BH Medical Dental Barrier Sleeves and Barrier Film
K190484 · Bh Medical Products Co., Ltd. · PEM · Sep 10, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190484
Device Name
BH Medical Dental Barrier Sleeves and Barrier Film
Applicant
Bh Medical Products Co., Ltd.
Product Code
PEM · General, Plastic Surgery
Decision Date
Sep 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4370
Device Class
Class 2
Indications for Use
BH Medical Dental Barrier Sleeves and Barrier Film are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.
Device Story
Disposable polyethylene barrier sleeves and films; designed to cover dental instruments and equipment (e.g., handpieces, curing lights, air/water syringes, dental chairs, instrument trays, x-ray heads, intraoral cameras, digital x-ray sensors). Used in dental clinics; applied by dental staff. Provides physical barrier against contamination; prevents cross-contamination between patients. Non-sterile; single-use. Protects equipment surfaces; does not interfere with device function (e.g., x-ray imaging).
Clinical Evidence
No clinical data. Bench testing only. Performance testing included synthetic blood penetration (ASTM F1670), viral penetration (ASTM F1671), tear resistance, tensile strength, and puncture resistance. Biocompatibility testing (ISO 10993-5, ISO 10993-10) confirmed non-cytotoxic, non-sensitizing, and non-irritating properties. Functional testing confirmed no impact on dental x-ray device performance.
Technological Characteristics
Materials: LLDPE (80%), LDPE (20%), blue pigment, adhesive. Form factor: Various sizes/shapes of polyethylene film covers. Thickness: 0.025mm. Standards: ASTM F1670, ASTM F1671, ASTM D882, ASTM F1342, ASTM D1004, ISO 10993-5, ISO 10993-10. Non-sterile, single-use.
Indications for Use
Indicated for use as a disposable barrier for dental instruments and equipment. Intended for single patient use only.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
Predicate Devices
Pac-Dent Barrier Sleeve, Cover-It Barrier Film (K151123)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
September 10, 2019
BH Medical Products Co., Ltd. Xiaohua Tao President No.90 Shenjiang Villagers' group, Zhangjiacunwei, Xilin Street, Zhonglou District Changzhou. 213024 Cn
Re: K190484
Trade/Device Name: BH Medical Dental Barrier Sleeves and Barrier Film Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape And Drape Accessories Regulatory Class: Class II Product Code: PEM Dated: August 6, 2019 Received: August 12, 2019
Dear Xiaohua Tao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190484
#### Device Name
BH Medical Dental Barrier Sleeves and Barrier Film
#### Indications for Use (Describe)
BH Medical Dental Barrier Sleeves and Barrier Film are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.
| Item # | Model | Designed for |
|---------|---------------------------------------|-------------------------------------------------------------------------------------------------------|
| #992514 | B tray sleeves | Dental instrument tray, 10 1/2"x14"(360x270mm) |
| #992515 | E tray sleeves | Dental instrument tray, 11 1/2"x16"(410x290mm) |
| #992516 | A tray sleeves | Dental instrument tray, 7 1/2"x10 1/2"(270x190mm) |
| #992446 | Headrest Cover | Dental chair headrest, 11"x9 1/2"(250x280mm) |
| #992447 | Headrest Cover | Dental chair headrest, 14"x9 1/2"(250x360mm) |
| #992513 | X-ray head sleeves | X-ray head, universal, 15"x26"(660x380mm) |
| #992481 | Full chair sleeves | Dental chair, universal, 29"x80"(2020x730mm) |
| #992479 | Half chair sleeves | Half dental chair, universal, 27 1/2"x24"(680x600mm) |
| #992486 | Syringe/HVE sleeve without tape | Air/water syringe HVE universal, 2 1/2"x10"(250x60x22mm) |
| #991372 | Digital x-ray sensor sleeve | Digital x-ray sensor, universal, large, 1 5/8"x8"(210x43mm) |
| #991371 | Digital x-ray sensor sleeve | Digital x-ray sensor, universal, large, 1 3/8"x8"(210x36mm) |
| #992492 | handpiece and pen sleeves-High speed | Dental high-speed handpiece universal(200x28mm) |
| #992493 | handpiece and pen sleeves-Lower speed | Dental low-speed handpiece universal(200x38mm) |
| #992499 | light handle sleeves | light handles, universal, T-shape(154x73mm) |
| #992508 | light handle sleeves | light handles, universal, Rectanglar(150x105mm) |
