# IKON 40 Manual Wheelchair (K190550)

_Broda Seating (Broda Lp) · IOR · Jul 1, 2019 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190550

## Device Facts

- **Applicant:** Broda Seating (Broda Lp)
- **Product Code:** [IOR](/productcode/IOR.md)
- **Decision Date:** Jul 1, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine

## Indications for Use

The standard IKON 40 is a manual wheelchair device that is intended to be used to provide mobility to persons ages 12 and over (adolescents and adults) with a weight capacity of 300 lbs. The device is to be used as a means of mobility for persons limited to a sitting position.

## Device Story

IKON 40 is a foldable, ultra-lightweight manual wheelchair; provides mobility for individuals limited to a sitting position. Device features aluminum folding frame, adjustable seat dimensions, swing-away/elevating legrests, and various wheel options. Operated by user or caregiver in clinical or home settings. Device provides physical mobility support; does not involve electronic processing, AI, or software. Substantial equivalence established via bench testing against ANSI/RESNA standards for strength, stability, and durability.

## Clinical Evidence

No clinical data. Bench testing only. Device performance validated against ANSI/RESNA and ISO 7176 standards for static stability, braking effectiveness, dimensions, mass, maneuvering space, and structural strength (static, impact, fatigue). Flammability testing conducted per CAL 117 and ISO 8191.

## Technological Characteristics

Aluminum folding frame; manual operation. Adjustable seat width (16-22"), depth (16-22"), and height (14-19"). Features swing-away/elevating legrests, flip-up footrests, and adjustable armrests. Compatible with 22"/24" spoke or composite mag wheels. Complies with ANSI/RESNA Wheelchair Standards (Vol 1, Sections 1, 3, 5, 7, 8, 15, 16) and flammability standards (CAL 117, ISO 8191). No electronic components, software, or energy source.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- MyOn HC™ Manual Wheelchair ([K152536](/device/K152536.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the agency's name on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs in the United States.

July 1, 2019

Broda Seating (Broda LP) Jeremy Ballard Product Manager 560 Bingemans Centre Drive Kitchener, ON, N2B3X9 CA

Re: K190550

Trade/Device Name: IKON 40 Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: March 28, 2019 Received: March 28, 2019

Dear Jeremy Ballard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Vivek Pinto, Ph.D. Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190550

Device Name IKON 40 Manual Wheelchair

Indications for Use (Describe)

The standard IKON 40 is a manual wheelchair device that is intended to be used to provide mobility to persons ages 12 and over (adolescents and adults) with a weight capacity of 300 lbs. The device is to be used as a means of mobility for persons limited to a sitting position.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

Date Prepared (revised): June 28,2019

K190550

Submitter:

Broda Seating (Broda LP)

560 Bingemans Centre Drive

Kitchener, ON, N2B3X9

Registration Number 8043392

Phone: 844-552-7632

Fax: 844-442-7632

#### Device:

Device Trade Name: IKON 40 Manual Wheelchair Common or Usual Name: Folding Ultra-Lightweight, Manual Wheelchair Classification Name: Mechanical Wheelchair 21 CFR §890.3850 Regulatory Class: I Product Code IOR: Wheelchair, mechanical

Predicate Device:

MyOn HC™ Manual Wheelchair (K152536); manufactured by Invacare®

(1525712)

{4}------------------------------------------------

### Device Description:

### Technical Specifications (* Standard Configuration):

The IKON 40 is a foldable, ultra-lightweight and fully adjustable manual wheelchair.

