# EZE SIT Valvulotome (K190267)

_Lemaitre Vascular · MGZ · Oct 30, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190267

## Device Facts

- **Applicant:** Lemaitre Vascular
- **Product Code:** [MGZ](/productcode/MGZ.md)
- **Decision Date:** Oct 30, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4885
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The EZE SIT Valvulotome is intended to render venous valves incompetent during in situ bypass procedures. This includes distal infrainguinal bypass when a non-anatomic position is required (e.g., profunda to anterior tibial artery), composite vein infrainguinal bypass, or aorto-renal bypass procedures.

## Device Story

Surgically invasive device for disrupting venous valves during in situ bypass procedures; used in patients with peripheral vascular disease. Device consists of a catheter with a luer-lock hub and a distal threaded connector for attaching interchangeable 2mm, 3mm, or 4mm diameter cutting heads. Physician selects appropriate cutter head size based on vessel diameter. Central lumen allows for irrigation during the procedure. Device operates via manual force applied by the physician to lyse valve cusps, enabling retrograde blood flow. Used in clinical settings by physicians. Benefits include restoration of blood flow to extremities in patients requiring bypass.

## Clinical Evidence

No clinical data. Bench testing only, including sterilization validation (ISO 11135-1:2014), shelf-life/aging studies (6-year accelerated aging), and biocompatibility testing.

## Technological Characteristics

Materials: 303/304 Stainless Steel, Polyurethane/Barium Sulfate, Nylon, Silicone, ABS, PET, PETG. Energy: Manual force. Form factor: Catheter with interchangeable cutting heads (2mm, 3mm, 4mm). Connectivity: None. Sterilization: Ethylene Oxide (EtO) to SAL 10^-6.

## Regulatory Identification

An external vein stripper is an extravascular device used to remove a section of a vein.

## Predicate Devices

- Tru-Incise Valvulotome ([K930011](/device/K930011.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 30, 2019

LeMaitre Vascular, Inc. Xiang Zhang VP of Regulatory 63 Second Avenue Burlington, Massachusetts 01803

Re: K190267

Trade/Device Name: EZE SIT Valvulotome Regulation Number: 21 CFR 870.4885 Regulation Name: External Vein Stripper Regulatory Class: Class II Product Code: MGZ Dated: September 26, 2019 Received: September 30, 2019

Dear Xiang Zhang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carmen Gacchina Johnson, Ph.D. Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190267

Device Name EZE SIT Valvulotome

Indications for Use (Describe)

The EZE SIT Valvulotome is intended to render venous valves incompetent during in situ bypass procedures. This includes distal infrainguinal bypass when a non-anatomic position is required (e.g., profunda to anterior tibial artery), composite vein infrainguinal bypass, or aorto-renal bypass procedures.

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for LeMaitre Vascular. The logo consists of a blue and white abstract shape on the left, followed by the words "LeMaitre" in a bold, teal font. Below "LeMaitre" is the word "VASCULAR" in a smaller, lighter blue font. The logo is clean and professional, and it conveys a sense of trust and reliability.

SECTION 8: 510(k) Summary

| I. SUBMITTTER              | LeMaitre Vascular, Inc.<br>63 Second Avenue<br>Burlington, MA 01803                                            |
|----------------------------|----------------------------------------------------------------------------------------------------------------|
| Establishment Registration | 1220948                                                                                                        |
| Phone:                     | 781-425-1685                                                                                                   |
| Fax:                       | 781-425-5049                                                                                                   |
| Contact Person:            | Xiang Zhang<br>VP of Regulatory<br>Email: xzhang@lemaitre.com                                                  |
| Date Prepared              | September 26, 2019                                                                                             |
| I. DEVICE                  |                                                                                                                |
| Device Name:               | EZE-SIT Valvulotome                                                                                            |
| Trade Name:                | EZE-SIT Valvulotome                                                                                            |
| Common Name:               | External Vein Stripper                                                                                         |
| Classification Panel:      | Cardiovascular                                                                                                 |
| Regulation Number:         | 21CFR §870.4885                                                                                                |
| Class:                     | II (2)                                                                                                         |
| Product Code:              | MGZ                                                                                                            |
| Prior Submissions:         | No prior submissions have been made for the changes<br>detailed in this traditional 510(k).                    |
| II. PREDICATE DEVICE       | Tru-Incise Valvulotome (K930011)<br>This predicate device has not been subject to a design-<br>related recall. |

K190267

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for LeMaitre Vascular. The logo features a stylized blue wave-like design on the left, followed by the words "LeMaitre" in a bold, teal font. Below "LeMaitre" is the word "VASCULAR" in a smaller, lighter blue font.

