K190568 · Morgan Scientific, Inc. · BZG · Jun 6, 2019 · Anesthesiology
Device Facts
Record ID
K190568
Device Name
ComPAS2 (Computerized Pulmonary Analysis System)
Applicant
Morgan Scientific, Inc.
Product Code
BZG · Anesthesiology
Decision Date
Jun 6, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The ComPAS2 software is intended to operate with the Screenstar pneumotachograph spirometer, Morgan Transflow test pulmonary function testing system and the Morgan transfer test benchmark pulmonary function testing system. ComPAS2 uses flow and volume from each of the devices to display the flow and volume information measured directly from patient effort. ComPAS2 also utilizes gas analyzer readings from the Transflow test and transfer test benchmark to display helium dilution lung volume data and single breath diffusion data measured directly from patient effort. This information is formatted for use in pulmonary function testing and reports.
Device Story
ComPAS2 is a software application designed to interface with previously cleared pulmonary function testing (PFT) hardware (e.g., Screenstar pneumotachograph, Transflow test system). It captures, displays, and manages clinical data from patient efforts, including flow, volume, and gas analyzer readings. The software runs on Windows 8.1/10 and utilizes a Microsoft SQL database for data storage, with an API for system interaction. It supports HL7 integration for patient data. Clinicians use the system to perform PFTs, manage patient records, and generate reports for interpretation. The software facilitates the measurement of lung disease effects on pulmonary function. It provides trending and reporting capabilities, aiding clinical decision-making by presenting standardized pulmonary metrics.
Clinical Evidence
Bench testing only. Validation included end-to-end testing with biological and mechanical controls, comparing results against the predicate ComPAS. Performance was validated using the ATS 26 waveform loops for flow/volume accuracy. Compliance with ATS/ERS standards for lung function testing, methacholine challenge, and single-breath CO uptake was verified.
Technological Characteristics
Software-based diagnostic system; runs on Windows 8.1/10 (64-bit); developed in .NET, C#, and WPF; uses Microsoft SQL database; connects via USB to external PFT hardware; supports HL7 for EMR integration; complies with IEC 62304, IEC 62366, EN 60601-1, and EN 60601-1-2.
Indications for Use
Indicated for use in pulmonary function testing and reporting, including static and dynamic spirometry, bronchial challenge, MVV, respiratory muscle strength, cough peak flow, lung volume sub-divisions (helium dilution, nitrogen washout, plethysmography), single breath diffusion, airway resistance, and lung clearance index closing volume.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 6, 2019
Morgan Scientific, Inc. Deborah Cornish Manager Quality & RC 151 Essex Street STE 8 Haverhill, Massachusetts 01832
Re: K190568
Trade/Device Name: ComPAS2 Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: May 6, 2019 Received: May 7, 2019
Dear Deborah Cornish:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190568
Device Name ComPAS2
### Indications for Use (Describe)
The ComPAS2 software is intended to operate with the Screenstar pneumotachograph spirometer, Morgan Transflow test PFT system and the Morgan transfer test benchmark PFT system. ComPAS2 uses flow and volume from each of the devices to display the flow and volume information measured directly from patient effort. ComPAS2 also utilizes gas analyzer readings from the Transfer test benchmark to display helium dilution lung volume data and single breath diffusion data measured directly from patient effort. This information is for use in pulmonary function testing and reports.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR 807.92
Date 510(k) summary prepared: May 06, 2019
#### I. SUBMITTER
| Submitter's Name: | Morgan Scientific, Inc. |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address: | 151 Essex Street STE 8<br>Haverhill, MA 01832<br>USA |
| Submitter's Phone: | (978) 521-4440 |
| Official Correspondent: | Deborah Cornish<br>Quality & RA Manager, CQT<br>Morgan Scientific, Inc.<br>151 Essex Street STE 8<br>Haverhill, MA 01832<br>Phone: 800.525.5002 978.521.4440, xt203<br>Email: deborah@morgansci.com |
| II. DEVICE | |
| Trade/proprietary name: | ComPAS2 |
| Common or Usual Name: | ComPAS2 |
| Regulation name: | Diagnostic spirometer |
| Classification: | 21 CFR 868.1840 (product Code: BZG) |
| Classification name: | Diagnostic Devices |
| Regulatory Class: | II |
| III. PREDICATE DEVICE | |
| | Substantial Equivalence is claimed with:<br>ComPAS - K021200 (Class 2 - Computerized Pulmonary Analysis |
System-product code BZG)
#### IV. DEVICE DESCRIPTION
ComPAS2 is designed to interface with various pieces of hardware (previously cleared pulmonary function testing devices, K142812, K042595, K022636, K953990, K013752) to capture clinical data in the performance of clinical testing. Those data can be reported directly to a printer or communicated with hospital information systems/electronic medical records. All data are preserved in
{4}------------------------------------------------
an SOL database, with key sub-systems of ComPAS2 interacting with the database through an API (Application Program Interface).
