R2GATE is intended for use as a software interface and image segmentation system for the transfer of imaging information from a CBCT scanner. It is also intended as pre-planning software for dental implant placement and surgical treatment.
Device Story
R2GATE is a web and desktop-based software application used by licensed dentists for pre-operative dental implant planning. It ingests CBCT scan data (DICOM) and 3D surface models (STL files). The software provides tools for image segmentation, nerve canal tracing, and 2D/3D visualization (axial, coronal, sagittal, cross-section, panorama). Dentists use the interface to manipulate and position virtual dental implants from a pre-loaded library, perform cephalometric analysis, and measure anatomical dimensions. The output consists of treatment plans and STL files compatible with 3D printers or milling machines for surgical guide fabrication. By enabling precise virtual placement and visualization of anatomical structures, the device assists clinicians in treatment evaluation and surgical planning, potentially improving implant placement accuracy and patient outcomes.
Clinical Evidence
Bench testing only. Validation included verification of CBCT data integrity (DICOM), comparison of physical measurements (distance/angle) against the predicate, and precision measurement of STL-based surgical guide models. All results met acceptance criteria.
Technological Characteristics
Web and desktop application (Windows OS). Programming languages: C#, C++. Functions include image segmentation, 2D/3D visualization, nerve canal tracing, and implant library management. Connectivity: imports DICOM and STL files; exports STL for CAM/3D printing. Software-based image processing and measurement tools.
Indications for Use
Indicated for use as a software interface and image segmentation system for transferring CBCT imaging data and as pre-planning software for dental implant placement and surgical treatment in patients requiring dental implant procedures.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MegaGen Implant Co., Ltd. % Mr. Dave Kim President Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025
June 26, 2019
Re: K190096
Trade/Device Name: R2GATE Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: June 4, 2019 Received: June 4, 2019
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190096
Device Name R2GATE
Indications for Use (Describe)
R2GATE is intended for use as a software interface and image segmentation system for the transfer of imaging information from a CBCT scanner. It is also intended as pre-planning software for dental implant placement and surgical treatment.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|
| <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| | | 510(k) SUMMARY |
|----|------------------------|---------------------------------------------|
| | | K190096 |
| | | R2GATE Picture Archiving and Communication |
| | | System |
| 1. | Submitter Information: | MEGAGEN IMPLANT CO., LTD |
| | | 41940, 10F, 12 Gong Pyeong Ro |
| | | Jung Gu Daegu, 41940 |
| | | Republic of Korea |
| | Contact Person: | HyeJin Jung |
| | Telephone Number: | +82-70-4352-1120 |
| | Fax Number: | +82-70-7469-1120 |
| | Date Prepared: | 1/13/2019 |
| 2. | Device Name: | |
| | Proprietary Name: | R2GATE |
| | Classification Name: | Picture Archiving and Communication System |
| | CFR Number: | 21 CFR §892.2050 |
| | Device Class: | Class II |
| | Product Code: | LLZ |
| 3. | Predicate Device: | |
| | Trade/Device Name: | Blue Sky Bio Plan |
| | 510k Number: | K090607 |
| | Regulation Number: | 21 CFR 892.2050 |
| | Regulation Name: | Picture archiving and communications system |
| | Regulatory Class: | Class II |
| | Product Code: | LLZ. |
#### 4. Description of Device:
R2GATE is a web application and is intended for pre-operative planning to create and review plans for dental implant placement and surgical treatment, using MEGAGEN Implants. A dental implant plan can be edited with R2GATE Editor, a desktop software application, by a licensed dentist with clinical experience in implant surgery and medical image review.
The implant surgery plan can be used for manufacturing a surgical guide or for evaluation of treatment options by a licensed dentist.
5. Indications for Use:
R2GATE is intended for use as a software interface and image segmentation system for the transfer of imaging information from a CBCT scanner. It is also intended as pre-planning software for dental implant placement and surgical treatment.
