K190119 · 4l Health Co., Ltd. · OMP · Nov 1, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190119
Device Name
Foryou NPWT Dressing Kit
Applicant
4l Health Co., Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
Nov 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Foryou NPWT Dressing Kit is intended to be used with Foryou NPWT Device to manage wounds through removal of fluids, and the NPWT system is indicated for patients who would benefit from a suction device, particularly as it may promote wound healing by removal of fluids, excess exudate, infectious material, and tissue debris. Foryou NPWT Dressing Kit is indicated for use in patients with the following wounds: traumatic, dehisced wounds, partial-thickness burns, chronic wounds including pressure ulcers, diabetic foot ulcers and venous leg ulcers, acute wounds, flaps and grafts.
Device Story
Foryou NPWT Dressing Kit is a single-use, sterile wound dressing system used with the Foryou NPWT Device to provide negative pressure wound therapy. The kit includes reticulated flexible polyurethane foam, a transparent polyurethane film with acrylic adhesive, and a tube unit (drainage tube, port, clip, strap). The system is applied by healthcare professionals to manage wounds by removing fluids, excess exudate, infectious material, and tissue debris. The negative pressure environment promotes wound healing. The device is intended for clinical use in various wound types, including chronic and acute wounds. The kit components are sterilized via ethylene oxide and packaged in medical pouches. The device is substantially equivalent to the Carilex VT Dressing Kits.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including pressure conduction, exudate removal performance, package integrity testing (per sterilization validation), and biocompatibility testing (ISO 10993). Shelf-life was established via accelerated aging per ASTM F1980-16.
Technological Characteristics
Components: Reticulated flexible polyurethane foam, polyurethane film with acrylic adhesive, PVC/TPU/ABS tube unit. Sterilization: Ethylene Oxide. Biocompatibility: ISO 10993 compliant. Shelf life: 3 years (ASTM F1980-16). Form factor: Sterile, single-use kit in various sizes. Connectivity: None (mechanical interface with NPWT pump).
Indications for Use
Indicated for patients requiring suction for wound management, including traumatic, dehisced, acute, and chronic wounds (pressure, diabetic foot, venous leg ulcers), partial-thickness burns, flaps, and grafts. Contraindicated for necrotic tissue, malignancy, untreated osteomyelitis, untreated malnutrition, exposed arteries/veins/nerves/organs, anastomotic sites, non-enteric/unexplored fistulas, and exposed bone or tendon.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 1, 2019
4L Health Co., Ltd. Jackie Jian President No. 108, Dongxin Rd, DongJiang Hi-tech Dist. Huizhou, 516000 China
Re: K190119
Trade/Device Name: Foryou NPWT Dressing Kit Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: September 28, 2019 Received: October 1, 2019
Dear Jackie Jian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Cynthia J. Chang, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190119
Device Name
Foryou NPWT Dressing Kit
## Indications for Use (Describe)
Foryou NPWT Dressing Kit is intended to be used with Foryou NPWT Device to manage wounds through removal of fluids, and the NPWT system is indicated for patients who would benefit from a suction device, particularly as it may promote wound healing by removal of fluids, excess exudate, infectious material, and tissue debris. Foryou NPWT Dressing Kit is indicated for use in patients with the following wounds: traumatic, dehisced wounds,
partial-thickness burns, chronic wounds including pressure ulcers, diabetic foot ulcers and venous leg ulcers, acute wounds, flaps and grafts.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Submitter's Information 1.
| Company: | 4L Health Co., Ltd. |
|----------------------|----------------------------------------------------------------|
| Address: | No. 108, Dongxin Road, Dong Jiang Hi-tech Dist., Huizhou City, |
| | Guangdong, 516000, China |
| Registration number: | 3009425221 |
| Contact Person: | Jackie JIAN |
| Tel: | +86-752-5300546 |
| E-mail: | info@ForyouNPWT.com |
Date Prepared: October 31, 2019
#### Trade Name of the Device 2.
Foryou NPWT Dressing Kit
#### Common or Usual Name 3.
Negative Pressure Wound Therapy Dressing Kit
#### 4. Classification Name
| Classification Name: | Negative Pressure Wound Therapy Powered Suction Pump |
|-------------------------|------------------------------------------------------|
| Review Panel: | General & Plastic Surgery |
| Regulation Description: | Powered suction pump |
| Product Code: | OMP |
| Regulation Number: | 21 CFR 878.4780 |
| Regulation Class: | II |
#### Predicate Device Information 5.
Carilex VT Dressing Kits, Carilex Medical, K172725
#### 6. Device Description
Foryou NPWT Dressing Kit is the wound dressing kit to be used with Foryou NPWT Device for negative pressure wound therapy. It consists of three key components: the foam dressing, the tube unit, and the transparent film. The components are kitted into medical pouch and sterilized, or individually packaged and sterilized. Foryou NPWT Dressing Kit is single-use and sterile. It is available in different sizes: small, medium, large and extra large. The foam dressing should not be used for more than 24 hours and should be replaced with other dressings compatible with Foryou NPWT pump as stated in the Foryou NPWT pump labeling.
