K190494 · Ningbo David Medical Device Co., Ltd. · FMZ · Nov 25, 2019 · General Hospital
Device Facts
Record ID
K190494
Device Name
Infant Incubator
Applicant
Ningbo David Medical Device Co., Ltd.
Product Code
FMZ · General Hospital
Decision Date
Nov 25, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The device is an infant incubator and an infant warmer. The two functions can be converted through the canopy lifting system. The incubator provides an enclosed, temperature-controlled environment and the warmer provides infrared heat in an open environment. The warmer provides heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. The incubator and warmer may also be used for short periods of time to facilitate a neonate's transition from the uterus to the external environment.
Device Story
The YP-3000 is a dual-function infant incubator and radiant warmer; functions are converted via a canopy lifting system. In incubator mode, it uses a convective heater and servo-control to maintain air/skin temperature and humidity. In warmer mode, it uses an infrared heating source with a parabolic reflector to provide heat to the patient, controlled via skin temperature feedback. The device includes a mattress, electronic scale, oxygen control system, observation lamp, tray, and IV pole. It is used in clinical settings for neonatal care, including resuscitation and emergency treatment. Healthcare providers monitor the device via integrated sensors and alarms; output allows for precise thermal and environmental management, supporting neonatal stability during transition or treatment.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), alarm systems (IEC 60601-1-8), and performance standards for infant incubators (IEC 60601-2-19) and radiant warmers (IEC 60601-2-21). Biocompatibility testing was performed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Dual-mode infant incubator/warmer; convective heater (incubator) and infrared radiant heater (warmer). Servo-controlled temperature (air/skin), humidity, and oxygen concentration. Includes electronic scale, RS232 connectivity, and multi-port transducer connectors. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-19, IEC 60601-2-21, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for neonates, including low birth weight infants and critically sick babies, who are unable to thermoregulate. Used for constant temperature incubation, body temperature resuscitation, transfusion, oxygen therapy, emergency treatment, and hospitalization.
Regulatory Classification
Identification
A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 25, 2019
Ningbo David Medical Device Co., Ltd Lin Dingyu Vice General Manager No.2, Keyuan Road, Shipu Technology Park, Xiangshan Ningbo. 15731 China
Re: K190494
Trade/Device Name: Infant Incubator Regulation Number: 21 CFR 880.5400 Regulation Name: Neonatal incubator Regulatory Class: Class II Product Code: FMZ, FMZ, FMT Dated: October 23, 2019 Received: October 25, 2019
Dear Lin Dingyu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190494
Device Name YP-3000 Infant Incubator
#### Indications for Use (Describe)
The device is an infant incubator and an infant warmer. The two functions can be converted through the canopy lifting system. The incubator provides an enclosed, temperature-controlled environment and the warmer provides infrared heat in an open environment. The warmer provides heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. The incubator and warmer may also be used for short periods of time to facilitate a neonate's transition from the uterus to the external environment.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------|--|
| <span style="white-space: nowrap;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> | |
| <span style="white-space: nowrap;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K190494 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
(As Required by 21 CFR 807.92)
### 1. Date Prepared [21 CFR 807.92(a)(1)]
November 22, 2019
### 2. Submitter's Information [21 CFR 807.92(a)(1)]
| Company Name: | Ningbo David Medical Device Co., Ltd. |
|------------------|---------------------------------------------------------------------------------------------|
| Company Address: | NO. 2, KEYUAN ROAD, SHIPU TECHNOLOGY PARK, XIANGSHAN, NINGBO, ZHEJIANG PROVINCE, P.R. China |
| Fax: | +86-574-87800008 |
**Fax:**
### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | |
|--------------------|--|
| Models: | |
| Product Codes: | |
| Regulation Number: | |
| Regulation Name: | |
| Device Class: | |
Infant Incubator YP-3000 FMZ, FMT 21 CFR 880.5400 Neonatal incubator Class II
### 4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[
The identification of the predicate device within this submission is as follows:
Manufacturer: Trade Name: Product Codes: Regulation Name: Regulation Number: FDA 510 (k) #:
OHMEDA MEDICAL. OmniBed FMZ, FMT Neonatal incubator 21 CFR 880.5400 K993407
### 5. Description of the Device [21 CFR 807.92(a)(4)]
{4}------------------------------------------------
The YP-3000 has the functions of an infant incubator and an infant radiant warmer. The infant radiant warmer is optional, the two functions can be converted through the canopy lifting system. For the infant incubator, the temperature control system implements servo control on the temperature in the incubator (air temperature / skin temperature) and the device performs proportional heating control based on the setting of the temperature and the actual measured temperature. The internal air is adjusted through thermal convection to create an environment with proper temperature and humidity for infant incubation. It is intended to be used for constant temperature incubation of low birth weight infants, critically sick babies and neonates, as well as neonatal body temperature resuscitation, transfusion, oxygen therapy, emergency treatment, and hospitalization for neonates. The Infant radiant warmer transfers heat to the patient from the infrared radiation heating source through the parabolic reflecting hood transmitting heat to the patient on the mattress. proportional heating is controlled based on actual skin temperature. The device is equipped with a mattress, an electronic scale, and an oxygen control system. Other accessories include an observation lamp, tray, and an IV pole.
