SpaceFlex Knee is a mold intended to assist in the creation of a temporary disposable cement spacer for a total knee replacement (TKR) for skeletally mature patients undergoing a two-stage procedure due to a septic process.
Device Story
SpaceFlex Knee system consists of disposable polypropylene molds used in the operating room to create a temporary total knee replacement (TKR) prosthesis. The surgeon fills the molds with low-viscosity polymethyl methacrylate (PMMA) bone cement; once hardened, the resulting cement spacer is implanted into the patient's knee joint. The device is intended for patients undergoing a two-stage procedure to address periprosthetic joint infection. The spacer acts as a temporary joint replacement and provides local antibiotic delivery (via gentamicin-loaded cement). The device is single-use and non-sterile; it is used by orthopedic surgeons in a clinical setting. The output is a patient-specific temporary prosthesis that maintains joint space and aids in infection management during the interval between the removal of an infected implant and the placement of a permanent replacement.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included rupture tests, fatigue tests, gentamicin elution tests, wear tests, molding temperature analysis, and residual analysis on the polymer matrix. Results demonstrated that the device meets performance specifications and exhibits gentamicin release profiles equivalent to the OsteoRemedies Knee Modular Spacer (K112470) at 1-day, 1-week, and 2-week intervals.
Technological Characteristics
Disposable cement spacer molds constructed from medical-grade polypropylene. Operates by molding low-viscosity PMMA bone cement into a temporary knee prosthesis. Single-use; non-re-sterilizable. Mechanical properties and gentamicin elution profiles validated against international standards.
Indications for Use
Indicated for skeletally mature patients undergoing a two-stage total knee replacement procedure due to a septic process. Temporary prosthesis implantation period is 180 days or less. Patients must consistently use mobility assist devices (e.g., crutches, walkers) throughout the implant period due to mechanical limitations of the bone cement material.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Predicate Devices
StageOne Disposable Cement Spacer Molds for Temporary Knee Prosthesis, Biomet, Inc. (K161273)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
G21 S.r.l % Manjusha Bharadwaj Regulatory/Quality Consultant RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, Massachusetts 01913
Re: K190216
Trade/Device Name: G21 s.r.1 SpaceFlex Knee Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: MBB, JWH Dated: March 5, 2019 Received: March 11, 2019
Dear Manjusha Bharadwaj:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative. Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K190216
Device Name G21 s.r.1 SpaceFlex Knee
### Indications for Use (Describe)
Disposable cement spacer molds are indicated for use to mold a temporary total knee replacement (TKR) for sketetally mature patients undergoing a two stage procedure due to a septic process. The molded temporary knee prosthesis is indicated for an implantation period of 180 days or less. Because of inherent mechanical limitations of the device material (G3A Bone Cement), the molded temporary prosthesis is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers) throughout the implant period.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
{3}------------------------------------------------
K190216 Page 1 of 3
# 510(k) SUMMARY
G21 S.r.l. 's SpaceFlex Knee 510k Submission
#### l. SUBMITTER
G21 S.r.l. Via S. Pertini, 8 41039 S. Possidonio (MO) Italy
Phone: +39 0535 30312 +39 0535 417332 Fax:
Contact Person: Manjusha Bharadwaj Senior Quality/Regulatory Consultant RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, MA 01913 Phone: 1-(978) 358-7307 Fax: 1 (978) 477-0206
Date Prepared: 01 October 2018
#### II. DEVICE
| Trade/Device Name: | SpaceFlex Knee |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Temporary Knee Prosthesis |
| Classification Name: | Polymethylmethacrylate (PMMA) bone cement, antibiotic<br>Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis |
| Regulation Number: | 21 CFR 888.3027, 21 CFR 888.3560 |
| Regulatory Class: | Class II |
| Product codes | MBB, JWH |
#### lll. PREDICATE DEVICES
Primary Predicate: StageOne Disposable Cement Spacer Molds for Temporary Knee Prosthesis, Biomet, Inc. (K161273)
Additional Predicates: SPACER-K Temporary Knee Prosthesis, Tecres S.p.A. (K062274), OsteoRemedies Knee Modular Spacer (K112470, filed under Garventis LLC Ronald Reagan and International Trade Center)
{4}------------------------------------------------
#### IV. DEVICE INFORMATION
## Device Description:
G21 SpaceFlex Knee system are disposable cement spacer molds made of polypropylene (PP) intended to be filled with low-viscosity polymethyl methacrylate bone cement directly in the operating room. Once the bone cement has hardened, the SpaceFlex Knee system creates a polymethyl methacrylate based bone cement spacer for patients undergoing a two-stage procedure following a periprosthetic joint infection. The device can be used in either the left or right knee joint. The SpaceFlex Knee molds are single-use and cannot be re-sterilized.