ComPAS2 is designed to operate with pulmonary function testing hardware by manufacturers offering the capability to measure key pulmonary functions including: static and dynamic spirometry, bronchial challenge, maximum voluntary ventilation (MVV), respiratory muscle strength, cough peak flow, lung volume sub-divisions (by helium dilution, nitrogen washout and plethysmography), single breath diffusion, airway resistance, distribution with lung clearance index closing volume. Other features include a task manager in order to manage patient data for reporting, manual entry in order to input additional information and historical data in order to analyze data for trending and reporting.
The user performs daily quality checks/calibrations prior to performing any tests. Once the test subject enters the Pulmonary Function Testing lab, foundation biographical information (Unique ID, Name, DOB) are either entered by the user, recalled or received in and order via Health Level 7(HL7) message from the information system. The functionality of ComPAS2 software remains the same as the predicate and provides the end user with the same experience.
Encounter information is added at the time of testing such as height, weight, diagnosis and physician. Test capability depends upon the Pulmonary Function Testing device being employed. Completed test information can be printed and handed to the physician for interpretation or data can be routed to the EMR (Electronic Medical Record).
ComPAS2 software connects using USB-powered desktop pulmonary function testing devices for the purpose of creating, adding and recalling subjects and performing pulmonary function testing on those subjects to aid in the measuring of the effect of lung disease on pulmonary function.
#### V. INDICATIONS FOR USE
The ComPAS2 software is intended to operate with the Screenstar pneumotachograph spirometer, Morgan Transflow test pulmonary function testing system and the Morgan transfer test benchmark pulmonary function testing system. ComPAS2 uses flow and volume from each of the devices to display the flow and volume information measured directly from patient effort. ComPAS2 also utilizes gas analyzer readings from the Transflow test and transfer test benchmark to display helium dilution lung volume data and single breath diffusion data measured directly from patient effort. This information is formatted for use in pulmonary function testing and reports.
{5}------------------------------------------------
### VI. TECHNICAL AND SUBSTANTIALLY EQUIVALENT CHARACTERISTICS COMPARISON WITH PREDICATE DEVICE
Technical Characteristics
The characteristics of the ComPAS2 Software are similar and substantially equivalent to those of the predicate device ComPAS listed in the comparison table. The similarities are as follows:
- Identical functionality ●
- Conforms to ATS guidelines ●
The differences between the predicate ComPAS and subject ComPAS2 software application are as follows:
- . Windows 8.1 or 10 (64 bit)
- Language capability supports localization
- .NET, C#, WPF ●
The primary difference between the modified device and the predicate device is the conversion of ComPAS software from Visual Basic version 6 into .Net, C#, and WPF. The functionality and intended use of the product remain the same. The difference is not visible to the end user as a result of the coding update.
No new risks have been identified since the previously cleared predicate (ComPAS K021200). The risks have been evaluated for ComPAS2 software. No new risks have been identified for ComPAS2. The functionality of ComPAS2 is identical to its predicate ComPAS. The indications for use are the same. The diagnostic software device is to be used with previously cleared predicate hardware devices: , K142812, K042595, K022636, K953990, K013752. The software application functionality and indications for use remain unchanged.