6. Substantial Equivalence:
Table 6-1 Substantial equivalence comparison table
{4}------------------------------------------------
| Criteria | Predicate Devices | Device under<br>Consideration | K190096<br>Equivalency |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Element of<br>Comparison | Blue Sky Bio, LLC (K090607) | MegaGen Implant Co., Ltd, | Equivalency |
| Device Name | Blue Sky Bio Plan | R2GATE Windows | |
| Intended Use | Review and approve implant plan, Edit implant<br>plan for approval in online case review. | Create, edit and approve implant<br>plan. For approval in online case<br>review. | Same |
| Indications for Use | Blue Sky Plan is intended to be used as<br>conversion software for Computed<br>Tomography (CT) generated DICOM images<br>into a format that allows a dentist to assess the<br>anatomic topography of the maxilla and<br>mandible as well as location of important<br>structures. It allows the information to be used<br>for pre- surgical treatment planning of dental<br>implant procedures. The Blue Sky Plan<br>software is deployed on standard personal<br>computer hardware using a Windows operating<br>system. | R2GATE is intended for use as a software<br>interface and image segmentation system<br>for the transfer of imaging information<br>from a CBCT scanner. It is also intended<br>as pre-planning software for dental implant<br>placement and surgical treatment. | Same |
| Media for Delivery | - Software-File for download | - Software-File for download | Same |
| Principles of<br>operation | Online software application<br>Desktop software application | Online software application<br>Desktop software application | Same |
| Program language | C# C++ | C# C++ | Same |
| Operating System | Windows | Windows | Same |
| | software functionalities | | |
| Reorientation | Reorientation function to change the<br>direction of the CT image on display. | Reorientation function to change the<br>direction of the CT image on display. | Same |
| Nerve canal | Create a nerve carnal tracing model using the<br>tools provided on the 2D image. The cross<br>section of the model is displayed as a 3D image<br>on the volumetric rendering view. | Create a nerve carnal tracing model using<br>the tools provided on the 2D image. The<br>cross section of the model is displayed as a<br>3D image on the volumetric rendering<br>view. | Same |
| 3D model scan<br>implant planning | The scanned 3D model (STL file) can be<br>reconstructed and used for implant planning. | The scanned 3D model (STL file) can be<br>reconstructed and used for implant | Same |
| Image processing | Change the image displayed on the 2D or 3D<br>screen using a tool provided by the program. | Change the image displayed on the 2D or<br>3D screen using a tool provided by the<br>program. | Same |
| Printing output | Output data that can be used for implant surgical<br>guide fabrication. | Output data that can be used for implant<br>surgical guide fabrication. | Same |
| Output compatibility | The STL file generated can be loaded form<br>other software in conformity with the<br>standard. | The STL file generated can be loaded<br>form other software in conformity with<br>the standard. | Same |
| Screenshot | A screen view on a program can be stored as<br>an image file. | A screen view on a program can be<br>stored as an image file | Same |
| | | | K190096 |
| Measurement Tools | Measure the angle and length on a 2D<br>screen using the tool available in the<br>program. Measure the length using 2 points<br>and measure the angle between the two line<br>segments using 3 points. | Measure the angle and length on a 2D<br>screen using the tool available in the<br>program. Measure the length using 2<br>points and measure the angle between the<br>two line segments using 3 points. | Same |
| Implant Control | Implant model can be added, deleted, moved<br>and rotated on the screen. Only the implant<br>models in the implant library can be used. | Implant model can be added, deleted,<br>moved and rotated on the screen. Only<br>the implant models in the implant<br>library can be used. | Same |
| Anchor pin Control | Anchor-pins can be added, deleted, moved and<br>rotated on the screen. The user may use the<br>product provided by the program or enter the<br>diameter and length setting. | Anchor-pins can be added, deleted,<br>moved and rotated on the screen. The<br>user may use the product provided by<br>the program only. | Same |
| Cephalo analysis | The user can set points of the analysis<br>using the cephalon analysis method<br>provided by the program. | The user can set points of the analysis<br>using the cephalon analysis method<br>provided by the program. | Same |
| 2D Image View | The CT image data can be loaded for axial,<br>coronal, sagittal, cross section and<br>panorama view mode. | The CT image data can be loaded for<br>axial, coronal, sagittal, cross section and<br>panorama view mode. | Same |
