← Product Code [GXI](/productcode/GXI) · K190259

# Vesta RF Cannula (K190259)

_Biomerics · GXI · Feb 7, 2020 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190259

## Device Facts

- **Applicant:** Biomerics
- **Product Code:** [GXI](/productcode/GXI.md)
- **Decision Date:** Feb 7, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4725
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Vesta RF Cannula, in combination with an RF generator and probe, are intended for use in radiofrequency (RF) heat lesion procedures for relief of pain.

## Device Story

Vesta RF Cannula is an insulated cannula with an active tip used in radiofrequency (RF) heat lesion procedures for pain relief. It functions by directing RF energy into target soft tissues to create lesions. The device is used in hospital surgical suites by physicians familiar with RF lesion techniques. It is designed for use with compatible pain management generators and probes (max voltage ≤280 V). The cannula is a single-use, sterile device. By enabling precise RF energy delivery, it facilitates targeted tissue necrosis to alleviate pain.

## Clinical Evidence

No clinical testing was performed. Substantial equivalence is supported by bench testing and biocompatibility assessments. Bench testing confirmed compliance with ISO standards for needle diameter, cannula verification, tip inspection, luer acceptance, and sterilization. Biocompatibility testing (ISO 10993) for limited contact (≤24 hours) demonstrated the device is non-irritant, non-sensitizing, non-cytotoxic, non-toxic, and non-pyrogenic.

## Technological Characteristics

Insulated cannula with active tip; stainless steel needle with polymer insulation; thermoplastic hubs. Dimensions: 18-22G diameter, ≤15cm length, 5-10mm active tip. Energy source: external RF generator. Sterilization: Ethylene Oxide (EtO). Standards: IEC 60601-2-2, ISO 7864, ISO 9626, ISO 10993 series, ISO 11135, ISO 11607, ISO 15223-1, ISO 80369-7.

## Regulatory Identification

A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.

## Predicate Devices

- Nimbus Electrosurgical Radiofrequency Multitined Expandable Electrode ([K121773](/device/K121773.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 7, 2020

Biomerics Jake Wakley Vice President of Quality and Regulatory 6030 West Harold Gatty Drive Salt Lake City, Utah 84116

Re: K190259

Trade/Device Name: Vesta RF Cannula Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency Lesion Probe Regulatory Class: Class II Product Code: GXI Dated: January 3, 2020 Received: January 8, 2020

Dear Jake Wakley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew Krueger, M.S.E. Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190259

Device Name Vesta RF Cannula

Indications for Use (Describe)

The Vesta RF Cannula, in combination with an RF generator and probe, are intended for use in radiofrequency (RF) heat lesion procedures for relief of pain.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Biomerics. The logo consists of a blue abstract symbol on the left and the word "BIOMERICS" in blue on the right. The abstract symbol is made up of interconnected circles, and the word "BIOMERICS" is written in a bold, sans-serif font.

510(k) Summary (21 CFR 807.92)

#### SUBMITTER I.

Submitter Name: Address:

Telephone: Fax:

Biomerics 6030 West Harold Gatty Drive Salt Lake City, UT 84116 801-355-2705 801-355-3045

Contact Person: Email of Contact: Date Prepared:

Jake Wakley jwakley@biomerics.com 01/24/2020

#### II. DEVICE

Device Name: Common/Usual Name: Classification Name: Classification Panel: Regulation Device Name: Regulation Number: Regulatory Class: Product Code:

Vesta RF Cannula Cannula, Radiofrequency Lesion Radiofrequency Lesion Probe Neurology Radiofrequency Lesion Probe 21 CFR 882.4725 2 GXI

## III. PREDICATE DEVICE

| Predicate Name:         | Nimbus Electrosurgical Radiofrequency Multitined<br>Expandable Electrode |
|-------------------------|--------------------------------------------------------------------------|
| Common/Usual Name:      | Probe, Radiofrequency Lesion                                             |
| Classification Name:    | GXI – Radiofrequency Lesion Probe                                        |
| Premarket Notification: | K121773                                                                  |
| Manufacturer:           | Biomerics                                                                |

Per the FDA Medical Device Recalls Database, this predicate has not been subject to a design related recall.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for BIOMERICS. The logo consists of a blue abstract symbol on the left and the word "BIOMERICS" in blue on the right. The abstract symbol is made up of interconnected circles, and the word "BIOMERICS" is written in a bold, sans-serif font.

