The Engage™ Partial Knee System is intended for medial unicompartmental knee arthroplasty to treat one or more of the following conditions: - Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis. - Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty - As an alternative to tibial osteotomy in patients with Unicompartmental osteoarthritis. The femoral component and tibial tray are intented fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application. When the tibial tray is implanted without the use of bone cement, the tibial anchor should be used. When the tibial tray is implanted with bone cement, the tibial anchor should not be used.
Device Story
Engage™ Partial Knee System is a unicompartmental knee arthroplasty implant for medial compartment replacement. System components include femoral component, tibial tray, tibial insert, and tibial anchor. Device used in orthopedic surgery; implanted by surgeons. Components allow cemented or cementless fixation; porous surfaces on femoral/tibial components facilitate biological fixation in cementless applications. Tibial anchor provides supplemental fixation for cementless tibial tray implantation; anchor omitted for cemented applications. Implant geometry optimized for medial compartment; femoral/tibial sizes are interchangeable. System restores joint function, reduces pain, and addresses arthritis or failed prior procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing including tibial baseplate fatigue strength (ASTM F3140), range of motion, contact pressure, and tibiofemoral articulation (ASTM F2083), constraint testing (ASTM F1223), tibial insert dislodgement resistance (ASTM F1814), biomechanical evaluation of tibial tray/anchor assembly, porous surface fixation evaluation, additive manufacturing process validation, packaging/shelf-life validation, and bacterial endotoxin testing.
Indicated for patients with moderately disabling joint disease of the knee due to painful osteoarthritis or post-traumatic arthritis, patients requiring revision of previous unsuccessful surgical procedures (including prior unicompartmental knee arthroplasty), and patients with unicompartmental osteoarthritis as an alternative to tibial osteotomy.
Regulatory Classification
Identification
A knee joint femorotibial (uni-compartmental) metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surface. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra-high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial baseplate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
{0}------------------------------------------------
November 21, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Engage Surgical Mr. Nicholas Slater Chief Operating Officer 3225 N. Colorado St Chandler, Arizona 85225
Re: K190439
Trade/Device Name: Engage™ Partial Knee System Regulation Number: 21 CFR 888.3535 Regulation Name: Knee Joint Femorotibial (Uni-Compartmental) Metal/Polymer Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: NJD, HSX Dated: October 21, 2019 Received: October 22, 2019
Dear Mr. Slater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for William Jung, Ph.D. Director (Acting) DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190439
Device Name The Engage™ Partial Knee System
#### Indications for Use (Describe)
The Engage™ Partial Knee System is intended for medial unicompartmental knee arthroplasty to treat one or more of the following conditions:
- Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis.
- Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty
- As an alternative to tibial osteotomy in patients with Unicompartmental osteoarthritis.
The femoral component and tibial tray are intented fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application. When the tibial tray is implanted without the use of bone cement, the tibial anchor should be used. When the tibial tray is implanted with bone cement, the tibial anchor should not be used.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY for The Engage™ Partial Knee System
In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of substantial equivalence of The Engage™ Partial Knee System.
