← Product Code [BTR](/productcode/BTR) · K190213

# Yikang Endotracheal Tube (K190213)

_Jiangxi Yikang Medical Instrument Group Co., Ltd. · BTR · Sep 6, 2019 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190213

## Device Facts

- **Applicant:** Jiangxi Yikang Medical Instrument Group Co., Ltd.
- **Product Code:** [BTR](/productcode/BTR.md)
- **Decision Date:** Sep 6, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5730
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The device is intended for oral or nasal intubation and for airway management.

## Device Story

Yikang Endotracheal Tube is a sterile, single-use medical grade PVC device for oral or nasal intubation and airway management. It consists of a tube shaft, cuff, inflation line, pilot balloon, one-way valve, and a 15mm conical connector; includes a radiopaque line for X-ray visualization and a Murphy eye. Available in cuffed and uncuffed configurations with internal diameters (I.D.) ranging from 2.0mm to 10.0mm. The device is inserted into the trachea to convey gases to and from the lungs; the cuff seals the airway to protect against aspiration and provide an unobstructed path during ventilation. Used in clinical settings by healthcare professionals. Benefits include secure airway maintenance and ventilation support.

## Clinical Evidence

Bench testing only. Performance testing conducted per ISO 5361:2016, including cuff diameter, cuff herniation, cuff leakage, tube collapse, kink resistance, and tracheal seal. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, subchronic toxicity, and implantation.

## Technological Characteristics

Materials: Medical grade PVC. Design: Cuffed/uncuffed, Murphy eye, radiopaque line, 15mm conical connector. Sizes: I.D. 2.0mm–10.0mm. Sterilization: ETO (SAL 10^-6). Standards: ISO 5361:2016 (performance), ISO 10993-1 (biocompatibility).

## Regulatory Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

## Predicate Devices

- Well Lead Endotracheal Tube ([K042683](/device/K042683.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 6, 2019

Jiangxi Yikang Medical Instrument Group Co., Ltd Nicholas Su Regulatory Affairs Specialist 188 Aihua Road, Jinxian County Nanchang, 331725 Cn

Re: K190213

Trade/Device Name: Yikang Endotracheal Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: Class II Product Code: BTR Dated: August 2, 2019 Received: August 12, 2019

Dear Nicholas Su:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190213

Device Name Yikang Endotracheal Tube

| Indications for Use (Describe) | The device is intended for oral or nasal intubation and for airway management. |
|--------------------------------|--------------------------------------------------------------------------------|
|--------------------------------|--------------------------------------------------------------------------------|

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;">☑</div> <div>Prescription Use (Part 21 CFR 201 Subpart D)</div> </div> | <div style="display:flex; align-items:center;"> <div style="margin-right:5px;">☐</div> <div>Over-The-Counter Use (21 CFR 201 Subpart C)</div> </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a circular emblem with two distinct symbols inside. The symbol on the left resembles a stylized depiction of a plant or a cluster of leaves, while the symbol on the right appears to be a character from a non-Latin alphabet, possibly Chinese or Japanese. The emblem is rendered in black against a white background, giving it a clear and contrasting appearance.

No.188 LiduAihua Ave.. Jinxian County, Nanchang, Jiangxi, China 331725 Tel: +86-791-85632226 Fax: +86-791-85632226

#### 510(k) Summary

(as requested by 21 CFR 807.92) K190213

| Submitter of 510(k): | Jiangxi Yikang Medical Instrument Group Co., Ltd  |
|----------------------|---------------------------------------------------|
| Address:             | 188 Aihua Road,Jinxian County, Nanchang, Jiangxi, |
|                      | China 331725                                      |
| Phone:               | +86 791 8563 2226                                 |
| Fax:                 | +86 791 8563 2226                                 |

Contact Person: Nicholas Su Regulatory Affairs Specialist E-mail: andy_smith@yeah.net

Date of summary: January 29, 2019

#### Proposed device:

| Trade Name:          | Yikang Endotracheal Tube |
|----------------------|--------------------------|
| Common Name:         | Endotracheal Tube        |
| Classification Name: | Tracheal tube            |
| Product Code:        | BTR                      |
| Panel:               | 73(Anesthesiology)       |
| Classification:      | Class II                 |
| Regulation Number:   | 868.5730                 |

