K-Shield Advantage Port Access Infusion Set (PAIS)
K190233 · Kawasumi Laboratories, Inc. · FPA · May 2, 2019 · General Hospital
Device Facts
Record ID
K190233
Device Name
K-Shield Advantage Port Access Infusion Set (PAIS)
Applicant
Kawasumi Laboratories, Inc.
Product Code
FPA · General Hospital
Decision Date
May 2, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The K-Shield Advantage Port Access Infusion Set (PAIS) is a safety port access device used to administer solutions to surgically implanted port.
Device Story
Safety port access infusion set; utilizes non-coring Huber needle to access surgically implanted ports; administers solutions. Features integral anti-needlestick protector (ANSP) to cover needle tip upon removal, preventing accidental needlestick injuries and bloodborne pathogen exposure. Used in hospital/healthcare settings by clinicians. Modifications from predicate include strengthened safety feature, improved gripping point for cannulation, and reduced needle cap bulkiness. Device is disposable, ethylene oxide sterilized, and includes tubing (micro or standard bore), clamp, female luer, and optional needleless access connector (NAC).
Clinical Evidence
No clinical data. Bench testing only. Performance verified via functional testing (tensile strength, leak tests, chemical tests per ISO 8536-4/ISO 7864) and simulated use testing for sharps injury prevention (n=599, 100% successful ANSP activation). Packaging and transportation testing (ISTA 2A, ISO 11607-1) confirmed integrity.
Technological Characteristics
Materials: Stainless steel needle, PVC hub/tubing (TOTM plasticized), PE needle cap, PP ANSP wing, epoxy glue. Energy: None (gravity feed). Connectivity: None. Sterilization: Ethylene Oxide (ETO). Form factor: Infusion set with Huber needle, tubing, and safety protector. Standards: ISO 8536-4, ISO 7864, ISO 11607-1, ISO 23908.
Indications for Use
Indicated for patients requiring administration of solutions to a surgically implanted port.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
K-Shield Advantage Port Access Infusion Set (K123344)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is written in blue letters next to it.
May 2, 2019
Kawasumi Laboratories, Inc. % Valerie Followell Regulatory Affairs Consultant Regulatory Compliance Associates, Inc. 10411 Corporate Drive, Suite 102 Pleasant Prairie, Wisconsin 53158
Re: K190233
Trade/Device Name: K-Shield Advantage Port Access Infusion Set (PAIS) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA, PTI Dated: April 1, 2019 Received: April 4, 2019
Dear Valerie Followell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tina Kiang, Ph.D Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
K-Shield Advantage Port Access Infusion Set (PAIS)
Indications for Use (Describe)
The K-Shield Advantage Port Access Infusion Set (PAIS) is a safety port access device used to administer solutions to surgically implanted port.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 5. 510(k) SUMMARY | K190233 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DATE: | March 28, 2019 |
| OWNER: | Kawasumi Laboratories, Inc.<br>Shinagawa Intercity Tower B<br>2-15-2, Konan,<br>Minato-ku, Tokyo, JAPAN 108-6109 |
| CONTACT PERSON: | Valerie Followell<br>Regulatory Affairs Consultant<br>Regulatory Compliance Associates®, Inc.<br>10411 Corporate Drive, Suite 102<br>Pleasant Prairie, Wisconsin 53158<br>Telephone: 847-400-6187<br>Email: v.followell@rcainc.com |
| DEVICE NAME: | Trade name: K-Shield Advantage Port<br>Access Infusion Set (PAIS)<br>Common name: Set, IV Fluid Transfer<br>set, administration,<br>intravascular<br>Classification:<br>Regulation<br>Number: 21 CFR 880.5440<br>Class: Product Class II<br>Code: FPA<br>Secondary<br>Product Code: PTI |
| PREDICATE DEVICE(S): | Previously cleared 510(k) for Kawasumi<br>Laboratories, Inc., K-Shield Advantage Port Access<br>Infusion Set |
| Predicate | Device | Indication | Clearance date | Company |
|-----------|----------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------|--------------------------------|
| 510(k) | Name | | | |
| K123344 | K-Shield Advantage<br>Port Access Infusion<br>Set (PAIS) | a safety port access device<br>used to administer solutions<br>to surgically implanted port. | 03/07/2013 | Kawasumi<br>Laboratories, Inc. |
{4}------------------------------------------------