{6}------------------------------------------------
| | Predicate ComPAS software | ComPAS2 software | Difference |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Model Name | ComPAS (K021200) | ComPAS2 (K190568) | - |
| 510(k)<br>Number | K021200 | K190568 | - |
| Classification | Class II Device 21 CFR 868.1840 | Class II Device 21 CFR 868.1840 | - |
| Indications for<br>use | The ComPAS software is intended to<br>operate with the Screenstar<br>pneumotachograph spirometer, Morgan<br>Transflow test pulmonary function<br>testing system and the Morgan transfer<br>test benchmark pulmonary function<br>testing system. ComPAS uses flow and<br>volume from each of the devices to<br>display the flow and volume information<br>measured directly from patient effort.<br>ComPAS also utilizes gas analyzer<br>readings from the Transflow test and<br>transfer test benchmark to display<br>helium dilution lung volume data and<br>single breath diffusion data measured<br>directly from patient effort. This<br>information is formatted for use in<br>pulmonary function testing and reports. | The ComPAS2 software is intended to<br>operate with the Screenstar<br>pneumotachograph spirometer, Morgan<br>Transflow test pulmonary function<br>testing system and the Morgan transfer<br>test benchmark pulmonary function<br>testing system. ComPAS2 uses flow and<br>volume from each of the devices to<br>display the flow and volume information<br>measured directly from patient effort.<br>ComPAS2 also utilizes gas analyzer<br>readings from the Transflow test and<br>transfer test benchmark to display<br>helium dilution lung volume data and<br>single breath diffusion data measured<br>directly from patient effort. This<br>information is formatted for use in<br>pulmonary function testing and reports. | - |
| Description | ComPAS is designed to operate with<br>pulmonary function testing hardware by<br>manufacturers offering the capability to<br>measure key pulmonary functions<br>including: static and dynamic<br>spirometry, maximum voluntary<br>ventilation, respiratory muscle strength,<br>cough peak flow, lung volume sub-<br>divisions (by helium dilution, nitrogen<br>washout and plethysmography), single<br>breath diffusion, airway resistance,<br>distribution with lung clearance index<br>closing volume. Other features include a<br>task manager in order to manage patient<br>data for reporting, manual entry in order<br>to input additional information and<br>historical data in order to analyze data<br>for trending and reporting. | ComPAS2 is designed to operate with<br>pulmonary function testing hardware by<br>manufacturers offering the capability to<br>measure key pulmonary functions<br>including: static and dynamic<br>spirometry, maximum voluntary<br>ventilation, respiratory muscle strength,<br>cough peak flow, lung volume sub-<br>divisions (by helium dilution, nitrogen<br>washout and plethysmography), single<br>breath diffusion, airway resistance,<br>distribution with lung clearance index<br>closing volume. Other features include a<br>task manager in order to manage patient<br>data for reporting, manual entry in order<br>to input additional information and<br>historical data in order to analyze data<br>for trending and reporting. | - |
| Standards and<br>References | ATS/ERS 2005 {ATS/ERS Task Force:<br>Standardization of Lung Function<br>Testing}<br>EN ISO 14971<br>EN 60601-1<br>EN 60601-1-2<br>IEC 62366<br>IEC 62304<br>21 CFR 820<br>13485:2016<br>SOR 98-282 | ATS/ERS 2005 {ATS/ERS Task Force:<br>Standardization of Lung Function<br>Testing}<br>EN ISO 14971<br>EN 60601-1<br>EN 60601-1-2<br>IEC 62366<br>IEC 62304<br>21 CFR 820<br>13485:2016<br>SOR 98-282 | - |
| Function | ComPAS (K021200) | ComPAS2 (K190568) | Difference |
| Flow measurement | Pneumotachograph | Pneumotachograph | - |
| Flow Range | -18 L/s to + 18 L/s | -18 L/s to + 18 L/s | - |
| Volume Accuracy | +/- 1% | +/- 1% | - |
| Flow Accuracy | +/- 2.5% | +/- 2.5% | - |
| Sampling Rate | 100 - 300 samples per second | 100 - 300 samples per second | - |
| Number of Tests per<br>Session | 8 – Pre bronchodilator<br>8 - Post Bronchodilator<br>Multiple Challenge Levels | 8 – Pre bronchodilator<br>8 - Post Bronchodilator<br>Multiple Challenge Levels | - |
| Flow Calibration | 3L calibration syringe | 3L calibration syringe | - |
| Units | Metric or Traditional | Metric or Traditional | - |
| ATS/ERS Review of<br>Acceptability | Yes | Yes | - |