# IV. DEVICE DESCRIPTION

The Vesta RF Cannula consists of an insulated cannula with an active tip for use in radiofrequency heat lesion procedures for the relief of pain. It is designed to be used with compatible pain management generators and probes that have a maximum voltage rating less than or equal to 280 V. The Vesta RF Cannula is sterilized and intended for single use only.

#### INDICATIONS FOR USE V.

The Vesta RF Cannula, in combination with an RF generator and probe, are intended for use in radiofrequency (RF) heat lesion procedures for relief of pain.

#### VI. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Vesta RF Cannula device is substantially equivalent to the predicate device, the Nimbus Electrosurgical Radiofrequency Multitined Expandable Electrode. Both the subject and the predicate device consist of an insulated cannula with an active tip that directs RF energy into target tissues. Both are intended for use in radiofrequency (RF) heat lesion procedures for relief of pain. Both have the same intended use, indications for use, similar design, similar sizes, similar or the same materials and similar properties. See comparison table below.

| Attribute                                              | Predicate Device                                                                                                                                                                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                           |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Device Name                                            | Nimbus Electrosurgical<br>Radiofrequency<br>Multitined Expandable<br>Electrode                                                                                                                                                                 | Vesta RF Cannula                                                                                                                                                                                                                                                                                                                                                                                                                                    | Substantially<br>Equivalent or<br>Identical to<br>Predicate                                               |
| 510(k)<br>Number                                       | K121773                                                                                                                                                                                                                                        | K190259 (Pending)                                                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                       |
| 510(k)<br>Submitter                                    | Biomerics                                                                                                                                                                                                                                      | Biomerics                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical to<br>Predicate                                                                                 |
| Trade Name                                             | Nimbus Electrosurgical<br>Radiofrequency<br>Multitined Expandable<br>Electrode                                                                                                                                                                 | Vesta RF Cannula                                                                                                                                                                                                                                                                                                                                                                                                                                    | Substantially<br>Equivalent to<br>Predicate                                                               |
| Classification Name                                    | Probe, Radiofrequency<br>Lesion                                                                                                                                                                                                                | Probe, Radiofrequency<br>Lesion                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical to<br>Predicate                                                                                 |
| Regulatory<br>Class                                    | 2                                                                                                                                                                                                                                              | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical to<br>Predicate                                                                                 |
| Classification<br>Regulation                           | 21 CFR 882.4725                                                                                                                                                                                                                                | 21 CFR 882.4725                                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical to<br>Predicate                                                                                 |
| Attribute                                              | Predicate Device                                                                                                                                                                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                      | Substantially Equivalent or Identical to Predicate                                                        |
| Device Name                                            | Nimbus Electrosurgical<br>Radiofrequency<br>Multitined Expandable<br>Electrode                                                                                                                                                                 | Vesta RF Cannula                                                                                                                                                                                                                                                                                                                                                                                                                                    | Substantially<br>Equivalent or<br>Identical to<br>Predicate                                               |
| Product Code                                           | GXI                                                                                                                                                                                                                                            | GXI                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical to<br>Predicate                                                                                 |
| Review Panel /<br>Division                             | Neurology                                                                                                                                                                                                                                      | Neurology                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical to<br>Predicate                                                                                 |