| Specification                                                   | IKON 40                                              |                                                      |
|-----------------------------------------------------------------|------------------------------------------------------|------------------------------------------------------|
| Cross bar                                                       | *1 x 1                                               | 2 x 1                                                |
| Seat widths                                                     | 16", *18"                                            | 20", 22"                                             |
| Seat depths                                                     | 16"- 22" Adjustable                                  | 16"- 22" Adjustable                                  |
| Seat height                                                     | 14", 15", 16", 17", 18", 19"<br>Adjustable           | 14", 15", 16", 17", 18", 19"<br>Adjustable           |
| Max user weight                                                 | 300 lbs.                                             | 300 lbs.                                             |
| Chair length                                                    | 40"                                                  | 40"                                                  |
| Chair width                                                     | 26"                                                  | 26"                                                  |
| Length (no legrests)                                            | 30"                                                  | 30"                                                  |
| Folded height                                                   | 34.5"                                                | 34.5"                                                |
| Folded width                                                    | 13.5"                                                | 13.5"                                                |
| Total weight (Including legrests, armrests and<br>anti-tippers) | 34 lbs.                                              | 34 lbs.                                              |
| Front seat height                                               | 16"                                                  | 16"                                                  |
| Backrest height                                                 | 16.5"                                                | 16.5"                                                |
| Legrests                                                        | *Swing Away<br>Elevating                             | Swing Away<br>Elevating                              |
| Legrest range                                                   | 13.5"- 19" Adjustable                                | 13.5"- 19" Adjustable                                |
| Legrest angle (°)                                               | 70°                                                  | 70°                                                  |
| Footrests                                                       | Flip Up and Angle Adjustable                         | Flip Up and Angle Adjustable                         |
| Armrests                                                        | *Flip Back Adjustable Height<br>Fixed Non-Adjustable | *Flip Back Adjustable Height<br>Fixed Non-Adjustable |
| Armrest height range                                            | 9"- 11" Adjustable                                   | 9"- 11" Adjustable                                   |
| Rear wheel                                                      | *22", 24" Spoke<br>22", 24" Composite Mag            | *22", 24" Spoke<br>22", 24" Composite Mag            |
| Front wheel                                                     | 5", 6", *7", 8" Castors                              | 5", 6", *7", 8" Castors                              |

# Indications for Use:

The standard IKON 40 is a manual wheelchair device that is intended to brovide mobility to persons ages 12 and over (adolescents and adults) with a weight capacity of 300 lbs. The device is a means of mobility for persons limited to a sitting position.

{5}------------------------------------------------

# Technological Characteristics Comparison to the Predicate Device:

The following device characteristics comparison demonstrates that the IKON 40 is substantially equivalent in design, construction, operational uses and functionality for persons limited to a sitting position as the previously cleared MyOn HC™ Manual Wheelchair (K152536).

| Characteristics                                                             | IKON 40                                                                                                                                                                                                                                                                                        | MyOn HC™ (K152536)                                                                                                                                                                                                                                                                                       |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>(IFU)                                                | The IKON 40 is a manual wheelchair device<br>that is intended to be used to provide<br>mobility to persons ages 12 and over<br>(adolescents and adults) with a weight<br>capacity of 300 lbs. The device is to be used<br>as a means of mobility for persons limited to<br>a sitting position. | The MyOn HC™ Manual Wheelchair is<br>intended to provide mobility to persons<br>over 12 years of age (adolescents and<br>adults) with a weight capacity of 220 & 290<br>lbs. depending on the seat width. The<br>device is indicated to provide mobility to<br>persons restricted to a sitting position. |
| Cross brace bars                                                            | 1 x 1 Aluminum Folding Frame                                                                                                                                                                                                                                                                   | 1 x 1 Aluminum Folding Frame<br>2 x 1 Aluminum Folding Frame                                                                                                                                                                                                                                             |
| Folding Method                                                              | Collapsible Cross-Brace                                                                                                                                                                                                                                                                        | Collapsible Cross-Brace                                                                                                                                                                                                                                                                                  |
| Warranty                                                                    | Lifetime on Frame                                                                                                                                                                                                                                                                              | Lifetime on Frame                                                                                                                                                                                                                                                                                        |
| Seat widths                                                                 | 16", 18", 20", 22"                                                                                                                                                                                                                                                                             | 15", 16", 17", 18",19", 20", 22", 24"                                                                                                                                                                                                                                                                    |
| Seat depths                                                                 | 16"- 22"<br>Adjustable                                                                                                                                                                                                                                                                         | 16"- 20"<br>Adjustable                                                                                                                                                                                                                                                                                   |
| Max user weight                                                             | 300 lbs. (1 x 1)<br>300 lbs. (2 x 1)                                                                                                                                                                                                                                                           | 275 lbs. (1 x 1)<br>350 lbs. (2 x 1)                                                                                                                                                                                                                                                                     |
| Wheelchair weight<br>(Including legrests,<br>armrests and anti-<br>tippers) | 34 lbs.                                                                                                                                                                                                                                                                                        | 31 lbs.                                                                                                                                                                                                                                                                                                  |
| Seat height                                                                 | 14", 15", 16", 17", 18", 19"<br>Adjustable                                                                                                                                                                                                                                                     | 14", 15", 16", 17", 18",19", 20" Adjustable                                                                                                                                                                                                                                                              |
| Legrests                                                                    | Swing Away<br>Elevating                                                                                                                                                                                                                                                                        | Swing Away<br>Elevating                                                                                                                                                                                                                                                                                  |
| Legrest range                                                               | 13.5"- 19" Adjustable                                                                                                                                                                                                                                                                          | 13.25"- 18.25" Adjustable                                                                                                                                                                                                                                                                                |
| Legrest angle (°)                                                           | 70°                                                                                                                                                                                                                                                                                            | 70°, 80°, 90°                                                                                                                                                                                                                                                                                            |
| Footrests                                                                   | Flip Up and Angle Adjustable                                                                                                                                                                                                                                                                   | Flip Up and Angle Adjustable                                                                                                                                                                                                                                                                             |
| Armrests                                                                    | Flip Back Adjustable Height<br>Fixed Non-Adjustable                                                                                                                                                                                                                                            | Flip Back Adjustable Height<br>T Arm Adjustable Height                                                                                                                                                                                                                                                   |
| Armrest height range                                                        | 9"- 11" Adjustable                                                                                                                                                                                                                                                                             | 9"- 11" Adjustable                                                                                                                                                                                                                                                                                       |
| Rear wheel                                                                  | 22", 24" Spoke<br>22", 24" Composite Mag                                                                                                                                                                                                                                                       | 24", 25", 26" - Spoke<br>20", 22", 24" - Composite Mag<br>20", 24", 25", 26" - Performance                                                                                                                                                                                                               |
| Front wheel                                                                 | 5", 6", 7", 8" Castors                                                                                                                                                                                                                                                                         | 5", 6", 8" Castors                                                                                                                                                                                                                                                                                       |
| Anti-tippers                                                                | Yes, Optional                                                                                                                                                                                                                                                                                  | Yes, Optional                                                                                                                                                                                                                                                                                            |
| Standards - Strength,<br>Stability and Durability                           | ANSI / RESNA Wheelchair Standard Vol. 1<br>Section 1,3,5,7,8,15 &16                                                                                                                                                                                                                            | ANSI / RESNA Wheelchair Standard Vol. 1<br>Section 1,3,5,7,8,15 &16                                                                                                                                                                                                                                      |
| Standards — FR                                                              | CAL 117 / ISO 8191                                                                                                                                                                                                                                                                             | CAL 117 / ISO 8191                                                                                                                                                                                                                                                                                       |