# Legally Marketed Predicate Device:

| Trade Name:                 | Well Lead Endotracheal Tube        |
|-----------------------------|------------------------------------|
| 510(k) Number:              | K042683                            |
| Submitter of 510(k)/holder: | Well lead Medical Instruments Ltd. |

# Device Description:

Yikang endotracheal tube is a device intended for oral or nasal intubation and for airway management. It is a sterile and single use device, it is made from medical grade PVC. The Yikang endotracheal tube has cuffed and uncuffed, and is available in size I.D. 2.0mm through I.D. 10.0mm, and has a Murphy eye. It basically consists of tube shaft, cuff, inflation line, pilot balloon, one-way valve, and a male 15mm conical connector. A radiopaque line is incorporated into the full length of the tracheal tube.

When used tracheal tube is inserted through the trachea to convey gases to and from the trachea. Cuffed tracheal tube is intended to seal and protect aspiration of secretions, and to provide an unobstructed airway in patients during ventilation. A radiopaque line helps to

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a circular emblem that is divided into two halves. The left half contains a symbol that resembles a stylized plant or flower with three distinct sections at the top. The right half features a more complex symbol that appears to be a stylized character, possibly Chinese or Japanese, with vertical and horizontal strokes forming a grid-like structure. The emblem is presented in black and white, with the symbols standing out against the background.

No.188 LiduAihua Ave., Jinxian County, Nanchang, Jiangxi, China 331725 Tel: +86-791-85632226 Fax: +86-791-85632226

Comparison

Same

Same

Same

Similar

determine proper placement of the tracheal tube with the characterization of visibility of tracheal tube in X-rays.

Size I.D. 2.0mm through 4.5mm are intended for pediatric use (12 years old to less than 21years old), Size I.D. 5.0mm through 10.0mm are intended for adult use.

#### Intended Use:

The device is intended for oral or nasal intubation and for airway management.

#### Device Performance:

Testing results indicated that the product meets specified performance requirements, and complied with ISO5361:2016.

#### Comparison of Technological Characteristics:

Yikanq's endotracheal tube has the same technological characteristics as the predicate device. The design, material, form, fit, function and method of operation are similar.

| Substantial Equivalence   |                                                                                                            |                                                                                                            |
|---------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Element of<br>Comparison  | Submission Device                                                                                          | Predicate Device<br>K042683                                                                                |
| Intended Use              | The device is intended for<br>oral or nasal intubation and<br>for airway management.                       | The device is intended for oral<br>or nasal intubation and for<br>airway management.                       |
| Principle of<br>Operation | Normal                                                                                                     | Normal                                                                                                     |
| Type                      | Oral/nasal cuffed, Oral/nasal<br>uncuffed, Nasal cuffed, Nasal<br>uncuffed, Oral cuffed, Oral<br>uncuffed. | Oral/nasal cuffed, Oral/nasal<br>uncuffed, Nasal cuffed, Nasal<br>uncuffed, Oral cuffed, Oral<br>uncuffed. |
| Size                      | Oral/nasal cuffed (I.D 2.0-10.0)                                                                           | Oral/nasal cuffed (I.D 5.0-10.0)                                                                           |
|                           | Oral/nasal uncuffed (I.D 2.0-10.0)                                                                         | Oral/nasal uncuffed (I.D 2.0-10.0)                                                                         |
|                           | Nasal cuffed (I.D 3.0-10.0)                                                                                | Nasal cuffed (I.D 5.0-10.0)                                                                                |
|                           | Nasal uncuffed (I.D 3.0-10.0)                                                                              | Nasal uncuffed (I.D 2.0-10.0)                                                                              |
|                           | Oral cuffed (I.D 3.0-10.0)                                                                                 | Oral cuffed (I.D 5.0-10.0)                                                                                 |

