Renovis A200 PS Knee System as part of the Renovis A200 Knee System
Applicant
Renovis Surgical Technologies
Product Code
JWH · Orthopedic
Decision Date
Jun 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The A200 Knee System is intended for use in total knee arthroplasty for the following indications: 1. Painful and disabled knee joint resulting from osteoarthritis, traumatic arthritis, traumatic arthritis where one or more compartments are involved. 2. Correction of varus, valgus, or posttraumatic deformity. 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure. This device is for cemented use only.
Device Story
Renovis A200 PS Knee System is a posterior stabilized (PS), tri-compartmental knee replacement prosthesis; replaces knee joint in patients with degenerative diseases where posterior cruciate ligament is absent. System comprises femoral components (CoCrMo), tibial inserts (UHMWPE/Vit E), and patellar components (UHMWPE/Vit E); intended for cemented fixation. Femoral components available in nine sizes; tibial inserts in three footprints with 10-23 mm thicknesses and two constraint levels (PS, PS Plus); patellar components in sombrero and domed designs. Used in orthopedic surgery; implanted by surgeons. Provides mechanical joint replacement to restore function and reduce pain. Reusable instrumentation (PPSU, silicone, stainless steel) included for surgical implantation.
Clinical Evidence
Bench testing only. Performance evaluated via tibial post fatigue strength, tibial tray locking mechanism testing, tibial-femoral constraint testing, tibial-femoral contact area/stress testing, and patello-femoral lateral subluxation/contact area/stress testing. Results were acceptable and consistent with contemporary PS knee systems.
Technological Characteristics
Materials: CoCrMo (ASTM F75-18), UHMWPE (ASTM F648-14), UHMWPE with Vitamin E (ASTM F2695-12), Ti6Al4V, and various stainless steels (ASTM A564, F899, A276, A240). Design: Posterior stabilized, tri-compartmental, cemented. Components: Femoral, tibial tray, tibial insert, patella. Sterilization: Gamma or Ethylene Oxide (ISO 11135, ISO 11137-2). Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients requiring total knee arthroplasty due to painful/disabled knee joints from osteoarthritis, rheumatoid arthritis, or traumatic arthritis; or for correction of varus, valgus, or posttraumatic deformity; or revision of failed osteotomy, arthrodesis, or previous joint replacement. Contraindicated for non-cemented use.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are displayed side-by-side.
June 20, 2019
Renovis Surgical Technologies % Sharyn Orton Senior Consultant MEDIcept, Inc. 200 Homer Avenue Ashland, Massachusetts 01721
Re: K190122
Trade/Device Name: Renovis A200 PS Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OIY Dated: May 16, 2019 Received: May 20, 2019
Dear Sharyn Orton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For: CAPT Raquel Peat, PhD, MPH, USPHS Director Office of Health Technology 6 Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190122
Device Name Renovis A200 PS Knee System
## Indications for Use (Describe)
The A200 Knee System is intended for use in total knee arthroplasty for the following indications:
1. Painful and disabled knee joint resulting from osteoarthritis, traumatic arthritis, traumatic arthritis where one or more compartments are involved.
2. Correction of varus, valgus, or posttraumatic deformity.
3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.
This device is for cemented use only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Traditional 510(k) Summary as required by 21 CFR 807.92(a) K190122
- A ) Submitted by: Renovis Surgical Technologies Inc. 1901 W. Lugonia Ave, Ste 340 Redlands, CA 92374 Phone: 909-557-2360 Fax: 909-307-8571
- Official Contact: Anthony DeBenedictis Vice President of Quality Assurance
- Consultant: Sharyn Orton, Ph.D. MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721
- Date prepared: 5/16/2019
- Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, B) Device Name: polymer/metal/polymer; Prosthesis, knee, patellofemorotibial. semi-constrained, cemented, polymer + additive/metal/polymer +
- Proprietary Name: Renovis A200 PS Knee System
- Device Class: Class II
- Regulation number: 21 CFR 888.3650
- Product codes: JWH, OIY
- Classification panel: Orthopedic
- C) Predicate: Primary: K120038 Renovis A200 Knee System K113550 Biomet Vanguard Complete Knee System
## D) Device Description:
The Renovis A200 PS Knee System is a patellofemorotibial polymer/metal/polymer semiconstrained prosthesis intended to be implanted to replace a knee joint. The system consists of femoral, tibial and patella components intended for use with bone cement.
