VASERlipo System

K190551 · Solta Medical · QPB · May 1, 2019 · General, Plastic Surgery

Device Facts

Record IDK190551
Device NameVASERlipo System
ApplicantSolta Medical
Product CodeQPB · General, Plastic Surgery
Decision DateMay 1, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.5040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The VASERlipo System is intended for the fragmentation of subcutaneous fatty tissues for aesthetic body contouring. The VentX console is intended for the suction of fluids and tissue during surgical procedures. The VentX console is designed to operate with the VASERlipo System or as a stand-alone system. The VASERlipo System is intended for the fragmentation, emulsification of subcutaneous fatty tissue for the purpose of aesthetic body contouring. The VASERlipo System is also indicated for use in the following surgical specialties for the fragmentation, emulsification and aspiration of soft tissues: - Neurosurgery - Gastrointestinal and affiliated organ surgery - Urological surgery - Plastic and reconstructive surgery - General surgery - Orthopedic surgery - Gynecological surgery - Thoracic surgery and - Laparoscopic surgery

Device Story

VASERlipo System is an ultrasonic surgical device for fragmentation, emulsification, and aspiration of soft tissues; used for aesthetic body contouring and various surgical specialties. System comprises VASER Ultrasonic Amplifier and VentX Infiltration and Aspiration Console; components operate independently or as integrated system. Amplifier utilizes ultrasonic handpiece and probe to fragment/emulsify tissue; Console uses sterile tubing, handle, and cannula for fluid infiltration and tissue aspiration. Operated by surgeons in clinical/surgical settings. Output allows for precise tissue removal; assists in body contouring procedures; benefits patients through targeted soft tissue management.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1:2005/AM1:2012 and ISO 10079-1:2015 demonstrated. Performance testing included cannula deformation resistance, suction vacuum (20 inHg at 5,000 ft), and infiltration rates (50-550 ml/min).

Technological Characteristics

Ultrasonic surgical system; 36kHz nominal operating frequency. Components: VASER Ultrasonic Amplifier and VentX Infiltration/Aspiration Console. Cannula dimensions: 1-6 mm diameter, 7-40 cm length. Infiltration rate: 50-550 ml/min. Max suction vacuum: 20 inHg at 5,000 ft. Electrical safety per IEC 60601-1; suction performance per ISO 10079-1.

Indications for Use

Indicated for fragmentation, emulsification, and aspiration of subcutaneous fatty tissue for aesthetic body contouring, and soft tissue fragmentation/emulsification/aspiration in neurosurgery, GI, urological, plastic/reconstructive, general, orthopedic, gynecological, thoracic, and laparoscopic surgery.

Regulatory Classification

Identification

A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.

