K190401 · Boston Scientific Corporation · DQY · Mar 15, 2019 · Cardiovascular
Device Facts
Record ID
K190401
Device Name
MAMBA and MAMBA Flex Microcatheters
Applicant
Boston Scientific Corporation
Product Code
DQY · Cardiovascular
Decision Date
Mar 15, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The MAMBA and MAMBA Flex Microcatheters are intended to provide support to facilitate the placement of guidewires in the coronary vasculature, and can be used to exchange one guidewire for another. The microcatheters are also intended to assist in the delivery of contrast media into the coronary vasculature.
Device Story
MAMBA and MAMBA Flex are percutaneous microcatheters used in coronary vasculature; provide platform for guidewire exchange without losing position; assist in delivery of contrast media and therapeutic agents. Device features inner lumen for 0.014 in. guidewires; coil-reinforced shaft for torque; hydrophilic coating on distal portion for advancement; radiopaque marker at distal tip for visualization. Used by physicians in clinical settings. Operator advances catheter over guidewire to target lesion; provides support for wire placement; enables exchange of wires; allows injection of contrast media via proximal luer hub. Benefits include improved access to difficult coronary lesions and procedural efficiency.
Clinical Evidence
No clinical data. Substantial equivalence based on non-clinical bench testing. Performance criteria evaluated: dimensional requirements, delivery/retraction, wire movement, shaft/manifold torque, wire lumen burst, tip bond, guidewire exchange, shaft flex/kink, and dye flow. Results demonstrate device meets requirements for intended use.
Technological Characteristics
Percutaneous catheter; coil-reinforced shaft; hydrophilic distal coating; radiopaque distal marker; proximal luer hub. Compatible with 0.014 in. guidewires. Available in 135cm and 150cm lengths. Sterilized device. No software or electronic components.
Indications for Use
Indicated for patients requiring additional guidewire support to reach lesions in the coronary vasculature, facilitating guidewire placement, guidewire exchange, and delivery of contrast media.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
March 15, 2019
Boston Scientific Corporation Kevin Catalano Sr. Regulatory Affairs Specialist Three Scimed Place Maple Grove, Minnesota 55311
Re: K190401
Trade/Device Name: MAMBA and MAMBA Flex Microcatheters Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: February 15, 2019 Received: February 19, 2019
# Dear Kevin Catalano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Image /page/1/Picture/7 description: The image shows a digital signature. The signature is from Lydia S. Glaw -S. The date of the signature is 2019.03.15, and the time is 16:24:23 -04'00'.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190401
Device Name
MAMBA and MAMBA Flex Microcatheters
Indications for Use (Describe)
The MAMBA and MAMBA Flex Microcatheters are intended to provide support to facilitate the placement of guidewires in the coronary vasculature, and can be used to exchange one guidewire for another. The microcatheters are also intended to assist in the delivery of contrast media into the coronary vasculature.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
per 21 CFR §807.92
| Sponsor: | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Kevin Catalano<br>Three Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-2413<br>Fax: 763-494-2222<br>e-mail: kevin.catalano@bsci.com |
| Prepared | 15 February 2019 |
| Proprietary Name | MAMBA™ and MAMBA™ Flex<br>Microcatheters |
| Common Name | Percutaneous Catheter |
| Product Code | DQY |
| Classification | Class II, 21 CFR Part 870.1250 |
| Predicate Device | MAMBA™ and MAMBA™ Flex<br>Microcatheters<br>K171452 (21 Aug 2017) |
| Reference Device | Acuity Pro Guide catheter<br>K171612 (30 Jun 2017) |
{4}------------------------------------------------
## Device Description
The MAMBA™ and MAMBA™ Flex devices (referred to as the "MAMBA Microcatheters") are designed to provide a platform for the user to exchange quidewires without losing their position in the vasculature.
The MAMBA device is available in a 135cm length, while MAMBA Flex is available in 135cm and 150cm lengths.
The MAMBA Flex models have a smaller outer diameter and a more flexible distal portion than the MAMBA device.
Marks on the proximal portion of the catheter shaft and a radiopaque marker at the distal tip facilitate placement of the device.
The device's inner lumen permits the use of 0.014 in./0.36 mm or smaller quidewires. The shaft consists of a coil which allows the device to be torqued to assist in advancement. The distal portion of the shaft is coated hydrophilic coating to assist in advancement.
The proximal end has a single luer hub for flushing or injecting through the inner lumen and is for guide wire lumen access during a guidewire exchange.
The MAMBA Microcatheters are to be used in patients where vessel conditions require additional wire support to reach the lesion and will deliver radiopaque media and therapeutic agents to selected sites in coronary vasculature.
#### Indications for Use / Intended Use
The MAMBA™ and MAMBA™ Flex Microcatheters are intended to provide support to facilitate the placement of guidewires in the coronary vasculature, and can be used to exchange one guidewire for another.
The microcatheters are also intended to assist in the delivery of contrast media into the coronary vasculature.
## Comparison of Technological Characteristics
The MAMBA Microcatheters incorporate substantially equivalent design, packaging, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate, MAMBA Microcatheters, K171452.
{5}------------------------------------------------
# Non-clinical Performance Data
Determination of substantial equivalence is based on an assessment of non-clinical performance bench testing data.
# Bench Testing:
Bench testing was performed to evaluate physical integrity, functionality and performance of the MAMBA Microcatheters. Performance criteria includes: dimensional requirements, delivery and retraction, wire movement, shaft and manifold torque, wire lumen burst, tip bond, guidewire exchange, shaft flex and shaft kink, and dye flow.
The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing; therefore, this device is considered to be substantially equivalent to the predicate device.
## Conclusion
Based on the indications for use, technological characteristics, performance testing, the MAMBA Microcatheters have been shown to be appropriate for their intended use and are considered to be substantially equivalent to MAMBA Microcatheters, K171452.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.