K190093 · Nxthera (A Boston Scientific Company) · KNS · Mar 25, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K190093
Device Name
Rezum Delivery Device, Rezum Generator
Applicant
Nxthera (A Boston Scientific Company)
Product Code
KNS · Gastroenterology, Urology
Decision Date
Mar 25, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm3 and ≤ 80cm3. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.
Device Story
Rezum System uses convective thermal energy to treat benign prostatic hyperplasia (BPH). System consists of reusable RF generator and single-use delivery device. Procedure involves converting sterile water into vapor outside body; vapor delivered to prostate tissue via needle in delivery device. Thermal energy from vapor ablates targeted prostate tissue. Generator controls vapor delivery and saline flush for urethral cooling. Used in clinical settings by physicians. Output is thermal ablation of prostate tissue; reduces symptoms and obstructions associated with BPH.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included biocompatibility (ISO 10993-1), tensile strength, functional tests, calorimetry, hardware/software verification, and IEC 60601 compliance. Modified device performance compared against predicate specifications after 24-month aging.
Technological Characteristics
Thermal ablation via water vapor. Components: RF generator (G2200) and delivery device (D2201). Materials include medical-grade resins and plastics with added plasticizers. Sterilization: Ethylene oxide (EO) to SAL 10-6. Connectivity: Standalone system. Software: Embedded control for RF energy and saline flush.
Indications for Use
Indicated for men ≥ 50 years of age with BPH, prostate volume ≥ 30cm3 and ≤ 80cm3, including hyperplasia of the central zone and/or median lobe.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 25, 2019
NxThera (A Boston Scientific Company) Justin Kapitan Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K190093
> Trade/Device Name: Rezum System Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS Dated: February 25, 2019 Received: February 27, 2019
Dear Justin Kapitan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Angel A. Solergarcia -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <b>Indications for Use</b> | |
| 510(k) Number (if known)<br>To be determined K190093 | |
| Device Name<br>Rezum System | |
| Indications for Use (Describe) | |
| The Rezūm System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm3 and ≤ 80cm3. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe. | |
| Type of Use (Select one or both, as applicable) | |
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740<br>EF |
|----------------------|-------------|----------------------------------------------|
|----------------------|-------------|----------------------------------------------|
{3}------------------------------------------------
#### 510(k) Summary for the Rezum System
#### A. Sponsor
NxThera Inc (a wholly-owned indirect subsidiary of Boston Scientific Corporation) Urology and Women's Health Division 100 Boston Scientific Way Marlborough, MA 01752
#### B. Contact
Justin Kapitan Sr. Regulatory Affairs Specialist 508-683-4518 justin.kapitan@bsci.com
or
Anand Patel Regulatory Affairs Specialist II 508-683-5320 anand.patel(@bsci.com
Date Prepared: February 25, 2019
# C. Proposed Device
| Trade name: | Rezūm System |
|------------------------|-----------------------------------------------------------------|
| Common/usual name: | Vapor Ablation Device |
| Classification Number: | 21 CFR 876.4300 |
| Classification Name: | Endoscopic electrosurgical unit and accessories |
| Classification: | Class II |
| Product Code: | KNS |
| Product Code Name: | Unit, Electrosurgical, Endoscopic (With Or Without Accessories) |
| Model Names/Number: | Rezum Delivery Device-D2201; Rezum C2 Generator-G2200 |
#### D. Predicate Device
| Trade name: | Rezūm System |
|------------------------|-----------------------------------------------------------------|
| Common/usual name: | Vapor Ablation Device |
| Classification Number: | 21 CFR 876.4300 |
| Classification Name: | Endoscopic electrosurgical unit and accessories |
| Classification: | Class II |
| Product Code: | KNS |
| Product Code Name: | Unit, Electrosurgical, Endoscopic (With Or Without Accessories) |
| Model Names/Number: | Rezum Delivery Device-D2201; Rezum C2 Generator-G2200 |
{4}------------------------------------------------
Identification of Predicate Device: NxThera Rezum System, K180237
# E. Device(s) Description
The reusable Rezūm Generator is provided with the following reusable components:
- Generator
- o One Power Cord
The single-use Rezum Delivery Device contains the following disposable components:
- One sterile Delivery Device with integrated cable and tubing ●
- One sterile Syringe
- One sterile Spike Adaptor
- One 50 ml Sterile Water Vial ●
## F. Intended Use/Indications for Use
The Rezum System is intended to relieve symptoms, obstructions, and reduce prostate tissue associated with BPH. It is indicated for men ≥ 50 years of age with a prostate volume ≥ 30cm3 and ≤ 80cm³. The Rezūm System is also indicated for treatment of prostate with hyperplasia of the central zone and/or a median lobe.