| STL file<br>Management | Load a STL file or save a 3D model<br>created in the program as an STL file. | Load a STL file or save a 3D model<br>created in the program as an STL file. | Same |
| View Management | 2D/3D view can be zoomed, moved,<br>rotated. | 2D/3D view can be zoomed, moved,<br>rotated. | Same |
| Project Management | Proceed to a new project. Save the current<br>project or recall the previous project. | Proceed to a new project. Save the<br>current project or recall the previous<br>project. | Same |
| Wax Up Fabrication | Create the desired model by adjusting the<br>position, size, and direction of the basic wax<br>up model provided by the program. | No function to make a wax up model | Difference |
| Surgical Guide<br>Fabrication | Using the virtual STL model and implant<br>model, the surgical guide is constructed as a<br>point-set area on the surface of the STL model. | This SW does not provide the surgical<br>guide fabrication function. | Difference |
| Report Template | Includes a report template via PDF export and<br>print setting. A screen shot can be saved for<br>display. | No report document template. A screen<br>shot can be saved for display. | Partially<br>equivalent |
| Printing Output | Dental Surgical Guide | Dental Surgical Guide | Same |
| Output<br>Compatibility | Compatible with dental 3D printers,<br>milling machines and CAM equipment<br>that support the STL file format. | Compatible with dental 3D printers,<br>milling machines and CAM equipment<br>that support the STL file format. | Same |
{5}------------------------------------------------
{6}------------------------------------------------
### Analysis of differences
The proposed device has similar indications for use and mostly same basic functionalities in comparison with the predicate device. The functions, view, and module for both the subject and predicate device are used for review of radiographic images, creation of a dental treatment plan and implant surgical guide. The differences include the proposed implant devices that each program depends on. For an implant surgical guide and treatment simulation, R2GATE utilizes a pre-loaded implant device library featuring MEGAGEN IMPLANT family whereas Blue Sky Bio may have different implant library other than MEGAGEN Implant.
## 7. Software testing
Software verification and validation was conducted to ensure the functionality and compatibility of all system components and to support the safety and effectiveness of the proposed devices.
The verification and validation tests consist of the following activities:
- Unit test
- . Peer Code Review
- Integration test ●
- Internal release test
- Formal system test
- . Acceptance test
Verification and validation tests confirm that all user needs and SW performance requirements are fulfilled according to the design input.
Also the following comparison tests confirm the functionality, safety and efficacy of the proposed devices.
#### 8. Non-Clinical Performance Data
Non-clinical tests have been developed in-house including acceptance criteria. The test results demonstrated substantial equivalence between the subject device and the predicate device. (Document No: R2W-D34-PER-001)
(i) Validation of CBCT data (DICOM Image Data): Physical measurement values and numerical values (Hounsfield unit, HU) of distance and angle between specific points of CBCT data of R2GATE Windows and Blue Sky Plan Software. The distance and angle value between the Blue Sky Plan and R2GATE Windows were measured within the acceptable range
(ii) Verification of integrity of CBCT data (DICOM Image Data): Experiments on modification of CBCT data of R2GATE Windows and Blue Sky Plan Software. There is no modification or conversion of the DICOM file between the Blue Sky Plan and R2GATE Windows.
(iii) Verification of the surgical guide model of the dental implant procedure using the STL file. The degree of bonding or precision measurement between the design file and the actual structure were reviewed and compared., The differences were within the acceptable range.
{7}------------------------------------------------
The test results met the acceptance criteria and demonstrated equivalence between the subject and predicate device.
## 9. Conclusion Regarding Substantial Equivalence
The R2GATE has similar indications for use and incorporate the same fundamental functions, views and module as the predicate device Blue Sky Bio Plan cleared under premarket notification K090607. The performance test for both the subject and the predicate device has been conducted and the test outcome support substantial equivalence between R2GATE and Blue Sky Bio Plan without raising any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.