### III. DEVICE DESCRIPTION:

The EZE SIT Valvulotome is a medical device utilized for disruption of venous valves, and is used in patients that suffer from peripheral vascular disease, for which an attending physician determines the relative benefit of in situ bypass to restore blood flow to the extremities. The EZE-SIT Valvulotome is comprised of a catheter with a luer-lock hub at the proximal end and a threaded connector at the distal end. The catheter is designed with a lumen that extends throughout its entire length. The threaded connector allows attachment of 2 mm, 3 mm and 4 mm diameter cutting heads. These cutting heads are also configured with a central lumen. This design allows the physician to select the cutter head diameter that is best suited for the particular vessel being prepared. The central lumen allows irrigation during the procedure. The cutting head design minimizes vessel wall contact while effectively disrupting valve leaflet tissue. The EZE SIT Valvulotome is surgically invasive and intended for less than 24-hour use. This medical device has no software components. This device is not an in vitro diagnostic device.

## IV. INDICATIONS FOR USE:

## V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The EZE SIT Valvulotome is intended to render venous valves incompetent during in situ bypass procedures. This includes distal infrainguinal bypass when a nonanatomic position is required (e.g., profunda to anterior tibial artery), composite vein infrainguinal bypass, or aorto-renal bypass procedures.

Overall Description, Device Design: The EZE SIT Valvulotome comprises a catheter with a luer lock hub at the proximal end, a threaded connector at the distal end, and a lumen that extends throughout its entire length. The threaded connector allows attachment of 2mm, 3mm and 4mm diameter cutting heads, which also feature a central lumen. The design allows the physician to select the cutter head diameter best suited for a particular vein being prepared. The central lumen allows irrigation during the procedure. The cutting heads minimize vessel wall contact while effectively disrupting valve leaflet tissue. The design of the EZE SIT Valvulotome has not changed since the original clearance (K930011).

Materials: The materials of the EZE SIT Valvulotome are identical to those used for the Tru-Incise Valvulotome (K930011). All components are patient-

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for LeMaitre Vascular. The logo features a stylized blue and white graphic on the left, resembling a wave or a stylized letter 'L'. To the right of the graphic is the company name, "LeMaitre," in a bold, teal font, with the registered trademark symbol next to it. Below "LeMaitre" is the word "VASCULAR" in a smaller, lighter teal font.

contacting unless indicated otherwise. EZE SIT is
surgically invasive and intended for <24h use.

| Part Name                                                             | Material                            |
|-----------------------------------------------------------------------|-------------------------------------|
| EZE SIT Threaded<br>Leader, 2mm                                       |                                     |
| EZE SIT Threaded<br>Leader, 3mm                                       | 303 Stainless Steel                 |
| EZE SIT Threaded<br>Leader, 4mm                                       |                                     |
| EZE SIT Connecting<br>Cannula, 2mm                                    |                                     |
| EZE SIT Connecting<br>Cannula, 3mm                                    | 304 Stainless Steel                 |
| EZE SIT Connecting<br>Cannula, 4mm                                    |                                     |
| EZE SIT Cutter Head<br>with Edge, 2mm                                 |                                     |
| EZE SIT Cutter Head<br>with Edge, 3mm                                 | 303 Stainless Steel                 |
| EZE SIT Cutter Head<br>with Edge, 4mm                                 |                                     |
| EZE SIT Introduction<br>Head                                          | 303 Stainless Steel                 |
| EZE SIT Vein Sleeve                                                   | Polyurethane 75D/<br>Barium Sulfate |
| EZE SIT Head Holder T<br>(non patient-contacting)                     | Nylon                               |
| EZE SIT Small Head<br>Holder Tubing (non-<br>patient contacting)      |                                     |
| EZE SIT Large Head<br>Holder Tubing (non-<br>patient contacting)      | Silicone                            |
| EZE SIT Extra Large<br>Head Holder Tubing<br>(non-patient contacting) |                                     |
| EZE SIT Vein Sleeve<br>Tether Tubing (non-<br>patient contacting)     |                                     |
| EZE SIT Tributary<br>Sleeve, 2mm                                      | Polyurethane/ Barium<br>Sulfate     |
| EZE SIT Tributary<br>Sleeve, 4mm                                      |                                     |
| EZE SIT Shaft Spring                                                  | 304 Stainless Steel                 |
| EZE SIT Threaded                                                      | 303 Stainless Steel                 |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for LeMaitre Vascular. The logo features a stylized blue and white graphic to the left of the company name. The company name is in a bold, sans-serif font, with "LeMaitre" in a darker blue and "VASCULAR" in a lighter blue below it.

