br>differences |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number<br>Decision Date | K190604 | K163508 | |
| Manufacturer | Same | Velano Vascular, Inc. | |
| Classification | Class II | Class II | |
| Product Code | JKA | JKA | |
| Regulation | 21 CFR 862.1675 | 21 CFR 862.1675 | |
| Indications for Use | The PIVO™ device<br>attaches to a peripheral IV<br>catheter system for use as<br>a direct blood draw device<br>into a vacuum tube or a<br>syringe. | The PIVO™ device is<br>attached to a peripheral IV<br>catheter for use as a direct<br>blood draw device into a<br>vacuum tube or a syringe. | Wording change was<br>made for clarity and<br>does not change the<br>intended use. |
| Intended Use | Same | Venous blood drawing | |
| Patient Interface | Same | Separately placed<br>commercially available<br>peripheral IV catheter | |
| PIV Attachment | Same | Clip-to-Connect | |
| Blood Collection<br>Attachment | Same | Female Luer to Blood<br>Transfer Device or Syringe | |
| Blood Control<br>Mechanism | Same | Cap on female luer and<br>clamp on flexible tubing | |
| Tubing<br>(distal) | Same | Transparent<br>Flexible | |
| Proximal Tube<br>Material<br>(no patient contact) | Pebax Vestimide | Pebax | Performance testing<br>(leak testing, joint<br>strength, hemolysis)<br>demonstrated that<br>this material change<br>does not affect the |
| | | | effectiveness of the<br>subject device |
| | | | |
| Compatible PIV<br>Sizes | 24G | 14G-22G | PIV compatibility<br>demonstrates that the<br>subject device is<br>compatible with<br>smaller (24G) PIV<br>systems and there is<br>no additional risk to<br>safety and<br>effectiveness. There<br>is no change to<br>compatibility with<br>larger G devices. |
| Color | Yellow | 20G Pink<br>22G Blue | |
| Inner Tubing<br>Length | 141.2mm | Same | |
| Outer Diameter<br>(OD) of Distal Inner<br>Tubing | 24 gauge = 0.400mm max | 20 gauge = 0.709mm max<br>22 gauge = 0.543mm max | PIV compatibility<br>testing and<br>dimensional testing<br>demonstrate that the<br>difference does not<br>create additional risk<br>to safety and<br>effectiveness of the<br>subject device |
| Wall Thickness of<br>Distal Inner Tubing | 24 G wall = 0.0575mm +/- 0.01mm<br>20 G wall = 0.070mm +/- 0.01mm<br>22 G Same | 20 G wall = 0.089mm +/- 0.01mm<br>22 G wall = 0.0635mm +/- 0.01mm | Performance testing<br>(hemolysis, flow<br>rate, kink resistance)<br>demonstrate that<br>there are no<br>additional risks to<br>safety and<br>effectiveness of the<br>subject device. |
| Sample collection | Same | Device attaches to female<br>luer of PIV system, tube<br>inserted into PIV, blood is<br>drawn through tube into a<br>blood transfer device | |
| Complete<br>Retraction | Same | Yes | |
| Sterilization Method | Same | Gamma | |
| Single Use Only | Same | Yes | |
{5}------------------------------------------------
{6}------------------------------------------------
| Risk Analysis: | A risk analysis was performed for the modifications done for<br>the subject device, in accordance to ISO 14971:2012 Medical<br>Devices - Applications of Risk Management to Medical<br>Devices and Velano Vascular Risk Management SOP.<br>Possible risks were identified which resulted from the gauge<br>size. Based on risk identification, verification and validation<br>activities were carried out to ensure the risk acceptability<br>criteria have been met and the risks have been mitigated. All<br>testing was performed on sterilized product. |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing: | Based on the risks identified, the following tests were<br>performed on the 24G:<br>• Leak testing<br>• Dimensional testing<br>• Joint strength testing<br>• PIV compatibility testing<br>• Flow rate testing<br>• Hemolysis testing<br>• Shelf life testing<br>• Biocompatibility testing per ISO 10993-1 Biological<br>evaluation of medical devices – Part 1: Evaluation and<br>testing within a risk management process |
| Summary of Substantial<br>Equivalence: | The changes made to the previously cleared PIVOTM devices<br>does not raise different questions regarding the safety and<br>effectiveness of the device. 24G PIVOTM is substantially<br>equivalent to the predicate 20G and 22G devices. This<br>conclusion is based upon the devices' identical intended use,<br>similar indications for use, principles of operation,<br>fundamental scientific technology, patient contacting materials,<br>and performance specifications. The changes made were tested<br>using the same acceptance criteria as the predicate device and<br>demonstrated that there are no new risks and the device is<br>substantially equivalent. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.