The ENConnect 40 mm Screw Cap to ENPlus Adapter is intended to connect enteral giving sets, equipped with an ENPlus connector, to enteral nutrition crown cap bottles or wide neck reservoirs in order to deliver nutritional fluids.
Device Story
Non-sterile, single-use adapter; connects enteral giving sets (ENPlus connector) to wide neck reservoirs or crown cap bottles containing nutritional fluids. Device consists of 40 mm screw cap, ENPlus port, vent, and gasket. Used in clinical settings to facilitate enteral feeding. Healthcare providers attach the adapter to the nutrition reservoir and connect the feeding set to the ENPlus port to enable fluid delivery. Benefits include secure interconnection between standard enteral nutrition reservoirs and feeding sets.
Clinical Evidence
No clinical data. Performance demonstrated via non-clinical bench testing, including biocompatibility (ISO 10993-1), flow rate, fluid leakage, and ISO 18250-3 mechanical testing (pressure, stress cracking, tensile, and separation resistance).
Technological Characteristics
Non-sterile, single-use mechanical adapter. Components: 40 mm screw cap, ENPlus port, vent, gasket. Materials tested for biocompatibility per ISO 10993-1. Complies with ISO 18250-3 for enteral connectors. No electrical or radiation components.
Indications for Use
Indicated for patients requiring enteral nutrition delivery via crown cap bottles or wide neck reservoirs using enteral giving sets with ENPlus connectors.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
U Deliver Bolink ENFit Enteral Feeding Sets (K170555)
Submission Summary (Full Text)
{0}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
April 9, 2019
Cedic S.r.1. % Crystal Koelper President Koelper Consulting, LLC 268 Biltmore Drive North Barrington, IL 60010
Re: K190327
> Trade/Device Name: ENConnect 40 mm Screw Cap to ENPlus Adapter Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: PIF Dated: February 12, 2019 Received: February 13, 2019
Dear Crystal Koelper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Section 5: 510(k) Summary of Safety and Effectiveness
## Submitter / 510(k) Holder
| Company: | Cedic s.r.l. |
|-----------------|-----------------------------------------------------------------|
| Address: | Via Liberazione, 63/9<br>20068 Peschiera Borromeo (MI)<br>Italy |
| Phone: | +1-847-345-6920 |
| Contact Person: | Crystal Koelper |
| Date Prepared: | April 4, 2019 |
#### Device Name and Classification
| Trade Name: | ENConnect 40 mm Screw Cap to ENPlus Adapter |
|----------------------|---------------------------------------------------------|
| Model Number(s): | A520 |
| Classification Name: | Gastrointestinal Tube and Accessories (21 CFR 876.5980) |
| Product Code: | PIF |
| Class: | II |
| Predicate Device(s) | |
U Deliver Bolink ENFit Enteral Feeding Sets (K170555)
#### Device Description
The ENConnect 40 mm Screw Cap to ENPlus Adapter (the proposed device) is a non-sterile, single use device consisting of screw cap with a 40 mm thread, an ENPlus port, vent, and gasket. The proposed device is intended to connect enteral giving sets, equipped with an ENPlus connector, via the ENPlus port to wide neck reservoirs and/or crown cap bottles containing enteral nutrition.
#### Indications for Use
The ENConnect 40 mm Screw Cap to ENPlus Adapter is intended to connect enteral giving sets, equipped with an ENPlus connector, to enteral nutrition crown cap bottles or wide neck reservoirs in order to deliver nutritional fluids.
{4}------------------------------------------------
### Product Comparison Summary
The proposed and predicate enteral feeding devices are both intended to connect enteral nutrition reservoirs to enteral devices (enteral feeding tubes or sets). Both the proposed and predicate devices operate by providing a means of interconnecting enteral reservoirs to other enteral feeding devices (tubes and/or sets). The designs of both the proposed and predicate devices are different, but the differences do not impact the safety and effectiveness of the proposed device. A comparison of the proposed and predicate devices is summarized below:
| Criteria for Comparison | Proposed Device Compared to Predicate<br>Device |
|---------------------------------|-------------------------------------------------|
| Intended Use | |
| Indications | Similar and SE |
| Clinical Settings | Same |
| Target Population | Same |
| Design Characteristics | |
| Screw Cap (including<br>gasket) | Similar and SE. |
| Tubing | Different, proposed does not have tubing. |
| Distal Connector | Similar and SE. |
| Sterility | Same |
| Materials | |
| Screw Cap | Same |
| Gasket | Same |
| Vent | Different, predicate does not have a vent. |
| Tubing | Different, proposed does not have tubing. |
| Distal Connector | Different |
| Performance | |
| Use/Human Factors | Same |
| Theory of operation | Same |
| Risk assessment | Same |
| Safety | |
| Mechanical | Similar and SE |
| Electrical | Same - Not an electrical device. |
| Radiation | Same - Not a radiation emitting device. |
#### Non-Clinical Performance Testing
Non-clinical performance testing was conducted to demonstrate that the ENConnect 40 mm Screw Cap to ENPlus Adapter is substantially equivalent to the predicate device and is safe and effective for its intended use. In addition, ENConnect 40 mm Screw Cap to ENPlus Adapters that underwent accelerated aging conditioning were tested to substantiate the product shelf life.
The following tests were performed:
- Biocompatibility testing per ISO 10993-1
- Flow Rate testing with Enteral Nutrition
- Fluid Leakage Testing
{5}------------------------------------------------
K190327 Page 3 of 3
- ISO 18250-3 Evaluations: ●
- Dimension Evaluation O
- O Positive Pressure Liquid Leakage Test
- Sub-atmospheric Pressure Air Leakage Test O
- O Stress Cracking
- O Resistance to Separation from Axial Load
- Resistance to Separation from Unscrewing O
- Resistance to Overriding O
- Disconnection by Unscrewing O
- Tensile Testing
- Distribution Simulation
- Packaging Tests:
- o Distribution Simulation
- o Burst Test
- o Peel Test
- O Dye Penetration Test
#### Clinical Performance Testing
Clinical performance testing was not required to demonstrate the performance of the ENConnect 40 mm Screw Cap to ENPlus Adapter. The ENConnect 40 mm Screw Cap to ENPlus Adapter performance has been adequately demonstrated by the completion of non-clinical performance testing.
#### Animal Testing
Animal testing was not required to demonstrate the performance of the ENConnect 40 mm Screw Cap to ENPlus Adapter. The ENConnect 40 mm Screw Cap to ENPlus Adapter performance has been adequately demonstrated by the completion of non-clinical performance testing.
#### Conclusion
The results of the non-clinical performance testing demonstrate that the ENConnect 40 mm Screw Cap to ENPlus Adapter is as safe and effective as the predicate device. Therefore, the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.