### 6. Indications for Use [21 CFR 807.92(a)(5)]
The device is an infant incubator and an infant warmer. The two functions can be converted through the canopy lifting system. The incubator provides an enclosed, temperature-controlled environment and the warmer provides infrared heat in an open environment. The warmer provides heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. The incubator and warmer may also be used for short periods of time to facilitate a neonate's transition from the uterus to the external environment.
### 7. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92(b)(2)]
| | YP-3000<br>Infant Incubator | OmniBed | Differences Discussion |
|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| State | Subject Device | Predicate Device | N/A |
| | | | |
| Standards | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-19<br>IEC 60601-2-21<br>ISO 10993-5<br>ISO 10993-10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-19<br>IEC 60601-2-21<br>ISO 10993-5<br>ISO 10993-10 | Same |
| Indications for<br>Use | The device is an infant<br>incubator and an<br>infant warmer. The<br>two functions can be<br>converted through the<br>canopy lifting system.<br>The incubator<br>provides an enclosed,<br>temperature-<br>controlled<br>environment and the<br>warmer provides<br>infrared heat in an<br>open environment.<br>The warmer provides<br>heat in a controlled<br>manner to neonates<br>who are unable to<br>thermoregulate based<br>on their own<br>physiology. The<br>incubator and warmer<br>may also be used for<br>short periods of time<br>to facilitate a<br>neonate's transition<br>from the uterus to the<br>external environment. | The OmniBed is a<br>combination of an<br>infant incubator<br>(incubator) and an<br>infant warmer<br>(warmer). Incubators<br>and warmers provide<br>heat in a controlled<br>manner to neonates<br>who are unable to<br>thermo-regulate<br>based on their own<br>physiology. Incubators<br>provide an enclosed,<br>temperature-<br>controlled<br>environment and<br>warmers provide<br>infrared heat in an<br>open environment.<br>They may also be<br>used for short periods<br>of time to facilitate the<br>neonate's transition<br>from the uterus to the<br>external environment. | Similar - same intended<br>use, users, and use<br>environment; minor<br>grammatical differences<br>and device trade name is<br>not mentioned. |
| Electrical Description | | | |
| Power supply | AC 220-230V/50Hz;<br>AC 110-120V/60Hz | 100V, 115V, 220V,<br>230V, 240V, 50/60Hz | Similar-1.<br>We performed safety tests<br>and performance tests<br>based on IEC 60601-1,<br>IEC 60601-2-19 and IEC |
| | | | 60601-2-21 standards to<br>cover both voltage ranges.<br>The test results<br>demonstrate that the<br>subject device YP-3000<br>met the standard<br>requirements, so this<br>difference does not raise<br>new questions of safety<br>and effectiveness. |
| Maximum<br>current/ power<br>consumption | 1300VA | ≈1100VA | Similar-2.<br>We conducted safety tests<br>and performance tests<br>based on the IEC 60601-1<br>standard. The test results<br>demonstrate that the<br>subject device YP-3000<br>met the standard<br>requirements, so this<br>difference does not raise<br>new questions of safety<br>and effectiveness. |
| Electrical safety | According to IEC<br>60601-1 | According to IEC<br>60601-1 | Same |
| Principles of Operation | | | |
| General | Radiant Warmer and<br>Infant Incubator | Radiant Warmer and<br>Infant Incubator | Same |
| Control Modes | Incubator mode;<br>Warmer mode | Incubator mode;<br>Warmer mode | Same |
| Probe Type &<br>Connectors | Air temperature<br>sensor | Air temperature<br>sensor | Same |
