K190105 · Taiwan Biomaterial Co., Ltd. · JCX · Feb 21, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190105
Device Name
ZOOM Aspiration Pump
Applicant
Taiwan Biomaterial Co., Ltd.
Product Code
JCX · General, Plastic Surgery
Decision Date
Feb 21, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ZOOM Aspiration Pump The ZOOM Aspiration Pump is intended for general suction use in hospitals or clinics. ZOOM Canister The ZOOM Canister is intended to collect aspirated fluids for disposal and prevent fluid ingress from damaging the ZOOM Aspiration Pump. ZOOM Aspiration Tubing The ZOOM Aspiration Tubing Set is intended to connect a suction catheter to the ZOOM Canister of the ZOOM Aspiration Pump and to allow the user to control the fluid flow.
Device Story
AC-powered oil-less positive displacement pump; delivers vacuum (0-29 inHg) for removal of bodily fluids. Components: pump housing with power button, vacuum regulator, analog pressure gauge, sliding canister shelf; 1000 mL single-use canister; sterile single-use aspiration tubing. Used in hospitals or clinics by healthcare professionals. User connects suction catheter to canister via tubing; adjusts vacuum level via regulator; monitors pressure via analog gauge. Canister collects fluids; overflow protection valve and filter prevent fluid ingress into pump. Benefits: provides consistent vacuum for fluid removal during clinical procedures.
Clinical Evidence
Bench testing only. No clinical data. Performance testing conducted per FDA guidance for Powered Suction Pumps, including IEC 60601-1, IEC 60601-1-2, ISO 10079-1, and IEC 60529. Biocompatibility testing performed per ISO 10993-1. All tests passed acceptance criteria.
Indicated for general suction use in hospitals or clinics for the removal of bodily fluids.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 21, 2019
Taiwan Biomaterial Co., Ltd. Monoj Kalita Deputy Manager, Regulatory Affairs 6F, No. 26-1, Sec.2, Shengyi Rd. Zhubei. 30261 Taiwan
Re: K190105
Trade/Device Name: ZOOM Aspiration Pump Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: JCX Dated: January 18, 2019 Received: January 22, 2019
Dear Monoj Kalita:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by David David Krause -S Krause -S Krause -S
Date: 2019.02.21 14:10:37 -05'00
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190105
Device Name
ZOOM Aspiration Pump
Indications for Use (Describe)
ZOOM Aspiration Pump The ZOOM Aspiration Pump is intended for general suction use in hospitals or clinics.
ZOOM Canister
The ZOOM Canister is intended to collect aspirated fluids for disposal and prevent fluid ingress from damaging the ZOOM Aspiration Pump.
ZOOM Aspiration Tubing
The ZOOM Aspiration Tubing Set is intended to connect a suction catheter to the ZOOM Canister of the ZOOM Aspiration Pump and to allow the user to control the fluid flow.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K190105
## ZOOM™ Aspiration Pump
Special 510(k)
## 510(K) SUMMARY
#### 1.1. APPLICANT INFORMATION
| Applicant Name | ---- Taiwan Biomaterial Co., Ltd. (TWBM) |
|----------------|--------------------------------------------------------------------------------------------------------------------------------|
| Address | ---- 6F, No. 26-1, Sec.2, Shengyi Rd., Zhubei City, Hsinchu County<br>30261, Taiwan. |
| Telephone | ---- +886-3-6683088 |
| Fax | ---- +886-3-6683099 |
| Contact Person | ---- Monoj Mon Kalita, Ph.D.<br>Deputy Manager, Regulatory Affairs<br>Phone: +886-3-6683088 ext. 309<br>Email: mon@twbm.com.tw |
| Date Prepared | ---- February 5, 2019 |
#### SUBJECT DEVICE 1.2.
| Trade Name | ---- ZOOM™ Aspiration Pump |
|---------------------|---------------------------------------------------------|
| Common/Usual Name | ---- Powered Suction Pump |
| Device Class | ---- Class II |
| Regulation Number | ---- 21 CFR 878.4780 |
| Classification Name | ---- Apparatus, Suction, Ward Use, Portable, AC-Powered |
| Product Code | ---- JCX |
### PREDICATE DEVICE NAME 1.3.
TWBM Pump (K180115)
#### 1.4. INTENDED USE
### ZOOM Aspiration Pump
The ZOOM Aspiration Pump is intended for general suction use in hospitals or clinics.
### ZOOM Canister
The ZOOM Canister is intended to collect aspirated fluids for disposal and prevent fluid ingress from damaging the ZOOM Aspiration Pump.
### ZOOM Aspiration Tubing
The ZOOM Aspiration Tubing Set is intended to connect a suction catheter to the ZOOM Canister of the ZOOM Aspiration Pump and to allow the user to control the fluid flow.
{4}------------------------------------------------
Special 510(k)
# ZOOM™ Aspiration Pump
#### 1.5. DEVICE DESCRIPTION
The ZOOM™ Aspiration Pump (ZOOM Pump) is an AC powered oil-less positive displacement pump capable of consistently delivering vacuum (0-29 inHg) for the removal of bodily fluids. The ZOOM Pump housing is fitted with a power (ON/OFF) button, one vacuum regulator, an analog pressure gauge and sliding canister shelf. The ZOOM Pump is provided non-sterile.
There are two accessories for the ZOOM Pump, the ZOOM™ Canister and ZOOM™ Aspiration Tubing. The ZOOM™ Canister is a non-sterile, single use 1000 mL canister which collects the removed bodily fluids. The ZOOM Canister is positioned in the sliding canister shelf and connected to the ZOOM Pump via the canister tubing and connectors. The ZOOM Aspiration Tubing is sterile, single use and connects to the blue connector on the canister lid.