Special Controls

*Classification.* Class II (special controls). Consensus standards and labeling restrictions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 9, 2021 Solta Medical % Manal Morcos Senior Director Regulatory Affairs, Surgical Equipment & Devices Manal Morcos, MS BME, MBA 400 Somerset Corporate Boulevard Bridgewater, New Jersey 08807 Re: K190551 Trade/Device Name: VASERlipo System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB Dear Manal Morcos: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 1, 2019. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov. Sincerely, # Cindy Chowdhury -S Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {1}------------------------------------------------ Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 1, 2019 Solta Medical % Ms. Marci Halevi Senior Director Regulatory Affairs, Surgical Equipment & Devices 400 Somerset Corporate Boulevard Bridgewater, New Jersey 08807 Re: K190551 Trade/Device Name: VASERlipo System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: March 5, 2019 Received: March 5, 2019 Dear Ms. Halevi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/2/Picture/5 description: The image shows the name "Joseph A. Nielsen" in large, bold font on the left side. On the right side, there is a digital signature block that includes information such as the signer's name, organization (FDA), email address, and date. The date of the signature is May 1, 2019, at 08:49:14 -04'00'. For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. ## Indications for Use 510(k) Number (if known) #### K190551 #### Device Name VASERlipo System Indications for Use (Describe) The VASERlipo System is intended for the fragmentation of subcutaneous fatty tissues for aesthetic body contouring. The VentX console is intended for the suction of fluids and tissue during surgical procedures. The VentX console is designed to operate with the VASERlipo System or as a stand-alone system. The VASERlipo System is intended for the fragmentation, emulsification of subcutaneous fatty tissue for the purpose of aesthetic body contouring. The VASERlipo System is also indicated for use in the following surgical specialties for the fragmentation, emulsification and aspiration of soft tissues: - · Neurosurgery - · Gastrointestinal and affiliated organ surgery - · Urological surgery - · Plastic and reconstructive surgery - · General surgery - · Orthopedic surgery - · Gynecological surgery - · Thoracic surgery and - · Laparoscopic surgery Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) {4}------------------------------------------------ # 510(k) SUMMARY K190551 ### 1. General Information Submitter: Contact Person: | Solta Medical Inc. | Marci Halevi | |----------------------------|----------------------------------| | 11720 North Creek Pkwy N., | Director, Regulatory Affairs | | Suite 100 | 400 Somerset Corporate Boulevard | | Bothell, WA 98011 | Bridgewater, New Jersey 08807 | | Tel: 510-259-5299 | Phone: 908 952 5174 | | | Marci.Halevi@bauschhealth.com | Preparation Date: 25 April 2019 - 2. Names | Device Name | VASERlipo System | |------------------------|-------------------------------------------| | Classification Name | System, Suction, Lipoplasty | | Common Name: | Suction lipoplasty system | | CFR References: | 21 CFR 878.5040 | | Product Codes: | MUU | | Performance Standards: | No performance standards for this device. | - 3. Predicate Device K110306 VASERlipo System # 4. Product Description The subject of this Special 510(k) submission is for the VASERlipo System which is substantially equivalent to the predicate VASERlipo System. The VASERlipo System is an ultrasonic surgical system that fragments, emulsifies, and aspirates soft tissues. The VASERlipo System is comprised of two components: the VASER Ultrasonic Amplifier and the VentX Infiltration and Aspiration Console. Both components are designed to operate independently, and may be sold or used separately/together, or as a system. The Amplifier uses an ultrasonic surgical handpiece and probe to fragment and emulsify the soft tissue. The Console uses sterile tubing, a handle, and cannula to infiltrate the tissue with fluids and aspirate the fluids/soft tissue. {5}------------------------------------------------ - 5. Indications for Use The VASERlipo System is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. The VentX console is intended for the suction or aspiration of fluids and tissue during surgical procedures. The VentX console is designed to operate with the VASERlipo System or as a stand-alone system. The VASERlipo System is intended for the fragmentation, emulsification and aspiration of subcutaneous fatty tissue for the purpose of aesthetic body contouring. The VASERlipo System is also indicated for use in the following surgical specialties for the fragmentation, emulsification and aspiration of soft tissues: - · Neurosurgery - · Gastrointestinal and affiliated organ surgery - · Urological surgery - · Plastic and reconstructive surgery - · General surgery - Orthopedic surgery - Gynecological surgery - Thoracic surgery and - · Laparoscopic surgery - 6. Summary of Technological Characteristics The technological characteristics of the VASERlipo System are substantially equivalent to those of the predicate device. | Characteristic | Subject Device<br>VASERlipo System | Predicate<br>K110306 | |-------------------------|---------------------------------------------------------------------------------------------------------------------|----------------------------------------| | Intended Use | Intended for the fragmentation and<br>emulsification of subcutaneous fatty tissues<br>for aesthetic body contouring | Identical to subject device | | Operating<br>Frequency | 36kHz Nominal | Identical to subject device | | Infiltration<br>Rates | 50-550 ml/min | Identical to subject device | | Suction<br>Vacuum (max) | 20 inHg at 5,000 ft | Identical to subject device | | Cannula | 1-6 mm diameter<br>7-40 cm length | 2.4-4.6 mm diameter<br>17-34 cm length | {6}------------------------------------------------ - 7. Safety and Effectiveness Information The review of the indications for use and technical characteristics provided demonstrates the VASERlipo System is substantially equivalent to the predicate device. - 8. Brief Summary of Nonclinical Tests and Results Safety tests of the VASERlipo System have demonstrated its compliance with applicable requirements of the following electrical standards: | IEC 60601-1:2005/AM1:2012 | Medical Electrical Equipment - Part 1: General<br>Requirements For Basic Safety And Essential<br>Performance | |---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------| | ISO 10079-1:2015 | Medical suction equipment -- Part 1: Electrically<br>powered suction equipment | | Cannula deformation resistance | $20°±5°$ and straightening without the shaft exhibiting<br>any fractures, sharp edges, or loss of suction<br>performance | | Suction Vacuum (max) 20 inHg at<br>5,000 ft | 20 inHg at 5,000 ft | | Infiltration Rates | 50-550 ml/min | - 9. Conclusion The VASERlipo System shares the same indications for use, design features, and functional features, and thus is substantially equivalent to the predicate device. Non-clinical test results demonstrate that the VASERlipo System is substantially equivalent to the predicate device and no new issues of safety or effectiveness have been raised.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...