### A. SUBMITTERS INFORMATION
| Submitter Name: | Engage Surgical |
|---------------------|--------------------------------------------------|
| Submitter Address: | 3225 N. Colorado Street, Chandler, AZ 85225 USA |
| Contact Person: | Nick Slater |
| Title: | Chief Operating Officer |
| Phone Number: | 570-713-9227 |
| Fax Number: | 407-641-8611 |
| Date of Submission: | October 21, 2019 |
## B. MANUFACTURER & DEVICE IDENTIFICATION
| Manufacturer Name: | Engage Surgical |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer Address: | 3505 Lake Lynda Drive, Suite 206<br>Orlando, Florida 32817 |
| Device Trade Name: | The Engage™ Partial Knee System |
| Device Common Name: | Unicompartmental Knee |
| Classifications: | 21 CFR 888.3535 – Knee joint femorotibial (uni-compartmental)<br>metal/polymer porous-coated uncemented prosthesis<br>21 CFR 888.3520 – Knee joint femorotibial metal/polymer non-<br>constrained cemented prosthesis |
| Product Codes: | NJD and HSX |
| Classification Panel: | Orthopedic |
| C1. PRIMARY<br>PREDICATE DEVICE<br>K133811 | Mako Surgical Inc.'s Restoris Porous Partial Knee System |
| C2. ADDITIONAL<br>PREDICATE DEVICES<br>K033810<br>K163700 | Centerpulse Orthopedics, Inc.'s Natural-Knee II Unicompartmental<br>Knee System<br>BodyCAD Unicompartmental Knee System |
| C3. REFERENCE DEVICES | |
| K142066<br>K033363 | Stryker Spine AERO-LL Lumbar Cage System<br>Zimmer Unicompartmental Knee System (ZUK) |
{4}------------------------------------------------
### D. DEVICE DESCRIPTION
The Engage™ Partial Knee System is a unicompartmental knee system comprised of tibial trays, tibial inserts, tibial anchors, and femoral components. The implant geometry is optimized for the medial compartment. The Engage™ Partial Knee System permits all sizes of femur to be interchangeable with all sizes of tibial trays and associated tibial insert combinations. The tibial anchor are composed of Ti-6AI-4V alloy, the tibial insert is composed of UHMWPE, and the femoral component is composed of CoCrMo alloy. The tibial tray includes pegs and an engineered porous surface, while the femoral component includes pegs and a sintered porous coating; both of which are intended for biological fixation when used without bone cement and to enhance fixation when used with PMMA bone cement. The Engage™ Partial Knee System is intended for cemented or uncemented replacement of the medial compartment of the knee. The tibial anchors provide for supplemental fixation of the tibial tray when implanted without bone cement. The tibial anchor is not used when the tibial tray is implanted with bone cement.
#### E. INDICATIONS FOR USE
The Engage™ Partial Knee System is intended for medial unicompartmental knee arthroplasty to treat one or more of the following conditions:
- · Moderately disabling joint disease of the knee resulting from painful osteo- or post-traumatic arthritis.
- · Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty
- · As an alternative to tibial osteotomy in patients with unicompartmental osteoarthritis.
The femoral component and tibial tray are intended for cemented fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application. When the tibial tray is implanted without the use of bone cement, the tibial anchor should be used. When the tibial tray is implanted with bone cement, the tibial anchor should not be used.
#### F. PERFORMANCE DATA
Pre-clinical performance testing was performed for the Engage™ Partial Knee System per the FDA Guidance Document "Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA", ASTM and ISO consensus standards, and publicly available information. The performance data presented in this 510(k) notification for the Engage™ Partial Knee System demonstrate its substantial equivalence to the identified predicate device unicompartmental knee device.
- Tibial baseplate fatigue strength per three-point bend method as described in ASTM F3140.
- Range of Motion (ROM), contact pressure, and area of tibiofemoral articulation of UHMWPE ● inserts per ASTM F2083. Constraint testing / analysis per ASTM F1223.
- Resistance to dislodgement of the tibial insert per ASTM F1814.
- Evaluation of the tibial baseplate and tibial anchor assembly.
{5}------------------------------------------------
- Biomechanical evaluation of the tibial tray and anchor assembly under cyclic loading. .
- . Evaluation of the porous fixation surfaces per the FDA guidance document "Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement".
- Validation of additive manufacturing process for the tibial baseplate. ●
- . Validation studies of packaging and shelf life testing.
- Bacterial endotoxins test (BET), also known as Limulus Amebocyte Lysate (LAL) testing. .
- Validation of reusable instrument reprocessing parameters including sterilization with ONE ● TRAY® Sealed Container System (K052567).
#### G. SUBSTANTIAL EQUIVALENCE
The Engage™ Partial Knee System is substantially equivalent to the identified predicate device based on similarities in indications for use, materials, design, size, and performance data presented in this 510(k) notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.