| #992495 | light cure sleeves | Dental curing light, medium,20mm |
| #992496 | light cure sleeves | Dental curing light, large,25mm |
| #992497 | light cure sleeves, curved tip | Dental curing light, medium,(340x53x16mm) |
| #992498 | light cure sleeves, curved tip | Dental curing light, large,(340x84x16mm) |
| #992582 | light cure sleeves, pen style | Dental curing light, pen type,tip:40x27mm,total:250x50mm |
| #992583 | light cure sleeves, pen style | Dental curing light, pen type,,tip:55x26mm,total:250x52mm |
| #992584 | light cure sleeves, pen style | Dental curing light, pen type,tip:40x30mm,total:250x55mm |
| #991361 | Barrier envelopes size 0 | Digital x-ray sensor(Kodak 6100 size 0), inner :36x23mm |
| #991362 | Barrier envelopes size 1 | Digital x-ray sensor(Kodak 6100 size 1), inner :42x26mm |
| #991363 | Barrier envelopes size 2 | Digital x-ray sensor(Kodak 6100 size 2), inner :45x33mm |
| #991364 | Barrier envelopes size 3 | Digital x-ray sensor(Kodak 6100 size 3), inner :56x30mm |
| #991365 | Barrier envelopes size 4 | Digital x-ray sensor(Kodak 6100 size 4), inner :80x60mm |
| #992489 | Tubing sleeves roll (2") | Dental unit tubing, 2",(50mm) |
| #992490 | Tubing sleeves roll (4") | Dental unit tubing, 4",(100mm) |
| #992483 | Barrier film white | Surface area that might be touched during procedure<br>4"x6", white, 10cmx15cmx180m |
| #992482 | Barrier film blue | Surface area that might be touched during procedure<br>4"x6", blue, 10cmx15cmx180m |
| #992585 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, DMD Telicam,<br>228x43x13.5mm |
| #992586 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Siemens Ceracam/minicam<br>Ultra, 228x43x24mm |
| #992587 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Dentsply Acucap |
| | | Concept, 228x43x16.6mm |
| #992588 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Camsight OptumII,<br>228x25x17mm |
| #992589 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, RF System SS21, 228x57x22mm |
| #992590 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Suni Sunic-am,228x48x31mm |
| #992591 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Digital Doc Iris,228x41x19mm |
| #992592 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Vision Tools Spectra Vu,<br>228x57x12mm |
| #992593 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Imagin Hot Rod,<br>228x50x17.5mm |
| #992594 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Carestream STV Pro I&II Kodak<br>1000, 228x35x18mm |
| #992595 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Video Dental Concepts<br>Miniscope & Microscope TD-2000,228x53x29mm |
| #992596 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, DMD Telicam Digital Doc<br>Aurora, 228x35x22mm |
| #991367 | Barrier envelopes size 0 | Digital x-ray sensor (size 0),inner :25x37mm |
| #991368 | Barrier envelopes size 1 | Digital x-ray sensor (size 1),inner :26.5x43mm |
| #991369 | Barrier envelopes size 2 | Digital x-ray sensor (size 2),inner :44x33mm |
| #991374 | Sensor sleeves | Dental instrument tray, 100x43mm |
| #991373 | Sensor sleeves | Dental instrument tray, 250x63mm |
{3}------------------------------------------------
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Section 5 510(k) Summary
# [As Required by 21 CFR 807.92]
Submitter: BH Medical Products Co., Ltd. No.90 Shenjiang Villagers' group, Zhangjiacunwei, Xilin Street, Zhonglou District, Changzhou City, Jiangsu Province, People's Republic of China (Establishment registration number: 3009307523)
Contact Person: Xiaohua TAO President Tel: 86-519-83978705 Fax: 86-519-83978717 Email: tao@bhmedical.com.cn
US Agent: Name : ISMILE DENTAL PRODUCTS INC. Address : 4201 SIERRA POINT DR STE 102 Sacramento California, 95834 UNITED STATES Contact Person: Bryan Shields Tel: (916) 718-0795 Email: bshields@ismiledp.com
Device/Trade Name: BH Medical Dental Barrier Sleeves and Barrier Film Common Name: Dental Barriers and Sleeves Device Panel: General and Plastic Surgery Basis for Submission: New Device Regulation Name: Surgical Drape and Drape Accessories
Device Classification: Class I
Regulation Number: 21 CFR 878.4370 Product Code: PEM
### Predicate Device to BH Medical Dental Barrier Sleeves and Barrier Film:
Trade Name: Pac-Dent Barrier Sleeve, Cover-It Barrier Film 510(k) Number: K151123 Manufactured by: Pac-Dent International, Inc.