| Variants                                               | • NIM-050-10BB<br>• NIM-100-10BB<br>• NIM-150-10BB<br>• NIM-100-10BB-CS                                                                                                                                                                        | • NLT-100-22-10-BB<br>• NLT-100-22-05-BB<br>• NLT-100-22-10-CS<br>• NLT-150-22-10-BB<br>• NLT-150-22-10-CS<br>• NLT-50-22-10-BB<br>• NLT-100-20-10-BB<br>• NLT-100-20-10-CS<br>• NLT-100-20-10-CH<br>• NLT-150-20-10-BB<br>• NLT-150-20-10-CS<br>• NLT-100-18-10-BB<br>• NLT-100-18-10-CS<br>• NLT-150-18-10-BB<br>• NLT-150-18-10-CS<br>• NLT-100-18-10-CH<br>• NLT-150-20-10-CH<br>• NLT-150-18-10-CH<br>• NLT-100-22-10-CH<br>• NLT-150-22-10-CH | Substantially<br>Equivalent to<br>Predicate                                                               |
| Intended Use                                           | Intended for use in<br>radiofrequency (RF) heat<br>lesion procedures for<br>relief of pain.                                                                                                                                                    | Intended for use in<br>radiofrequency (RF) heat<br>lesion procedures for<br>relief of pain.                                                                                                                                                                                                                                                                                                                                                         | Identical to<br>Predicate (The<br>Intended Use has<br>not changed as a<br>result of the<br>Modifications) |
| Attribute                                              | Predicate Device                                                                                                                                                                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                      | Substantially Equivalent or Identical to Predicate                                                        |
| Device Name                                            | Nimbus Electrosurgical<br>Radiofrequency<br>Multitined Expandable<br>Electrode                                                                                                                                                                 | Vesta RF Cannula                                                                                                                                                                                                                                                                                                                                                                                                                                    | Substantially<br>Equivalent or<br>Identical to<br>Predicate                                               |
| Indications for<br>Use                                 | The Nimbus<br>Electrosurgical<br>Radiofrequency<br>Multitined Expandable<br>Electrode, in combination<br>with an RF generator and<br>probe, are intended for<br>use in radiofrequency<br>(RF) heat lesion<br>procedures for relief of<br>pain. | The Vesta RF Cannula, in<br>combination with an RF<br>generator and probe, are<br>intended for use in<br>radiofrequency (RF) heat<br>lesion procedures for<br>relief of pain.                                                                                                                                                                                                                                                                       | Identical to<br>Predicate                                                                                 |
| Clinical<br>Mechanism of<br>Action                     | Insulated cannula with an<br>active tip that directs RF<br>energy into target tissue.                                                                                                                                                          | Insulated cannula with an<br>active tip that directs RF<br>energy into target tissue.                                                                                                                                                                                                                                                                                                                                                               | Identical to<br>Predicate                                                                                 |
| Anatomical<br>Sites                                    | Necrosis of soft tissues                                                                                                                                                                                                                       | Necrosis of soft tissues                                                                                                                                                                                                                                                                                                                                                                                                                            | Identical to<br>Predicate                                                                                 |
| Where Used                                             | Hospital, Surgical Suite                                                                                                                                                                                                                       | Hospital, Surgical Suite                                                                                                                                                                                                                                                                                                                                                                                                                            | Identical to<br>Predicate                                                                                 |
| Active Tip                                             | Deployable tines to<br>modify the shape of the<br>lesion.                                                                                                                                                                                      | Multiple active tip lengths<br>available.                                                                                                                                                                                                                                                                                                                                                                                                           | Substantially<br>Equivalent to<br>Predicate                                                               |
| Specific Drug<br>Use                                   | None                                                                                                                                                                                                                                           | None                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical to<br>Predicate                                                                                 |
| Human<br>Factors                                       | Restricted to use by<br>physicians familiar with<br>radio-frequency lesion<br>techniques.                                                                                                                                                      | Restricted to use by<br>physicians familiar with<br>radio-frequency lesion<br>techniques.                                                                                                                                                                                                                                                                                                                                                           | Identical to<br>Predicate                                                                                 |