# Intended Use:
SpaceFlex Knee is a mold intended to assist in the creation of a temporary disposable cement spacer for a total knee replacement (TKR) for skeletally mature patients undergoing a two-stage procedure due to a septic process.
## Indications for Use:
SpaceFlex Knee disposable cement spacer molds are indicated for use to mold a temporary total knee replacement (TKR) for skeletally mature patients undergoing a two-stage procedure due to a septic process. The molded temporary knee prosthesis is indicated for an implantation period of 180 days or less. Because of inherent mechanical limitations of the device material (G3A Bone Cement), the molded temporary prosthesis is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers) throughout the implant period.
#### V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technological features of the SpaceFlex Knee are similar to the predicates with one difference in the materials used for the manufacturing of the material used in the SpaceFlex Knee molds is medical grade polypropylene widely used in medical syringes and the predicate device molds are made of medical grade silicone.
Performance testing was completed on the SpaceFlex Knee device in accordance with established international standards to establish the mechanical properties and gentamicin release profile of the device. The results of the performance testing demonstrated that the SpaceFlex Knee device possess mechanical and elution characteristics equivalent to those of the predicate device. Specifically, the gentamicin release profile for the SpaceFlex Knee spacers was compared to the predicate device OsteoRemedies Knee Modular Spacer (K112470). The cumulative gentamicin release from SpaceFlex Knee was compared at time intervals of day 1, week 1 and week 2 to the predicate device's (OsteoRemedies Knee Modular Spacer (K112470)) gentamicin release profile available from the predicate labeling documents. The comparison of this data evaluated that the
{5}------------------------------------------------
cumulative release of gentamicin from the SpaceFlex Knee spacer is substantially equivalent to the additional predicate device OsteoRemedies Knee Modular Spacer (K112470).
Based on the comparison data of the mechanical properties and the gentamicin release profile, the subject device and predicate device are similar in design, fundamental scientific technology and comparable gentamicin elution profile. The minor differences between the subject and predicates do not lead to new safety or effectiveness issues.
### VI. PERFORMANCE DATA
The mechanical properties of the SpaceFlex knee System were tested in accordance with applicable international standards. In all instances the device functioned as intended and all results were satisfactory and met all performance specifications. The testing performed includes,
- 1. Rupture Test
- 2. Fatigue Test
- 3. Gentamicin Elution Test
- 4. The Wear Test Mechanical performance
- 5. Molding Temperature Analysis
- 6. Residual Analysis on Polymer Matrix
### VII. CONCLUSION
The SpaceFlex knee disposable cement spacer is substantially equivalent to other legally marketed cement spacers indicated for use to mold a temporary total knee replacement. The SpaceFlex knee has the same general intended use and substantially the same indications for use, technological characteristics, and principles of operation as the previously cleared primary predicate StageOne Disposable Cement Spacer Molds for Temporary Knee Prosthesis, Biomet, Inc. (K161273) and additional predicates SPACER-K Temporary Knee Prosthesis, Tecres S.p.A. (K062274) and OsteoRemedies Knee Modular Spacer (K112470).
The substantial equivalence discussion included in the submission demonstrates the substantial equivalence of the SpaceFlex Knee System (the subject device) and its predicate devices as well as describing the minor differences in the technological characteristics, which do not raise any new questions of safety or efficacy. The indications for use, performance testing mechanical/bench testing and gentamicin elution release profile of the Spacer are comparable to the primary as well as the additional predicates listed above and the data presented demonstrate that the SpaceFlex Knee system is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.