# Substantial Equivalence

|                  |                         | Oral/nasal uncuffed (I.D 2.0-10.0)       | Oral/nasal uncuffed (I.D 2.0-10.0)                          | Same          |
|------------------|-------------------------|------------------------------------------|-------------------------------------------------------------|---------------|
|                  |                         | Nasal cuffed (I.D 3.0-10.0)              | Nasal cuffed (I.D 5.0-10.0)                                 | Similar       |
|                  |                         | Nasal uncuffed (I.D 3.0-10.0)            | Nasal uncuffed (I.D 2.0-10.0)                               | Similar       |
|                  |                         | Oral cuffed (I.D 3.0-10.0)               | Oral cuffed (I.D 5.0-10.0)                                  | Similar       |
|                  |                         | Oral uncuffed (I.D 3.0-10.0)             | Oral uncuffed (I.D 2.0-10.0)                                | Similar       |
| Tube             | I.D                     | Oral/nasal cuffed ±5mm                   | Oral/nasal cuffed ±3mm                                      | /             |
| Length           | 2.0                     | 170                                      | /                                                           | Meet          |
|                  | 2.5                     | 170                                      | /                                                           | requirements  |
|                  | 3.0                     | 190                                      | /                                                           | of            |
|                  |                         |                                          |                                                             |               |
| 3.5              | 215                     | /                                        | ISO5361:2016                                                |               |
| 4.0              | 215                     | /                                        | /                                                           |               |
| 4.5              | 240                     | /                                        | /                                                           |               |
| 5.0              | 265                     | 245                                      | Similar, and                                                |               |
| 5.5              | 275                     | 275                                      | meet                                                        |               |
| 6.0              | 285                     | 285                                      | requirements                                                |               |
| 6.5              | 305                     | 295                                      | of                                                          |               |
| 7.0              | 310                     | 310                                      | ISO5361:2016.                                               |               |
| 7.5              | 315                     | 315                                      |                                                             |               |
| 8.0              | 330                     | 330                                      |                                                             |               |
| 8.5              | 330                     | 330                                      |                                                             |               |
| 9.0              | 330                     | 330                                      |                                                             |               |
| 9.5              | 330                     | 330                                      |                                                             |               |
| 10.0             | 330                     | 330                                      |                                                             |               |
| I.D              | Oral/nasal uncuffed±5mm | Oral/nasal uncuffed ±3mm                 | /                                                           |               |
| 2.0              | 170                     | 140                                      | Similar, and                                                |               |
| 2.5              | 170                     | 145                                      | meet                                                        |               |
| 3.0              | 190                     | 165                                      | requirements                                                |               |
| 3.5              | 215                     | 185                                      | of                                                          |               |
| 4.0              | 215                     | 210                                      | ISO5361:2016.                                               |               |
| 4.5              | 240                     | 225                                      |                                                             |               |
| 5.0              | 265                     | 245                                      |                                                             |               |
| 5.5              | 275                     | 275                                      |                                                             |               |
| 6.0              | 285                     | 285                                      |                                                             |               |
| 6.5              | 305                     | 295                                      |                                                             |               |
| 7.0              | 310                     | 310                                      |                                                             |               |
| 7.5              | 315                     | 315                                      |                                                             |               |
| 8.0              | 330                     | 330                                      |                                                             |               |
| 8.5              | 330                     | 330                                      |                                                             |               |
| 9.0              | 330                     | 330                                      |                                                             |               |
| 9.5              | 330                     | 330                                      |                                                             |               |
| 10.0             | 330                     | 330                                      |                                                             |               |
| I.D              | Nasal cuffed, ±5mm      | Nasal cuffed, ±3mm                       | /                                                           |               |
| 2.0              | /                       | /                                        | /                                                           |               |
| 2.5              | /                       | /                                        | /                                                           |               |
| 3.0              | 290                     | /                                        | Meet                                                        |               |
| 3.5              | 290                     | /                                        | requirements                                                |               |
| 4.0              | 310                     | /                                        | of                                                          |               |
| 4.5              | 310                     | /                                        | ISO5361:2016.                                               |               |
|                  |                         |                                          |                                                             |               |
| 5.0              | 330                     | 330                                      | Similar, and<br>meet<br>requirements<br>of<br>ISO5361:2016. |               |
| 5.5              | 330                     | 330                                      |                                                             |               |
| 6.0              | 360                     | 360                                      |                                                             |               |
| 6.5              | 360                     | 360                                      |                                                             |               |
| 7.0              | 380                     | 380                                      |                                                             |               |
| 7.5              | 380                     | 380                                      |                                                             |               |
| 8.0              | 390                     | 390                                      |                                                             |               |
| 8.5              | 390                     | 390                                      |                                                             |               |
| 9.0              | 400                     | 400                                      |                                                             |               |
| 9.5              | 400                     | 400                                      |                                                             |               |
| 10.0             | 400                     | 400                                      |                                                             |               |
| I.D              | Nasal uncuffed, ±5mm    | Nasal uncuffed, ±3mm                     | /                                                           |               |