## G. Technological Characteristics Compared to Predicate
The principles of operation are identical between the predicate and subject devices: The Rezūm System converts water into vapor outside of the body and the vapor is delivered to the prostate tissue via a needle within the sterile Delivery Device. The vapor ablates the targeted tissue within the prostate via thermal ablation as energy is transferred from the vapor to the prostate tissue. The amount of vapor delivered is controlled by an RF Generator, which also controls the amount of saline flush used to cool the urethra during the procedure.
The differences between the subject device and the predicate are minor. Differences include:
- Second source for resin used to manufacture plastic handpiece components -. Additional supplier of equivalent resin to reduce supply disruptions
{5}------------------------------------------------
- Drain Line Addition of plasticizer to current material. The tubing is in indirect ● patient contact.
- Flush and Water Lines- Addition of plasticizer to the Flush Line and material ● changes to the Water Line indirect patient contacting components to facilitate extended shelf life.
- Drip Chamber Material change to improve durability and remove the need for ● primer during the bonding process. This component is in indirect patient contact.
- Shelf life extension in subject device
The technological characteristics remain equivalent to the predicate device because the modifications that are the subject of this submission are limited to improvements to the existing design.
# H. Substantial Equivalence
The modified NxThera Rezūm System is substantially equivalent to the NxThera Rezūm System (K180237). It has the same intended use for thermal ablation of BPH tissue and the same indications for use. The system design and principles of operation remain the same.
# I. Biocompatibility
Biocompatibility testing was performed to show that all patient contacting materials meet applicable biocompatibility standards per ISO 10993-1:2009 and the FDA guidance: Use of International Standard ISO 10993-1 "Biological evaluation of medical devices. Evaluation and testing within a risk management process."
The following testing was performed with passing results to support the biocompatibility of the device:
- Cytotoxicity ●
- Sensitization
- Irritation and Intractaneous Reactivity ●
- Accute Systemic Toxicity ●
- Materials Mediated Pyrogenicity ●
## J. Performance Testing
The predicate Rezum System has been tested and successfully meets all its physical and performance specifications on the bench including:
- Dimensions
- Tensile strength tests
- Full functional tests ●
- Calories tests ●
{6}------------------------------------------------
- Hardware tests ●
- Software verification and validation ●
- Packaging tests ●
- Distribution tests .
- Applicable IEC 60601 tests ●
Results of this testing demonstrate that the device meets the same performance criteria as the predicate and is therefore substantially equivalent to the predicate.
The modifications to the subject Rezūm System have been tested in the same manner to ensure compliance to the initial specifications. Based on the change assessment including risk analysis, the following design verification tests were repeated. The test methods used were the same as those submitted for the predicate.
- . Delivery Device tests before and after 24-month aging:
- Dimensions O
- Tensile strength tests o
- Full Functionality (temperature, pressure, etc.) O
- Tubing (compliance, kink, burst, etc.) o
- Calorimetry о
- Corrosion Resistance O
- Packaging and Distribution O
- Applicable IEC 60601 tests o
- Software verification and validation ●
- o Biocompatibility
The device continues to be sterilized by ethylene oxide (EO) to an SAL 106 level.
The conclusion of the assessments demonstrates that the modified device continues to function as intended in a manner equivalent to the predicate device. The modified device raises no new issues of safety or effectiveness compared to the predicate.
## K. Conclusion
Based on the test data, the same intended use, and same indications for use, the modified Rezūm System is substantially equivalent to its predicate, K180237.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.