| ATS/ERS Review of<br>Repeatability | Yes | Yes | - |
| Standards Compliance | ATS, ERS, SSD & OSHA | ATS, ERS, SSD & OSHA | - |
| Predicted Values | Standard Sets include: GLI, NHANES<br>III, Knudson, Rosenthal and others.<br>Predicted formulas editable through<br>script. | Standard Sets include: GLI,<br>NHANES III, Knudson, Rosenthal<br>and others.<br>Predicted formulas editable<br>through script | - |
| Report Formats | Multiple with Report Editing available | Multiple with Report Editing available | - |
| Quality Control | Data stored for reporting and tracking<br>purposes | Data stored for reporting and<br>tracking purposes | - |
| Hardware<br>Communication | USB | USB | - |
| Database | Microsoft SQL | Microsoft SQL | - |
| Operating System | Windows 7 & 10 | Windows 8.1 & 10 (64 bit) | Updated<br>from<br>Windows 7<br>& 10 to<br>Windows<br>8.1 & 10 (64<br>bit) |
| Test capability | Static and dynamic spirometry,<br>bronchial challenge, maximum<br>voluntary ventilation, respiratory<br>muscle strength, cough peak flow,<br>lung volume sub-divisions (by helium<br>dilution, nitrogen washout and<br>plethysmography), single breath<br>diffusion, airway resistance,<br>distribution with lung clearance index<br>closing volume. Other features<br>include a task manager in order to<br>manage patient data for reporting,<br>manual entry in order to input<br>additional information and historical<br>data in order to analyze data for<br>trending and reporting. | Static and dynamic spirometry,<br>bronchial challenge, maximum<br>voluntary ventilation, respiratory<br>muscle strength, cough peak flow,<br>lung volume sub-divisions (by<br>helium dilution, nitrogen washout<br>and plethysmography), single<br>breath diffusion, airway resistance,<br>distribution with lung clearance<br>index closing volume. Other<br>features include a task manager in<br>order to manage patient data for<br>reporting, manual entry in order to<br>input additional information and<br>historical data in order to analyze<br>data for trending and reporting. | - |
| Function | ComPAS (K021200) | ComPAS2 (K190568) | Difference |
| Patient Height<br>Recording | Standing height, arm span, demi arm<br>span, forearm and knee | Standing height, arm span, demi<br>arm span, forearm and knee | - |
| Data backup | Automated through user-configurable<br>scheduler | Automated through user-<br>configurable scheduler | - |
| Test History/ Data<br>Trending | Available during testing and reporting | Available during testing and<br>reporting | - |
| System Security | Through configurable login rules | Through configurable login rules | - |
| Language Capability | English Only | Supports Localization | Updated to<br>include<br>localization |
| HTML Help | Yes | Yes | - |
## ComPAS (K021200) and ComPAS2 (K190568) Substantial Comparison:
{7}------------------------------------------------
Software Function Substantial Comparison:
{8}------------------------------------------------
#### VII. TESTING AND VALIDATION
Measurement activities conducted for ComPAS2 are in accordance with the set of current standards issued by the American Thoracic Society and European Respiratory Society's Standardization for Lung Function Testing (e.g., Laszlo, 2006; Macintyre et.al., 2005; Miller, Crapo, Hankinson, et.al., 2005; Pellegrino, et.al., 2005; Wanger et.al., 2005). In addition, the newly issued standards on methacholine challenge testing (ERS/ATS, 2017) and standards for single-breath carbon monoxide uptake in the lung (ERS/ATS, 2017) were also applied.
ComPAS2 software testing activities consisted of developing test cases and test runs for the performance of end to end testing with both biological and mechanical controls. Results of these tests were then validated against existing results from ComPAS, the aforenamed predicate device. Requirements and the design specification were reviewed for consistency and accuracy. Final validation was accomplished through system level testing to ensure that the product is capable of meeting the intended use. As an example of the testing and validation activities described above, results from ComPAS2 testing of the American Thoracic Society's 26 waveform loops for flow/volume validation (Hankinson & Crapo, 1995) are provided with this submission.
#### VIII. CONCLUSION:
Section 10 - Page | 6
+
ComPAS2 was found to be substantially equivalent to the predicate device ComPAS.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.