| 2.0              | /                       | 140                                      | /                                                           |               |
| 2.5              | /                       | 145                                      | /                                                           |               |
| 3.0              | 190                     | 165                                      | Similar, and<br>meet<br>requirements                        |               |
| 3.5              | 215                     | 185                                      |                                                             |               |
| 4.0              | 215                     | 210                                      | requirements<br>of<br>ISO5361:2016.                         |               |
| 4.5              | 240                     | 225                                      |                                                             |               |
| 5.0              | 265                     | 245                                      |                                                             |               |
| 5.5              | 275                     | 275                                      |                                                             |               |
| 6.0              | 285                     | 285                                      |                                                             |               |
| 6.5              | 305                     | 295                                      |                                                             |               |
| 7.0              | 310                     | 310                                      |                                                             |               |
| 7.5              | 315                     | 315                                      |                                                             |               |
| 8.0              | 330                     | 330                                      |                                                             |               |
| 8.5              | 330                     | 330                                      |                                                             |               |
| 9.0              | 330                     | 330                                      |                                                             |               |
| 9.5              | 330                     | 330                                      |                                                             |               |
| 10.0             | 330                     | 330                                      |                                                             |               |
| I.D              | Oral cuffed, ±5mm       | Oral cuffed, ±3mm                        | /                                                           |               |
| 2.0              | /                       | /                                        | /                                                           |               |
| 2.5              | /                       | /                                        | /                                                           |               |
| 3.0              | 170                     | /                                        | Meet                                                        |               |
| 3.5              | 180                     | /                                        | requirements                                                |               |
| 4.0              | 190                     | /                                        | of                                                          |               |
| 4.5              | 215                     | /                                        | ISO5361:2016.                                               |               |
| 5.0              | 225                     | 230                                      | Similar, and                                                |               |
| 5.5              | 240                     | 245                                      | meet                                                        |               |
| 6.0              | 255                     | 255                                      | requirements                                                |               |
|                  |                         |                                          |                                                             |               |
|                  | 6.5                     | 265                                      | 265                                                         | of            |
|                  | 7.0                     | 290                                      | 295                                                         | ISO5361:2016. |
|                  | 7.5                     | 310                                      | 300                                                         |               |
|                  | 8.0                     | 325                                      | 310                                                         |               |
|                  | 8.5                     | 335                                      | 320                                                         |               |
|                  | 9.0                     | 350                                      | 335                                                         |               |
|                  | 9.5                     | 350                                      | 350                                                         |               |
|                  | 10.0                    | 350                                      | 350                                                         |               |
|                  | I.D                     | Oral uncuffed, ±5mm                      | Oral uncuffed, ±3mm                                         | /             |
|                  | 2.0                     | /                                        | 125                                                         | /             |
|                  | 2.5                     | /                                        | 135                                                         | /             |
|                  | 3.0                     | 170                                      | 155                                                         | Similar, and  |
|                  | 3.5                     | 180                                      | 175                                                         | meet          |
|                  | 4.0                     | 190                                      | 200                                                         | requirements  |
|                  | 4.5                     | 215                                      | 215                                                         | of            |
|                  | 5.0                     | 225                                      | 230                                                         | ISO5361:2016. |
|                  | 5.5                     | 240                                      | 245                                                         |               |
|                  | 6.0                     | 255                                      | 255                                                         |               |
|                  | 6.5                     | 265                                      | 265                                                         |               |
|                  | 7.0                     | 290                                      | 295                                                         |               |
|                  | 7.5                     | 310                                      | 300                                                         |               |
|                  | 8.0                     | 325                                      | 310                                                         |               |
|                  | 8.5                     | 335                                      | 320                                                         |               |
|                  | 9.0                     | 350                                      | 335                                                         |               |
|                  | 9.5                     | 350                                      | 350                                                         |               |
|                  | 10.0                    | 350                                      | 350                                                         |               |
| Material         |                         | PVC                                      | PVC                                                         | Same          |
| Performance      |                         | Conforms to ISO5361                      | Conforms to ISO5361                                         | Same          |
| Biocompatibility |                         | Conforms to ISO10993-1                   | Conforms to ISO10993-1                                      | Same          |
| Labeling         |                         | Meet the requirements of 21 CFR Part 801 | Meet the requirements of 21 CFR Part 801                    | Same          |
| Sterility        |                         | Sterile, SAL: 1X10-6 ETO Sterilization   | Sterile, SAL: 1X10-6 ETO Sterilization                      | Same          |
| Single use       |                         | Yes                                      | Yes                                                         | Same          |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a circular emblem with two distinct halves. The left half contains a character that appears to be composed of stacked, box-like shapes, while the right half features a more complex character with vertical and angled strokes. The emblem is primarily black against a white background, and the characters are stylized and possibly symbolic.

