K190256 · Epimed International, Inc. · GXI · Mar 8, 2019 · Neurology
Device Facts
Record ID
K190256
Device Name
Rulo Radiofrequency Lesion Probe
Applicant
Epimed International, Inc.
Product Code
GXI · Neurology
Decision Date
Mar 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4725
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Rulo™ Radiofrequency Lesion Probe is an injection probe which may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made.
Device Story
Rulo Radiofrequency Lesion Probe is a single-use, surgically invasive device used by physicians in clinical settings to treat pain. It functions as an injection probe and RF electrode. The device is inserted into the spine using a flexible introducer accessory. Nerve localization is achieved via RF electrostimulation or contrast medium injection under radiography. Once localized, the device delivers RF energy to create a lesion on pain-conducting nerve tissue or delivers local anesthetic for nerve blocks. The probe features a directional active tip to focus energy and fluid delivery. The device connects to an external RF generator. By disrupting electrical signal conduction in targeted nerves, the device provides pain relief.
Clinical Evidence
Bench testing only. Performance testing included biocompatibility (ISO 10993-5, -10, -11), mechanical testing (ISO 594-1, ISO 594-2, ISO 7864), and safety/performance standards (ANSI/AAMI/EN 60601-1, BS 6196). No clinical data provided.
Technological Characteristics
Constructed of steel cannula with polymer insulation, hub, cable, connector, cap, and PVC fluid injection port. Features directional active tip for focused energy/fluid delivery. Connects to external RF generators. Dimensions: 16G and 18G models. Biocompatibility per ISO 10993. Standards: ISO 594-1/2 (Luer tapers), ISO 7864 (needles), ANSI/AAMI/EN 60601-1 (electrical safety), BS 6196 (epidural needles).
Indications for Use
Indicated for percutaneous nerve blocks with local anesthetic or radiofrequency lesioning to treat pain by disrupting nerve tissue conduction.
Regulatory Classification
Identification
A radiofrequency lesion probe is a device connected to a radiofrequency (RF) lesion generator to deliver the RF energy to the site within the nervous system where a lesion is desired.
Predicate Devices
Diros OWL Sterile Single Use R.F. Insulated Cannulae Models 466 and DHC (K141586)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 8, 2019
Epimed International, Inc. % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K190256
Trade/Device Name: Rulo Radiofrequency Lesion Probe Regulation Number: 21 CFR 882.4725 Regulation Name: Radiofrequency lesion probe Regulatory Class: Class II Product Code: GXI Dated: February 5, 2019 Received: February 7, 2019
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew C. Krueger -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190256
Device Name RuloTM Radiofrequency Lesion Probe
#### Indications for Use (Describe)
The RuloTM Radiofrequency Lesion Probe is an injection probe which may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "Epimed", a company that designs products for experts. The logo features a red triangle on the left side, followed by the company name in bold, black letters. Below the company name, there is a tagline that reads "Designed by Experts for Experts" in a smaller font.
# 510(k) Summary: K190256
#### Submitter Information:
| Name:<br>Address: | Epimed International, Inc.<br>13958 Diplomat Drive<br>Farmers Branch, Texas 75234 |
|--------------------------------------|-----------------------------------------------------------------------------------|
| Telephone Number: | (518) 725-0209 |
| Fax Number: | (518) 725-0207 |
| Contact Person:<br>Telephone Number: | Preston H. Frasier<br>Manager – QA/RA<br>(518) 725-0209 Ext. 1300 |
| Email Address: | prestonf@epimed.com |
| Date Prepared: | January 23, 2019 |
# Device Name & Classification:
| Device Trade Name: | Rulo™ Radiofrequency Lesion Probe |
|-----------------------|----------------------------------------------------|
| Common Name: | Radiofrequency Lesion Probe |
| Classification Names: | Radiofrequency Lesion Probe, GXI (21 CFR 882.4725) |
| Models/Sizes: | 300-1615 (16 gauge); 300-1815 (18 gauge) |
## Predicate Device(s) Information:
Predicate Device: Diros OWL Sterile Single Use R.F. Insulated Cannulae Models 466 and DHC (K141586)
# Device Description:
The Rulo™ Radiofrequency Lesion Probe (Rulo Probe) creates radiofrequency
(RF) lesions on nerves and can also create percutaneous nerve blocks by
delivering anesthetic to the site. It is a single-use, surgically invasive device. It is
Page 1
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for a company called "Epimed". The logo consists of a red triangle on the left side and the company name in black letters on the right side. Below the company name is the text "Designed by Experts for Experts" in a smaller font.
inserted with the use of a flexible introducer accessory to contact a nerve in an area of the spine for less than 15 minutes to treat pain. The RF energy creates a lesion on nerve tissue that conducts pain signals. The tissue's ability to conduct electrical signals is disrupted by the lesion. The targeted nerve can be localized either by using RF electrostimulation through the cannula or by injecting contrast medium through the cannula with the use of the fluid injection port and using radiography concomitantly. The nerve may be blocked by injecting local anesthetic solution or making an RF lesion. The device is used by a trained physician in a clinical setting.
