PF Nitrile Exam Glove with pH coating, Black-Gray & White-Black Colors, Tested for Use with chemotheraphy drugs and Fentanyl Citrate

K190466 · Shen Wei USA, Inc. · LZA · Sep 26, 2019 · General Hospital

Device Facts

Record IDK190466
Device NamePF Nitrile Exam Glove with pH coating, Black-Gray & White-Black Colors, Tested for Use with chemotheraphy drugs and Fentanyl Citrate
ApplicantShen Wei USA, Inc.
Product CodeLZA · General Hospital
Decision DateSep 26, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs and Fentanyl Citrate.

Device Story

Disposable nitrile examination glove; pH-coated; available in Black-Gray and White-Black colors. Worn by healthcare examiners to provide barrier protection against contamination between patient and examiner. Tested for permeation resistance against various chemotherapy agents and Fentanyl Citrate. Used in clinical settings; provides protection during patient examinations and handling of hazardous drugs. Benefits include reduced risk of cross-contamination and chemical exposure for the examiner.

Clinical Evidence

Bench testing only. Permeation testing performed for chemotherapy drugs and Fentanyl Citrate. Breakthrough detection times reported for various agents; specific warnings provided for Carmustine and Thiotepa due to low permeation times.

Technological Characteristics

Nitrile material; pH-coated; powder-free; disposable; non-sterile. Form factor: examination glove. Colors: Black-Gray and White-Black.

Indications for Use

Indicated for use as a disposable medical glove worn by examiners to prevent cross-contamination between patient and examiner. Suitable for use with specific chemotherapy drugs and Fentanyl Citrate. Contraindicated for use with Carmustine and Thiotepa.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 26, 2019 Shen Wei USA Inc. Cheryl Reep Compliance Director 33278 Central Ave Ste 102 Union City, California 94587 Re: K190466 Trade/Device Name: PF Nitrile Exam Glove with pH coating, Black-Gray & White-Black Colors, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC, ODO Dated: September 3, 2019 Received: September 9, 2019 Dear Cheryl Reep: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Elizabeth Claverie-Williams, MS Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) ## K190466 ### Device Name Powder-Free Nittle Examination Glove with pH conting, Black-Gray and White-Black Colors, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate #### Indications for Use (Describe) A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs and Fentanyl Citrate. | Chemotherapy Drug and Concentration | White-Black<br>Breakthrough Detection | Black-Gray<br>Time in Minutes | |--------------------------------------|---------------------------------------|-------------------------------| | | Carmustine (BCNU), 3.3mg/ml | 27.8 | | Cisplatin, 1.0mg/ml | >240 | >240 | | Cyclophosphamide (Cytoxan), 20mg/ml | >240 | >240 | | Cyclosporin A, 100 mg/ml | >240 | not tested | | Dacarbazine (DTIC) 10g mg/ml | >240 | >240 | | Doxorubicin Hydrochloride, 2.0 mg/ml | >240 | >240 | | Etoposide (Toposar), 20mg/ml | >240 | >240 | | Fluorouracil 50mg/ml | >240 | >240 | | Paclitaxel (Taxol), 6.0mg/ml | >240 | >240 | | Thiotepa, 10mg/ml | 77.9 | 27.4 | | Vincristine Sulfate, 1.0mg/ml | >240 | not tested | | Fentanyl Citrate, 100mcg/2ml | >240 | >240 | WARNING: Do not use with Carmustine and Thiotepa. Please note that Carmustine (BCNU) has low permeation of 27.8 minutes for White-Black color and 27.1 minutes for Black-Gray color, Thiotepa has 77.9 minutes for White-Black color and 27.4 minutes for Black-Gray color. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D). × Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect. of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...