RIVAL View Plating Systems and Reduce Fracture Plating Systems
Applicant
Orthofix Srl
Product Code
HRS · Orthopedic
Decision Date
Jun 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
RIVAL™ View Plating System™ and Reduce Fracture Plating System™ are intended for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The plates are available for use with Rival locking and non-locking bone screws. Plates and screws are intended for single use only. Screws are not intended for use in the spine. Examples of these procedures for which the RIVAL™ View Plating Systems™ may be used, but are not limited to: - First metatarsal osteotomies for hallux valgus correction including: - Opening base wedge osteotomy - Closing base wedge osteotomy - Crescentic osteotomy - Proximal Chevron osteotomy - Distal Chevron osteotomy (Austin) - First metatarsal fracture fixation - Arthrodesis of the first metatarsalcuneiform joint (Lapidus Fusion) - Arthrodesis of the first metatarsophalangeal joint (MTP) including: - Primary MTP Fusion due to hallux rigidus and/or hallux valgus - Revision MTP Fusion - Revision of failed first MTP Arthroplasty Implant - Calcaneal-Cuboid Fusion - Talonavicular Fusion - Evans Osteotomy Example of indications for which the RIVAL™ Reduce Plating Systems™ may be used, but are not limited to: - Metatarsal or metacarpal fractures and osteotomies - Phalanges fractures and osteotomies - Lapidus Fusion - Lisfranc Arthrodesis - 1st MPJ Arthrodesis - Forefoot osteotomies
Device Story
RIVAL View and Reduce Fracture Plating Systems consist of bone plates and locking/non-locking screws; used for stabilization, fixation, and reconstruction of small bones in extremities (hand, feet, wrist, ankles, fingers, toes). Devices are manufactured from Ti6Al4V titanium alloy. Intended for single-use; implanted by orthopedic surgeons during surgical procedures. System provides mechanical support for bone healing in fractures, osteotomies, and arthrodesis. Output is physical stabilization of bone segments. Benefits include restoration of skeletal alignment and support for bone fusion.
Clinical Evidence
No clinical data; substantial equivalence supported by mechanical performance testing, engineering assessment, and review of clinical literature on similar devices. Pyrogenicity testing was conducted.
Technological Characteristics
Materials: Ti6Al4V titanium alloy. Components: Bone plates and locking/non-locking screws. Design: Identical thickness, geometry, width, and plate angles to predicates. Sterilization: Same as predicates. Connectivity: None (mechanical implant).
Indications for Use
Indicated for stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers, and toes. Contraindicated for use in the spine.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
EDGE Orthopaedics VIEW and REDUCE Plating Systems (K140876)
Reference Devices
Orthopaedics REDUCE Fracture Plating System Line Extension (K142135)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol of the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
June 7, 2019
Orthofix Srl % Cheryl Wagoner Consultant Wagoner Consulting LLC PO Box 15729 Wilmington, North Carolina 24408
Re: K190388
Trade/Device Name: RIVAL View Plating Systems and Reduce Fracture Plating Systems Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: February 8, 2019 Received: February 19, 2019
Dear Cheryl Wagoner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190388
#### Device Name
RIVAL™ View Plating System™ and Reduce Fracture Plating System™
#### Indications for Use (Describe)
RIVAL™ View Plating System™ and Reduce Fracture Plating System™ are intended for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The plates are available for use with Rival locking and non-locking bone screws.
Plates and screws are intended for single use only. Screws are not intended for use in the spine.
Examples of these procedures for which the RIVAL™ View Plating Systems™ may be used, but are not limited to:
- · First metatarsal osteotomies for hallux valgus correction including:
- Opening base wedge osteotomy
- Closing base wedge osteotomy
- Crescentic osteotomy
- Proximal Chevron osteotomy
- Distal Chevron osteotomy (Austin)
- First metatarsal fracture fixation
- · Arthrodesis of the first metatarsalcuneiform joint (Lapidus Fusion)
- · Arthrodesis of the first metatarsophalangeal joint (MTP) including:
- Primary MTP Fusion due to hallux rigidus and/or hallux valgus
- Revision MTP Fusion
- Revision of failed first MTP Arthroplasty Implant
- · Calcaneal-Cuboid Fusion
- · Talonavicular Fusion
- · Evans Osteotomy
Example of indications for which the RIVAL™ Reduce Plating Systems™ may be used, but are not limited to:
- · Metatarsal or metacarpal fractures and osteotomies
- Phalanges fractures and osteotomies
- · Lapidus Fusion
- · Lisfranc Arthrodesis
- · 1st MPJ Arthrodesis
- · Forefoot osteotomies
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
## 510(k) K190388
### 510(k) Summary
A 510(k) Summary of the safety and effectiveness information upon which this substantial equivalence determination is based, is provided in this section.