{4}------------------------------------------------
#### 7. Indications for Use
Foryou NPWT Dressing Kit is intended to be used with Foryou NPWT Device to manage wounds through removal of fluids, and the NPWT system is indicated for patients who would benefit from a suction device, particularly as it may promote wound healing by removal of fluids, excess exudates, infectious material, and tissue debris.
Foryou NPWT Dressing Kit is indicated for use in patients with the following wounds: traumatic, dehisced wounds, partial-thickness burns, chronic wounds including pressure injuries, diabetic foot ulcers and venous leg ulcers, acute wounds, flaps and grafts.
#### 8. Technological Comparison to Predicate Device
The comparison between Proposed Device and Predicate Device is as follows:
| Elements of<br>Comparison | Proposed Device | Predicate Device |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Foryou NPWT Dressing Kit | Carilex VT Dressing Kits |
| 510K number | K190119 | K172725 |
| Manufacturer | 4L Health Co., Ltd | Carilex Medical, Inc. |
| Classification | Class II | Class II |
| Regulation<br>Number | 21 CFR 878.4780 | 21 CFR 878.4780 |
| Product Code | OMP | OMP |
| Target Area | For management of chronic, acute,<br>traumatic, subacute and dehisced wounds,<br>partial-thickness burns, ulcers (such as<br>diabetic or pressure), flaps and grafts. | For management of chronic, acute,<br>traumatic, subacute and dehisced<br>wounds, partial-thickness burns, ulcers<br>(such as diabetic or pressure), flaps and<br>grafts. |
| Review Panel | General & Plastic Surgery | General & Plastic Surgery |
| Prescription<br>Medical<br>Device | Yes | Yes |
| Compatible NPWT<br>Pump | Foryou NPWT Device<br>(OMP: K113236) | Carilex VT Device<br>(OMP: K112853, K161410) |
| Intended Use | Foryou NPWT Dressing Kit is intended to be<br>used with Foryou NPWT Device to manage<br>wounds through removal of fluids, and the<br>NPWT system is indicated for patients who<br>would benefit from a suction device,<br>particularly as it may promote wound healing<br>by removal of fluids, excess exudates,<br>infectious material, and tissue debris. | Carilex VT Dressing Kits is intended to<br>be used with the Carilex VT device (S1<br>series) to manage wounds through<br>removal of fluids, and the NPWT<br>system is indicated for patients who<br>would benefit from a suction device,<br>particularly as the device may promote<br>wound healing by removal of fluids,<br>excess exudates, infectious material,<br>and tissue debris. |
| Indications for Use | Foryou NPWT Dressing Kit is intended to be<br>used with Foryou NPWT Device to manage<br>wounds through removal of fluids, and the<br>NPWT system is indicated for patients who<br>would benefit from a suction device,<br>particularly as it may promote wound healing<br>by removal of fluids, excess exudates,<br>infectious material, and tissue debris. | Carilex VT Dressing Kits is intended to<br>be used with the Carilex VT device (S1<br>series) to manage wounds through<br>removal of fluids, and the NPWT<br>system is indicated for patients who<br>would benefit from a suction device,<br>particularly as the device may promote<br>wound healing by removal of fluids,<br>excess exudates, infectious material,<br>and tissue debris. |
| Elements of<br>Comparison | Proposed Device | Predicate Device |
| | use in patients with the following wounds:<br>traumatic, dehisced wounds, partial-<br>thickness burns, chronic wounds including<br>pressure injuries, diabetic foot ulcers and<br>venous leg ulcers, acute wounds, flaps and<br>grafts. | and tissue debris.<br>The Carilex VT Dressing Kits is<br>indicated for use for patients with the<br>following wounds:<br>Traumatic<br>Dehisced wounds<br>Partial-thickness burns<br>Chronic wounds including pressure<br>ulcers, diabetic foot ulcers and venous<br>leg ulcers<br>Acute wounds<br>Flaps and grafts |
