← Product Code [MAX](/productcode/MAX) · K190483

# SPIRA Open Matrix ALIF and LLIF (K190483)

_Camber Spine Technologies · MAX · Sep 23, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190483

## Device Facts

- **Applicant:** Camber Spine Technologies
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Sep 23, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Camber Spine Technologies SPIRA™ Open Matrix ALIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA™ Open Matrix ALIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA™ Open Matrix ALIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft. The Camber Spine Technologies SPIRA Open Matrix LLIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA Open Matrix LLIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA Open Matrix LLIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.

## Device Story

SPIRA Open Matrix ALIF and LLIF are intervertebral body fusion devices designed for spinal stabilization. ALIF features arched support members; LLIF features spiral support members; both utilize an open matrix design to facilitate bone growth throughout the implant. Surfaces are roughened to minimize migration. Devices are implanted by surgeons during spinal fusion procedures to treat DDD. They require use with supplementary fixation systems and autogenous or allogenic bone graft. The primary purpose of this submission is to expand the indications to include the use of allogenic bone graft.

## Clinical Evidence

Bench testing only. Finite Element Analysis (FEA) was conducted to evaluate design changes, specifically assessing static axial compression, shear, and surface area for subsidence and expulsion equivalence. Results confirmed that design changes did not create new worst-case scenarios.

## Technological Characteristics

Intervertebral body fusion devices with open matrix architecture. ALIF utilizes arched support members; LLIF utilizes spiral support members. Roughened surface texture on non-perimeter surfaces. Designed for use with autogenous or allogenic bone graft.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- SPIRA Open Matrix LLIF ([K180724](/device/K180724.md))
- SPIRA Open Matrix ALIF ([K162986](/device/K162986.md))
- Centinel Spine STALIF TT70, STALIF MIDLINE®, MIDLINE II™, and MIDLINE II-Ti™ ([K150643](/device/K150643.md))

## Reference Devices

- SPIRA-C Open Matrix Cervical Interbody ([K172446](/device/K172446.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

September 23, 2019

Camber Spine Technologies % Matthew Carey Junior Regulatory Affairs/Quality Consultant RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, Massachusetts 01913

Re: K190483

Trade/Device Name: SPIRA Open Matrix ALIF and LLIF Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: August 16, 2019 Received: August 27, 2019

Dear Matthew Carey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190483

Device Name SPIRA Open Matrix ALIF and LLIF

### Indications for Use (Describe) SPIRA Open Matrix ALIF

The Camber Spine Technologies SPIRA™ Open Matrix ALIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA™ Open Matrix ALIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA™ Open Matrix ALIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.

### SPIRA Open Matrix LLIF

The Camber Spine Technologies SPIRA Open Matrix LLIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA Open Matrix LLF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA Open Matrix LLIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.

| Type of Use (Select one or both, as applicable)  |                                             |  |
|--------------------------------------------------|---------------------------------------------|--|
| X   Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K190483 Page 1 of 3

## 510(k) SUMMARY

Camber Spine Technologies SPIRA Open Matrix ALIF and LLIF

#### SUBMITTER l.

Camber Spine Technologies 501 Allendale Road King of Prussia, PA 19406

Phone: 484.427.7060

Contact Person: Matthew Carey RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, MA 01913 Phone: 1-(978) 834-5703 Fax: 1 (978) 477-0206

Date Prepared: 19 September 2019

#### II. DEVICE

| Trade/Device Name:    | SPIRA Open Matrix ALIF and SPIRA Open Matrix LLIF    |
|-----------------------|------------------------------------------------------|
| Common or Usual Name: | Intervertebral Fusion Device with Bone Graft, Lumbar |
| Classification Name:  | Intervertebral body fusion device                    |
| Regulation Number:    | 21 CFR 888.3080                                      |
| Regulatory Class:     | Class II                                             |
| Product codes         | MAX                                                  |

#### lll. PREDICATE DEVICES

| Primary Predicates:   | SPIRA Open Matrix LLIF (K180724)<br>SPIRA Open Matrix ALIF (K162986)                      |
|-----------------------|-------------------------------------------------------------------------------------------|
| Additional Predicate: | Centinel Spine STALIF TT70, STALIF MIDLINE®, MIDLINE II™, and<br>MIDLINE II-Ti™ (K150643) |
| Reference Predicate:  | SPIRA-C Open Matrix Cervical Interbody, Camber Spine<br>Technologies (K172446)            |

{4}------------------------------------------------

#### IV. DEVICE DESCRIPTION

SPIRA Open Matrix ALIF is an interbody fusion device with an open matrix design consisting primarily of arched support members to permit bone growth (i.e., interbody fusion) throughout the implant, SPIRA Open Matrix ALIF. With the exception of the perimeter, all surfaces have a roughened texture to help prevent movement of the device.

SPIRA Open Matrix LLIF is an interbody fusion device with an open matrix design consisting primarily of spiral support members to permit bone growth (i.e., interbody fusion) throughout the implant, SPIRA-L Open Matrix LLIF

The purpose of this submission is the addition of the allogenic bone graft indication to both SPIRA ALIF and SPIRA LLIF devices that have been cleared under K162986 and K180724 respectively.

## Indications for Use:

## SPIRA™ Open Matrix ALIF

The Camber Spine Technologies SPIRA™ Open Matrix ALIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA™ Open Matrix ALIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA™ Open Matrix ALIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.

### SPIRA™ Open Matrix LLIF

The Camber Spine Technologies SPIRA Open Matrix LLIF is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of nonoperative treatment prior to treatment with the devices. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Camber Spine Technologies SPIRA Open Matrix LLIF is intended to be used with additional FDA-cleared supplementary fixation systems. The Camber Spine Technologies SPIRA Open Matrix LLIF system must be used with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.

{5}------------------------------------------------

#### V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject SPIRA Open Matrix LLIF and SPIRA Open Matrix ALIF device was demonstrated to be substantially equivalent to primary predicates SPIRA Open Matrix LLIF (K180724) and SPIRA Open Matrix ALIF (K162986) with respect to indications, design, materials, function, manufacturing, and/or performance.

#### VI. PERFORMANCE DATA

A Finite Element Analysis (FEA) was performed to evaluate the effects of design changes to the SPIRA Open Matrix LLIF on the worst-case implant and determine if the design changes created new worst-case scenarios. The testing performed was static axial compression shear, and surface area comparison (subsidence and expulsion equivalence). Testing proved that the new design did not create a new worst-case under any testing condition.

#### VII. CONCLUSION

Camber Spine Technologies provided sufficient information to demonstrate the SPIRA Open Matrix LLF and SPIRA Open Matrix ALIF device is substantially equivalent to predicates SPIRA Open Matrix LLIF (K180724), SPIRA Open Matrix ALIF (K162986), and Centinel Spine STALIF TT70, STALIF MIDLINE®, MIDLINE II™, and MIDLINE II-Ti™ (K150643) with respect to indications, design, materials, function, manufacturing, and performance.

---

**Source:** [https://fda-staging.innolitics.com/device/K190483](https://fda-staging.innolitics.com/device/K190483)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
