K190368 · Canon, Inc. · MQB · Mar 13, 2019 · Radiology
Device Facts
Record ID
K190368
Device Name
Enhanced Feature Software Pack CXDI Series
Applicant
Canon, Inc.
Product Code
MQB · Radiology
Decision Date
Mar 13, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
As a part of the CXDI series radiography system, the CXDI Control Software when used with a compatible CXDI detector is intended to provide digital image capture, processing, and display for conventional film/screen radiographic examinations. This device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures including specialist areas like intensive care, trauma, and pediatric work. This device is not intended for fluoroscopic, angiographic, or mammography applications.
Device Story
The Enhanced Feature Software Pack for CXDI Series is a software update for Canon CXDI radiography systems. It introduces 'One Shot Long Length' imaging, which captures images across multiple compatible flat panel detectors (CXDI-710C Wireless, CXDI-410C Wireless) in a single X-ray exposure. The software automatically stitches these images into a single long-length image (up to 120cm) and provides tools for manual adjustment of stitch positions. The device is used in clinical settings by healthcare professionals to replace traditional film/screen systems. The output is a digital radiographic image displayed for diagnostic review. The software also retains previously cleared scatter correction features. The system benefits patients by enabling long-length imaging with a single exposure, potentially reducing radiation dose and procedure time compared to multi-exposure methods.
Clinical Evidence
Bench testing only. The device modification was evaluated using FMEA methodology to analyze risks and hazardous impacts. Risk control and protective measures were implemented and verified. The system complies with established electrical safety and electromagnetic compatibility standards (IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-32) and FCC/IC standards for wireless detectors.
Technological Characteristics
Software-based image processing for digital radiography. Features automatic and manual image stitching for long-length imaging (up to 120cm). Compatible with Canon CXDI-710C and CXDI-410C Wireless detectors. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-32, 47CFR 2.1093, FCC Part 15 Subpart B, and ICES-003.
Indications for Use
Indicated for digital radiographic image capture, processing, and display in general purpose diagnostic procedures, including intensive care, trauma, and pediatric applications. Not indicated for fluoroscopic, angiographic, or mammography use.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Canon DIGITAL RADIOGRAPHY CXDI Series Detectors (K170332)
Canon DIGITAL RADIOGRAPHY CXDI Series Detectors (K171270)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Canon, Inc. % Ms. Katie Reneson Regulatory Specialist Ken Block Consulting 800 East Campbell Road. Suite 202 RICHARDSON TX 75081
March 13, 2019
Re: K190368
Trade/Device Name: Enhanced Feature Software Pack for CXDI Series Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: February 14, 2019 Received: February 15, 2019
Dear Ms. Reneson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website
(http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Thalia Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K190368
#### Device Name
Enhanced Feature Software Pack for CXDI Series
Indications for Use (Describe)
As a part of the CXDI series radiography system, the CXDI Control Software when used with a compatible CXDI detector is intended to provide digital image capture, and display for conventional film/screen radiographic examinations. This device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures including specialist areas like intensive care, trauma, and pediatric work. This device is not intended for fluoroscopic, angiographic, or mammography applications.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <div> <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the word "Canon" in red font. The font is bold and slightly italicized. The word is easily readable and takes up most of the image.
### 5. Special 510(k) SUMMARY
| Submitter: | Canon, Inc.<br>30-2 Shimomaruko, 3-chrome<br>Ohta-ku, Tokyo 146-8501 Japan |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Tatsuya Yamazaki<br>General Manager<br>TEL: 81-3-3758-2111;<br>FAX: 81-3-5482-3960<br>yamazaki.tatsuya@mail.canon |
| Date Prepared: | February 14, 2019 |
| Submission Type: | Special 510(k) Submission |
| Trade Name: | Enhanced Feature Software Pack for CXDI Series |
| Common Name: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| Regulatory Number: | 21 CFR 892.1680 |
| Classification Name: | Stationary X-ray system |
| Product Code: | MQB |
| Primary Predicate #1: | MQB K153312 Scatter Correction for CXDI Series |
| Predicate #2: | MQB Canon DIGITAL RADIOGRAPHY CXDI Series Detectors<br>510(k) # Models<br>K170332 CXDI-710C Wireless<br>K171270 CXDI-410C Wireless |
The subject of this Special 510(k) submission is a change to the CXDI Control Device Description: Software to incorporate the ability to capture an automatically stitched long length image in a single exposure. The addition of the One Shot Long Length to the cleared Scatter Correction of CXDI Series is the subject of this Special 510(k) Submission. The One Shot Long Length stitches long length images into a single image which is accomplished by using multiple detectors, a single exposure, and the automatic stitching software. The One Shot Long Length software feature along with the Scatter Correction cleared under K153312 make up the Enhanced Feature Software Pack for CXDI Series. The Scatter Correction clearance and compatible detectors are not impacted by this submission. This submission adds the One Shot Long Length feature to limited compatible FPDs. The One Shot Long Length feature included in the Enhanced Feature Software Pack for CXDI Series also includes features for manual stitching to allow for users to manually adjust and fine-tune the stitch positions after the automatic stitch operation. By incorporating the One Shot Long Length imaging into the CXDI Control Software, images up to 120cm in length can be acquired in one exposure. The firmware within the flat panel detectors did not require updating. The flat panel detectors that have been previously cleared by FDA and are compatible with the Enhanced Feature Software Pack for CXDI Series have not been physically modified and performance have not changed.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Canon" in red font. The font is bold and slightly italicized. The word is easily readable and takes up most of the image.
#### Indications for Use Statement
The intended use of the modified device, as described in the labeling, has not changed as a result of the modification(s).