The device is constructed with a steel cannula and uses a variety of polymers for the insulation, hub, cable, connector, cap and fluid injection port
The flexible introducer accessory was cleared in K051860. The fluid injection port was cleared as an extension set in K020926.
## Statement of Intended Use:
The Rulo™ Radiofrequency Lesion Probe is an injection probe which may be used either for percutaneous nerve blocks with local anesthetic solution or for radiofrequency lesioning. A nerve is localized either by using electrostimulation through the needle or by injecting contrast medium through the needle and using radiography concomitantly. The nerve may then be blocked by injecting local anesthetic solution or a radiofrequency lesion may be made.
The Rulo Radiofrequency Lesion Probe Intended Use/Indications for Use statement is essentially identical to that of the predicate device, Diros OWL Sterile Single Use R.F. Insulated Cannulae Models 466 and DHC (K141586).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "Epimed". The logo consists of a red triangle on the left, followed by the word "Epimed" in bold, black letters. Below the word "Epimed" is the tagline "Designed by Experts for Experts" in a smaller, lighter font. The logo appears to be for a company or organization that specializes in expert-designed products or services.
# Technological Characteristics and Substantial Equivalence:
The Rulo Probe is similar in design and material composition to the predicate device, Diros OWL Sterile Single Use R.F. Insulated Cannulae Models 466 and DHC (K141586). Both have an insulated steel needle (cannula) where the opening in the insulation at the tip of the device delivers radiofrequency energy and fluid to the tissue it contacts. Both devices connect to radiofrequency generators and have a fluid port. The uninsulated delivery site on the cannula of the Rulo Probe is directional and can focus the energy and fluid to the desired tissue target while the Diros predicate device does not have the directional feature. The Diros device has a sharp tip for direct insertion into the tissue while the Rulo Probe is inserted with a flexible introducer accessory.
### Nonclinical Testing:
The results of the performance testing, demonstrate that the Epimed Rulo™ Radiofrequency Lesion Probe performs comparably to and is substantially equivalent to the predicate device (K141586).
There is not a harmonized standard or quidance document that directly applies to radiofrequency lesion devices. Applicable tests and requirements have been adopted from the following standards to compile the suite of performance testing:
ISO 594-1 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements
ISO 594-2 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: Lock Fittings
BS 6196 British Standard Specification for Sterile Epidural Catheters and Introducer Needles for Single Use
ANSI/AAMI/ EN 60601-1 Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance
ISO 7864 Sterile Hypodermic Needles for Single Use
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "Epimed", which is written in bold, black font. To the left of the word is a red triangle. Below the word "Epimed" is the phrase "Designed by Experts for Experts" in a smaller, lighter font.
In accordance with the applicable sections recommended per ISO 10993-1, the relevant patient contacting components have been shown to meet the necessary biocompatibility requirements. The biocompatibility of the materials used to manufacture the Rulo Probe have been tested and analyzed with discussion in the Biocompatibility Assessment. The biocompatibility testing which has been performed includes the following:
- . Cytotoxicity (per ISO 10993-5)
- Sensitization (per ISO 10993-10) .
- . Irritation or Intracutaneous Reactivity (per ISO 10993-10)
- Systemic Toxicity (per ISO 10993-11) .
The flexible introducer accessory is made of stainless steel and was cleared in K051860. The fluid port is made of PVC and was cleared as an extension set in K020926.
# Conclusions:
Epimed's Rulo Radiofrequency Lesion Probe and the predicate device, Diros OWL Sterile Single Use R.F. Insulated Cannulae Models 466 and DHC (K141586), have the same intended use/indications for use and the probes are very similar in technological characteristics, including the overall design, dimensions, delivery of RF energy and fluids, and choice of materials.
There are some minor differences in technological characteristics that do not raise new questions of safety or efficacy. The predicate device has an active tip that does not direct the energy or fluids in a specific direction. The Rulo Probe has a focused active tip for directionality in delivery of RF energy, and this direction is indicated by the handle. The predicate device has a sharper tip and can be inserted directly into the tissues, the while Rulo Probe has a blunt tip, and is inserted into the tissue with the use of a flexible introducer accessory. This accessory has been cleared in in the marketplace for several years, and does not present any new questions of safety or efficacy.
Page 4
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "Epimed". The logo consists of a red triangle on the left, followed by the word "Epimed" in bold, black letters. Below the word "Epimed" is the phrase "Designed by Experts for Experts" in a smaller font.
In additional to the same intended use, the technological characteristics are the same or similar, and where there are minor differences, there are not new or different questions of safety or efficacy, as demonstrated by bench and laboratory testing. Epimed's Rulo Radiofrequency Lesion Probe is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.