#### Device Description:
BH Medical Dental Barrier Sleeves and Barrier Film consist of various sizes and shapes of clear polyethylene covers which are positioned on various small hand-held dental instruments such as hand pieces, curing lights, air/water syringes and similar hand instruments. In other forms, they are used to cover various devices such as dental chairs,
{5}------------------------------------------------
dental instrument trays, x-ray heads, and others. The products are sold non-sterile, prepackaged, and are disposable, single use only.
# Indications for Use:
BH Medical Dental Barrier Sleeves and Barrier Film are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.
| Item | Model | Description | Designed for |
|------|---------|------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| 1 | #992514 | B tray sleeves | Dental instrument tray, 10 1/2"x14"(360x270mm) |
| 2 | #992515 | E tray sleeves | Dental instrument tray, 11 1/2"x16"(410x290mm) |
| 3 | #992516 | A tray sleeves | Dental instrument tray, 7 1/2"x10<br>1/2"(270x190mm) |
| 4 | #992446 | Headrest Cover | Dental chair headrest, 11"x9 1/2"(250x280mm) |
| 5 | #992447 | Headrest Cover | Dental chair headrest, 14"x9 1/2"(250x360mm) |
| 6 | #992513 | X-ray head sleeves | X-ray head, universal, 15"x26"(660x380mm) |
| 7 | #992481 | Full chair sleeves | Dental chair, universal, 29"x80"(2020x730mm) |
| 8 | #992479 | Half chair sleeves | Half dental chair, universal, 27<br>1/2"x24"(680x600mm) |
| 9 | #992486 | Syringe/HVE sleeve<br>without tape | Air/water syringe HVE universal, 2<br>1/2"x10"(250x60x22mm) |
| 10 | #991372 | Digital<br>x-ray sensor<br>sleeve | Digital x-ray sensor, universal, large, 1<br>5/8"x8"(210x43mm) |
| 11 | #991371 | Digital<br>x-ray sensor<br>sleeve | Digital x-ray sensor, universal, large, 1<br>3/8"x8"(210x36mm) |
| 12 | #992492 | handpiece and pen<br>sleeves-High speed | Dental high-speed handpiece universal(200x28mm) |
| 13 | #992493 | handpiece and pen<br>sleeves-Lower speed | Dental low-speed handpiece universal(200x38mm) |
| 14 | #992499 | light handle sleeves | light handles, universal, T-shape(154x73mm) |
| 15 | #992508 | light handle sleeves | light handles, universal, Rectangular(150x105mm) |
| 16 | #992495 | light cure sleeves | Dental curing light, medium,20mm |
| 17 | #992496 | light cure sleeves | Dental curing light, large,25mm |
| 18 | #992497 | light cure sleeves,<br>curved tip | Dental curing light, medium,(340x53x16mm) |
| 19 | #992498 | light cure sleeves,<br>curved tip | Dental curing light, large,(340x84x16mm) |
| 20 | #992582 | light cure sleeves, pen<br>style | Dental curing light, pen<br>type, tip:40x27mm, total:250x50mm |
| 21 | #992583 | light cure sleeves, pen style | Dental curing light, pen type,,tip:55x26mm,total:250x52mm |
| 22 | #992584 | light cure sleeves, pen style | Dental curing light, pen type,tip:40x30mm,total:250x55mm |
| 23 | #991361 | Barrier envelopes size 0 | Digital x-ray sensor(Kodak 6100 size 0 ),inner :36x23mm |
| 24 | #991362 | Barrier envelopes size 1 | Digital x-ray sensor(Kodak 6100 size 1 ),inner :42x26mm |
| 25 | #991363 | Barrier envelopes size 2 | Digital x-ray sensor(Kodak 6100 size 2 ),inner :45x33mm |