{4}------------------------------------------------
# ENOVIS
The subject of this 510k Premarket Notification are additional offerings to the Renovis A200 Knee System (K120038) resulting from customer requests to better meet patient need (anatomy and surgical technique). The additional offerings are part of the FDA cleared A200 Knee System and are called the A200 PS Knee System. The PS Knee System is a primary, posterior stabilized (PS), tri-compartmental knee replacement system to treat knee disorders resulting from degenerative diseases where the posterior cruciate ligament is absent.
The new A200 PS Knee System includes:
- Femoral components in nine (9) sizes (1, 2, 3, 3N, 4, 4N, 5, 6, and 7) in a posterior ● stabilized (PS) design, each for left and right sides, manufactured from CoCrMo in compliance with ASTM F75-18 Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants
- Tibial insert components in three (3) footprint sizes; each for the left and right sides; with ● articulating thicknesses ranging from 10 mm to 23 mm; in two levels of constraint - PS and PS Plus. The tibial inserts have a mechanical lock to the tray (snap mechanism). Tibial inserts are manufactured from highly cross-linked UHMWPE (with and without 0.1% by wt Vit E), and 1020 UHMWPE in compliance with ASTM F648-14 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants and ASTM F2695 -12 Standard Specification for Ultra-High Molecular Weight Polyethylene Powder Blended With Alpha-Tocopherol (Vitamin E) and Fabricated Forms for Surgical Implant Applications
- Patellar components in five (5) sizes in sombrero and five (5) sizes in domed design. ● Patellar components are manufactured from highly cross-linked UHMWPE (with and without 0.1% by wt Vit E), and 1020 UHMWPE in compliance with ASTM F648-14 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants and ASTM F2695 -12 Standard Specification for Ultra-High Molecular Weight Polyethylene Powder Blended With Alpha-Tocopherol (Vitamin E) and Fabricated Forms for Surgical Implant Applications
- FDA cleared (K120038) Tibial tray (baseplate) components in (9) sizes, each for the left ● and right sides, and in each in two designs, a conical design (CoCrMo) and a keeled design (Ti6A14V)
The system also includes reusable instrumentation. The instruments are manufactured from medical grade Polyphenylsulfone (PPSU) and silicone, as well as Medical-Grade Stainless Steel (SS), in compliance with:
- . 17-4 SS per ASTM A564-2010
- 420 SS per ASTM F899-2012 ●
- 455 SS per ASTM A564-2010
- 465 SS per ASTM A564-2010 .
- 301 SS per ASTM A276-2017 or F899-2012 ●
- 302 SS per ASTM A276-2017 or F899-2012 .
- 304 per ASTM A240-2017, A276-2017, or F899-2012 ●
- 316 SS per ASTM A240-2017, A276-2017, or F899-2012 ●
- Nitronic 60 SS per ASTM F899-2012 .
{5}------------------------------------------------
## E) Indications For Use:
There is no change in the Indications for Use as the A200 PS Knee System is part of the A200 Knee System.
The A200 Knee System is intended for use in total knee arthroplasty for the following indications:
- 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valgus, or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure
This device is for cemented use only.
## F) Substantial Equivalence Discussion
Compared to the predicate devices, the Renovis A200 PS Knee System has the same Indications for Use.
Compared to the primary predicate Renovis A200 Knee System (K120038), the Renovis A200 PS Knee System:
- Has the same: .