| DEVICE DESCRIPTION: | The K-Shield Advantage Port Access Infusion Set (PAIS)<br>is a safety port access device used to administer solutions<br>to a surgically implanted port. This device is designed<br>utilizing a non-coring Huber needle to access the implanted<br>port. The K-Shield Advantage PAIS has an integral safety<br>device intended to protect against accidental needle stick<br>injuries and infection caused by blood borne pathogens.<br>Upon removal from the port, the anti-needlestick protector<br>(ANSP) covers the needle tip protecting against accidental<br>needle stick injuries.<br>The devices are disposable ethylene oxide sterilized<br>medical devices which are constructed from non-coring<br>needle (Huber needle), wing, tubing (either micro bore or<br>standard bore), clamp, female conical fitting and locking<br>cap. The device has optional injection site (needle<br>injection, needleless access connector (NAC), or no<br>injection site) and anti-needle stick protector (ANSP). |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INDICATIONS FOR USE<br>STATEMENT: | The K-Shield Advantage Port Access Infusion Set (PAIS)<br>is a safety port access device used to administer solutions<br>to surgically implanted port. |
| TECHNOLOGICAL<br>CHARACTERISTICS: | The K-Shield Advantage Port Access Infusion Set (PAIS)<br>is substantially equivalent to the predicate device with<br>regard to technological characteristics, performance, and<br>intended use.<br><br>The proposed K-Shield Advantage Port Access Infusion<br>Set has minor enhancements to the original design of the<br>predicate device. These modifications include<br>strengthening the safety device feature, improving gripping<br>point for easy cannulation and reducing bulkiness of the<br>needle cap. The modifications do not affect the intended<br>use of the device or alter the fundamental scientific<br>technology of the device. |
{5}------------------------------------------------
| | Features | Subject Device | Predicate Device | |
|---------------|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Regulatory | Device Name | K-Shield Advantage Port Access<br>Infusion Set (PAIS) | K-Shield Advantage Port Access<br>Infusion Set (PAIS) | Same |
| | 510(k)Number | K190233 | K123344 | Same |
| | Applicant | Kawasumi Laboratories, Inc. | Kawasumi Laboratories, Inc. | Same |
| | Class | 2 | 2 | Same |
| | Product code | FPA, PTI | FPA | Same |
| | C.F.R. Section | 880.5440 | 880.5440 | Same |
| | Intended use | This product is a safety port<br>access device used to administer<br>solutions to surgically implanted<br>port. | This product is a safety port<br>access device used to administer<br>solutions to surgically implanted<br>port. | Same |
| | Packaging | Blister rigid tray with heat sealed<br>lid | Blister rigid tray with heat sealed<br>lid | Same |
| | Size | Needle Gauge:<br>19G, 20G, 22G<br>Needle Length:<br>1 inch, 3/4 inch<br>Tube Length:<br>200mm (w/o NAC),<br>180mm x 100mm(NAC) | Needle Gauge:<br>19G, 20G, 22G<br>Needle Length:<br>1 inch, 3/4 inch<br>Tube Length:<br>200mm (w/o NAC),<br>180mm x 100mm(NAC) | Same |
| | Priming volume<br>(mL) | Standard bore : 1.3<br>Micro bore(w/o NAC): 0.4<br>Micro bore(with NAC): 0.9 | Standard bore : 1.3<br>Micro bore(w/o NAC): 0.4<br>Micro bore(with NAC): 0.9 | Same |
| | Needle | Huber Needle | Huber Needle | Same |
| | Connection Luer | Female | Female | Same |
| | Tubing | Standard bore/ Micro bore | Standard bore/ Micro bore | Same |
| Specification | Antineedle Stick<br>Protector (ANSP) | ANSP<br>*Minor modification to ANSP | ANSP | Different<br>(but<br>equivalent) |
| | Injection site | needleless access connector<br>(NAC) / none | needleless access connector<br>(NAC) / none | Same |
| | Component &<br>Material | See Table 2 | See Table 2 | Different<br>(but<br>equivalent) |
| | Sterilization | Ethylene Oxide Gas (ETO) | Ethylene Oxide Gas (ETO) | Same |
| | Shelf Life | 3 years | 3 years | Same |
{6}------------------------------------------------
| Component | Fluid Path | Material<br>(Subject device) | Material<br>(Predicate device K123344) |
|------------|----------------------|-----------------------------------|----------------------------------------|
| Needle | Y | Stainless Steel | Stainless Steel |
| Needle Cap | N | Polyethylene (PE) | Poly-propylene (PP) |
| ANSP Wing | Y<br>(*Skin contact) | Poly-propylene (PP) | Poly-propylene (PP) |
| Glue | N | Epoxy | Epoxy |
| Hub | Y | Poly-vinyl chloride (PVC) | Poly-vinyl chloride (PVC) |
| Tubing | Y | TOTM Poly-vinyl chloride<br>(PVC) | TOTM Poly-vinyl chloride<br>(PVC) |
### Table 5-2. Component and Material
### ASSESSMENT OF NONCLINICAL DATA:
Bench testing was performed and confirms that the device meets design requirements and specifications. The devices comply with the International standard ISO 8536-4:2010-Infustion equipment for medical use -Part 4: Infusion sets for single use, gravity feed [Including: Amendment 1 (2013)] mainly, to be used primarily in a hospital / healthcare facility setting. The devices were evaluated to be safe and effective as a medical device based on design verification test results.