| 26 | #991364 | Barrier envelopes size 3 | Digital x-ray sensor(Kodak 6100 size 3 ),inner :56x30mm |
| 27 | #991365 | Barrier envelopes size 4 | Digital x-ray sensor(Kodak 6100 size 4 ),inner :80x60mm |
| 28 | #992489 | Tubing sleeves roll (2") | Dental unit tubing, 2",(50mm) |
| 29 | #992490 | Tubing sleeves roll (4") | Dental unit tubing, 4",(100mm) |
| 30 | #992483 | Barrier film white | Surface area that might be touched during procedure 4"x6", white, 10cmx15cmx180m |
| 31 | #992482 | Barrier film blue | Surface area that might be touched during procedure 4"x6",blue,10cmx15cmx180m |
| 32 | #992585 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, DMD Telicam,228x43x13.5mm |
| 33 | #992586 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Siemens Ceracam/minicam Ultra,228x43x24mm |
| 34 | #992587 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Dentsply Acucap Concept, 228x43x16.6mm |
| 35 | #992588 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Camsight OptumII, 228x25x17mm |
| 36 | #992589 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, RF System SS21,228x57x22mm |
| 37 | #992590 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Suni Sunic-am,228x48x31mm |
| 38 | #992591 | Intraoral Camera Sleeves | Digital intraoral camera sleeves, Digital Doc Iris 228x41x19mm |
| 39 | #992592 | Intraoral Camera<br>Sleeves | Digital intraoral camera sleeves, Vision Tools<br>Spectra Vu,228x57x12mm |
| 40 | #992593 | Intraoral Camera<br>Sleeves | Digital intraoral camera sleeves, Imagin Hot<br>Rod,228x50x17.5mm |
| 41 | #992594 | Intraoral Camera<br>Sleeves | Digital intraoral camera sleeves, Carestream STV<br>Pro I&II Kodak 1000,228x35x18mm |
| 42 | #992595 | Intraoral Camera<br>Sleeves | Digital intraoral camera sleeves, Video Dental<br>Concepts Miniscope & Microscope TD-<br>2000,228x53x29mm |
| 43 | #992596 | Intraoral Camera<br>Sleeves | Digital intraoral camera sleeves, DMD Telicam<br>Digital Doc Aurara,228x35x22mm |
| 44 | #991367 | Barrier envelopes size 0 | Digital x-ray sensor (size 0),inner :25x37mm |
| 45 | #991368 | Barrier envelopes size 1 | Digital x-ray sensor (size 1),inner :26.5x43mm |
| 46 | #991369 | Barrier envelopes size 2 | Digital x-ray sensor (size 2),inner :44x33mm |
| 47 | #991374 | Sensor sleeves | Dental instrument tray, 100x43mm |
| 48 | #991373 | Sensor sleeves | Dental instrument tray, 250x63mm |
Table for the product names and model numbers:
{6}------------------------------------------------
{7}------------------------------------------------
# Comparison of Technological Characteristics:
| Item | Predicate Device | Subject Device | Comment |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device | Pac-Dent Barrier Sleeve<br>510(K) Number: K151123 | BH Medical Dental Barrier<br>Sleeves and Barrier Film | / |
| Indications for Use | Pac-Dent Barrier Sleeve are<br>intended to be used as a barrier to<br>cover dental instruments. It is<br>non-sterile and is intended for<br>single patient use only. | BH Medical Dental Barrier<br>Sleeves and Barrier Film are<br>intended to be used as a<br>disposable barrier for dental<br>instruments and equipment. This<br>device is non-sterile and intended<br>for single patient use only. | Same |