| | Skin temperature<br>sensor | Skin temperature<br>sensor | Same |
| | air flow temp. sensor | air flow temp. sensor | Same |
| | O2 Sensors | O2 Sensors | Same |
| | Humidity Sensor | Humidity Sensor | Same |
| | Isolated air<br>temperature sensor | Isolated air<br>temperature sensor | Same |
| | Multi-port connector<br>connects 2 | Multi-port connector<br>connects 2 | Same |
| | transducers | transducers | |
| | RS232 connector | RS232 connector | Same |
| Warmer Operation | | | |
| Warmer | Radiant Warmer | Radiant Warmer | Same |
| Temperature<br>Control Modes | Pre-warm mode;<br>Manual mode;<br>Baby mode (Baby<br>temperature<br>control) | Manual mode;<br>Baby mode;<br>(Pre-warm mode); | Same |
| Temperature<br>Control Range<br>Under the Skin<br>temp. Mode | 34.5°C~37.5°C in 0.1<br>increments | 35~37.5°C in 0.1<br>increments | Similar-3.<br>Compared with the<br>predicate device, the<br>temperature control range<br>of the subject device YP-<br>3000 is slightly wider, and<br>under this temperature<br>control range, the<br>temperature control<br>accuracy of the device, the<br>consistency of the bed<br>temperature and the<br>predicate device are<br>equivalent. This control<br>range satisfies the<br>requirements of<br>201.15.4.2.2.101 of<br>IEC60601-2-21. This<br>difference does not raise<br>new questions of safety<br>and effectiveness. |
| Difference<br>Between Skin<br>Temperature<br>Sensor and<br>Control<br>Temperature | ≤0.5°C | ≤0.5°C | Same |
| Temperature<br>Uniformity of<br>Bed Surface | ≤2°C | ≤2°C | Same |
| Incubator Operation | | | |
| Heater | Convective heater | Convective heater | Same |
| Modes under the<br>incubator mode | Air mode;<br>Baby mode | Air mode;<br>Baby mode | Same |
| Air Temperature<br>Control range | 25~39℃ in<br>0.1increments | 20~39℃ in 0.1<br>increments | Similar-4.<br>Neonatal patients in the<br>ambient temperature<br>should not be less than 25<br>℃, so YP-3000 product air<br>control temperature range<br>is designed to 25℃~39℃.<br>Under this temperature<br>control range, the<br>temperature control<br>accuracy, temperature<br>stability and predicate<br>equipment of the<br>equipment are equivalent.<br>This control range satisfies<br>the requirements of<br>201.15.4.2.2.101 in<br>IEC60601-2-19. Therefore,<br>this difference does not<br>raise new questions of<br>safety and effectiveness. |
| Skin<br>Temperature<br>Control range | 34~38℃ in 0.1<br>increments | 35~37.5℃ in 0.1<br>increments | Similar-5.<br>The comparison with the<br>predicate device, YP-<br>3000's temperature control<br>range is slightly wider, and<br>under this temperature<br>control range, the<br>temperature control<br>accuracy, temperature<br>stability and predicate<br>equipment of the device<br>are all equivalent. This<br>control range satisfies the<br>requirements of<br>201.15.4.2.2.102 in |
| | | | |
| | | | IEC60601-2-19. Therefore,<br>this difference does not<br>raise new questions of<br>safety and effectiveness. |
| Temperature<br>Rise Time | ≤40min | <50min | Similar-6.<br>Compared with the<br>predicate device, the<br>temperature rise time of<br>the YP-3000 is faster, and<br>the product can be<br>warmed up to the<br>predetermined<br>temperature and put into<br>use. The performance of<br>this article meets the<br>requirements of<br>201.12.1.107 in IEC60601-<br>2-19, so this difference<br>does not raise new<br>questions of safety and<br>effectiveness. |
| Difference<br>Between<br>Average<br>Incubator<br>Temperature<br>and Control<br>Temperature | ≤1.0°C | ≤1.0°C | Same |
| Difference<br>Between<br>Incubator<br>Temperature<br>and Average<br>Incubator<br>Temperature | ≤0.5°C…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.