Paradigm All-Suture Anchor 1.8mm, White/Black #2

K190333 · Conmed Corporation · MBI · Mar 18, 2019 · Orthopedic

Device Facts

Record IDK190333
Device NameParadigm All-Suture Anchor 1.8mm, White/Black #2
ApplicantConmed Corporation
Product CodeMBI · Orthopedic
Decision DateMar 18, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor systems thereby stabilize the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

Device Story

Paradigm™ All-Suture Anchor is a non-absorbable orthopedic fixation device; used to reattach soft tissue to bone. Device consists of high-strength flat-braided suture co-braided with nylon monofilament; preloaded on a disposable driver. Surgeon performs a pilot hole in bone; inserts anchor; anchor expands to provide fixation. Used in arthroscopic or open procedures; provides stabilization of soft tissue during healing. Single-use; EO sterilized.

Clinical Evidence

Bench testing only. No clinical data provided. Verification testing included reliability, ultimate fixation strength, cyclic testing, transportation, pyrogen/endotoxin testing, biocompatibility, shelf-life, post-aging functional testing, and MR safety. Validation testing included user validation, packaging, labeling, and sterilization.

Technological Characteristics

Materials: High-strength flat-braided suture co-braided with nylon monofilament. Design: Expandable all-suture anchor, 1.8mm diameter. Insertion: Drilled pilot hole, press-fit. Sterilization: Ethylene Oxide (EO). Connectivity: None (mechanical device).