As a part of the CXDI series radiography system, the CXDI Control Software when used with a compatible CXDI detector is intended to provide digital image capture, processing, and display for conventional film/screen radiographic examinations. This device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures including specialist areas like intensive care, trauma, and pediatric work. This device is not intended for fluoroscopic, angiographic, or mammography applications.
## Summary of Technological Characteristics:
Comparisons with the predicate devices show the characteristics of the proposed addition of the One Shot Long Length feature to the Enhanced Feature Software Pack for CXDI Series to be substantially equivalent to the predicate device(s). The One Shot Long Length feature of the Enhanced Feature Software Pack for CXDI Series can only be used with compatible Canon CXDI detectors CXDI-710C Wireless, CXDI-410C Wireless.
| Trade Name | Proposed Device<br>Enhanced Feature Software Pack<br>for CXDI Series | Predicate<br>Control Software V2.14<br>Scatter Correction | |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k)<br>Submitter<br>[Number] | Canon, Inc.<br>[K190368] | Canon, Inc.<br>[K153312] | IDENTICAL |
| Indication<br>for Use | As a part of the CXDI series<br>radiography system, the CXDI<br>Control Software when used<br>with a compatible CXDI<br>detector is intended to provide<br>digital image capture,<br>processing, and display for<br>conventional film/screen<br>radiographic examinations. This<br>device is intended to replace<br>radiographic film/screen systems<br>in all general purpose diagnostic<br>procedures including specialist<br>areas like intensive care, trauma,<br>and pediatric work. This device<br>is not intended for fluoroscopic.<br>angiographic, or mammography<br>applications. | As a part of the CXDI series<br>radiography system, the CXDI<br>Control Software when used<br>with a compatible CXDI<br>detector is intended to provide<br>digital image capture,<br>processing, and display for<br>conventional film/screen<br>radiographic examinations.<br>This device is intended to<br>replace radiographic<br>film/screen systems in all<br>general purpose diagnostic<br>procedures including specialist<br>areas like intensive care,<br>trauma, and pediatric work.<br>This device is not intended for<br>fluoroscopic, angiographic, or<br>mammography applications. | IDENTICAL |
| Scatter<br>Correction | An image can be created (with<br>high contrasts) by using software<br>algorithm, in clinical field,<br>without a grid | An image can be created (with<br>high contrasts) by using<br>software algorithm, in clinical<br>field, without a grid | IDENTICAL |
| One Shot<br>Long Length<br>Imaging | One exposure to obtain images<br>across multiple detectors,<br>automatically stitched together<br>with ability to make manual<br>adjustments after the automatic<br>stitching. | N/A | MODIFIED |
| Software<br>/Version | CXDI Control Software V2.17 | CXDI Control Software V2.14<br>(highest version cleared<br>previously) | MODIFIED |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Canon" in red, bold letters. The font is sans-serif and slightly italicized. The word is centered and takes up most of the frame.
| Trade Name | Proposed Device<br>Canon DIGITAL<br>RADIOGRAPHY CXDI Series<br>Detectors containing Enhanced<br>Feature Software Pack for CXDI<br>Series | Predicate Device<br>Canon DIGITAL<br>RADIOGRAPHY CXDI<br>Series Detectors | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k)<br>Submitter<br>[Number] | Canon, Inc.<br>[K190368] | Canon, Inc.<br>[K170332, K171270] | IDENTICAL |
| Indication<br>for Use | The DIGITAL<br>RADIOGRAPHY CXDI Series<br>Detectors provides digital image<br>capture for conventional<br>film/screen radiographic<br>examinations. This device is<br>intended to capture for display<br>radiographic images of human<br>anatomy, and to replace<br>radiographic film/screen systems<br>in all general purpose diagnostic<br>procedures. This device is not<br>intended for mammography<br>applications. | The DIGITAL<br>RADIOGRAPHY CXDI<br>Series Detectors provides<br>digital image capture for<br>conventional film/screen<br>radiographic examinations.<br>This device is intended to<br>capture for display<br>radiographic images of human<br>anatomy, and to replace<br>radiographic film/screen<br>systems in all general purpose<br>diagnostic procedures. This<br>device is not intended for<br>mammography applications. | IDENTICAL |
| Scatter<br>Correction | An image can be created (with<br>high contrasts) by using software<br>algorithm, in clinical field,<br>without a grid | An image can be created (with<br>high contrasts) by using<br>software algorithm, in clinical<br>field, without a grid | IDENTICAL |
| One Shot<br>Long Length<br>Imaging | One exposure to obtain images<br>across multiple detectors,<br>automatically stitched together<br>with ability to make manual<br>adjustments after the automatic<br>stitching. | N/A | MODIFIED |
| Firmware<br>/Version | 04-02-0c-00 | 04-02-0b-00 | MODIFIED |
Performance: The hardware within the Canon Digital Radiography CXDI Series Detectors has not been modified, the detectors have the same performance, biocompatibility, effectiveness, and safety and is substantially equivalent to the predicate device. The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted. As reported in prior submissions to FDA, the compatible detectors comply with the U.S. Performance Standard for radiographic equipment and with relevant voluntary safety standards for Electrical Safety and Electromagnetic Compatibility testing, specifically IEC standards 60601-1, 60601-1-2, 60601-1-3, and 60601-2-32. The wireless detectors also comply with the FCC test standard for SAR, specifically 47CFR 2.1093 and for EMI test regulations FCC Part 15 Subpart B:2012 Class A and ICES-003 Issue 5:2012 Class A. Canon, Inc. considers the Enhanced Feature Software Pack for CXDI Series to be Conclusion: substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.