| Standards Met                                          | • ASTM D4169-16<br>• AAMI TIR28<br>• BS EN 15223-1<br>• IEC 60601-2-2<br>• ISO 7864<br>• ISO 10993-1<br>• ISO 10993-5<br>• ISO 10993-7<br>• ISO 10993-10<br>• ISO 11135<br>• ISO 11607-1                                                       | • IEC 60601-2-2<br>• ISO 7864<br>• ISO 9626<br>• ISO 10993-1<br>• ISO 10993-5<br>• ISO 10993-7<br>• ISO 10993-10<br>• ISO 11135<br>• ISO 11607-1<br>• ISO 13485<br>• ISO 14971                                                                                                                                                                                                                                                                      | Substantially<br>Equivalent to<br>Predicate                                                               |
| Attribute                                              | Predicate Device                                                                                                                                                                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                           |
| Device Name                                            | Nimbus Electrosurgical<br>Radiofrequency<br>Multitined Expandable<br>Electrode                                                                                                                                                                 | Vesta RF Cannula                                                                                                                                                                                                                                                                                                                                                                                                                                    | Substantially<br>Equivalent or<br>Identical to<br>Predicate                                               |
|                                                        | • ISO 11607-2<br>• ISO 13485<br>• ISO 14971<br>• ISO 80369-7                                                                                                                                                                                   | • ISO 15223-1<br>• ISO 80369-7<br>• ISTA 2A                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                           |
| Physical<br>Description<br>(Diameter,<br>Length, etc.) | Diameter: 17G<br>Length: ≤15cm<br>Active Length: 4-10mm                                                                                                                                                                                        | Diameter: 18-22G<br>Length: ≤15cm<br>Active Length: 5-10mm                                                                                                                                                                                                                                                                                                                                                                                          | Substantially<br>Equivalent to<br>Predicate                                                               |
| Components /<br>Materials                              | Hubs:<br>Thermoplastics<br>Cannula/Needle:<br>Stainless steel with<br>polymer insulation<br>Tines:<br>Stainless Steel                                                                                                                          | Hubs:<br>Thermoplastics<br>Cannula/Needle:<br>Stainless steel with<br>polymer insulation                                                                                                                                                                                                                                                                                                                                                            | Substantially<br>Equivalent to<br>Predicate                                                               |
| General<br>Materials                                   | Biocompatible<br>(Polyethylene,<br>Polycarbonate, Stainless<br>Steel, Nylon, Silicone<br>and Polyethylene<br>Terephthalate (PET),<br>Biocompatible Colorants)                                                                                  | Biocompatible<br>(Polycarbonate, Stainless<br>Steel, and Polyethylene<br>Terephthalate (PET), &<br>Biocompatible Colorants)                                                                                                                                                                                                                                                                                                                         | Substantially<br>Equivalent to<br>Predicate                                                               |
| Insulation                                             | IEC 60601 compliant<br>PET to protect user and<br>patient from errant<br>electrical discharge.                                                                                                                                                 | IEC 60601 compliant<br>PET to protect user and<br>patient from errant<br>electrical discharge.                                                                                                                                                                                                                                                                                                                                                      | Substantially<br>Equivalent to<br>Predicate                                                               |
| Electrical<br>Safety                                   | Conforms to IEC 60601-<br>2-2                                                                                                                                                                                                                  | Conforms to IEC 60601-<br>2-2                                                                                                                                                                                                                                                                                                                                                                                                                       | Identical to<br>Predicate                                                                                 |
| Single-Use                                             | Yes                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical to<br>Predicate                                                                                 |
| Sterilization                                          | EtO                                                                                                                                                                                                                                            | EtO                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical to<br>Predicate                                                                                 |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Biomerics. The logo consists of a blue abstract symbol on the left and the word "BIOMERICS" in blue on the right. The abstract symbol is made up of interconnected circles, and the word "BIOMERICS" is written in a bold, sans-serif font.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Biomerics. The logo consists of a blue abstract symbol on the left and the word "BIOMERICS" in blue on the right. The abstract symbol is made up of interconnected circles, and the word "BIOMERICS" is written in a bold, sans-serif font.