| Device<br>Description | Pac-Dent Barrier Sleeves consist<br>of various sizes and shapes of<br>clear polyethylene covers which<br>are positioned on various small<br>hand-held dental instruments such<br>as handpieces, curing lights,<br>air/water syringes and similar<br>hand instruments. In other forms,<br>they are used to cover various<br>devices such as dental chairs, | BH Medical Dental Barrier<br>Sleeves and Barrier Film consist<br>of various sizes and shapes of<br>clear polyethylene covers which<br>are positioned on various small<br>hand-held dental instruments such<br>as hand pieces, curing lights,<br>air/water syringes and similar<br>hand instruments. In other forms,<br>they are used to cover various | Same |
| | | | |
| | dental instrument trays, x-ray heads, and others. | devices such as dental chairs, dental instrument trays, x-ray heads, and others. The products are sold non-sterile, prepackaged, and are disposable, single use only. | |
| Classification<br>Product Code | PEM | PEM | Same |
| Material | Polyethylene film | Polyethylene film | Same |
| Material<br>Composition | LLDPE(80%),LDPE(20%),Blue pigment(Item#100B,C101B),Adhesive(Item#C101,C101B) | LLDPE(80%),LDPE(20%),Blue pigment(#992482),Adhesive(#992482,#992483) | Same |
| Specifications<br>and Tolerances | Paper backing: some of model;<br>Film thickness:0.02-0.06mm;<br>Tolerance:<0.01mm; | Paper Backing: No<br>Film Thickness: 0.025mm<br>Tolerance:<0.01mm | Similar |
| Labeling | Single use device | Single use device | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Performance<br>Properties | Synthetic Blood Penetration test (ASTM F1670) -Pass | Synthetic Blood Penetration test (ASTM F1670) -Pass | Same |
| | Viral Penetration test (ASTM F1671) -Pass | Viral Penetration test (ASTM F1671)-Pass | |
| Biocompatibility | Non-cytotoxic<br>Non-sensitizing<br>Non-irritating | Biocompatibility tests were conducted in compliance with ISO 10993-5 and ISO 10993-10.<br>Test results indicate the device is non-cytotoxic, non-sensitizing and non-irritating. | Same |
| Dimensions | Determined by the size and shape of dental instruments and equipment | Difference in dimensions is due to the size of instruments and equipment they cover | Same |
| FDA-<br>Recognized<br>Standards | ASTM F1670<br>ASTM F1671<br>ASTM D882<br>ASTM F1342<br>ASTM D1004<br>ISO 10993-5<br>ISO 10993-10 | ASTM F1670<br>ASTM F1671<br>ASTM D882<br>ASTM F1342<br>ASTM D1004<br>ISO 10993-5<br>ISO 10993-10 | Same |
{8}------------------------------------------------
#### Tests:
Performance test including resistance to penetration, tear resistance, tensile strength, puncture resistance was performed. Biocompatibility test in accordance with ISO 10993-5, ISO 10993-10 has been performed. Effectiveness of x-ray devices covered with BH Dental Barrier Sleeves was performed to demonstrate the subject device does
{9}------------------------------------------------
not impact the function of dental x-ray devices.
The biocompatibility test result indicates that the barrier sleeves do not have the potential cytotoxicity, sensitization and irritation.
## Clinical Performance Data:
NA
## Conclusion
Based on the tests performed the subject device BH Medical Dental Barrier Sleeves and Barrier Film is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.