- Materials:
- Femoral components from CoCrMo in compliance with ASTM F75-18 Standard . Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants
- Tibial inserts and patella components from highly cross-linked UHMWPE with and o without 0.1% by wt Vit E) in compliance with ASTM F648-14 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants and ASTM F2695-12 Standard Specification for Ultra-High Molecular Weight Polyethylene Powder Blended With Alpha-Tocopherol (Vitamin E) and Fabricated Forms for Surgical Implant Applications
- Tibial trays (from K120038) O
- Sombrero patella design and dimensions o
- Locking mechanism (from K120038) O
- Gamma or EO Sterilization O
- Similar tibial insert sizes and dimensions ●
- Additionally offers:
- Posterior stabilized design o
- Tibial inserts and patella components manufactured from UHMWPE 1020 in O compliance with ASTM F648-14 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants
- A Tibial Insert with a 23 mm articulating thickness o
- Domed patella component, but with the same dimensions and number of pegs as the O sombrero patella component
{6}------------------------------------------------
Similar to the predicate Biomet Vanguard Complete Knee System (K113550, PS design), the Renovis A200 Knee System offers:
- Posterior stabilized design ●
- . Domed, three peg patella
Testing of the Renovis A200 PS Knee System was conducted (and compared to data from contemporary PS designs or the literature, where applicable). The results for all testing were found to be acceptable for the system design.
G) Performance Testing - Bench
Testing of the Renovis A200 PS Knee System components was conducted and found to be acceptable for the following:
- Tibial Post Fatigue Strength (static and fatigue) ●
- Tibial Tray Locking Mechanisms Testing ●
- Tibial-Femoral Constraint Testing
- Tibial-Femoral Contact Area/Contact Stress Testing
- Patello-Femoral Lateral Subluxation and Contact Area/Contact Stress Testing ●
## Conclusion
It has been demonstrated that A200 PS Knee System testing is consistent with the performance testing of other knee systems, and will perform as expected. The Renovis A200 PS Knee System has the same Indications for Use as the predicate devices. Differences between Renovis A200 PS Knee System and predicate devices do not raise or cause different issues of safety or effectiveness. The Renovis A200 PS Knee System is substantially equivalent to the predicate devices.
### Consensus Standards H)
The Renovis A200 PS Knee complies with the following additional standards:
- . ISO 7207-2:2011 Implants For Surgery - Components For Partial And Total Knee Joint Prostheses - Part 2: Articulating Surfaces Made Of Metal, Ceramic And Plastics Materials [Including AMENDMENT 2 (2016)]
- ASTM F2083:2012 Standard Specification for Knee Replacement Prosthesis ●
- ANSI/AAMI ST72:2011 (R2016) Bacterial Endotoxins Test Methods, Routine ● Monitoring, And Alternatives To Batch Testing
- BS EN ISO 11137-2:2015 Sterilization of health care products - radiation - part 2: establishing the sterilization dose
- AAMI/ANSI/ISO 11737-1:2006 R(2011) Sterilization of health care products -● microbiological methods - part 1: determination of the population of microorganisms on product
- AAMI/ANSI/ISO 10993-7:2008(R)2012, Biological Evaluation of Medical Devices -● Part 7: Ethylene Oxide Sterilization Residuals
{7}------------------------------------------------
- ANSI/AAMI/ISO 11135:2014, Sterilization of health care products Ethylene oxide -. Requirements for development, validation, and routine control of a sterilization process for medical devices
- ISO 11607-1:2006/Amd 1:2014 Packaging for terminally sterilized medical devices -● Part 1: Requirements for materials, sterile barrier systems and packaging systems.
- ISO 11607-2:2006/Amd 1:2014 Packaging for terminally sterilized medical devices . Part 2: Validation requirements for forming sealing and assembly processes.
- ASTM F88-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F1886-16 Standard Test Method for Determining Integrity of Seals for Flexible ● Packaging by Visual Inspection
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for ● Medical Devices
- ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.