| Contraindications | Presence of Necrotic Tissue<br>Malignancy<br>Untreated Osteomyelitis<br>Untreated Malnutrition<br>Exposed Arteries, Veins, Nerves,<br>or Organs<br>Use Over Anastomotic Sites<br>Use in Non-enteric and Unexplored Fistulas,<br>as well for Use Over Exposed Bone or<br>Tendon | Presence of Necrotic Tissue<br>Malignancy<br>Untreated Osteomyelitis<br>Untreated Malnutrition<br>Exposed Arteries, Veins, Nerves,<br>or Organs<br>Use Over Anastomotic Sites<br>Use in Non-enteric and Unexplored<br>Fistulas, as well for Use Over Exposed<br>Bone or Tendon |
| Technology<br>Principle of<br>Operation | Foryou NPWT Dressing Kit is used with<br>Foryou NPWT Device to manage wound with<br>negative pressure and removal of exudate. | Carilex VT Dressing Kits is intended to<br>be used with the Carilex VT device (S1<br>series) to manage wounds through<br>removal of fluids. |
| User | Healthcare professional only | Healthcare professional only |
| Single Use or<br>Reusable | Single Use | Single Use |
| Sterile | YES | YES |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide |
| Biocompatibility | All components comply with ISO 10993 | All components comply with ISO 10993 |
| Packaging | Components are kitted into medical pouch<br>and sterilized, or individually packaged and<br>sterilized. | Individually packaged and sterilized<br>components are kitted into Tyvek<br>pouch. |
| Shelf Life | 3 years | 2 years |
| Dressing Kits<br>Components | Black Foam Dressing<br>Transparent Film<br>Tube Unit (Composed of drainage tube, port,<br>circle drape, clip, and strap) | Black Foam, Film Drape, Port Set<br>(composed of Circle Drape, Connecting<br>Tube, Port, Clip, Lockable Connector) |
| Foam Dimension | Small: 75x100x30 mm<br>Medium: 125x180x30 mm<br>Large: 150x260x30 mm<br>Extra Large: 250x350x15 mm | Small: 80x100x30 mm<br>Medium: 125x200x30 mm<br>Large: 150x250x30 mm |
| Foam Material | Reticulated Flexible Polyurethane Foam | Reticulated Flexible Polyether Based<br>Polyurethane Foam |
| Film<br>Dimension | 300x200mm , 350x320mm | 203x305mm |
| Film Quantity | Extra Large: 350x320mm, 4 pcs<br>Large Kits: 350x320mm, 2 pcs<br>Medium Kits: 350x320mm, 2 pcs<br>Small Kits: 300x200mm, 1 pc | Large Kits: 3 pcs<br>Medium Kits: 2 pcs<br>Small Kits: 1 pc |
| Film Material | Polyurethane Film Coated with Acrylic<br>Adhesive | Polyurethane Film Coated with Acrylic<br>Adhesive |
| Tube Unit Material | Polyvinyl chloride drainage tube,<br>Thermoplastic urethane port, | Unknown |
| Elements of<br>Comparison | Proposed Device | Predicate Device |
| | Acrylonitrile butadiene styrene clip,<br>Paper&polyethylene terephthalate strap. | |
| Packaging<br>Material | 83g coated medical paper pouch | Unknown |
{5}------------------------------------------------
# 4L Health Co., Ltd
Premarket Notification 510(k): Traditional
ーK190110
Page 3 / 4
{6}------------------------------------------------
### 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The following tests were performed to determine substantial equivalence:
- a. Package Integrity Tests
The components of Foryou NPWT Dressing Kit are kitted into medical pouch and sterilized, or individually packaged and sterilized.
The pouch tests have been done during the validation of the sterilization cycle. The tests results show that the pouch can meet the requirements of package integrity.
- b. Shelf-life Test
In accordance with ASTM F1980-16, the stability of the Proposed Device was established from the results of the testing data from an accelerated aging test. Based on the evaluation of the results of the testing data, the Shelf life of Foryou NPWT Dressing Kit is 3 years.
#### ﻦ Biocompatibility Tests
The biocompatibility tests were performed with Foryou NPWT Dressing Kits, which showed that Foryou NPWT Dressing Kits are biocompatible for the intended use for a 'limited' contact duration.
# d. Performance tests
The Foryou NPWT Dressing Kit met all pre-defined acceptance criteria and passed the tests for pressure conduction and exudate removal.
# 10. Clinical Test
Clinical tests were not performed.
# 11. Conclusion
The proposed device is substantially equivalent to the predicate device. The differences between the proposed device and the predicate device do not raise different questions regarding its safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.