Kawasumi Laboratories, Inc. conducts risk analysis according to the requirements of ISO 14971:2007 Medical Devices-Application of Risk Management to Medical Devices for the modifications to the device. The identified risks were adequately mitigated and verified by means of bench testing.
The device continues to meet the same material testing standards and sterilization process standards as the predicate devices. Device verification testing of performance has been verified through functional and simulate use testing.
{7}------------------------------------------------
| Test item | Standard | Result |
|-------------------------------------------------------------|-----------------------------|--------|
| Transportation | ISTA 2A | Pass |
| Packaging test<br>- Seal strength<br>- Dye penetration test | ISO 11607-1:2006(Amd 2014) | Pass |
| Visual inspection | In house standard | Pass |
| Functionality of Anti-needle stick protector | In house standard | Pass |
| Test for tensile strength<br>(ISO 8536-4) | ISO 8536-4:2010 (Amd 2013) | Pass |
| Test for tensile strength<br>(In house) | In house standard | Pass |
| Test for tensile Strength<br>(ISO 7864) *Cannula and hub | ISO 7864:2016 | Pass |
| Leak test | ISO 8536-9:2016 | Pass |
| Chemical test | ISO 8536-4:2010 (Amd. 2013) | Pass |
## Table 5-3. List of nonclinical tests
Given the needle cap is not patient contacting, the biocompatibility testing of the modified feature was determined not to be necessary.
| ORMANCE | Modified K-Shield Advantage PAIS products | |
|------------|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ING CAP | (Evaluation of cap functionality) | |
| OVAL FORCE | Purpose: | The needle cap used for the subject device<br>(modified K-Shield Advantage PAIS) is changed<br>into straw CAP, from T-CAP of predicate device<br>(K123344). In order to make sure that the<br>function of the needle cap has not changed,<br>measurement of cap removal force is performed<br>with the new and old caps. |
| | Method: | The removal force was measured when removing<br>the cap by hand from the product fixed on a<br>force gauge.<br>Sample size: N= 30 for each cap. |
| | Acceptance<br>Criteria: | The removal resistance of straw cap is lower<br>than T-cap.<br>(In-house criteria) |
| | Result: | Conforms |
| | Discussion<br>and<br>Conclusion: | Cap removal force was lower with straw cap<br>compared to T-CAP. Moreover, as per the<br>transportation test already performed there was<br>no cap detachment due to the transportation.<br>As a conclusion, it was confirmed that there was<br>no change in the function of the needle cap due<br>to this modification. |
# PERFC TESTI REMO
{8}------------------------------------------------
| Simulated Use Testing<br>(Sharps Injury<br>Prevention testing): | K-Shield Advantage Port Access Infusion Set includes<br>sharps injury prevention feature (Anti-needle Stick<br>Protector; ANSP). As ANSP of the device was<br>modified, simulated use testing was conducted to<br>evaluate the safety and effectiveness of its function.<br>The test was conducted as per ISO 23908:2011 and<br>FDA Guidance – Medical Device with Sharp Injury<br>Prevention Features. |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | For all of effective samples (n=599) evaluated, proper<br>activation of ANSP were observed. This result exceeds<br>the pre-defined acceptance criteria. The result<br>confirmed that the anti- needle stick function of the<br>modified K-Shield Advantage PAIS works as safely<br>and effectively as the predicate device. |
| CONCLUSIONS: | The K-Shield Advantage Port Access Infusion Set<br>(PAIS) is substantially equivalent to the predicate<br>device. Testing against established standards and<br>guidelines for its intended use and results of the risk<br>analysis demonstrate that the proposed device is<br>substantially equivalent to the predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.