| Connector                                               |                                                                 |
|---------------------------------------------------------|-----------------------------------------------------------------|
| EZE SIT Catheter Hub                                    | Acrylonitrile Butadiene<br>Styrene (ABS)                        |
| <b>Packaging Materials</b> (all non-patient contacting) |                                                                 |
| EZE SIT Head Holder<br>Pouch                            | Tyvek/Polyethylene<br>Terephthalate (PET)/<br>Polyethylene (PE) |
| EZE SIT Device Pouch                                    | Polyethylene (PE)                                               |
| EZE SIT Tyvek Lid                                       | Tyvek                                                           |
| EZE SIT Packaging Tray                                  | Polyethylene<br>Terephthalate Glycol<br>(PETG)                  |

Chemical Composition: All of the components of the EZE SIT Valvulotome remain unchanged from the original device design cleared in K930011.

Energy Source: The EZE SIT Valvulotome is a device that transduces manual force provided by the user into the lysis of valve cusps, thus allowing retrograde blood flow in a vein. EZE SIT uses and provides no other source of energy of electromechanical, radiological or other origin. The aspect of energy source supplied to the medical device remains unchanged from the predicate device cleared in K930011.

Summary: All aspects of the medical device, including the design, materials, chemical composition and energy source of the subject device remain unchanged from the original predicate device cleared in K930011, and thus support substantial equivalence of the subject to the predicate device.

### VI. PERFORMANCE DATA

N/A. This submission is intended to simply indicate a new manufacturer for a currently cleared device design that has not changed since its clearance (K930011). This submission includes 3 types of nonclinical data, including Sterilization, Shelf Life/Aging, and Biocompatibility.

Sterilization: The EZE SIT is Ethylene Oxide (EtO) sterilized to a Sterility Assurance Level (SAL) of 10-6. Sterilization is validated with half-cycle using ISO 11135-1: 2014. Residuals measured at 48h after sterilization indicate that the device has a maximum of EtO of 0.6 mg, and a maximum of <0.1 mg of Ethylene Chlorohydrin (ECH) (VP-160032). Endotoxin is measured by LAL method.

Shelf Life/Aging studies provided here confirmed the shelf life of 2.5 years as reported for the predicate

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is the logo for LeMaitre Vascular. The logo features a stylized blue and white symbol on the left, resembling a curved ribbon or wave. To the right of the symbol is the company name, "LeMaitre," in a bold, teal font, with a registered trademark symbol next to it. Below "LeMaitre" is the word "VASCULAR" in a smaller, lighter blue font.

device (K930011), but also extended the shelf life to 6 years, all based on accelerated aging. Thus, the Shelf Life/Aging studies support the conclusion of substantial equivalence.

Biocompatibility: Upon acquisition of the right to manufacture the device for sale outside the US market, LeMaitre Vascular performed a biocompatibility battery that recapitulated the same tests as performed for the predicate device cleared in K930011. Like the predicate device, the subject device detailed here passed all tests in the biocompatibility battery, thus supporting substantial equivalence of these devices. None.

LeMaitre Vascular has demonstrated that the EZE SIT Valvulotome is substantially equivalent to the predicate device based on its intended use and fundamental scientific technology, and that the subject device is as safe, as effective, and performs as well as the predicate device.

Summary of non-clinical and clinical Studies: VII. CONCLUSION:

---

**Source:** [https://fda-staging.innolitics.com/device/K190267](https://fda-staging.innolitics.com/device/K190267)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).