← Product Code [MQV](/productcode/MQV) · K190371

# Morpheus-C (K190371)

_Biogennix, LLC · MQV · Aug 9, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190371

## Device Facts

- **Applicant:** Biogennix, LLC
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Aug 9, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

BIOGENNIX Morpheus-C is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. It is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides bone void filler that resorbs and is replaced with bone during the healing process.

## Device Story

Morpheus-C is a moldable, resorbable, osteoconductive bone graft substitute. It consists of 1-2mm biphasic calcium salt granules (osteoSPAN) suspended in a biocompatible organic binder (polymer and type I collagen). The device is packed into bony voids or gaps by a surgeon. The organic binder is rapidly absorbed in-situ, leaving the osteoconductive scaffold of granules and collagen to facilitate bone healing and resorption. It is intended for non-structural skeletal defects in the extremities and pelvis.

## Clinical Evidence

No human clinical data. Bench testing included biocompatibility (ISO 10993-1: cytotoxicity, sensitization, systemic toxicity, pyrogenicity, genotoxicity, implantation). In vivo testing used a critically sized defect model evaluated via histology, histomorphometry, x-ray, and micro-CT; results confirmed new bone formation, bone remodeling, and implant resorption with no adverse reactions.

## Technological Characteristics

Composition: 1-2mm biphasic calcium salt granules (calcium carbonate/calcium phosphate) and organic binder (biocompatible polymer and type I collagen). Granules: ~65% porous, 500-micron interconnected pores. Form: Moldable graft substitute. Sterilization: Validated. Biocompatibility: ISO 10993-1 compliant.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- Morpheus ([K132377](/device/K132377.md))
- NovaBone Macroform ([K140946](/device/K140946.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Biogennix, LLC % Elaine Duncan President Paladin Medical, Inc. P.O. Box 560 Stillwater, Minnesota 55082

Re: K190371

Trade/Device Name: Morpheus-C Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: July 9, 2019 Received: July 10, 2019

#### Dear Ms. Duncan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K190371

Device Name Morpheus-C

Indications for Use (Describe)

Indications for Use

BIOGENNIX Morpheus-C is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. It is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities, and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides bone void filler that resorbs and is replaced with bone during the healing process.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

### I. SUBMITTER

### Submitted on behalf of:

| Company Name:<br>Address: | BIOGENNIX, LLC<br>1641 McGaw Ave.<br>Irvine, CA 92614                                                  |
|---------------------------|--------------------------------------------------------------------------------------------------------|
| Telephone:<br>Fax:        | 949-253-0094<br>949-266-5800                                                                           |
| by:                       | Elaine Duncan, M.S.M.E., RAC<br>President, Paladin Medical, Inc.<br>PO Box 560<br>Stillwater, MN 55082 |
| Telephone:<br>Fax:        | 715-549-6035<br>715-549-5380                                                                           |
| Contact Person:           | Elaine Duncan                                                                                          |
| Date Prepared:            | July 9, 2019                                                                                           |

### II. SUBJECT DEVICE

| Trade Name:        | Morpheus-C                                      |
|--------------------|-------------------------------------------------|
| Common Name(s):    | Bone void filler, Bone graft substitute         |
| Regulation Number: | 21 CFR88.3045                                   |
| Regulation Name:   | Resorbable calcium salt bone void-filler device |
| Product Code:      | MQV                                             |
| Regulatory Class:  | II                                              |

### III. PREDICATE DEVICE

The contents of this submission have demonstrated that Morpheus-C is substantially equivalent to its primary predicate Morpheus (K132377) and secondary predicate NovaBone Macroform (K140946) when used as a bone graft substitute.

## IV. DEVICE DESCRIPTION

Morpheus-C is a moldable, resorbable osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules suspended in a biocompatible organic binder to facilitate shaping and containment of the implant.

## Biogennix, LLC.

{4}------------------------------------------------

The osteoSPAN granules in Morpheus-C are approximately 65% porous, biphasic calcium salts with interconnected pores having a nominal cross-section of 500 microns. The primary composition of each granule is calcium carbonate with a thin laver of calcium phosphate throughout its entire porosity.

The organic binder in Morpheus-C is a combination of a biocompatible polymer and type I collagen fibers. The polymer is rapidly absorbed in-situ, leaving behind osteoSPAN granules and collagen fibers as an osteoconductive scaffold. The collagen in Morpheus-C provides improved intraoperative handling.

# V. INDICATIONS FOR USE

BIOGENNIX Morpheus-C is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. It is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., the extremities and pelvis). These defects may be surqically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides bone void filler that resorbs and is replaced with bone during the healing process.

### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The function, intended use and technological characteristics of the subject device are substantially equivalent to the predicate devices cleared under 510(k) premarket notifications K132377 and K140946.

# VII. PERFORMANCE DATA

Biogennix followed the "Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device: Guidance for Industry and FDA. June 2, 2003, as well as the company's own design controls and risk analysis procedures to ensure that Morpheus-C is safe and effective for use.

Biocompatibility evaluation of Morpheus-C was conducted in accordance with "Use of International Standard ISO 10993-1. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." As a tissue/bone permanent implant device, Morpheus-C satisfied a battery of tests assessing the following biological effects:

- . Cytotoxicity
- Sensitization
- Acute Systemic Toxicity ●
- Material-Mediated Pvrogenicity ●
- Subacute/Subchronic toxicity ●
- Genotoxicity

# Biogennix, LLC.

{5}------------------------------------------------

### 510(k) Summary-Continued

- . Implantation
In vivo testing of Morpheus-C was conducted using a critically sized defect model. Device performance was evaluated at multiple time points against the primary predicate and negative controls using histology, histomorphometry, x-ray, and micro-CT analyses. The critical nature of the model was validated by the negative controls. No adverse reactions were noted at the implant site or in distant organs: new bone formation, bone remodeling, and implant resorption for the test materials were confirmed with time. Based on the endpoints and results of this study, Morpheus-C was concluded to be substantially equivalent to the primary predicate Morpheus.

Bench testing results for various physical and chemical characteristics of the primary predicate apply to the subject device since the addition of the type I collagen does not affect the granule chemistry, crystallinity, porosity, pore interconnectivity, or density. Sterilization validation and shelf-life aging studies support Morpheus-C labeling statements.

# VIII. CONCLUSIONS

The non-clinical data presented in this submission demonstrate that Morpheus-C is substantially equivalent to its predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K190371](https://fda-staging.innolitics.com/device/K190371)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).