6030 West Harold Gatty Drive, • Salt Lake City, UT 84116 • Telephone: 801-355-2705 • Fax: 801-355-3045 • Website: www.biomerics.com

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Biomerics. The logo consists of a blue abstract symbol on the left and the word "BIOMERICS" in blue on the right. The abstract symbol is made up of interconnected circles, and the word "BIOMERICS" is written in a bold, sans-serif font.

Table 1: Technological Characteristics with the Predicate Device

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Biomerics. The logo consists of a blue abstract symbol on the left and the word "BIOMERICS" in blue capital letters on the right. The abstract symbol is made up of interconnected circles, and the word "BIOMERICS" is written in a bold, sans-serif font.

## VII. PERFORMANCE DATA

### Bench Testing

Bench tests are listed below in Table 2.

| Standard                                | Test Name                      | Result |
|-----------------------------------------|--------------------------------|--------|
| ISO 6009                                | Needle Diameter                | Passed |
| ISO 9626                                | Cannula Verification           | Passed |
| ISO 7864                                | Needle Tip Inspection          | Passed |
| ISO 80369-7                             | Luer Acceptance                | Passed |
| ISO 15223-1                             | Label Proof Approval           | Passed |
| ISO 11135, ISO 10993-<br>7, ISO 11138-2 | Sterilization Assessment       | Passed |
| ISO 11607-1                             | Stability and Aging Assessment | Passed |

Table 2: Vesta RF Cannula Bench Testing

### Clinical Testing

No clinical testing was performed.

### Biocompatibility

All Vesta RF Cannula subject devices tested for biocompatibility were the same configuration: 150mm length, 18G diameter, 10mm active tip, with both green and blue hub variants either pooled or tested separately. The subject device was evaluated per ISO 10993 as a limited contact device (≤24 hours), and was found to be biocompatible for its intended use. The subject device is therefore equivalent to the predicate device.

| Test                                                                   | Result                                                                                                                                                                                                                                                                                               | Conclusions    |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Irritation –<br>Intracutaneous<br>Injection (Pooled<br>Blue/Green Hub) | The test article sites did not show a<br>significantly greater biological reaction<br>than the sites injected with the control<br>article.                                                                                                                                                           | Non-irritant   |
| Sensitization –<br>Kligman<br>Maximization (Pooled<br>Blue/Green Hub)  | The USP 0.9% Sodium Chloride for<br>Injection (NaCl) and Cottonseed Oil<br>(CSO) extracts of the test article, NLT-<br>150-18-10-CS Blue Hub Lot 18121-1<br>NLT-150-18-10-CS Green Hub lot<br>18121.1, elicited no reaction at the<br>challenge (0% sensitization), following<br>an induction phase. | Non-sensitizer |
| Cytotoxicity – MEM<br>Elution (Blue Hub)                               | Cell monolayers treated with test sample<br>exhibited no reactivity (Grade 0).                                                                                                                                                                                                                       | Non-cytotoxic  |
| Cytotoxicity – MEM<br>Elution (Green Hub)                              | Cell monolayers treated with test sample<br>exhibited no reactivity (Grade 0).                                                                                                                                                                                                                       | Non-cytotoxic  |

6030 West Harold Gatty Drive, • Salt Lake City, UT 84116 • Telephone: 801-355-2705 • Fax: 801-355-3045 • Website: www.biomerics.com

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Biomerics. The logo consists of a blue abstract graphic to the left of the word "BIOMERICS" in a dark blue, sans-serif font. The graphic is made up of interconnected circles and lines, creating a network-like design.

| Test                             | Result                                                                                                                                                                                                                                           | Conclusions                                            |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Systemic Toxicity<br>(Blue Hub)  | None of the test article treated animals were observed with clinical signs consistent with toxicity at any of the observation periods.                                                                                                           | Non-toxic                                              |
| Systemic Toxicity<br>(Green Hub) | None of the test article treated animals were observed with clinical signs consistent with toxicity at any of the observation periods.                                                                                                           | Non-toxic                                              |
| EO Residuals                     | EO and ECH residuals were tested per ISO 10993-7 Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals and were below allowable limits established in Section 4.3 for a limited exposure device (≤24 hours). | EO sterilization is appropriate for the subject device |
| Material Mediated<br>Pyrogen     | The material-mediated pyrogenicity risk of the Vesta RF Cannula was assessed and was found to be acceptable                                                                                                                                      | Non-pyrogenic                                          |

Table 3: Vesta RF Cannula Biocompatibility Testing

## VIII. CONCLUSION

Through performance bench testing results, it has been demonstrated that the subject device is substantially equivalent to the predicate device, Nimbus Electrosurgical Radiofrequency Multitined Expandable Electrode, K121773.

---

**Source:** [https://fda-staging.innolitics.com/device/K190259](https://fda-staging.innolitics.com/device/K190259)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).