{6}------------------------------------------------

### Performance Testing Data:

### Non-Clinical Testing:

Testing was performed by a non-clinical laboratory on the IKON 40 to determine substantial equivalence:

- ANSI / RESNA - ISO 7176-1:2014 Wheelchairs - Section 1: Determination of static stability
- ANSI / RESNA ISO 7176-3:2012 Wheelchairs Section 3: Determination of effectiveness of brakes ●
- ANSI / RESNA - ISO 7176-5:2008 Wheelchairs - Section 5: Determination of dimensions, mass and maneuvering space
- ANSI / RESNA - ISO 7176-7:1998 Wheelchairs — Section 7: Measurement of seating and wheel dimensions
- ANSI / RESNA - ISO 7176-8:2014 Wheelchairs - Section 8: Requirements and test methods for static, impact and fatigue strengths
- ANSI / RESNA - ISO 7176-11 Wheelchairs- Section 11: Test Dummies
- ANSI / RESNA – ISO 7176-13 Wheelchairs- Section 13: Determination of coefficient of friction
- ANSI / RESNA - ISO 7176-15:1996 Wheelchairs — Section 15: Requirements for information disclosure, documentation and labelling
- ANSI / RESNA WC/Volume 1 2009, Section16: Resistance to Ignition of Upholstered Parts .
- CAL117:2013, Section 1: Flammability Testing
- ISO 8191-1:1987 & 8191-2:1988: Flammability Testing

Animal Study:

Animal testing was not required for this submission

Clinical Testing:

Clinical testing was not required for this submission

#### Conclusions:

The IKON 40, this 510(k) subject device, has the same intended use, safety and similar technological characteristics as the predicate device, the MyOn HC™. Non-clinical laboratory test results support the subject device and demonstrate the IKON 40 should perform as intended in the specified use conditions. Broda has determined that therefore, the IKON 40 Manual Wheelchair is substantially equivalent to the predicate device, MyOn HC™, identified above throughout this document.

---

**Source:** [https://fda-staging.innolitics.com/device/K190550](https://fda-staging.innolitics.com/device/K190550)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).