No.188 LiduAihua Ave., Jinxian County, Nanchang, Jiangxi, China 331725 Tel: +86-791-85632226 Fax: +86-791-85632226

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a circular emblem with Chinese characters inside. The emblem is divided into two halves, with each half containing a different character. The characters are written in a traditional style, and the emblem has a simple, clean design. The emblem is black and white.

No.188 LiduAihua Ave., Jinxian County, Nanchang, Jiangxi, China 331725 Tel: +86-791-85632226 Fax: +86-791-85632226

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a circular emblem with two distinct symbols within it. The symbol on the left appears to be composed of vertical lines and a curved shape at the top, while the symbol on the right features a more complex structure with both vertical and horizontal lines. The emblem is enclosed by a thick, dark border, and the symbols inside are also dark, creating a strong contrast against the lighter background.

No.188 LiduAihua Ave., Jinxian County, Nanchang, Jiangxi, China 331725 Tel: +86-791-85632226 Fax: +86-791-85632226

# Discussion of Similarities and Differences

As indicated by above comparison table our Yikang Endotracheal Tube has same intended use, and similar design, material, technical characteristics, principle of operation and performance as predicate device, Well Lead Endotracheal Tube under k042683.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a circular emblem with two distinct calligraphic characters. The emblem is divided vertically, with one character on each side. The characters are stylized and appear to be of East Asian origin, possibly Chinese or Japanese. The emblem has a traditional and symbolic feel.

No.188 LiduAihua Ave.. Jinxian County, Nanchang, Jiangxi, China 331725 Tel: +86-791-85632226 Fax: +86-791-85632226

The additional size endotracheal tubes (oral/nasal cuffed (I.D2.0-4.5), nasal cuffed(I.D 3.0-4.5), and oral cuffed (I.D 3.0-4.5)) of propose device when compared to predicate device all complied with the requirements specified by ISO5361: 2016, compliance testing indicated in section 18: performance testing of this submission. The tube length of the proposed device for every type and size was similar as the predicate device, and it met the tube length requirements specified by ISO5361: 2016. Therefore, the differences between our proposed device and legally market predicate device (additional size and tube length) did not raise different questions of safety or effectiveness.

Based on the comparison of intended use, design, material, operation principle, and performance, our YiKang Endotracheal Tube is substantial equivalent to its predicate device which approved by FDA under K042683.

# Summary of Performance testing

Performance testing have been conducted on YiKang Endotracheal Tube per ISO5361: 2016, and all the testing results meet requirements of ISO5361: 2016 and defined acceptance criteria. Detailed testing conducted as follows:

- Cuff diameter ●
- Cuff herniation
- Cuff leakage ●
- Tube collapse
- Kink resistance
- Tracheal seal ●

# Biocompatibility

The YiKang Endotracheal Tube passed biocompatibility testing of cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, subchronic toxicity, and implantation per ISO10993-1.

# Conclusion

Based on the information included in this premarket notification submission, and in accordance with 21 CFR Part 807 (E), Jianqxi Yikang Medical Instrument Group Co., LTD. concludes that the YiKang Endotracheal Tube is substantially equivalent to the predicate device cleared under K042683 without raising different questions of safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K190213](https://fda-staging.innolitics.com/device/K190213)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