Indications for Use

Indicated for reattachment of soft tissue (ligaments, tendons, joint capsules) to bone in orthopedic surgical procedures (arthroscopic or open). Contraindicated in patients with bone/soft tissue pathology affecting fixation, conditions limiting activity restriction, foreign body sensitivity, active sepsis/infection, or for posterior spinal element fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below. March 18, 2019 ConMed Corporation Orjada Dervishleri Regulatory Affairs Specialist 525 French Road Utica, New York 13502 Re: K190333 Trade/Device Name: Paradigm™ All-Suture Anchor 1.8mm. White/Black #2 Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: February 13, 2019 Received: February 14, 2019 Dear Orjada Dervishleri: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Laurence D. Coyne -S For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | K190333<br>Page 1 of 1<br>Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number (if known) | K190333 | | Device Name | Paradigm™ All-Suture Anchor 1.8mm, White/Black #2 | | Indications for Use (Describe)<br>Intended Use | The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. | | Indications for Use | The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor systems thereby stabilize the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period. | | Type of Use (Select one or both, as applicable) | | | <div style="display:flex; align-items:center;"><img src="data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAwAAAAMCAYAAABWdVkkAAAAAXNSR0IArs4c6QAAAARnQU1BAACxjwv8YQUAAAAJcEhZcwAADsMAAA7DAcdvqGQAAABLSURBVEhL7c0BCoAgEAVB/6FHFhc4iCoS8QkJwgfggQKckw9gQv+8hx9EuKzH+WduKjKnEhHl5MEnQRGwQFYAAAAASUVORK5CYII=" style="width:12px; height:12px; margin-right:5px;"/>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><img src="data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAwAAAAMCAYAAABWdVkkAAAAAXNSR0IArs4c6QAAAARnQU1BAACxjwv8YQUAAAAJcEhZcwAADsMAAA7DAcdvqGQAAABLSURBVEhL7c0BCoAgEAVB/6FHFhc4iCoS8QkJwgfggQKckw9gQv+8hx9EuKzH+WduKjKnEhHl5MEnQRGwQFYAAAAASUVORK5CYII=" style="width:12px; height:12px; margin-right:5px;"/>Over-The-Counter Use (21 CFR 801 Subpart C)</div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRASraff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the Conmed Corporation logo. The logo consists of a blue square with a white curved shape inside, followed by the word "Conmed" in black, with a blue line underneath. Below the line, the word "Corporation" is written in smaller black letters. ## 510(k) SUMMARY In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CONMED Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number __K190333 ### I. SUBMITTER CONMED Corporation 11311 Concept Blvd Largo, Florida 33773 Phone: 727-399-5491 Fax: 727-399-5264 Contact Person: Orjada Dervishleri Date Prepared: February 13th, 2019 ### II. DEVICE NAME | Device Name: | Paradigm™ All-Suture Anchor | |----------------------|------------------------------------------------| | Common Name: | Nonabsorbable Suture Anchor System | | Classification Name: | Fastener, fixation, nondegradable, soft tissue | | Regulatory Class: | Class II, per 21 CFR Part 888. 3040 | | Product Codes: | MBI | ### III. PREDICATE/ LEGALLY MARKET DEVICE | Device Name: | ConMed Linvatec Y-Knot™ All-Suture Anchor | |---------------|-------------------------------------------| | Company Name: | ConMed Linvatec | | 510(k) #: | K111779 | - IV. REFERENCE DEVICE Device Name: Y-Knot Flex All-Suture Anchor, w/Two #2 (5 Metric) Hi-Fi Sutures, 1.8mm ConMed Linvatec Company Name: 510(k) #: K131035 ## V. DEVICE DESCRIPTION The CONMED Paradigm™ All-Suture Anchor manufactured from High Strength Flat Braided Suture, co-braided with a nylon monofilament, and threaded with one #2 (5 metric) Hi-Fi® suture strand. The ParadigmTM All-Suture Anchor is provided single-use, sterile, and preloaded on disposable driver. The anchor, suture, and disposable driver are EO Sterilized. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a blue square on the left side with a white curved shape inside. To the right of the square is the word "CONMED" in black, with a horizontal blue line underneath. Below the line is the word "CORPORATION" in smaller, black letters. ## VI. INTENDED USE/ INDICATIONS FOR USE The non-absorbable suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period. ## VII. COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The following table represents a summary of the technological characteristics between the proposed and the predicate device. | | Paradigm™ All-Suture Anchor<br>Proposed Device | ConMed Linvatec Y-Knot®<br>All-Suture Anchor<br>Predicate Device | Y-Knot Flex All-Suture<br>Anchor, w/Two #2 (5 Metric)<br>Hi-Fi Sutures, 1.8mm<br>Reference Device | |--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| | Device Description | The CONMED Paradigm™ All-Suture<br>Anchor manufactured from High Strength<br>Flat Braided Suture, co-braided with a nylon<br>monofilament, and threaded with one #2 (5<br>metric) Hi-Fi® suture strand. The<br>Paradigm™ All-Suture Anchor is provided<br>single-use, sterile, and preloaded on<br>disposable driver. The anchor, suture, and<br>disposable driver are EO Sterilized. | Same | Same | | Intended Use | The non-absorbable suture anchors are<br>intended to reattach soft tissue to bone in<br>orthopedic surgical procedures. | Same | Same | | Indication for Use | The device may be used in either<br>arthroscopic or open surgical procedures.<br>After the suture is anchored to the bone, it<br>may be used to reattach soft tissue, such as<br>ligaments, tendons, or joint capsules to the<br>bone. The suture anchor system thereby<br>stabilizes the damaged soft tissue, in<br>conjunction with appropriate postoperative<br>immobilization, throughout the healing<br>period. | Same | Same | | Contraindications | 1. Pathological conditions of bone which<br>would adversely affect the Paradigm™<br>All-Suture Anchor.<br>2. Pathological conditions in the soft tissue to<br>be repaired or reconstructed which would<br>adversely affect suture fixation.<br>3. Physical conditions that would eliminate,<br>or tend to eliminate, adequate implant<br>support or retard healing.<br>4. Conditions which tend to limit the patient's<br>ability or willingness to restrict activities | Same | Same | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Conmed Corporation. The logo consists of a blue square with a white curved shape inside, followed by the word "Conmed" in black, with a blue line underneath. Below the line is the word "Corporation" in smaller, black letters. | | or follow directions during the healing<br>period.<br>5. Attachment of artificial ligaments or other<br>implants.<br>6. Foreign body sensitivity, known or<br>suspected allergies to implant and/or<br>instrument materials.<br>7. This device is not approved for screw<br>attachment or fixation to the posterior<br>elements (pedicles) of the cervical,<br>thoracic or lumbar spine.<br>8. Patients with active sepsis or infection. | | | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|--------------------------------------------------------------------------------------------------------------| | Components | All-Suture Anchor<br>Disposable Driver<br>Suture<br>Handle | Same | Same | | Technological<br>Characteristics | Flat-braid suture anchor<br>One #2 suture<br>Drilled pilot hole<br>Press-fit insertion<br>Expandable design | Same | Flat-braid suture anchor<br>Two #2 sutures<br>Drilled pilot hole<br>Press-fit insertion<br>Expandable design | ## VIII. PERFORMANCE DATA Testing has been completed to demonstrate that the Paradigm™ All-Suture Anchor performs as intended and is substantially equivalent to the predicate device. Bacterial endotoxin testing was conducted and met the endotoxin limits. Completed testing includes the following: ## Verification Testing - Reliability - Ultimate Fixation Strength ● - Transportation ● - Cyclic ● - Pyrogen ● - Biocompatibility - Shelf-life - Post Aging Functional Testing - MR Safety Testing ● # Validation Testing - User Validation - Packaging ● - Labeling ● - Sterilization ● ## IX. CONCLUSION The Paradigm™ All-Suture Anchor is either substantially equivalent or identical in design, materials, intended use, principles of operation, and technical characteristics to the predicate ConMed Linvatec Y-Knot™ All-Suture Anchor. Based upon the findings of our performance testing, the differences present no new issues of safety and efficacy, and the Paradigm™ All-Suture Anchor is substantially equivalent to the ConMed Linvatec Y-Knot™ All-Suture Anchor (K111779).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...