High V+ Bone Cement, CarboClear® Fenestrated Pedicle Screws
Applicant
CarboFix Orthopedics , Ltd.
Product Code
PML · Orthopedic
Decision Date
Sep 19, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used in conjunction with the CarboClear Fenestrated Pedicle Screws, High V+ Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+ Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
Device Story
System comprises CarboClear Fenestrated Pedicle Screws and High V+ Bone Cement; used for spinal stabilization in patients with advanced tumors. Screws are cannulated, polyaxial, made of carbon fiber-reinforced PEEK with titanium shell and tantalum marker; feature lateral fenestrations at distal tip. Bone cement is self-curing, radiopaque PMMA; supplied as powder (PMMA, barium sulfate, hydroxyapatite, benzoyl peroxide) and liquid (methylmethacrylate monomer, N,N-dimethyl-p-toluidine, hydroquinone). Components mixed to form paste; injected through fenestrated screws into vertebral body to augment fixation. Used by surgeons in clinical settings. Output is mechanical stabilization of spinal column. Benefits include restored structural integrity in palliative oncology patients.
Clinical Evidence
Bench testing only. Mechanical testing included axial pullout strength (ASTM F2193, ASTM F543), insertion/removal torque, and static bending/torsion tests. Compatibility and performance testing characterized cement flow and bolus formation.
Indicated for patients with advanced stage tumors of the thoracic and lumbar spine with insufficient life expectancy for fusion; intended to restore spinal column integrity where structural integrity is not severely compromised.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
September 19, 2019
CarboFix Orthopedics Ltd. Ms. Hila Wachsler-Avrahami Regulatory Affairs 11 Ha'hoshlim Street Herzeliya, 4672411 ISRAEL
Re: K190526
Trade/Device Name: High V+ Bone Cement , CarboClear® Fenestrated Pedicle Screws Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: PML, NKB Dated: August 6, 2019 Received: August 12, 2019
Dear Ms. Wachsler-Avrahami:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190526
Device Name High V+ Bone Cement
Indications for Use (Describe)
When used in conjunction with the CarboClear Fenestrated Pedicle Screws, High V+ Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+ Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190526
#### Device Name
CarboClear® Fenestrated Pedicle Screws
#### Indications for Use (Describe)
When used in conjunction with High V+ Bone Cement, the CarboClear Fenestrated Pedicle Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. CarboClear Fenestrated Pedicle Screws augmented with High V+ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary CarboFix Orthopedics Ltd. CarboClear® Fenestrated Pedicle Screws and High V+ Bone Cement
### Applicant Name
CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939
### Contact Person
Hila Wachsler-Avrahami
### Date Prepared
February 28, 2019
| Trade / Proprietary<br>Name | High V+ Bone Cement | CarboClear® Fenestrated<br>Pedicle Screws |
|-----------------------------|----------------------------------------------|--------------------------------------------|
| Common Name | Polymethylmethacrylate<br>(PMMA) Bone Cement | Pedicle Screw System |
| Device Classification | Class II | Class II |
| Product Code | PML | NKB |
| Regulation Description | Bone Cement, Posterior<br>Screw Augmentation | Thoracolumbosacral Pedicle<br>Screw System |
| Regulation Number | 21 CFR §888.3027 | 21 CFR §888.3070 |
| Panel | Orthopaedic | Orthopaedic |
### Device Regulation and Classification
### Predicate Devices
The subject High V+ Bone Cement is substantially equivalent to the primary predicate device - High V+ (Teknimed® SAS; K161114); and to the additional predicate device - High V+ Bone Cement (NuVasive, Incorporated; K180498).
{5}------------------------------------------------
The subject CarboClear Fenestrated Pedicle Screws are substantially equivalent to the predicate device - CarboClear® Pedicle Screws (CarboFix Orthopedics Ltd.; K173487, K182377); and to the additional predicate devices - NuVasive® Reline® Fenestrated Screws (NuVasive, Incorporated; K180498).
#### Indications for Use
#### High V+ Bone Cement:
When used in conjunction with the CarboClear Fenestrated Pedicle Screws, High V+ Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+ Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
#### CarboClear Fenestrated Pedicle Screws:
When used in conjunction with High V+ Bone Cement, the CarboClear Fenestrated Pedicle Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. CarboClear Fenestrated Pedicle Screws augmented with High V+ Bone Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
#### System Description
The CarboClear Fenestrated Pedicle Screws are cannulated polyaxial pedicle screws in various dimensions, with lateral fenestrations near screws' distal tip, which allow controlled delivery of polymethylmethacrylate (PMMA) bone cement (High V+ Bone Cement) into the vertebral body. The CarboClear Fenestrated Screws are implanted with components of the CarboClear Pedicle Screw System.
The CarboClear Fenestrated Screws are made of carbon fiber-reinforced polyetheretherketone (CFR-PEEK). The threaded portion of the screws is encased within a thin titanium shell, and includes a small tantalum marker.
The CarboClear Fenestrated Screws are supplied sterile, and are intended for single use.
{6}------------------------------------------------
The High V+ Bone Cement is a self-curing, high viscosity, radiopaque PMMA based bone cement. It is provided sterile in two components: 20 grams of powder and 8.6 grams of liquid. The powder component consists of polymethylmethacrylate, with barium sulfate and hydroxyapatite as radiopacifier, and benzoyl peroxide as an initiator. The liquid component comprises methylmethacrylate monomer, with N.Ndimethyl-p-toluidine as a promoter, and hydroquinone as a stabilizer. The powder and liquid components are mixed into homogenous paste, to initiate the polymerization reaction.
# Summary Comparison of Technological Characteristics and Intended Use with Predicate Devices
The subject CarboClear Fenestrated Pedicle Screws are similar to the cleared CarboClear Pedicle Screws (K173487, K182377) in their material, design, dimensions, production processes and sterilization, and additionally include lateral fenestrations at their distal portion. Furthermore, the CarboClear Fenestrated Pedicle Screws are implanted with the components of the cleared CarboClear Pedicle Screw System. The CarboClear Fenestrated Pedicle Screws are intended for the treatment of the same population as the cleared CarboClear Pedicle Screws, however their fixation is further augmented with bone cement.
In addition, the intended use, design, dimensions and principles of operation of the CarboClear Fenestrated Pedicle Screws are substantially equivalent to those of the additional predicate devices.
The subject High V+ Bone Cement is identical in composition production processes including sterilization, and its manufacturer to the primary predicate device - the High V+ bone cement (Teknimed, K161114; and NuVasive, K180498). Its delivery method (via fenestrated screws) and indications for use (patients with advanced stage spinal tumors and short life expectancy) are also similar to those of the High V+ Bone Cement (NuVasive, K180498).
Additionally, the indications for use, cement composition and principles of operation of the subject High V+ Bone Cement are substantially equivalent to those of the additional predicate devices.
### Performance Data
Comparative bench testing was provided to support substantial equivalence of the subject devices and their predicate devices. Mechanical testing included axial pullout strength test according to ASTM F2193 and ASTM F543, insertion and removal torque tests, and static bending and torsion tests. In addition, compatibility and
{7}------------------------------------------------
performance test was conducted for CarboClear fenestrated screws and High V+ Bone Cement, to verify proper operation of the components and to characterize cement flow and bolus formation.
# Conclusion
Based on the information provided in this Premarket Notification, the subject CarboClear Fenestrated Pedicle Screws and High V+ Bone Cement are substantially equivalent to their legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.