## 510(k) Summary
(21 CFR 807.92)
#### Submitter information
| Submitter Name | Orthofix Srl |
|----------------|------------------------------------------------------|
| Address | Via delle Nazioni, 9 37012 Bussolengo (VR) - Italy |
| Telephone | + 39 045 6719.000 |
| Fax | + 39 045 6719.380 |
| Contact Person | Gianluca Ricadona<br>Sr. Quality & Regulatory Affairs Manager |
|----------------|---------------------------------------------------------------|
| Address | Via delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy |
| Telephone | + 39 045 6719 000 |
| Fax | + 39 045 6719 380 |
| email | GianlucaRicadona@orthofix.it |
| Date Prepared | June 7, 2019 |
#### Trade Name, Common Name, Classification
| Trade Name | RIVAL™ View Plating System™ and Reduce Fracture Plating<br>System™ |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Common Name | Plating System |
| Panel Code | Orthopedic |
| Class | Class II |
| Classification Name and<br>Product code | Plate, Fixation, Bone (21 CFR 888.3030, Product Code HRS)<br>Screw, Fixation, Bone (21 CFR 888.3040, Product Code HWC) |
#### Predicate device
| Predicate Device | 510(k) Number | Manufacturer |
|---------------------------------------------------------------|---------------|--------------|
| Primary Predicate | | |
| EDGE Orthopaedics VIEW and<br>REDUCE Plating Systems | K140876 | ORTHOFIX SRL |
| Reference device | | |
| Orthopaedics REDUCE Fracture<br>Plating System Line Extension | K142135 | ORTHOFIX SRL |
| Device description | RIVAL™ View Plating System™ and Reduce Fracture Plating<br>System™ has been designed to support multiple indications within<br>the forefoot and mid-foot. The Subject device introduces new line<br>extension codes, in sterile and non-sterile configuration.<br>Subject device is manufactured by the same material (Ti6Al4V) and<br>by the same manufacturer Orthofix Srl as the predicate EDGE<br>Orthopaedics VIEW and REDUCE Plating System. |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{4}------------------------------------------------
# Additional Information (AI) to the 510(k) K190388
| Indications for use | As the predicate EDGE Orthopaedics VIEW and REDUCE Plating<br>System, the Subject device are intended for use in stabilization and<br>fixation of fresh fractures, revision procedures, joint fusion, and<br>reconstruction of small bones of the hand, feet, wrist, ankles, fingers<br>and toes. The plates are available for use with Orthofix locking and<br>non-locking bone screws.<br>Plates and screws are intended for single use only. Screws are not<br>intended for use in the spine.<br>Examples of these procedures for which the Rival View Plating<br>System may be used, but are not limited to:<br>• First metatarsal osteotomies for hallux valgus correction including:<br>- Opening base wedge osteotomy<br>-Closing base wedge osteotomy<br>- Crescentic osteotomy<br>- Proximal Chevron osteotomy<br>- Distal Chevron osteotomy (Austin)<br>• First metatarsal fracture fixation<br>• Arthrodesis of the first metatarsal-cuneiform joint (Lapidus Fusion)<br>• Arthrodesis of the first metatarsophalangeal joint (MTP) including:<br>Primary MTP Fusion due to hallux ridgidus and/or hallux valgus |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | - Revision MTP Fusion<br>- Revision of failed first MTP Arthroplasty Implant<br>• Calcaneal-Cuboid Fusion<br>• Talonavicular Fusion<br>• Evans Osteotomy<br>Example of indications for which the Rival Reduce Fracture Plating<br>System may be used, but are not limited to: |
| | • Metatarsal or metacarpal fractures and osteotomies<br>• Phalanges fractures and osteotomies<br>• Lapidus Fusion<br>• Lisfranc Arthrodesis<br>• 1st MPJ Arthrodesis<br>• Forefoot osteotomies |
| Technological<br>Characteristics and<br>Intended Use | The fundamental scientific principles and technological characteristic,<br>including the intended use, material and general design are the same<br>as, or similar to, the predicate devices. Summary of the technological<br>characteristics:<br>✓ Plate thickness, geometry and width(s) are identical to the<br>predicates.<br>✓ Range of plate angles are identical to predicates.<br>✓ Identical material (Titanium Alloy) to the cited predicates<br>✓ Locking and non-locking screws same as the predicates,<br>✓ Sterile method, same as the predicates<br>✓ Indications for Use and anatomical site, , operating principles,<br>conditions of use are substantially equivalent to predicates.<br>The technological characteristics of the subject device and the<br>predicates are substantially equivalent to the predicates. |
| Performance<br>Analysis | Subject device has similar configuration, material, sizes and design<br>as the predicate devices. Mechanical performance and engineering<br>assessment with worst case of subject device and corresponding<br>predicate devices, confirm that subject devices have equivalent or<br>better strength and resistance.<br>Any potential hazard have been evaluated and controlled through<br>Risk Management activities.<br>The review of clinical literatures on similar devices still support the<br>clinical data of the Subject devices with no additional clinical<br>information.<br>Pyrogenicity testing was conducted in support of substantial<br>equivalence. |
| Conclusion | Based upon equivalences in: intended use, site of application, ,<br>conditions of use, performance data, operating principles, and<br>according to the results of nonclinical testing, the subject devices<br>belonging to RIVAL™ VIEW Plating System™ and REDUCE<br>Fracture Plating System™ have been shown to be safe, as effective,<br>and performs as well as or better than the legally marketed predicate<br>device.<br>Therefore, the Subject device is expected to be substantially<br>equivalent to the legally marketed predicate devices. |
{5}------------------------------------------------
# Additional Information (AI) to the 510(k) K190388
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.