## 1.6.
## TECHNOLOGICAL AND REGULATORY ATTRIBUTES COMPARISON
The technological and regulatory attributes of the subject device are compared with the predicate device as illustrated in the table below:
| Attributes | TWBM Pump<br>(Predicate Device) | ZOOM™ Aspiration Pump<br>(Subject Device) |
|------------------------|-------------------------------------------------------|-------------------------------------------|
| 510(k) Number | K180115 | K190105 |
| Classification Name | Apparatus, Suction, Ward Use,<br>Portable, AC-Powered | Same as predicate |
| Class | II | Same as predicate |
| Product Code | JCX | Same as predicate |
| Intended Use | General suction use in hospitals<br>or clinics | Same as predicate |
| Product Type | AC-Powered suction pump | Same as predicate |
| Manufacturer | TWBM | Same as predicate |
| Electrical requirement | 110-120Vac<br>60Hz | Same as predicate |
| Flow Rate | 0-0.8 SCFM (0-23 LPM) | Same as predicate |
| Vacuum Range | 0-29 inHg (0-737mmHg) | Same as predicate |
| Noise level | <60dB | Same as predicate |
{5}------------------------------------------------
# ZOOM™ Aspiration Pump
Special 510(k)
| Attributes | TWBM Pump<br>(Predicate Device) | ZOOM™ Aspiration Pump<br>(Subject Device) |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimensions | 12.2in×6.2in×11.3in<br>(30.9cm×15.7cm×28.7cm) | 12.3in×7.2in×12.4in<br>(31.3cm×18.3cm×31.5cm) |
| Weight | 14.8 lb (6.7 kg) | 15.2 lb (6.9 kg) |
| Duty Cycle | 45 min (20%) ON/ 3 hours (80%)<br>OFF | 45 min (97.8%) ON/ 1 min (2.2%)<br>OFF |
| Accessories | Canister<br>• TWBM Canister 1000 mL,<br>• Canister tubing,<br>• Overflow protection valve,<br>• Filter (Microbial, Hydrophobic),<br>• Single Use Only,<br>• Non-Sterile | ZOOM™ Canister<br>• ZOOM™ Canister 1000 mL,<br>• Canister tubing,<br>• Overflow protection valve,<br>• Filter (Microbial, Hydrophobic),<br>• Single Use Only,<br>• Non-Sterile |
| | Aspiration Tubing<br>N/A | ZOOM™ Aspiration Tubing<br>• Class II Exempt<br>• EO Sterilized<br>• Non-pyrogenic<br>• Biocompatible<br>• Single Use Only<br>• ID: 0.110"<br>• Length: 264.2cm<br>• Flow clamp |
| Operating Environment | Temp: 65°F~75°F (18°C~ 24°C)<br>Humidity: <75% RH<br>Pressure: Sea Level - 6000 ft<br>(1829m) | Same as predicate |
| Storage Environment | Temp: -20°F~120°F (-29°C~<br>49°C)<br>Humidity: <95% RH | Same as predicate |
{6}------------------------------------------------
# ZOOM™ Aspiration Pump
#### 1.7. SUMMARY OF DEVICE TESTING
ZOOM™ Aspiration Pump was subjected to Electromagnetic Compatibility testing in accordance with IEC 60601-1-2 and performance bench testing in accordance with ISO 10079-1, IEC 60601-1 as well as IEC 60529. Testing for ZOOM™ Aspiration Pump were conducted by following the FDA's "Guidance Document for Powered Suction Pump 510(k)s" (Sept. 30, 1998). All the test results from the performance testing passed the acceptance criteria set forth by the respective standards.
#### 1.8. SUMMARY OF NON-CLINICAL TESTING
The table below highlights the non-clinical performance testing of the ZOOM™ Aspiration Pump and its accessories, conducted for the determination of substantial equivalence with the predicate TWBM Pump and its accessory.
| Attributes | Acceptance Criteria | Result<br>(YES/NO) |
|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|--------------------|
| ZOOM™ ASPIRATION PUMP | | |
| Compliance with ANSI/AAMI ES60601-1 and<br>CAN/CSA C22.2 NO. 60601-1 | 100% Pass | YES |
| Compliance with IEC 60601-1-2 | 100% Pass | YES |
| Compliance with ISO 10079-1 | 100% Pass | YES |
| Compliance with IEC 60529 for IP Code | The requirement for the IP<br>21 marking should be met. | YES |
| ZOOM™ CANISTER (COMPLIANCE WITH ISO 10079-1) | | |
| Design and Operational Requirements for<br>ZOOM™ Canister and its Components | 100% Pass | YES |
| ZOOM™ ASPIRATION TUBING | | |
| Operational Requirements for ZOOM™<br>Aspiration Tubing and its Components<br>(Compliance with ISO 10079-1) | 100% Pass | YES |
| Compliance ISO 10993-1 | Biocompatible | YES |
{7}------------------------------------------------
Special 510(k)
## ZOOM™ Aspiration Pump
## CONCLUSION
1.9.
Both the ZOOM™ Aspiration Pump as well as the predicate TWBM Pump are AC-Powered Suction Pumps with similar accessories for fluid containment. Additionally, the non-clinical bench testing results and the comparison of technological characteristics, intended use, the principle of operation and operating environment support the determination of ZOOM™ Aspiration Pump to be substantially equivalent to the predicate TWBM Pump. It can be concluded that the subject device performs as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.